Please read this manual carefully before using the SW-180. This manual is comprised as an essential part in the SW-180. Save this manual at a designated place for your reference whenever required.
INDEX For Safe and Correct Use
1
Note on Operation Manual
1
WARNING SYMBOL MARKS
1
INTENDED USE OF THE PRODUCT
2
CONTRAINDICATIONS
2
Accessories
4
Standard Accessories
4
Optional Accessories
5
Specifications
5
Cautions before operating the unit
6
Cautions for operation
7
Cautions after use
8
Cautions for maintenance and inspection
9
Maintenance and Check items
10
Medical treatment procedure
11
Installation
12
Operation
13
Preparations and Start of Treatment
15
End of Treatment
20
Error Screens and Warning Screens
21
EMC
22
For Safe and Correct Use
Read the manual carefully and follow the instructions to use the device safely and correctly
●Note on Operation Manual This manual is an integral part of the equipment and components purchased. It contains important information regarding equipment usage, method of operation, and safety cautions. Carefully read the manual to fully understand its contents.
●WARNING SYMBOL MARKS Following warning symbol marks are used in this manual. indicates a danger of death or serious injury of an operator or patient if the operator ignores this symbol and use the SW-180. indicates a possibility of death or serious injury of an operator or patient when the operator ignores this warning and use the SW-180. indicates a possibility of death or serious injury of an operator or property damage when the operator ignores this caution and use the SW-180. This symbol is intended to draw attention to "Danger, Warning, or Caution."
Sample Indication "Beware of Electric Shok."
This symbol indicates a "Prohibited" action
Sample Indication "Disassembly Prohibited."
This symbol is intended to "Force" an action to be taken.
Sample Indication "Pull out the plug from the wall outlet."
SW-180
1
●INTENDED USE OF THE PRODUCT 1.Intended benefits from using SW-180 are relief of pain, reduction of muscle spasm and joint contracture. 2.Disorder of musculo-skeletal and chronic inflammatory are indications for the use of shortwave diathermy.
●CONTRAINDICATIONS 1.Metallic objects placed internally and externally 2.Implanted medical electronic devices such as cardiac pacemakers 3.Pregnancy 4.Ischemic tissue 5.Malignant tumor 6.Hemorrhage 7.Testes 8.Excessive edema 9.Impaired sensation 10.Acute injury or inflammation
2
SW-180
For Safe and Correct Use Do not use SW-180 concurrently with following medical electronic devices as it may cause malfunction. 1.Cardiac pacemakers, other implant-type devices or electrodes 2.Artificial heart-lung or other devices for life-support 3.ECG or other devices mounted on the patient's body 4.Do not permit patient's entry with above devices to the treatment room. Do not press the electrode with excessive force against the body. The resultant pressure may reduce the blood flow in the area and cause burn. Ask the patient whether he or she feels the warmth comfortable or not 2 to 3 minutes after starting the radiation to avoid overheating everytime output power is increased. Install SW-180 apart from other machines (including other SW-180s), and do not share AC outlet with other machines to avoid mutual interference. Special attention should be paid when apply it to the ears, nose and throat. It should be used at low output and treatment time should be within 5 minutes. When treating ENT, special attention should be paid to avoid direct irradiation to the eyes. Do not use the unit in the vicinity of electric bed. It may cause malfunction of the bed.
In an emergency Turn off the unit immediately
SW-180
3
Accessories Standard Accessories ① Main Unit Condenser Disk-applicator with Cable, φ135, 2pcs/pack
② 010642 ③ 011198
④ ●
② ●
Arm Unit, 2×
④ B010020 Test Tube ⑤ Operational Manual * B180157 Power Supply Cord (220–240 V, Type F) or B180290 Power Supply Cord (110–120 V, Type A) ⑤ ●
Power SW
③
LCD control
Electrode connectors
①
Fuses
GND terminal
AC line input
CAUTION : Use only designated accessories.
