Supplement to Instructions for Use
2 Pages
Preview
Page 1
Ergänzung zur Gebrauchsanweisung Laserchirurgiegerät Limax® 120 Supplement to IFU Surgical Laser Unit Limax® 120
Please note the following changes in the IFU: 6.5.2
Testing the Purge Air Flow
To test the purge air flow, proceed as follows: • Switch Limax® 120 to single-pulse mode (10 W, pulse duration 0.5 s), see section 6.15.5 “Operating mode”, page 73. 6.15.5
Operating modes
Page 73: Depending on the selected operating mode, up to 2 pulse parameters can be set: • Pulse duration (0.5–10 s) Sets the length of one laser pulse • Pulse pause (0.5–10 s) Sets the pause between 2 pulses in pulse mode 6.15.13
11
AIR / VAC OFF
Technical Data
Designation Operating modes
Specification Limax® 120 / Limax® VAC Single pulse: pulse on-time: 0,5 – 10 s Pulse train, adjustable: pulse on-time: 0.5–10 s pulse off-time: 0.5–10 s
Gebrüder Martin GmbH & Co. KG Ein Unternehmen der KLS Martin Group KLS Martin Platz 1 · 78532 Tuttlingen · Germany Postfach 60 · 78501 Tuttlingen · Germany Tel. +49 7461 706-0 · Fax +49 7461 706-193 [email protected] · www.klsmartin.com
Page 2 of 2
Date of Release: 2016-08 · 90-636-52-30 · Printed in Germany · Copyright by Gebrüder Martin GmbH & Co. KG · Alle Rechte vorbehalten · Technische Änderungen vorbehalten · We reserve the right to make alterations · Cambios técnicos reservados · Sous réserve de modifications techniques · Ci riserviamo il diritto di modifiche tecniche · Revision 4