JOUAN
JOUAN C, CT, CR4-22 series User Manual Rev h Feb 1999
User Manual
69 Pages
Preview
Page 1
JOUAN CENTRIFUGEURS
C / CR / CT4.22 - G/GR / GT4.22 CENTRIFUGES
MANUEL D'UTILISATION USER'S MANUAL 89000187-h
LIRE CE MANUEL ATTENTIVEMENT AVANT D’INTERVENIR SUR VOTRE APPAREIL. LES INFORMATIONS CONTENUES DANS CE DOCUMENT SONT LA PROPRIETE DE JOUAN ; IL NE PEUT ETRE REPRODUIT OU COMMUNIQUE SANS SON AUTORISATION. L’APPLICATION DE LA GARANTIE EST SUBORDONNEE A L’OBSERVATION DES REGLES ET PRECAUTIONS D’UTILISATION DECRITES DANS CE DOCUMENT
CAREFULLY READ THIS MANUAL BEFORE OPERATING YOUR INSTRUMENT. INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF JOUAN ; IT MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT THE OWNER’S AUTHORIZATION. THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF THE INSTRUCTIONS AND PRECAUTIONS DESCRIBED IN THIS DOCUMENT.
C/G422_U
MISE A JOUR REVISION STATUS
INDICE INDEX
DATE DATE
PAGES MODIFIEES AMENDED PAGES
a
04/93
Remplace et complète les manuels 25990011 et 89000006 In replacement of former manuals 25990011 and 89000006
b
09/93
Liste de colisage Packing list
c
11/93
Liste de colisage Packing list
d
07/94
e
12/94
f
04/97
2-1, 2-2, chap. 3
Plage température, alimentation et sécurité électrique, energie, accessoires Temperature range, mains supply and electrical safety, energy, accessories
g
09/97
2-2, 4-1, 5-1, 5-2, 6-2
Bouton d'interverrouillage, risques biologiques Lid interlock button, biological risks
h
02/99
2-2, 2-3, 6-2
Ouverture manuel, risques biologiques Manual lid opening, biological risks
3-1, 6-1, 7-2, 7-3
NOTES NOTES
Remplacement des fusibles Fuse replacement Edition révisée Revised edition
C/G422_U
SOMMAIRE SUMMARY
- MANUEL FRANÇAIS ... FRENCH USER'S MANUAL ...
PARTIE PART
1 1
- MANUEL ANGLAIS ... ENGLISH USER'S MANUAL...
PARTIE PART
2 2
C/G422_U
Vitesse
3 - 7
Fin rotation
Début de freinage Commutation de KR
1200 1000 800 600 400 200
9 8 7 6 5
4
3
2
1
0
0 2 secondes
t
FIGURE 3.1 - PENTES DE FREINAGE
C/G422_U
GUARANTEE TERMS
JOUAN guarantees that this unit is free from defects in materials and workmanship when it leaves the factory, and undertakes to replace or repair the unit if it proves defective in normal use or during servicing for a period of ONE YEAR from the delivery. Our liability under this guarantee is limited to repairing the defective unit or any part of the unit providing it is sent, carriage paid, to an authorised service centre or the SAINT-HERBLAIN office. This guarantee is invalidated if the unit is incorrectly used, poorly serviced or neglected, mis-used or accidentally damaged. There is no explicit guarantee other than as stated above.
FOR FURTHER INFORMATION, ASSISTANCE OR SERVICING Please contact :
USA :
WINCHESTER, VIRGINIA
Tel. (540) 869 8623 - Fax. (540) 869 8626
ENGLAND :
ILKESTON, DERBYS
Tel. (0115) 944 7989 - Fax. (0115) 944 7080
GERMANY :
UNTERHACHING
Tel. (089) 611 4038 - Fax. (089) 611 3087
OTHER COUNTRIES :
Your local representative
or our CUSTOMER SERVICE CENTRE :
SOCIETE JOUAN Service Assistance Technique 10, rue Duguay Trouin 44800 SAINT HERBLAIN Tél. +33 (0) 2 28 03 20 20 Fax. +33 (0) 2 28 03 20 03
C/G422_U
PACKING LIST
QUANTITY 1 1 1 1 1 1 1 1 1 1 1 1 1 5 2 2
ITEM NAME Centrifuge User's manual Unpacking instructions Imbalance protection instructions Rotor removal tool Unlocking tool Mains cord 12 and 13 mm wrench 17 and 19 mm wrench (G series) 21 mm wrench 3 mm Allen key (C series) 10 mm box wrench (G series) Screwdriver (G series) Fuses (6 for CT/GT) Carbon brushes Immobilizing blocks (G series)
C/G422_U
CONTENTS
CHAPTER 1 - USE AND FUNCTION...