4
SW-180
Optional Accessories Specifications Optional Accessories ① 010643
Condenser Disk-applicator with Cable, φ70, 2pcs/pack
⑦ ●
② 010645
Coil Disk-applicator with Cable, φ70
③ 010644
Coil Disk-applicator with Cable, φ135
④ 010646
Condenser Plate-applicator-L with Cable, 2pcs/pack
⑧ ●
⑤ ● ② ●
⑤ 010829
Condenser Plate-applicator-M with Cable, 2pcs/pack
⑥ 010649
Coil Plate-applicator-S with Cable for ENT
④ ● ① ●
⑦ B010807 Strap (L), 45×1200mm ⑧ B010820 Strap (S), 45×600mm
③ ●
⑥ ● ⑨ ●
⑨ B080234 Core Filter
Specifications Power Supply
AC100V−120V±10% 50/60Hz or AC220V−240V±10% 50/60Hz
Power Consumption
① Continuous: 5W〜80W(5W step)
330VA
Safety class according to IEC 60601-1
ClassⅠ, Type BF
Timer 30-minute digital timer ± 10% Output Frequency
27.12MHz±162.72kHz
The number of memories : 28 Tuning Automatic Size Weight
835(H)×380(W)×320(D)mm 26kg
Ambient Temperature Relative Humidity Atmospheric pressure
② Fixed-Peak Pulsed Mode Pulse width table[unit : ms] Frequency Output 5W 10W 15W 20W 25W 30W
25Hz
50Hz
100Hz
200Hz
400Hz
600Hz
2.50 5.00 7.50 10.0 12.5 15.0
1.25 2.50 3.75 5.00 6.25 7.50
0.63 1.25 1.88 2.50 3.13 3.75
0.31 0.63 0.94 1.25 1.56 1.88
0.16 0.31 0.47 0.63 0.78 0.94
0.10 0.21 0.31 0.42 0.52 0.63
③ Variable Peak Pulsed Mode Average power output in Variable-Peak Pulsed Mode
Operation Condition Operation Environment 10 to 40℃ 30 to 85% 700 to 1060hPa
Output Power (Accuracy ± 30%)
Storage Environment −10 to 60℃ 30 to 85% 700 to 1060hPa
Transportation Condition −10 to 60℃ 30 to 95% 700 to 1060hPa
Peak Output 15W 35W 50W 65W 80W
450Hz 400μs 3W 6W 9W 12W 15W
90Hz 2ms
900Hz 400μs
180Hz 2ms
6W 12W 18W 24W 30W
SW-180
5
Cautions before operating the unit ■Caution before operating the unit
(
Allow the patient to become fully relaxed without being subjected to any mental or physical stress before receiving intended medical treatment.
)
1.Avoid using this unit in combination with other medical equipments. Otherwise, intrinsically correct diagnosis by the unit may be misinterpreted or confusing, and a dangerous situation may arise from the combined use of equipments. 2.The unit should be used by a lisensed medical practitioner. 3.Never operate the unit with wet hands. 1.For installation. (1)Install the unit at a place free from a splash of water. (2)Avoid installing the unit in the environment susceptible to undue humidity, temperature, dust, air containing salt or sulfur, or places exposed to direct sunbeam. (3)Install the unit at a flat stable place free from inclination, vibration or shock. (4)Do not use the unit in the inflammable environment containing oxygen, nitrous oxide, anesthetic gas, or inflammable disinfectant mixed with the air. (5)Do not install the unit at storage places of chemicals and pharmaceuticals, or at places subject to inflammable gas generation. (6)Observe the ratings of frequency, voltage and amperage(or consumption) of power. (7)Make sure that grounding is made firmly and correctly. 2.Check switches, keys, dials and their settings, and confirm that the unit functions properiy. 3.Check that all cables are correctly and safely connected. 4.Do not operate the unit jointly with other equipments other than the designated stimulator. 5.To ensure safety, never treat areas of a patient's ankle or knee with the electrode applied across the sides of the legs. Doing so will result in skin burns if the patient's legs touch even briefly.
