Sara 3000
TM
Preventive Maintenance Schedule
KKX81011-US KKX 52180.GB/2 Issue 1 Aug 2000 June 2003
Preventive Maintenance Schedule • Equipment operators and Service/Maintenance Personnel should hold to this Preventive Maintenance Schedule. • This Preventive Maintenance Schedule documents the basic steps to ensure product safety. ARJO offers comprehensive service support programs to maximise the long-term safety, reliability and value of this product. For more information and quotation on these services, contact your local ARJO representative. • The equipment is subject to wear and tear. The following maintenance procedures must be performed when specified to ensure that the equipment remains within its original manufacturing specification. • Please refer to the Operating Instructions No. KKX81010.US and the Parts List No. KKX81013.US which are available from your ARJO dealer. • Unauthorized modifications on any ARJO Ltd. equipment may affect its safety and are in breach of any equipment warranty. ARJO will not be held responsible for any accidents, incidents or lack of performance that occur as a result of any unauthorized modifications to its products. • ARJO strongly advise and warn that only ARJO company designated parts, which are designed for the purpose, should be used on equipment and other appliances supplied by ARJO, to avoid injuries attributable to the use of inadequate parts.
Description of symbols used Warning: Means: failure to understand and obey this warning may result in injury to you or to others. Caution: Means: failure to follow these instructions may cause damage to all or part of the system or equipment. •
Note: Means: this information is important for the correct use of this system or equipment.
2
Preventive Maintenance Schedule Matrix Before each and every use
Action/Check
Every week
Every 12 months
Every 2 years
Every 3 years
SLINGS
X
Examine the sling, straps and clips for damage or fraying as required.
S
PATIENT SUPPORT ARMS
X
Visually check exposed surfaces for damage, sharp edges, etc. See Note 1.
X X
Visually check sling attachment points. Do not use if damaged. Make sure all labels are attached. See Operating Instructions for location.
S
Check to make sure the handgrips are secure. Rebond if required.
BATTERY CHARGER
S
Examine the charger and wires for integrity and connections.
MAIN HOIST
X X X
Visually check exposed surfaces for damage, sharp edges, etc. See Note 1. Operate the lift through its full range. Visually check the handset and cable for damage. See Note 2.
S S S S S S S S
Perform a full function test on the hoist. Check operation of the Stop/Reset and System Lower Override Device. Check batteries for leakage and/or deterioration. Replace if required. Make sure all fixings, screws, nuts are secure and tight. Check all castor wheels for wear. Replace as required. Check that the covers fit correctly and are not damaged. Replace as required. Check for evidence of corrosion. Replace as necessary. Check that batteries will charge when in charging station.
S
Replace all batteries every 3 years from installation.
Key: X = Customer Actions
S = Qualified Service Technician Actions
3
Additional Notes Note 1: Visually check all exposed parts, especially where patients or users will come in personal contact. Ensure that no cracks or sharp edges have developed that could cause patient or user injury or have become unhygienic. If found damaged in any way, contact your local qualified service technician. Note 2: Examine the condition of the handset and its cable. If these are damaged, do not use them and replace them immediately with a new handset and cable assembly.
Recommended Spares For a full list of recommended spares and spares packages, please refer to the Parts List Catalog No. KKX81013.US. This catalog is available from your local ARJO distributor.
4
Notes
5
Notes
6
Notes
7
SVERIGE ARJO SCANDINAVIA AB Box 61, 241 21 Eslöv Tel. 0413-645 00 Fax 0413-645 83 E-post: [email protected] www.arjo.com DANMARK ARJO SCANDINAVIA Postboks 51, 4632 Bjæverskov Tel. 45 93 27 37 Fax. 45 93 27 39 E-post: [email protected] Internet: www.arjo.com NORGE ARJO SCANDINAVIA Enebakkveien 117, 0680 Oslo Tel. 98 28 11 70 Fax. 23 05 11 99 E-post: [email protected] Internet: www.arjo.com BELGIQUE / BELGIË ARJO Hospital Equipment nv/sa Ternesselei 248 2160 Wommelgem Belgium Tel + 32 3 353 91 00 Fax + 32 3 353 91 01 Email: [email protected] CANADA ARJO Canada Inc. 1575 South Gateway Road Unit C Mississauga, ONTARIO L4W 5J1 Tel. 1 800 665 4831 Fax.1 800 309 1116 www.arjo.com
CESKÁ REPUBLIKA ARJO Hospital Equipment s.r.o. Strmá 35 CZ-616 00 BRNO Tel. 549 254 252 Fax. 541 213 550 DEUTSCHLAND ARJO Systeme für Rehabilitation GmbH Christof-Ruthof-Weg 6 55252 Mainz-Kastel Tel.: 06134 186 0 Fax: 06134 186 160 E-mail: [email protected] www.arjo.com ESPAÑA ARJO SPAIN, S.A. C/ San Rafael, nº 6 28108 ALCOBENDAS (MADRID) SPAIN Tel. 00 34 91 490 0636 Fax. 00 34 91 490 0637 FRANCE ARJO Equipements Hospitaliers sa 45 Avenue de l'Europe Eurocit BP133 59436 Roncq Cedex Tél. 03 20 28 13 13 Fax. 03 20 28 13 14 Email: [email protected] GREAT BRITAIN ARJO St. Catherine Street Gloucester, GL1 2SL ENGLAND U.K. Tel. (08702)430430 Fax. (01452) 428344 HONG KONG ARJO Far East Ltd. 1001-3 APEC Plaza, 49 Hoi Yuen Road, Kwun Tong, Kowloon, HONG KONG Tel. 852 2508 9553 Fax. 852 2508 1416
ITALIA ARJO Italia S.p.A. Via Poggio Verde, 34 00148 ROMA Tel. 06 6566356 - 57 Fax. 06 65663212 [email protected] www.arjo.com NEDERLAND ARJO NEDERLAND B.V. De Blomboogerd 8 4003 BX TIEL Postbus 6116 4000 HC TIEL Tel. 0344 64 08 00 Fax. 0344 64 08 85 E-mail: [email protected] www.arjo.com ÖSTERREICH ARJO GmbH Föhrenweg 5 A-6065 THAUR Tel. 05223 493350 Fax. 05223 49335075 ARJO Poland Sp. z o.o. ul. Lirowa 27 02-387 Warsaw Poland Tel./fax office: +48 22 882-06-26..28 Tel./fax service: +48 22 882-06-29 SCHWEIZ / SVIZZERA ARJO-Sic AG Florenzstrasse 1d Postfach - 4023 Basel SCHWEIZ Tél. 061 337 97 77 Fax 061 311 97 42 E-mail: [email protected] www.arjo.com USA ARJO, Inc. 50 North Gary Avenue Roselle, IL 60172 Tel. 1-800-323-1245 Fax. 1-888-594-ARJO(2756) www.arjo.com