1-1
1.1 - Description... 1.2 - Relative centrifugal force...
1-1 1-2
CHAPTER 2 - INSTALLATION PROCEDURE...
2-1
2.1 - Lifting and transport... 2.2 - Unpacking... 2.3 - Environmental conditions... 2.4 - Location... 2.5 - Mains supply... 2.6 - Lid opening... 2.7 - Accessory installation... 2.8 - Balancing the load...
2-1 2-1 2-1 2-1 2-2 2-2 2-3 2-3
CHAPTER 3 - SPECIFICATIONS...
3-1
3.1 - Dimensions and weight... 3.2 - Electrical specifications... 3.3 - Performance and accessories...
3-1 3-1 3-2
CHAPTER 4 - OPERATING PRINCIPLES...
4-1
4.1 - Drive system... 4.2 - Safety interlock system... 4.3 - Imbalance detection system... 4.4 - Refrigeration system... 4.5 - CT/GT4.22 heating system... 4.6 - Integration...
4-1 4-1 4-1 4-2 4-2 4-2
CHAPTER 5 - INSTRUCTIONS FOR USE...
5-1
5.1 - Controls and indicators... 5.2 - Use...
5-1 5-2
CHAPTER 6 - HAZARDS, PRECAUTIONS AND LIMITATIONS OF USE...
6-1
6.1 - Speed control... 6.2 - Cautions... 6.3 - IEC 1010-2-020 standard... 6.4 - Operational limitations... 6.5 - Electrical... 6.6 - Improper use... 6.7 - Biological risks...
6-1 6-1 6-1 6-1 6-2 6-2 6-2
CHAPTER 7 - SERVICING AND PREVENTATIVE MAINTENANCE...
7-1
7.1 - Cleaning... 7.2 - Periodic preventative maintenance... 7.3 - Motor brush replacement... 7.4 - Rotor removal... 7.5 - Fuse replacement... 7.6 - Trunnion lubrication... 7.7 - Spare parts list...
7-1 7-1 7-1 7-2 7-2 7-3 7-3 C/G422_U
1 - 1
CHAPTER 1 USE AND FUNCTION
1.1 - DESCRIPTION The C/G4.22, CR/GR4.22 and CT/GT4.22 centrifuges are used for the separation of substances, of varied density, by centrifugal force. The C/G4.22 are ventilated units. The CR/GR4.22 are refrigerated units. The CT/GT4.22 are thermostated units. In the C/G4.22, there is a permanent air circulation which reduces the heat level in the rotor chamber. However the temperature of the accessories can reach 10 to 12°C above the ambient temperature. The CR/GR4.22 refrigerated units are equipped with a cooling unit which can maintain the rotor chamber temperature at or below the ambient temperature. The CT/GT4.22 thermostated models are equipped with a cooling unit and heating resistor which enable the control of the rotor chamber temperature below, above and at the ambient temperature.
FIGURE 1.1 - GR4.22 AND CR4.22 C/G422_U
1 - 2
1.2 - RELATIVE CENTRIFUGAL FORCE The C/G4.22, CR/GR4.22 and CT/GT4.22 are automated centrifuges designed for general purpose clinical and research laboratory uses that require the application of relative centrifugal force (R.C.F.). Suspended solids are separated by the Relative Centrifugal Force (RCF) applied to the sample. The effective force increases with the square of the speed of rotation and the distance from the axis of rotation. The following formulae permit the calculation of primary parameters and of transformations relating to changes in primary parameters. In fact, the control system of the instrument carries out and displays the results of all calculations between speed, radius and RCF. Note : The value introduced for the radius can be adjusted to allow for position within the tube such as at a boundary. Maximum radii are quoted in the specifications tables (chapter 3).