6
SW-180
Cautions for operation 6.Hold the plug, not the cord, when connecting the cord to the AC outlet. Do not pull directly the cord when disconnecting the unit ; this may cause cord breakage, short-circuiting the unit or cause shock. 7.To operate the unit, connect it to a dedicated outlet. Avoid placing multiple loads on one electrical outlet. 8.Do not place the unit near extreme heat. This may deform the unit or interfere with its operation. 9.Use the unit independently and safely distant from other equipments. Do not connect the unit to the same outlet being used by other equipments. Otherwise, possible equipment interaction may result in faulty operation by the other equipments or the unit itself. 10.If the pair of electrodes is not applied to the human body, the unit will not output adequate short-wave power. In this case, the Test Tube will not emit light, even when positioned close to the electrode.
■Cautions for operation 1.When using the unit under high output power, note that the electrode cable may get hot. Therefore, make sure that the electrode cable is positioned away from you and the patient during medical treatment. Otherwise, you and/or the patient may be burned. 2.Avoid unprotected (i.e., bare skin) and prolonged contact with a Strap during Strap use. The Strap is made from natural rubber, which in rare cases may cause such allergic reactions as itching, reddening of the skin, nettle rash, swelling, blistering, breathing difficulty or other asthma-like symptoms, low blood pressure, and shock. In case of such reactions, immediately stop using the unit, and provide the patient with immediate medical treatment. 3.Be sure to clearly understand the doctor's diagnosis of and subscription given to the patient to determine the inclusion of any special medical precautions or instructions. 4.To ensure safety, never treat areas of a patient's ankle or knee with the electrode applied across the sides of the legs. Even a towel may not be able to prevent contact between the feet, with the possibility of burns occurring at the point of contact. Before beginning treatment, please check the following points. Failure to do so could result in burn, or other injury. (1)Make sure that the patient's skin is dry. If the skin is wet, be sure to wipe it dry.
SW-180
7
Cautions after use (2)If there is any ointment on the patient's skin, wipe it completely away. (3)Tell the patient to let you know immediately if he or she feels anything unusual during treatment. (4)Instruct the patient to remove any clothing containing metallic fibers, any bandage containing wet materials, all metallic accessories, metal objects, plastic compound, conductive or partly conductive objects, hearing aids, and other objects. They might cause burns or electric shock. (5)Do not apply shortwave radiation over clothing(either outer or undergarments) that has an anti-static coating. This could cause burns. (6)Avoid radiating over an area where a diaper is worn, because burns may occur. 5.Do not direct the radiator toward the unit or any other nearby equipment. This may cause the equipment to malfunction. In addition, shortwaves reflected from nearby objects may cause other equipment to malfunction (e.g., therapy units, telephones, automatic door openers, fire sensors, personal computers, and other electronic devices). Change the direction of the radiator or move it further away if necessary. 6.Please make sure that the amount of short-wave radiation is not excessive. Treatment time, output, and distance all need to be taken into account in determining the appropriate amount of radiation. Heat is generated within the body, whereas temperature is sensed by the skin. Therefore, close attention must be paid to increases in the temperature of internal tissues. 7.Continuously monitor both the patient and the unit for any signs of abnormality. Should an abnormality occur, turn off the unit, check the status of the patient, and take whatever action is necessary. 8.If the patient shifts his or her body position or if the position of the electrode shifts during treatment, the patient in some cases may notice a reduction in warmth. In that case, terminate treatment (reset output to "0") one time, or press the up key more than three seconds to retune the unit.
Precautions ●When moving the unit, be sure to lift the main frame and not the arm or electrode. ●When moving the unit, be careful not to collide with other objects.
8
SW-180
Cautions for maintenance and inspection ■Caution after use Be sure to return all switches and levers to their respective home positions and turn off the power. Unplug the Power Supply Cord from the AC outlet. Make sure to pull the plug, not the cord, when disconnecting the unit. Put the unit in order, then store it at the specified location.
