Uroflow Transducer
Models SA0972AW & SA0972AC
OPERATOR’S MANUAL
M1078 Rev A
CAUTION: FEDERAL (USA) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A
PHYSICIAN.
Copyright © 2006, by Life-Tech®, Inc. All Rights reserved. No part of this manual may be reproduced in any manner,
in whole or in part without written permission of Life-Tech.
Page 3
Table of Contents
Introduction .......................................................................................................................................... 3
General Description ........................................................................................................................... 3
Formatting Conventions Used in this Manual.................................................................................... 3
Contact Information............................................................................................................................ 3
Precautions ........................................................................................................................................ 4
Illustrations in this Manual ................................................................................................................. 4
Safety................................................................................................................................................. 4
Environment....................................................................................................................................... 5
Environmental Protection................................................................................................................... 5
Service ............................................................................................................................................... 6
Meaning of Symbols .......................................................................................................................... 6
Specifications..................................................................................................................................... 7
Interface Cables................................................................................................................................. 7
Unpacking the Components .............................................................................................................. 8
Overview of Uroflow Transducer Components.................................................................................. 8
Installation and System Interface....................................................................................................... 9
Connecting the SA0972AC Uroflow Transducer to the M30/M30P Urodynamic Module ................. 9
Disconnecting the SA0972AC Uroflow Transducer from the M30/M30P Urodynamic Module....... 10
Connecting the SA0972AW Uroflow Transducer to a Patient Unit.................................................. 11
Disconnecting the SA0972AW Uroflow Transducer from a Patient Unit ......................................... 12
Using the Uroflow Transducer.......................................................................................................... 13
Preparation ...................................................................................................................................... 13
Operation ......................................................................................................................................... 13
Maintenance and Repair.................................................................................................................... 14
Product Return Information.............................................................................................................. 15
Uroflow Transducer Troubleshooting Guide.................................................................................... 16
Maintenance of the Uroflow Transducer.......................................................................................... 16
Cleaning & Disinfection..................................................................................................................... 17
Declaration of Conformity ................................................................................................................. 19
Index .................................................................................................................................................... 21
i
Page 4
Table of Figures
Figure 1 – Uroflow Transducer Components ............................................................................................ 8
Figure 2 – Connect the SA0972AC Uroflow Transducer(s) to the
M30/M30P Urodynamic Module .............................................................................................. 9
Figure 3 – Connecting a Circular Plug .................................................................................................... 10
Figure 4 – Disconnecting a Circular Plug ................................................................................................ 10
Figure 5 – Connect the SA0972AW Uroflow Transducer(s) to a Patient Unit......................................... 11
Figure 6 – Preparing the Uroflow Transducer for Use ............................................................................ 13
Figure 7 – Cleaning & Disinfection of the Uroflow Transducer ............................................................... 18
ii
Page 5
Introduction
General
Description
The Model SA0972AW and SA0972AC Uroflow Transducers measure the weight
increase as fluid enters a graduated beaker placed on top of the transducer.
The Uroflow Transducer is calibrated to produce an output that is proportional to the
change in volume. The unit does not incorporate a display of the signal output.
Rather, the signal output is displayed on a digital channel of the recording device to
which the transducer is connected.
Please read this manual and follow its instructions carefully. Throughout the manual,
Formatting
Conventions you will see paragraphs with the following headings:
Used in this
WARNING
Manual
CAUTION
NOTE
Carefully review the paragraph associated with any of these headings to ensure the
safe and effective operation of this product.
WARNING: A WARNING indicates that the personal safety of the patient or
equipment operator may be involved. Disregarding a WARNING could result in
injury to the patient or operator of the equipment.
CAUTION: A CAUTION indicates that particular service procedures or
precautions must be followed to avoid possible damage to the product.
Contact
Information
NOTE: A NOTE indicates special information to improve the ease of maintaining
or using the product, or to clarify important information.
Mailing Address: ..........................Life-Tech, Inc.