CENTRIFUGATION FORMULAE Legend :
R = radius (in millimetres) N = speed (in r.p.m.) ÷ 1000 RCF = gravitational acceleration ‘g’ M+ = add to memory MR = memory recall
Note : To calculate actual results, press the keys on a pocket calculator in the order shown. Primary calculations RCF (x g)
= 1.118 R N 2
Speed (r.p.m.) = 946 Radius (mm)
Key sequence
=
RCF R
RCF 1.118 R N2
Transformations
N x = x 1.118 x R = RCF ÷ R = √ x 946 = N x = x 1.118 = M+ RCF ÷ MR =
Key sequence
To determine actual ‘g’ achieved at a different speed :
( )
RCF2 = RCF1 N2 N1
2
N2 ÷ N1 = x = x RCF1 =
To determine actual speed required to achieve a different ‘g’ at the same radius : N2 = N1
RCF2 RCF1
RCF2 ÷ RCF1 = √ x N1 =
To determine actual speed required to achieve the same ‘g’ at a different radius : N2 = N1
C/G422_U
R1 R2
R1 ÷ R2 = √ x N1 =
2 - 1
CHAPTER 2 INSTALLATION PROCEDURE
2.1 - LIFTING AND TRANSPORT Due to the weight of the machine, all lifting and transporting must be carried out using proper handling equipment (eg : fork lift trolley) that complies with current regulations, and by people having undergone the necessary training. The machine must be supported from underneath. If it has to be transported without its pallet, for example on a staircase, professional handling assistance is required.
2.2 - UNPACKING Carefully remove the unit, the accessories and the documents from the packing. A minimum of two people are required to handle the machine. Check all items are present using the delivery note. Before connecting the machine to the mains, remove the foam pad locking the imbalance detector according to the specific instruction sheet included in the unit.
2.3 - ENVIRONMENTAL CONDITIONS This instrument is designed to operate safely under the following conditions : - Indoor use. - Temperature : 5°C to 40°C. - Maximum relative humidity of 80% for temperatures up to 22°C. - Maximum altitude : 2000 m. Maximum performance is assured across the following temperature range : 15°C to 25°C.
2.4 - LOCATION Place the centrifuge in a clean, non-corrosive environment to prevent damage to the centrifuge and its accessories. Install the centrifuge to allow at least 15 cm clearance between the back of the centrifuge and any obstruction. This clearance is necessary to get the highest cooling performance. The versions of the machine on wheels are immobilized by using the blocks provided. Slip the black plastic parts onto the rods of the immobilizing blocks and slide them underneath the 2 front feet of the machine. Attach them with the screws and washers provided. Tighten the screws.
C/G422_U
2 - 2
Lift the machine up slightly by turning the nut on the block. The wheels should be about 3 to 4 mm from the floor (see diagram).
3 to 4 mm
2.5 - MAINS SUPPLY C/G4.22
230 VAC ±10% - 50 Hz 120 VAC ±10% - 60 Hz
6 A single phase + ground 12 A single phase + ground
CR/GR4.22
230 VAC ±10% - 50 Hz 120 VAC ±10% - 60 Hz 208/230 VAC ±10% - 60 Hz
10 A single phase + ground 16 A single phase + ground 10 A single phase + ground
GT/GT4.22
230 VAC ±10% - 50 Hz 120 VAC ±10% - 60 Hz 208/230 VAC ±10% - 60 Hz
10 A single phase + ground 16 A single phase + ground 10 A single phase + ground
Remember that in order to respect the electrical safety standards related to protection of operators against indirect contact, the supply of power to the instrument must be via a power socket fitted with a protection device ensuring automatic cut-off in the case of an insulation fault. A supply fitted with a cicuit breaker of the correct rating complies with this requirement.
2.6 - LID OPENING Ensure the centrifuge has been switched ON. Press the button on the right hand side of the centrifuge and hold it in while pressing the "COVER" key : the lid is automatically unlocked and opens. At any other time, wait until display returns to the stand-by mode before pressing this key. In the case of a mains power cut, opening of the lid is prevented by the lid locking safety device. It is recommended to wait for the mains to be switched back on so that this safety device enables the lid to be unlocked. During rotation, the slightest leak from a sample is enough to create an aerosol. Rotation of the rotor can only be detected when the machine is switched on. Opening the lid manually if the rotor is still rotating, would disperse the aerosol in the environment.
C/G422_U
2 - 3
Rotating parts are also a risk as these could come in contact with the user or be ejected. There is particularly high risk of injury if : - The user tries to manually stop the rotor, - Any object falls inside the centrifugation chamber.