■Cautions for maintenance and inspection 1.Do not try to repair or modify the unit, as this may result in fire or accident. In case of a malfunction, turn off the unit and contact the nearest dealer or importer. 2.Never remove the cover of the unit. The high-voltage generated inside may cause electric shock. 3.Do not hit or jar the unit, turn it over, or subject it to shock or vibration. If such accidents occur, damage may develop over time, eventually causing accidents or malfunction. 1.When the unit is not to be used for a prolonged period (i.e., in the shutdown state), be sure to disconnect the power plug from the outlet. Otherwise, the insulation may deteriorate, resulting in the risk of electric shock and leak-derived fire. 2.The Power Supply Cord must be protected against damage, breakage, unauthorized alteration, forcible bending, stretching, twisting, and bundling. Moreover, never place the cord under heavy objects, pinch it between objects, or abruptly jerk it for any reason. Otherwise, the cord may be damaged, thus posing an hazardous risk of fire, electric shock, and other accidents. 3.Cables and electrodes eventually deteriorate after a long period of ordinary use or short periods of excessive use, thus posing a potential risk of accidents. Therefore, conduct an annual inspection and replace deteriorated cables and electrodes with new ones. 4.The uint should be stored at a specified location with regard to the following points. (1)Store the unit at a place free from a splash of water.
SW-180
9
Maintenance and Check items (2)Do not store the unit in the environment susceptible to undue humidity, temperature, dust, air containing salt or sulfur, or places exposed to direct sunbeam. (3)Store the unit at a flat stable place free from inclination, vibration or shock. (4)Do not store the unit in the inflammable environment containing oxygen, nitrous oxide, anesthetic gas, or inflammable disinfectant mixed with the air. (5)Do not store the unit at storage places of chemicals and pharmaceuticals, or at places subject to inflammable gas generation. 5.Be sure to make it a pre-operation rule to confirm proper operation of the unit and accessories. 6.Annually inspect the unit and accessories to ensure that original performance is maintained, and check for any deterioration and wear due to friction. 7.The accessories must be routinely checked for possible malfunction. Promptly replace any faulty accessory to prevent possible accidents. 8.Before restarting the unit after a prolonged period of shutdown, be sure to check the unit to confirm normal and safe operation.
CAUTION ●Never clean the unit and accessories using a cloth moistened with thinner, gasoline, polihing powder, hot water, or chemical agents. Only use a cloth moistened with lukewarm water or a neutral detergent and sufficiently wrung out.
■Disposable of equipment and accessories Dispose the unit and accessories according to local law to minimize the effect to enviroment.
10 SW-180
Medical treatment procedure ■Maintenance and Check Items Always check the unit and its accessories before using them. Also, inspect them once a year to cheak for preservation or degradation of the performance and the degree of wear. Item
Description
Appearance and markings
Method
Check for damage, panel distortion, and flickering indicators or displays.
Operation
Turn on the power switch and check whether the device functions normally.
Accessories
Check for damaged accessories and cables.
Safety function
Check whether the zero-start function works nomally.
Check visually Check by operating the device Check visually Check by operating the device
●Replacement Parts and Consumables Parts Name
Replacement Method
Time of Replacement
Electrode
Have your dealer or manufacturer service this component.
When inspection reveals deterioration.
Strap
Have your dealer or manufacturer service this component.
When inspection reveals deterioration.
1.Should a pre-operation check or inspection detect any trouble with the unit, be sure to contact the manufacturer and report the problems. 2.Should any trouble be detected during the operation, stop operation immediately, turn off power, then contact your authorized agent or the manufacturer. 3.Contact a dealer or an importer about calibration of the unit. Never calibrate it by yourself.