.....................................................PO Box 1849
.....................................................Stafford, Texas 77497-1849
.....................................................United States of America
Telephone....................................(281) 491-6600 or (800) 231-9841
FAX: Administrative / Financial...(281) 491-6646
Sales / Customer Service ............(281) 491-7197
Technical and Clinical Support ....(281) 491-6852
International Sales.......................(281) 491-6852
Internet.........................................http://www.life-tech.com
The Life-Tech representative for the European Community is:
FBI Fred Berninger Importe
Bergstrasse 12
82024 Taufkirchen, GERMANY
Tel: +49 89 61 453 453
Fax: +49 89 61 453 455
Email: info@fbi-medizintecnik.de
3
Page 6
Introduction, Continued
Precautions
WARNING: Read this instruction manual thoroughly and be familiar with its
contents prior to using this equipment. Read the entire instruction manual
before assembling or connecting the equipment.
Take special care to follow the warnings shown on the Uroflow Transducer
itself, as well as any safety suggestions in this manual.
CAUTION: Previous generations of Life-Tech Uroflow Transducers required
removal of the transducer (top) platter for cleaning and storage. This new
model of transducer does not require this practice. DO NOT remove the
transducer platter by pulling it off, as you will damage the transducer and void
its warranty.
Illustrations in
this Manual
Some of the illustrations contained in this manual may not represent the exact
construction details of the user’s specific equipment model. The illustrations in this
manual are provided as general information to aid in showing possible features
and components.
Safety
Please read these safety instructions carefully before using this instrument.
Before using the instrument it is imperative that you be acquainted with how it
operates and is controlled.
1.
WARNING: Do NOT touch the pins of the connectors marked with the
symbol without following the Electrostatic Discharge (ESD)
precautionary procedures detailed on page 9 of this manual.
Staff that could touch the pins of the connectors identified with the ESD
warning symbol should receive precautionary explanation and training. This
includes clinical / biomedical engineering and healthcare staff.
Electrostatic Discharge training should include an introduction to the physics
of electrostatic charge, the voltage levels that can occur in normal practice
and the damage that can be done to electronic components if they are
touched by an operator who is electrostatically charged. Further, an
explanation should be given of methods to prevent build-up of electrostatic
charge, and how and why to discharge one’s body to an earth ground or to
the frame of the equipment, or bond oneself by means of a wrist strap to the
equipment or to an earth ground prior to making a connection.
2.
WARNING: There is a risk of an explosion if the equipment is used in the
presence of flammable anesthetics.
3.
WARNING: To reduce the risk of electrical shock, do NOT open the
instrument’s inner housing. Refer servicing to qualified personnel only.
4.
Disassembly of equipment by unauthorized personnel will void the
instrument’s warranty.
5.
Power Source: The Model SA0972AC Uroflow Transducer is powered via its
connection to the M30/M30P Urodynamic Module.
The Model SA0972AW Uroflow Transducer is powered via its connection to
the following Patient Units: M22J; M22; M20, and M42; and Microflow 2001
and Spectrum 2200A.
Continued on next page...
4
Page 7
Introduction, Continued
Safety (Cont.)
Environment
6.
WARNING: Take care to keep the transducer where it cannot be stepped
on or kicked, which will damage the unit and present a tripping hazard for
persons in the area.
7.
The Uroflow Transducer is placed on the floor directly beneath the uroflow
chair or table as shown in Figure 6 – Preparing the Uroflow Transducer for
Use on page 13.
1.
Water and Moisture: Do NOT immerse the instrument in any fluid. This
instrument has an IP 21 rating.
2.
Magnetism: Position the instrument away from equipment or devices that
generate strong magnetic fields.
3.
Storage / Shipping Conditions:
Humidity: 10 to 90% (non-condensing)
Temperature: -4° to 140°F (-20° to 60°C)
Atmospheric Pressure: 50kPa to 106kPa
4.
Operating Conditions:
Humidity: 30 to 75% (non-condensing)
Temperature: 50° to 104°F (10° to 40°C)
CAUTION: Allow adequate air circulation to prevent internal heat build-up.
Do NOT install the unit in a location near heat sources such as radiators or air
ducts, or in a place subject to direct sunlight, excessive dust or humidity, or
mechanical vibration or shock. To avoid moisture condensation, do not install
the unit in a location where the temperature may rise rapidly.