MANUAL LID OPENING PROCEDURE Manual lid unlocking must only be done by someone informed of the danger and of the necessary precautions to be undertaken. If the machine has stoppped because of a mains power cut, the rotor could still be rotating. Wait 10 minutes before opening the lid. In spite of the absence of noise, the rotor could still be rotating when you need to open the lid manually. Always set the mains switch to the OFF position, even in the case of a mains power cut. - Insert the unlocking tool into the hole at the top on the left hand side of the instrument by piercing the protection sticker. - Insert the tool and push to unlock the lid. The protection sticker must be replaced after it has been pierced.
2.7 - ACCESSORY INSTALLATION - Remove the shaft nut (by turning the nut clockwise). - Check the cleanliness of the shaft and the rotor. - Place the rotor on the shaft. - Screw down the shaft nut and secure it by using the key provided in the tool kit. - In the case of a swing-out rotor, grease the trunnions. - Place the buckets on the trunnions after checking the weight of the buckets is identical (the weight is engraved on the bottom of each bucket).
2.8 - BALANCING THE LOAD When using fixed angle rotor, weight deviation between the tubes must not exceed ±5 grams. When using a swing-out rotor, the samples are placed in plastic inserts. The load must be statically and dynamically balanced : Static balancing consists of balancing the weight of the 2 inserts which are in opposite locations in the rotor (±15 to 20 grams). Dynamic balancing consists of placing the tubes in each insert symetrically according to the shaft axis of the rotor.
C/G422_U
2 - 4
The following pictures give examples of correct and incorrect loading :
CORRECT
CAUTION :
INCORRECT
Never use old accessories on a new centrifuge. Only use plastic ware or glassware designed for centrifugation.
3 - 1
CHAPTER 3 SPECIFICATIONS
3.1 - DIMENSIONS AND WEIGHT C4.22
G4.22
CR/CT4.22
GR/GT4.22
* Height (lid open)
mm/in
900/351/ 2
1380/54
900/35 1/ 2
1380/54
* Height (lid closed)
mm/in
430/17
900/35 1/ 2
420/16 1/ 2
900/35 1/ 2
* Width
mm/in
555/22
555/22
820/32
555/22
* Depth
mm/in
595/231/ 2
625/24 1/ 2
595/23 1/ 2
625/24 1/ 2
* Depth with line cord socket
mm/in
630/25
660/26
630/25
660/26
* Weight unpacked
kg/lb
67/148
100/220
95/210
130/287
* Weight packed
kg/lb
77/170
120/265
107/236
150/331
C4.22, CR4.22 and CT4.22 centrifuges are instruments to be installed on a bench. G4.22, GR4.22 and GT4.22 centrifuges are on wheels, designed to be placed directly on the floor.
3.2 - ELECTRICAL SPECIFICATIONS Voltage :
C4.22
G4.22
CR4.22
GR4.22
CT4.22
GT4.22
220/240 V ±10% - 50 Hz
11175318
11176915
11175330
11176917
11175332
11176919
120 V ±10% - 60 Hz
11175319
11176914
11175331
11176916
11175333
11176918
/
/
/
11176929
/
11176930
230 V - 60 Hz
Power consumption :
C/G4.22
CR/GR4.22
CT/GT4.22
220/240 V
5.5 A
7.5 A
9.5 A
110 V
11 A
14 A
15 A
Motor :
500 W 4500 r.p.m.
Refrigeration (CR/GR4.22 and CT/GT4.22) :
3 / 8 HP
C/G422_U
3 - 2
3.3 - PERFORMANCE AND ACCESSORIES The accessories used with the C/G4.22, CR/GR4.22 and CT/GT4.22 are : - 2 swing-out rotors (M4 Cat. N°. : 11175338 and T4 Cat. N°. : 11175366) which allow the centrifugation of many types of tubes in plastic inserts fitting standard or sealed buckets. - 6 fixed angle rotors for centrifugation at higher speed.
ANGULAR ROTORS - PERFORMANCE Rotor
Catalogue number
Capacity
Diameter (mm)
Depth (mm)
Min./max. tube ht (mm)
CD1
11175181
8 x 100 ml
44.5
101
106/115
134
7000
7340
7200
7760
CD2
11175183
10 x 50 ml
35
92
97/115
130
7300
7740
7500
8170
CD3
11175186
40 x 15 ml
18.5
93
96/110
132 145
6800
6820 7500
7000
7230 7940
CD4
11175180
40 x 15 ml
16.5
93
96/131
131 144
6800
6770 7440
7000
7180 7890
CD5 Sealed
11175187
40 x 15 ml
16.5
110 107
115/130 112/127
136 151
6500
6420 7130
6700
6820 7580
Eppendorf 11174217
30 x 1.5/2 ml
11
open
40
115
7000
6300
7200
6660
C/G422_U
Holes
Max. radius (mm)
C/G4.22 Max. r.p.m.