■Medical treatment procedure 1.Never conduct treatment with the patient sitting in a metal chair or on a metal bed. 2.Never apply the electrode directly onto the patient's skin; always apply it over a dry towel. 3.Refer to the danger and related warnings described on page 2 and 3.
SW-180 11
Installation ■Installation Fuses
1
Plug one end of the Power Supply Cord into the AC line input connector on the rear panel of the main unit; plug the other end into the AC outlet. Before doing this, make sure that the power switch is set OFF. AC line input
GND terminal
2
Adjust LCD intensity with the LCD control knob on the rear panel, if necessary.
Power SW LCD control
Electrode connectors
3
Select the type of electrode according to the area to be treated and the purpose of treatment. Connect the electrode cable to the electrode connector(s) on the rear panel of the Main Unit. Note: The antenna electrode should be installed to the arm. Install the left Holding Arm (also right Holding Arm) on the rear panel with the four screws attached.
4
Insert the electrode into the rod. Turn the electrode the dimple to be positioned under the screw and secure it. How to discriminate between types of electrodes: 1. The condenser-type electrode has one electrode cable. It is displayed on the screen as follows: 2. The coil-type electrode has two electrode cables. It is displayed on the screen as follows:
12 SW-180
Operation ■Key and Dial Operations ELECTRODE
MAX
FIXED PEAK
MODE
CW TIME
REPEAT
80 W 2
VARI. PEAK
PROG. PROG. SELECT MAKING
5 min 20 min
10 min 25 min
15 min 30 min
MODE C.
TIME C.
OK
3
3 2
4
1 The SW-180 is operated with the following four types of keys:
① ● Function Keys ② ● UP/DOWN Keys ③ ● RIGHT/LEFT keys ④ ● Output Dial ①Function Keys : Each function key corresponds to the item marked immediately above the key. The item varies with the screen.
②UP/DOWN Keys : Selection keys for modes, treatment times, Program No. and Items. ③RIGHT/LEFT keys : Keys to select the conditions of each item. ④Output Dial : 1) Turn it clockwise to start treatment at an output power selected. 2) Turn it counterclockwise to 0W to stop treatment. 3) Turn it clockwise to select an output power, then start treatment again in preset time after tuning.
SW-180 13
1
Turn the power switch on. Switching the power on displays the"Initial Check Screen."
Initial Check Screen
Error Screen The circuit, electrode, and
The screen shown at left will
ERROR
overall system are checked
Internal Check
appear if the electrode is not
here to determine whether
・Circuitry check ・Electrode check ・System check
normal or not.
inserted securety. No electrodes are connected
The same screen will appear
Solution : Connect electrodes.
if the electrode is pulled out of the Main Unit.
0%
50%
100%
Shortwave Therapy System
2
When processing is completed on the Initial Check Screen, the Initial Screen opens. When the machine has a coil-type electrode (large) connected: ELECTRODE
MAX
TIME
REPEAT
VARI. PEAK
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
ELECTRODE
80 W
FIXED PEAK
MODE
When the machine has a condensertype electrode (medium) connected:
This arrow is manipulated by using the UP/DOWN keys.
MAX
FIXED PEAK
MODE
TIME
REPEAT
80 W VARI. PEAK
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
① Initial settings: Mode: FIXED PEAK Time: 20 minutes ② The ELECTRODE indication varies with the type of electrode connected to the machine. This variation is automatically accompanied by a change in the maximum wattage. Max. wattage: 70φ Condenser electrode, 70φ Coil electrode, twin plate electrodes : 30W Other electrodes : 80W
③ Names and description of items: ● REPEAT :Invokes the previous treatment screen (provided that the Initial Screen is restored). ● MODE C. :(Mode Change): Treatment mode selection key ● TIME C. :(Time Change): Treatment time selection key ●
OK
:Key to sets up the parameters.
※ The MODE C. key and TIME C. key enable selection of an output mode and treatment time, regardless of where the item selection arrow (s) is positioned.