5.
Electromagnetic Compatibility: The Model SA0972AC Uroflow Transducer,
used in conjunction with a M30/M30P Urodynamic System, has been tested
and found to comply with the EMC limits for the Medical Device Directive
93/42 EEC (EN 55011 Class A and EN 60601-1-2:2001). Refer to the LifeTech M30/M30P Operator’s Manual for more information on EMC
compliance.
WARNING: It is prudent to separate all electrical equipment that is very close
in distance to the Uroflow Transducer. If it is essential to use the Uroflow
Transducer very close to other electrical equipment, it is prudent to determine,
by observation, if the performance of either product is affected by unintended
electromagnetic coupling.
Environmental
Protection
At the end of its useful life, you may return hardware components to Life-Tech for
disposal. If you choose not to return items, follow your local governing ordinances
and recycling plans regarding disposal or recycling of device components.
Continued on next page...
5
Page 8
Introduction, Continued
Service
1.
Damage Requiring Service: The instrument should be serviced by qualified
service personnel when:
•
•
•
•
2.
Any cable, cord, or plug has been damaged.
The instrument does not appear to operate normally or exhibits a marked
change in performance.
The instrument has been dropped, or the casing is damaged.
Fluid has been spilled on the instrument, or it has been immersed, and it
appears that fluid has entered the casing.
Service: The user should not attempt to service the instrument beyond that
described in this operator’s manual unless directed by Life-Tech Service
personnel. Refer all other servicing to qualified service personnel.
Circuit diagrams, itemized parts listings, descriptions, and other items of
available documentation can be ordered by suitably qualified user personnel
duly authorized by Life-Tech for their use in repairing those components of
the instrument that have been designated by Life-Tech as repairable. Supply
of such technical documentation relating to the equipment shall not be
construed as constituting Life-Tech’s authorization of user’s personnel,
regardless of their levels of technical training, to open or repair the
instrument. Explicitly exempted here are those maintenance and repair
operations described in this manual.
Meaning of
Symbols
REF
SN
Continued on next page...
6
Reference (model) number
Serial number
Page 9
Introduction, Continued
Specifications
Unit Size:
13.5 cm Diameter x 7.5 cm Height (5.3” x 2.9”)
Weight:
510 Gm (1.15 lbs)
Volume Accuracy
± 2% of full scale
Volume Range
0 – 1000 ml
Safety
When used in conjunction with the M30/M30P
Urodynamic Module, the Model SA0972AC Uroflow
Transducer complies with:
IEC/EN60601-1 (International/Europe)
C22.2 No.60601.1-M90 (Canada)
UL Std. No. 60601-1 (USA)
Product Classification:
Type BF Applied Parts
Class I equipment
Mode of operation: Continuous
This equipment is NOT suitable for use in the
presence of a flammable anesthetic mixture
with air or with oxygen or with nitrous oxide.
Ingress rating: IP21.
Interface
Cables
•
•
The SA0972AC Uroflow Transducer connects to the M30/M30P Urodynamic
Module via an interface cable terminating in a circular connector.
The SA0972AW Uroflow Transducer connects to the M22J, M22, M20 or
M42 Patient Unit via an interface cable terminating in a modular plug.
Continued on next page...
7
Page 10
Introduction, Continued
Unpacking the
Components
Unpack and identify the Uroflow Transducer before you are ready to begin using
the system. As you do so, examine it for obvious damage.
If you find any damage, observe anything unusual, or if any parts are missing
notify Life-Tech’s Technical Service Department at 800-231-9841 or 281-4916600 before attempting to use the instrument.
Overview of
Uroflow
Transducer
Components
The SA0972AC and SA0972AW Uroflow Transducers are a two-piece assembly
consisting of a transducer platter, screwed onto the transducer body. The
transducer body incorporates a hardwired interface cable, terminating in a circular
connector (SA0972AC) or a modular plug (SA0972AW).
The components of the Model SA0972AC are illustrated and described below:
Figure 1 – Uroflow Transducer Components
Component
Description
X Transducer Platter
Screws onto the top of the body and holds a graduated beaker.
o Body
Z Transducer Cable &
Houses the uroflow transducer circuitry.