CR/GR/CT/GT4.22
Max. RCF (g)
Max. r.p.m.
Max. RCF (g)
3 - 3
M4 SWING-OUT ROTOR - REF : 11175338 Bucket or carrier (set of 4)
Rack or insert (set of 4)
Rack or insert colour
Rack or insert capacity
Max. tube Ø (mm)
Min/max. tube height (mm)
Tube volume (ml)
C/G4.22 CR/GR4.22 CT/GT4.22 Max. radius Max. Max. Max. Max. Max. Max. (mm) rpm RCF(g) rpm RCF(g) rpm RCF(g)
11174204 Black/light blue
30
8.5
40
0.8
165
4000
2950
4500
3740
4300
3410
11174201 Black/light blue
30
11.5
40
1.5-2
160
4000
2860
4500
3600
4300
3310
Standard
11174206
Black/coral
12
16.5
40
4 flat bottom
184
4000
3290
4500
4160
4300
3800
buckets
11174205
Black/beige
9
18
40
4 flat bottom
184
4000
3290
4500
4160
4300
3800
11174154
11174157
Black/grey
37
12.5
100
5
172
4000
3070
4500
3900
4300
3560
PTFE coated
11174193
Black/grey
For decanting : Ø 11.7 to 12.5
172
4000
3070
4500
3900
4300
3560
62813600
11174158 Black/light blue
30
13.5
100
6
172
4000
3070
4500
3900
4300
3560
11174159
Black/yellow
20
13.5
107
7 vac.
172
4000
3070
4500
3900
4300
3560
Sealed
11174160
Black/red
19
16.5
107
10 vac.
172
4000
3070
4500
3900
4300
3560
buckets
11174219
White
16
17
60/120
5/10 flat
180
4000
3220
4500
4075
4300
3720
11174218
11174208
Black/orange
12
16.5
125
10 flat bottom
184
4000
3290
4500
4160
4300
3800
PTFE coated
11174188
Beige
9
18
120
10 flat bottom
172
4000
3070
4500
3900
4300
3560
62814801
11174161
Brown
7
17.7
134
15 vac.
185
4000
3310
4500
4190
4300
3820
11174162
Beige
9
18
122
15 conical
184
4000
3290
4500
4160
4300
3800
Spare lid
11174190
Orange
12
16.5
125
15
184
4000
3290
4500
4160
4300
3800
25144058
11174163
Coral
12
22.2
108
20
177
4000
3170
4500
4010
4300
3660
11174186
Dark red
7
30
115
25 flat bottom
172
4000
3070
4500
3900
4300
3560
Spare seal
11174164
Dark red
7
30
115
50
182
4000
3260
4500
4120
4300
3760
85241381
11174529
Black
7*
30
116
50 conical
185
4000
3310
4500
4190
4300
3820
10 spare
11174191
Light blue
4
30
120
50 conical
185
4000
3310
4500
4190
4300
3820
seals
11174199
Light blue
4
30
120
50 flat bottom
172
4000
3070
4500
3900
4300
3560
89000814
11174165
Dark blue
4
35
120
50
185
4000
3310
4500
4190
4300
3820
11174166
Dark green
2
45
118
100
185
4000
3310
4500
4190
4300
3820
Bottles (6)
11174167
Turquoise
1
63
134
250 flat bottom 185
4000
3310
4500
4190
4300
3820
750 ml
11174192
Navy blue
1
80.5
132
185
4000
3310
4500
4190
4300
3820
41193190
11174226 (set of 2)
Black
180
1570
500
1570
500
1570
500
11174203
600 ml stainless steel flat bottom bottle
190
4000
3400
4500
4300
4300
3930
11174169
Adaptor for blood bag (pediatric)
190
4000
3400
4500
4300
4300
3930
450
1 cyto bucket (11172499 : set of 2 cyto buckets)
Microtitration 11174168
3 plates
96 wells
158
3600
2290
4000
2830
3800
2550
Sealed µT 11174223
3 plates
96 wells
174
3600
2520
3850
2880
3750
2730
500
208
3500
2850
3900
3540
3800
3350
11174207
Rubber pad for flexible microtitration plate
11175371
250 ml conical bottle liner
11175370
1
11175372
Buckets for 250 ml conical bottle
213
2000
1000
2000
1000
2000
1000
11175378
Sealed buckets for 250 ml conical bottle
213
2000
1000
2000
1000
2000
1000
11175385
Sealed buckets + liner for plasma/red cell separation
190
4000
3400
4500
4300
4300
3930
* In open buckets only
C/G422_U
3 - 4
T4 SWING-OUT ROTOR - REF : 11175366 Bucket or carrier (set of 4)
