HF440  CLINICAL PROCEDURE  No. 1  TOPIC  Substitution – Dialysate Fluids Bag change.  CLINICAL NOTE  Fluids bags will require changing when they are empty. The frequency of this is related to the substitution and or dialysate settings (mL/hr). The machine scales weigh the bags and as weight decreases to the preset alarm level (gms) the alarm will sound. This level will require a different setting depending on a decision to hang 2 or 3 bags. The residual weight alarm level is set with consideration of the empty weight of the bags and line tubing. e.g. 2 bags or 10 L, set alarm level at 500 gms. For 3 bags or 15 L set the alarm at 650 gms. The alarm level can be changed when the alarm sounds and the alarm window displays using the ‘set levels’ touch button. This can also be set after priming, before treatment from the ‘change container’ touch button. To edit this during treatment, simply lift a bag off the scales to set-off the alarm. Silence the alarm, select the ‘adjust alarms’ touch button, change the ‘substitution empty at’ weight and validate, then validate again to continue treatment.  PROCEDURE  The HF 440 will alarm with a pop up window. Press the alarm silence (red bell) button. Change the bag(s) hanging with new full bags onto scales hooks. Make sure each bag is opened, no line clamps on. Stabilise the bags from movement. Now press the green tick  on the front screen.