•
Plug
•
The SA0972AC Uroflow Transducer connects into the input jack on the
bottom panel of the M30/M30P Urodynamic Module via an interface
cable terminating in a circular connector.
The SA0972AW Uroflow Transducer connects to the input jack on the
bottom panel of M22J, M22, M20 or M42 Patient Unit via an interface
cable terminating in a modular plug.
CAUTION: The transducer platter on Uroflow
Transducer Models SA0972AC or SA0972AW should
never need to be removed. If you must remove it,
unscrew it counterclockwise until it comes free of the
transducer body.
Pulling the transducer platter off instead of
unscrewing it will cause damage to both
components, which will void the unit’s warranty.
8
Page 11
Installation and System Interface
Connecting
the SA0972AC
Uroflow
Transducer to
the M30/M30P
Urodynamic
Module
The Model SA0972AC Uroflow Transducer is interfaced to the M30/M30P
Urodynamic Module.
WARNING: Do NOT touch the pins of the connectors marked with the
symbol without following ESD precautionary procedures, which include:
•
Methods to prevent build-up of electrostatic charge (e.g., air conditioning,
humidification, conductive floor coverings, and non-synthetic clothing).
• Discharging one’s body to the frame of the equipment or to an earth
ground, or a large metal object.
• Bonding oneself, by means of a wrist strap, to the equipment or to an
earth ground.
For more information, see Precautions on page 4.
Figure 2 – Connect the SA0972AC Uroflow Transducer(s) to the M30/M30P Urodynamic Module
Continued on next page…
9
Page 12
Installation and System Interface, Continued
Connecting the
SA0972AC
Uroflow
Transducer to
the M30/M30P
Urodynamic
Module (cont.)
To connect the SA0972AC Uroflow Transducer to the M30/M30P
Urodynamic Module:
1.
Confirm that the system computer, M30/M30P Urodynamic Module and all
accessory components are powered OFF.
2.
Connect the circular plug on the end of the Uroflow Transducer cable into
the input jack labeled Uroflow1 on the bottom panel of the M30/M30P
Urodynamic Module.
To connect a circular plug:
a. Orient the interface cable
plug with the receptacle, so
that the “keyed” (notched)
insulator faces align.
b. Push the plug into the
receptacle until the locking
tabs snap into the undercuts.
Figure 3 – Connecting a Circular Plug
Disconnecting
the SA0972AC
Uroflow
Transducer
from the
M30/M30P
Urodynamic
Module
3.
If your system purchase included a second SA0972AC transducer and the
Uroflow2 software option, connect the circular plug on the end of the cable of
the second transducer to the input jack labeled Uroflow2 on the bottom
panel of the M30/M30P Urodynamic Module.
4.
The SA0972AC is ready for use.
To disconnect the SA0972AC Uroflow Transducer from the M30/M30P
Urodynamic Module:
1. Confirm that all system hardware is powered OFF.
2. Squeeze the circular plug on
the Uroflow Transducer cable
at the finger grips to flex the
plastic coupling area. The
locking tabs will deflect
outward and unlock.
3. Pull back on the plug until it is
fully disengaged.
Figure 4 – Disconnecting a Circular Plug
Continued on next page...
10
Page 13
Installation and System Interface, Continued
Connecting
the SA0972AW
Uroflow
Transducer to
a Patient Unit
The Model SA0972AW Uroflow Transducer is interfaced to the Life-Tech
Urodynamic System via the following Patient Units: M22J, M22, M20, M42, and
Microflow 2001, Spectrum 2200A.
To connect the SA0972AW Uroflow Transducer to a Patient Unit:
1.
Confirm that the system computer and all accessory components are
powered OFF.
2.
Connect the modular plug end of the transducer cable into the Uroflow input
jack on the bottom panel of the Patient Unit.
3.
Insert the modular plug into the Uroflow connector until the latch clicks and
secures the cable in place.
Figure 5 – Connect the SA0972AW Uroflow Transducer(s) to a Patient Unit
Continued on next page…
11
Page 14
Installation and System Interface, Continued
Connecting the
SA0972AW
Uroflow
Transducer to a
Patient Unit
(cont.)