14 SW-180
Preparations and Start of Treatment 3
Select the desired mode on the Initial Screen. The selected item then appears high lighted in reverse video. (1) FIXED PEAK --- Fixed-Peak Pulsed Mode
(4) PROG.SELECT --- Treatment mode by registered program
(2) VARI.PEAK --- Variable-Peak Pulsed Mode (5) PROG.MAKING --- Program creation mode (3) CW --- Continuous Mode ●Mode selection is made by using the MODE C. key or RIGHT/LEFT keys. Note: When using the RIGHT/LEFT keys, the item pointed to by the selection arrow is changed. ●Time is changed by using the TIME C. key. Note: To change the time using the RIGHT/LEFT keys, first use the UP/DOWN keys to position the selection arrow at the desired time.
4
Once a selection item has been defined, press the [OK] key. ●Pressing the [OK] key after selecting a treatment mode opens the setup screen of the selected mode. ●Pressing the [OK] key after selecting a program opens the corresponding setup screen. Mode screen appears after pressing the [OK] key:
1. When FIXED PEAK is selected ELECTRODE
MAX
VARI. PEAK
FIXED PEAK
MODE
TIME
REPEAT
80 W
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
An 80W condenser-type electrode is connected.
Pressing the [OK] key on the above screen opens the following screen: ●Names and functions of the keys ELECTRODE MAX 80 W FIXED PEAK
OUTPUT MODE
FREQUENCY
25 Hz 200 Hz
50 Hz 400 Hz
100 Hz 600 Hz
Turn INTENSITY clockwise to begin therapy.
V. PEAK
CW
PROG.
V.PEAK … To Variable-Peak Pulsed Mode CW
… To Continuous Mode
PROG. … To selection of existing program MAIN W. … Opens the Initial Screen.
MAIN W.
Use the RIGHT/LEFT keys to select the desired frequency on the right of FREQUENCY, then turn the Output Dial. Treatment then begins.
SW-180 15
●The initial setting is [25 Hz]. If this is the desired frequency, you can start treatment right away by simply turning the Output Dial. ●Names and functions of the keys
●Tuning Screen MAX
V.PEAK … To Variable-Peak Pulsed Mode
PRESET FRQ 600 Hz WIDTH 0.63 ms TIME 30min
80 W
F.PEAK
5w Peak 80 w 28:58 min. AV
CW
PROG. … To selection of existing program TIME C. … Pressing this key lets you change
TUNING
treatment time at any moment.
Adjust power by INTENSITY dial
V. PEAK
… To Continuous Mode
CW
PROG.
TIME C.
●Even if tuning is lost during treatment, the Auto-Tuning function will activate to automatically establish correct tuning. MAX
PRESET FRQ 600 Hz WIDTH 0.63 ms TIME 30min
80 W
F.PEAK
30 w Peak 80 w 28:58 min. AV
●The screen is displayed when the unit is being tuned, or has lost tuning
ERROR
Tuning is not possible. Solution:Check the electrodes.
TUNING RETRY
CANCEL
Adjust power by INTENSITY dial
V. PEAK
CW
MAX
PROG.
PRESET FRQ 600 Hz WIDTH 0.63 ms TIME 30min
80 W
F.PEAK
30 w Peak 80 w 28:58 min. AV
CW
PROG.
ELECTRODE
MAX
TIME C.
16 SW-180
80 W
FIXED PEAK
MODE
REPEAT
This screen opens
when the unit has
when the correct
completed tuning
tuning cannot be maintained during treatment.
2. When VARI.PEAK is selected
TIME
●The screen is displayed
TUNING
Adjust power by INTENSITY dial
V. PEAK
●Tuning Error Screen
TIME C.
VARI. PEAK
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
Preparations and Start of Treatment Pressing the [OK] key on the above screen opens the following screen: ELECTRODE
MAX
80 W VARIABLE PEAK
OUTPUT MODE 90 Hz /2ms
FREQUENCY
180 Hz /2ms
450 Hz /0.4ms 900 Hz /0.4ms
Turn INTENSITY clockwise to begin therapy.