Rack or insert (set of 4)
Rack or insert colour
Rack or insert capacity
Max. tube Ø (mm)
Min/max. tube height (mm)
Tube volume (ml)
C/G4.22 CR/GR4.22 CT/GT4.22 Max. radius Max. Max. Max. Max. Max. Max. (mm) rpm RCF(g) rpm RCF(g) rpm RCF(g)
11175340
Black
42
8.5
35/-
0.5-0.8
174
3500
2380
3700
2660
3600
2520
11175341
Salmon
20
11
35/-
1.5-2.2
174
3500
2380
3700
2660
3600
2520
Standard
11175342
Beige
30
11
65/107
3/5
174
3500
2380
3700
2660
3600
2520
buckets
11175343
Grey
26
12.7
67/107
5
174
3500
2380
3700
2660
3600
2520
11175367
11175357
White
7
18
35/-
5 flat bottom
174
3500
2380
3700
2660
3600
2520
11175344
Light blue
20
13.5
65/107
6
174
3500
2380
3700
2660
3600
2520
11175345
Yellow
18
13.5
65/108
7 vac.
174
3500
2380
3700
2660
3600
2520
11175346
Red
12
16.5
65/112
10 vac.
174
3500
2380
3700
2660
3600
2520
Transparent
11175347
Cream
12
17.5
65/112 10 flat bottom 174
3500
2380
3700
2660
3600
2520
sealed lid
11175358
White
7
18
90/122 12 flat bottom 174
3500
2380
3700
2660
3600
2520
11175368
11175348
White
7
18
90/122
15 conical
176
3500
2410
3700
2690
3600
2550
11175349
Orange
7
16.5
90/127
15 vac.
174
3500
2380
3700
2660
3600
2520
11175350
Brown
4
16.5
90/133
15 vac.
176
3500
2410
3700
2690
3600
2550
11175351
Coral
8
22.2
90/116
20
176
3500
2410
3700
2690
3600
2550
Spare seal
11175352
Light yellow
5
25
90/120
25
176
3500
2410
3700
2690
3600
2550
25439074
11175375
Light yellow
5
25
90/120 25 flat bottom 176
3500
2410
3700
2690
3600
2550
11175374
Dark red
4
30
90/122 25/50 flat bottom 176
3500
2410
3700
2690
3600
2550
11175353
Dark red
4
30.5
90/122
50
176
3500
2410
3700
2690
3600
2550
11175354
Light blue
3
30
90/117
50 conical
176
3500
2410
3700
2690
3600
2550
11175376
Light blue
3
30
90/117 50 flat bottom 176
3500
2410
3700
2690
3600
2550
11175361
Dark blue
2
35
90/122
50
176
3500
2410
3700
2690
3600
2550
11175355
Green
1
45
90/132
100
176
3500
2410
3700
2690
3600
2550
11175356
Turquoise
1
63
65/125 250 flat bottom 172
3500
2350
3700
2630
3600
2490
1570
500
1570
500
1570
500
2045
3400
2045
11175359 (set of 2)
Black
11175360
5 sets of 4 decanting rubber pads (Ø 10.5 - 11 - 11.5 - 11.9 - 12.1 mm) for 11175343
Microtitration 11175377
1 cyto-bucket (11172499 : set of 2 buckets)
3 plates
181
96 wells
158
3400
2045
3400
JOUAN GLASSWARE Capacity
Material
Bottom
Type
Standard tube cat. number
Graduated tube cat. number
Insert cat. number
12 ml 12 ml 16 ml 24 ml
GLASS GLASS GLASS PP
conical round round round
O C O C
41191314 41191212 41191110 41193113
41191315
11174163 11174163 11174163 11174163
20 ml 27 ml 40 ml 47 ml 58 ml
GLASS GLASS GLASS GLASS PP
conical conical round round round
C O C O C
41191238 41191336 41191232 41191130 41193134
66 ml 95 ml 104 ml 130 ml
GLASS GLASS GLASS PP
conical round round round
C C O C
375 ml 485 ml
GLASS PP
round round
O C
PP : Polypropylene
C/G422_U
O : Open
C : Capped
41192111
Rotor model
11174165 11174151 11174165 11174165 11174165
CD2 CD2 CD2 CD2
41191466 41191262 41191160 41193164
11174166 11174166 11174166 11174166
CD1 CD1 CD1 CD1
41191186 41193189
11174192 11174192
41192337 41192131
3 - 5
SPEED CONTROL Range :
C/CR4.22 and G/GR4.22 : CT/GT4.22 :
Increments : Accuracy :
200 - 8000 r.p.m. 500 - 8000 r.p.m. 10 r.p.m. ± 10 r.p.m.