NOTE: Labeling and availability of multiple Uroflow input jacks varies with
Patient Unit models.
If using the Model…
Then…
M22J Patient Unit…
a. Connect the Uroflow Transducer to the input
jack labeled Uroflow1 on the bottom panel of
the patient unit.
b. If your system purchase included a second
SA0972AW transducer and the Uroflow2
software option, connect the second
transducer to the input jack labeled Uroflow2
on the bottom panel of the patient unit.
M22, M20 or M42
Patient Unit…
¾ Connect the Uroflow Transducer to the input
jack labeled Uroflow on the bottom panel of
the patient unit.
The SA0972AW transducer is ready for use.
Disconnecting
the SA0972AW
Uroflow
Transducer
from a Patient
Unit
12
To disconnect the SA0972AW Uroflow Transducer from a Patient Unit:
1. Confirm that all system hardware is powered OFF.
2. Release the locking mechanism on the modular plug.
3. Pull back on the modular plug until it is fully disengaged.
Page 15
Using the Uroflow Transducer
Preparation
To prepare the Uroflow Transducer for use:
WARNING: Take care to keep the transducer where it cannot be stepped on
or kicked, which will damage the unit and present a tripping hazard for
persons in the area.
Note
Refer to page 17 for instructions on cleaning and disinfection of the Uroflow
Transducer.
1.
Power the attached urodynamic system
components ON.
2.
Place the Uroflow Transducer on the floor,
centered directly beneath the opening of the
funnel in the Uroflow chair.
3.
Place an empty graduated beaker inside the
recessed rim on the Uroflow Transducer
platter.
NOTE: It is important that you position the
graduated beaker before starting an
urodynamic procedure, so that the
automatic zeroing function in the system
software will set the voided volume and flow
rate channel / displays at zero,
compensating for the weight of the empty
graduated beaker.
4.
Figure 6 – Preparing the Uroflow
Transducer for Use
Operation
The transducer is now ready for use.
NOTE: Refer to the Life-Tech Urodynamic
System Operator’s Manual for further
instruction on procedure setup.
To operate the Uroflow Transducer:
¾
The SA0972AC and SA0972AW Uroflow Transducers have no useroperated mechanical or electronic parts. Once connected to the
M30/M30P Urodynamic Module or patient unit, the transducer is ready for
operation.
13
Page 16
Maintenance and Repair
Introduction
Problems with the Model SA0972AC or SA0972AW Uroflow Transducer typically
fall into the following categories:
1.
Operating error during urodynamic procedures (software or hardware
setup).
2.
Interface cable is damaged or disconnected.
3.
Need for routine calibration / maintenance.
4.
Instrument component failure.
If you experience a problem, please refer to the Uroflow Transducer
Troubleshooting Guide on page 16. If you are able to identify the cause of the
problem, try the suggested actions.
For troubleshooting assistance, contact Life-Tech Technical Service at
800-231-9841 or 281-491-6600. For assistance outside the USA, contact your
local Life-Tech authorized dealer, or FAX us at 281-491-6646. You can also
contact us through our web site: http://www.life-tech.com, or email our service
department at serv@life-tech.com.
The Life-Tech representative for the European Community is:
FBI Fred Berninger Importe
Bergstrasse 12
82024 Taufkirchen, GERMANY
Tel: +49 89 61 453 453
Fax: +49 89 61 453 455
Email: info@fbi-medizintecnik.de
14
Page 17
Maintenance and Repair, Continued
Product Return
Information
If your system components must be returned to Life-Tech for service or
calibration, follow this procedure:
1.
Call Life-Tech Technical Service and request a return authorization (RMA)
number.
2.
Clean, disinfect, and dry the instrument before placing it in packaging. For
cleaning instructions, see the Cleaning & Disinfection section of this manual
starting on page 17.
3.
Place the transducer in its original packaging, or in a sturdy container, well
padded with soft packing material.
4
In the mailing container, place a written list of the following information:
•
•
•
•
•
5.
A detailed description of why you are returning the Volume Transducer.
Your name and address.
The name and telephone number of a contact person who is a user of
the equipment.