F. PEAK
CW
PROG.
●Names and functions of the keys F.PEAK … To Fixed-Peak Pulsed Mode CW
… To Continuous Mode
PROG. … To selection of existing program MAIN W. … Opens the Initial Screen.
MAIN W.
Use the RIGHT/LEFT keys to select the desired combination on the right of FREQ/WIDTH, then turn the Output Dial. Treatment then begins. ●The initial setting is [90Hz/2ms]. If this is the desired frequency, you can start treatment right away by simply turning the Output Dial. 3. When CW is selected ELECTRODE
MAX
VARI. PEAK
FIXED PEAK
MODE
TIME
REPEAT
80 W
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
An 80W condenser-type electrode is connected.
Pressing the [OK] key in the above state opens the following screen: ELECTRODE
MAX
OUTPUT MODE
80 W
●Names and functions of the keys F.PEAK … To Fixed-Peak Pulsed Mode V.PEAK … To Variable-Peak Pulsed Mode
CW
PROG. … To selection of existing program Turn INTENSITY clockwise to begin therapy.
F. PEAK
V. PEAK
PROG.
MAIN W. … Opens the Initial Screen.
MAIN W.
Turn the Output Dial clockwise. Treatment then begins, followed by tuning. MAX
80 W
CW
AV
30 w
PRESET FRQ Hz WIDTH ms TIME 30min
28:58 min. Adjust power by INTENSITY dial
F. PEAK
V. PEAK
PROG.
This display appears after tuning has been completed.
TIME C.
SW-180 17
4. When PROG.SELECT is selected
ELECTRODE
MAX
80 W VARI. PEAK
FIXED PEAK
MODE
TIME
REPEAT
CW
PROG. SELECT
PROG. MAKING
5 min. 20 min.
10 min. 25 min.
15 min. 30 min.
TIME C.
OK
MODE C.
Pressing the [OK] key on the above screen opens the following screen: PROG.SELECTION NO. 01 02 03 04 05 06 07
POSITION NECK TMJ
Use the UP/DOWN keys to select the desired Program
Page1
Choice
EL.
MODE 90Hz CW
30 20
450Hz CW 25Hz 600Hz
15 30 20 15
L. ABDO SHOULDER PERIPH JOINTS
Turn INTENSITY clockwise to begin therapy.
NEXT P.
DEL. W.
MODIF.W.
MAIN W.
No., then turn the Output Dial. Treatment then begins. ●Pressing a function key here opens the screen marked on that key. NEXT P. … To the next page DEL.W. … Opens the Program Deletion Screen. MODIF.W. … Opens the Program Modification Screen. MAIN W. … Opens the Initial Screen. ●Program Deletion Screen
PROG.DELETION
Page1
Choice
NO. 01 02 03 04 05 06 07
EL.
MODE 90Hz CW
30 20
450Hz CW 25Hz 600Hz
15 30 20 15
POSITION NECK TMJ L. ABDO SHOULDER PERIPH JOINTS
NEXT P.
DELETE
POSITION NECK TMJ L. ABDO SHOULDER PERIPH JOINTS
Prog. Choice
NEXT P.
REG.
you want to delete, then press the [DELETE] function key. The program of the No. selected is then deleted.
CANCEL
●Program Modification Screen
PROG.MODIFICATION Page1 NO. 01 02 03 04 05 06 07
Use the UP/DOWN keys to select the Program No.
EL.
MODE 90Hz CW
30 20
450Hz CW 25Hz 600Hz
15 30 20 15
Item Choice
MODIFY
PREV.W.
Use the UP/DOWN keys to select the Program No. you want to modify, then select the desired item. Scroll to select the data by moving with the RIGHT/LEFT keys, and make modifications for POSITION, EL, MODE and
with the [MODIFY]
function key, then press the REG. function key. The modification is then made.
18 SW-180