Page 1
Specifications Capacity & Graduation
300kg/0.1kg (660lb/0.2lb)
Power Requirements
Adaptor 120VAC-9VDC-50Hz / 230VAC–9VDC–50Hz
Environmental
Operating temperatures: 50°F to 95°F / 10°C to 35°C Storage temperatures: 32°F to 122°F / 0°C to 50°C Humidity: 85% Use in atmospheric pressure
Product Dimensions
850 mm x 630 mm x 965mm
Chair Dimensions
430 mm x 420 mm x 76mm
Features: Movement compensation technology BMI (Body Mass Index) function Zero / Hold / Reweigh / Clear Kg Lb / Print / Enter /Tare Up and Down arrows USB and RS232 Port for data transmission 5-digit ¾” LCD negative/positive FSTN
Power: The scales have rechargeable or non-rechargeable batteries. 6 x 1.5 AA batteries (cannot be
switched to rechargeable) Rechargeable battery (cannot be switched to AA batteries)
9-12 VDC, 500mA Auto Power – off Low battery indicatio Options: 230V AC adaptor / 120V AC adaptor
Battery indicator
Battery status
Flag indicator
Maximum battery value Medium 2 battery value Medium 1 battery value Low battery Minimum
3 bars are on 2 bars are on 1 bar is on Battery frame Flashing battery frame Scrolling "LOW bAt" and auto shutoff Off
Critical No Battery
H550-551 19-11-14
www.liden-weighing.se
• • •
If external power supply or USB power is connected the battery flag is turned off. Battery flag update. On startup the update is every 5 seconds for 12 cycles (overall 1 minute), after that the battery flag is updated once every 10 seconds. When on battery or USB power supply the backlight power is reduced to 60 percent.
The display Annunciators
ZERO, LOCK, HOLD, BMI, GROSS, NET
Indicators Battery (see info above)
H550-551 19-11-14
USB,
WIFI (Optional)
www.liden-weighing.se
KEY DESCRIPTION:
End user operation mode: Key Name On/Off Print Kg/Lb
Zero
Hold / Release
BMI UP/DOWN
CLEAR
ENTER
TARE
H550-551 19-11-14
Function Turns the scale on or off A long press will trigger a transmission of data sent out by RS232 (see RS232 protocol). When data is sent out from the port, the display flashes the word PRINT to acknowledge the data was sent. A short press allows toggling between pounds and kilograms (providing it is enabled on configuration mode). No toggle in BMI mode. Zeroes the weight on the scale, Attempts to zero for three seconds Works only if the weight is stable Zeroes up to 2% of full weight Pressing on this key while HOLD indicator is off will keep the most current weight value shown on the display along with the HOLD indicator Pressing on this key while HOLD indicator is on will release the weight and turn off the HOLD indicator This key is not active in BMI mode Enter BMI mode, only works if there is a locked weight shown on the display and the function is turned on in the configuration settings Adjust the value of the flashing digitnumber
In BMI height display causes the value to return to default (170.0 cm, 5 ft.7.5 in) In BMI final result display pressing this key will return the scale to weighing mode Accepts the value of the parameter last entered and moves to the next stage. A long press during scales start-up process will enter id display (pre-parameters mode) For weight under the limit (ASTART value), will enter manual tare, If above the limit will enter tare function NET & GROSS indicators are on according to the tare operation www.liden-weighing.se
BMI procedure KG mode Ensure that the scale is at zero. Have the person sit down on the scale to obtain a weight. Pressing on BMI key will enter BMI mode, BMI and CM indicators are on and the display shows the height default value 170.0 cm. while all digits are flashing (the height resolution is 0.5 cm). UP/DOWN key are used to adjust the entire value. Pressing it once will adjust by 0.5 cm. If you hold the key down it will keep incrementingdecrementing the values. Pressing ENTER will move to BMI value. BMI value is shown on the display, Pressing CLEAR to return to weighing mode, BMI indicator is off. * The last height value entered is saved until the zero flag is displayed or negative weight is detected, at this point the value returns to the default value. * The height value in cm. is independent to the value in ft. and vice versa. HOLDRELEASE FUNCTION 1. When the person is on the scale, press the HOLDRELEASE key. 2. When the person gets off the scale the weight will remain the same. The zero flag will not turn on. 3. At this stage the ZERO key will not work. The only way to return to zero from here is to press the HOLDRELEASE key once more. TARE FUNCTION: 1. When the weight is below the ASTART parameter value press the TARE key. 2. The default tare value is displayed (factory programmed at 33.0 lb.15.0 kg), while the zero is flashing. 3. UP/DOWN key is used to adjust the value. 4. Press enter start the tare function. NET annunciator will be turned on instead of the GROSS annunciator.
H550-551 19-11-14
www.liden-weighing.se
MANUAL TARE 1. When the weight is above the ASTART parameter value press the TARE key. 2. The tare will be activated with the value that was displayed at the time of the tare activation. 3. NET annunciator will be turned on instead of the GROSS annunciator.
Print button On pressing the print button the following data will be sent. In weight mode: PATIENT WEIGHT 60.1 kg In BMI (On calculated BMI display only): In kg: PATIENT WEIGHT 60.1 kg PATIENT HEIGHT 170.0 cm PATIENT BMI 20.8 In case of underweight or overweight the word 'UNDER' or 'OVER' will be sent accordingly.
H550-551 19-11-14