A purchase order number (for non-warranty repairs / calibration).
Return shipping address and return shipping instructions.
Label the outside of the box with your return address, the return
authorization number and Life-Tech’s mailing address:
Life-Tech, Inc.
4235 Greenbriar Drive
Attn: Technical Service
Stafford, TX 77477-3995
NOTE: For fastest service, use prepaid airfreight. Insure the shipment and
keep the name of the carrier and freight waybill number to trace the shipment
if necessary.
Continued on next page...
15
Page 18
Maintenance and Repair, Continued
Uroflow
Transducer
Troubleshooting
Guide
The troubleshooting tips in this table will help you identify the possible causes of
problems you may encounter when using the Uroflow Transducer.
If…
Then…
The amount of
fluid in the
graduated beaker
and the reported
voided volume
don’t match.
•
•
•
Maintenance of
the Uroflow
Transducer
If the volume discrepancy exceeds ± 20 ml, the Uroflow Transducer may
have been damaged and require recalibration. Contact Technical Service
to make arrangements to return the unit for service.
Consider the possibility that the patient kicked the graduated beaker with
sufficient force to simulate the onset of flow. This would have started
recording and, if the patient did not begin voiding before the recorder’s
Post Void Delay timeout, the unit would have ended recording and printed
before fluid entered the graduated beaker. This is usually seen as a sharp
spike on the Uroflow waveform, followed by no further activity. Refer to the
recording system’s operator’s manual for more information.
The patient may have voided intermittently with long pauses. If a pause is
longer than the Post Void Delay setting in the urodynamic recorder
software, the unit ends recording, analyzes and prints. Refer to the
recording system’s operator’s manual for more information.
WARNING: Disconnect mains AC power to the system before conducting
any regular maintenance.
To maintain your Uroflow Transducer in optimum operating condition, do the
following:
1.
Have the instrument tested and calibrated to the factory specifications shown
on page 7, every 12 months.
2.
Protect the instrument from fluid spills and keep it free of dust and particulate
residue from infusion solutions, body fluids, etc. This is particularly important
when using contrast medium. Contrast leaves a sticky, thick, corrosive
residue and must be cleaned immediately. The procedures for cleaning and
disinfecting equipment follow.
3.
Protect the Uroflow Transducer from physical damage such as getting
stepped on or kicked during use, or damaged during floor cleaning.
CAUTION: The transducer platter
on Uroflow Transducer Models
SA0972AC or SA0972AW should
never need to be removed. If you
must remove it, unscrew it
counterclockwise until it comes
free of the transducer body.
Pulling the transducer platter
off instead of unscrewing it will
cause damage to both
components, which will void the
unit’s warranty.
16
Page 19
Cleaning & Disinfection
Introduction
In this section, you will find instructions for cleaning and disinfection of the Uroflow
Transducer and cable.
NOTE: Life-Tech, Inc. makes no claims of the efficacy of disinfection when
this procedure is used. The cleaning and disinfecting solutions recommended
herein have been tested only for materials compatibility with the Uroflow
Transducer. Refer to published guidelines1,2,3 or solution manufacturers’
recommendations for additional information on disinfection of medical
equipment.
The Uroflow Transducer should be cleaned:
•
Anytime it has been or is suspected to have been contaminated by
biologically active materials.
•
Before using it on a patient if its condition warrants (e.g. if there is an
accumulation of dust or other visible debris or contamination).
•
Immediately after spillage of infusion medium since contrast is particularly
corrosive.
The solutions shown in the table below were selected from the Association for
Professionals in Infection Control and Epidemiology (APIC) guidelines (1996)1
and CDC2 recommendations for disinfection of noncritical items and are approved
for use on system components.
Solution
Concentration
Alcohol (ethyl or isopropyl)
70%
Sodium hypochlorite
1 part household bleach to 10 parts water3
Aqueous quaternary
ammonium compounds
0.1 – 0.2 % (e.g. Buckeye® Sani-Care
Quat-256)
WARNING: Disconnect mains AC power to the system before applying any
cleaning or disinfecting solutions to hardware.