SPEED READOUT Display : Increments : Accuracy :
0 - 9,999 r.p.m. 10 r.p.m. ± 20 r.p.m.
RCF CONTROL Range :
C/CR4.22 and G/GR4.22 : CT/GT4.22 :
Increments :
10 - 9500 g 50 - 9500 g 1 g step from 10 to 99 g 10 g step from 100 to 9500 g
The set value is cancelled if the speed calculation, according to the set radius value, exceeds 8000 r.p.m.
RCF READOUT Range : Increments :
0 - 9,999 g 1g
ENERGY Max. density : Max. load on drive shaft : Max. energy :
C4.22 and G4.22 : CR/CT4.22 and GR/GT4.22 :
1.2 g/ml 10 kg 29190 J 30810 J
The maximum rotational energy is not systematically that obtained with the maximum permissible load on the drive shaft.
TIMER Range : Increments : Accuracy :
1 s - 99 mn 59 s + hold position (0 min 00 sec) 1s ± 0.1 % of maximum time, or 6 sec
INTEGRATOR Range : Increments :
1 to 999 E9 m. rad. 2/s 1 Ex (0≤ x ≤9)
C/G422_U
3 - 6
TEMPERATURE CONTROL Range :
CR/GR4.22 : CT/GT4.22 :
-8°C +40°C -8°C +60°C
Note : If the set speed is below 1000 r.p.m. the maximum temperature will be automatically reduced to 50°C 1 °C ± 0.5°C at the probe location -9°C to +11°C
Increments : Accuracy : Temperature compensation
TEMPERATURE READOUT Display : Increments : Accuracy :
-10°C to + 65°C 1 °C ± 0.5°C
ACCELERATION RATE CONTROL - Range : 10 rates (0 to 9)
Value
0
1
2
3
4
5
6
7
8
9
Threshold of slope change
1000
1000
1000
1000
1000
1000
1000
1000
1000
1000
Duration 0 to 4500 r.p.m.
4 min 00 sec
3 min 40 sec
3 min 25 sec
3 min 05 sec
2 min 50 sec
2 min 30 sec
2 min 10 sec
1 min 55 sec
1 min 40 sec
1 min 20 sec
Time to reach threshold
2 min 55 sec
2 min 40 sec
2 min 20 sec
2 min 00 sec
1 min 45 sec
1 min 25 sec
1 min 10 sec
51 sec
38 sec
17 sec
Note : The values here above are average values. They may vary according to the load and the motor use.
BRAKING RATE CONTROL - Range : 10 rates (0 to 9)
Value
0
1
2
3
4
5
6
7
8
9
Threshold of slope change
1000
800
600
400
/
400
600
800
1000
1200
Duration 4500 to 0 r.p.m.
12 min 00 sec
11 min 30 sec
10 min 25 sec
8 min 50 sec
4 min 27 sec
3 min 12 sec
2 min 58 sec
2 min 48 sec
2 min 35 sec
2 min 26 sec
Time to reach threshold
1 min 20 sec
1 min 30 sec
1 min 47 sec
2 min 05 sec
/
2 min 05 sec
1 min 47 sec
1 min 30 sec
1 min 20 sec
1 min 10 sec
Note : The values here above are average values. They may vary according to the load and the motor use.
C/G422_U
Speed
3 - 7
End rotation
Start of braking Switching of KR
1200 1000 800 600 400 200
9 8 7 6 5
4
3
2
1
0
0 2 seconds
t
FIGURE 3.1 - BRAKING RATES
C/G422_U