Cleaning &
Disinfection
Procedure
Supplies:
•
Cleaning and disinfecting solution(s)
•
Clean, soft cotton cleaning cloths
CAUTION: Do NOT spray or pour solutions directly into the Uroflow
Transducer, nor submerge the instrument, system hardware, interface cables,
and power cords in any solution. Wipe the hardware with a clean cloth
dampened with cleaning and disinfecting solution. The cloth should be damp,
not dripping, to ensure that solution does not enter the component cases.
Continued on next page…
1
Dix, K. Choosing Surface Disinfectants, Infection Control Today, 9/1/2002. Available online @
http://www.infectioncontroltoday.com/articles/291clean.html
2
Sterilization or Disinfection of Patient-Care Equipment: HIV Related. Available online @
http://www.cdc.gov/ncidod/hip/sterile/hivsteri.htm
3
OSHA & CDC Current Issues, New OSHA Inspection Guidelines. Medical Environment.
http://www.medicalenvironment.com/subscribers/osha.html
Buckeye is a registered trademark of Buckeye International, Inc.
17
Page 20
Cleaning & Disinfection, Continued
Cleaning &
Disinfection
Procedure
(cont.)
CAUTION: The transducer platter
on Uroflow Transducer Models
SA0972AC or SA0972AW should
never need to be removed. If you
must remove it, unscrew it
counterclockwise until it comes
free of the transducer body.
Pulling the transducer platter off
instead of unscrewing it will
cause damage to both
components, which will void the
unit’s warranty.
To order
replacement
graduated beakers
contact the LifeTech Customer
Service department.
Figure 7 – Cleaning & Disinfection
of the Uroflow Transducer
To clean and disinfect the Uroflow Transducer and/or cable:
NOTE: Efficacy of the disinfection process is NOT guaranteed herein, as it is
dependent upon user adherence to published clinical guidelines or solution
manufacturer’s instructions for use.
1.
Turn the system computer and all mains AC powered accessory equipment
OFF and disconnect the power cords from the mains AC power source.
2.
Apply cleaning or disinfecting solution to a clean cloth and use the cloth to
wipe the outside casing of the transducer to remove visible contaminants.
CAUTION: Cloth should be damp, not dripping, to ensure that the solution
does not enter the Uroflow Transducer, hardware casing, or cable
connections.
3.
If using…
Then…
Alcohol or chlorine
bleach solution…
Allow the item to air dry up to 10 minutes. Wipe
with a clean dry cloth after that time to remove
any residual.
Using a commercial
germicidal detergent
solution
Follow the manufacturer’s recommendations.
When the cleaned components are confirmed completely dry, reconnect
power cables to the mains AC power source.
WARNING: Do NOT reconnect the equipment to the mains AC power supply,
or turn power ON until all components are completely dry (e.g., all the alcohol
vapors have dispersed). An electrical spark could ignite the vapors and cause
a fire.
18
Page 21
Declaration of Conformity
19
Page 22
20
Page 23
Index
Cabling
Precautions ...................................................... 4
Circular plugs.............................................. 10
Environmental .............................................. 5
Interface Cables ........................................... 7
ESD .............................................................. 4
Cleaning.......................................................... 17
Safety ....................................................... 4, 7
Connecting
Preparation..................................................... 13
SA0972AC .................................................... 9
Repair............................................................. 14
SA0972AW ................................................. 11
SA0972AC Components .................................. 8
Declaration of Conformity ............................... 19
Disconnecting
SA0972AC Transducer
Connecting ................................................... 9
SA0972AC .................................................. 10
Disconnecting............................................. 10
SA0972AW ................................................. 12
SA0972AW Components ................................. 8
Disinfection ..................................................... 17
SA0972AW Transducer
EC Representative ..................................... 3, 14
Connecting ................................................. 11
Environmental Protection ................................. 5
Disconnecting............................................. 12
General Description.......................................... 3
Service ............................................... 3, 6, 8, 14
Life-Tech Information........................................ 3
Specifications ................................................... 7
Maintenance ............................................. 14, 16
Symbols............................................................ 6
Operation ........................................................ 13
Troubleshooting Guide................................... 16
Placement................................................... 5, 13