Checklist
6 Pages
Preview
Page 1
,
Periodic preventive maintenance Safety check Manual for Clinicare 100HF (1VCKE) Date Version
01/2014 01
N/A
Check mains cable Visually inspect the electrical mains cable for any signs of damage. Replace if necessary
N/A
General visual check of SCU Visually inspect the SCU Housing, Front Panel, Display Membrane, Air Connector micro switches, rubber mounting feet. Replace where necessary
N/A
Check the Hook Visually inspect the Hook Assembly for mechanical damage. Check the function. Replace if necessary
N/A
Replace external air filter Remove and replace the SCU Filter. The filter must be replaced irrespective of whether it looks visually clean. Ensure that when fitting the replacement filter it’s securely fitted into the front cover aperture
N/A
SCU external
SCU external
SCU external
Identification Write the serial and model numbers onto the preventative check report.
SCU external
Tools
SCU external
Action
SCU external
Picture
SCU external
P
Electrical testing Using an Electrical Safety Tester ensure the SCU passes the 500 Volt Insulation
SDIU
Test
Check On/Off button Press On button and change mode.
N/A
Check CPR alarm Remove mattress connector from SCU - alarm should be heard and the CPR light should flash.
N/A
SCU internal SCU internal SCU internal SCU internal SCU internal SCU internal Performance test Performance test Performance test
Open SCU for internal check Remove the 6 rear cover screws as shown in figure. Gently lift off the rear cover
N/A
to expose the internal components
Check electrical wiring Visually inspect all of the wiring looms for signs of damage.
N/A
Replace if necessary
Check electrical plug Ensure all the electrical plug connections are pushed fully home and are secure
N/A
Check pipes Visually inspect all of the pipes for signs of damage and internal dirt / dust.
N/A
Replace if necessary.
Check fuses Visually inspect the 2 x 1 amp Ceramic Power Supply PCB fuses. Ensure they are
N/A
rated at the correct value. See figure for fuse positions.
Close SCU Re-fit the front cover and secure the front cover with the 6 retaining screws
N/A
Pressure and cycle check (APT Mode) Ensure all pressures do not drop by more than 10mmHg during the inflated period and the cells inflate in the correct order during the cycle. The total cycle time for all 3 cells to inflate and deflate is 7,5 minutes.
Pressure check (Static Mode)
Check power-fail alarm Je tam? Switched on SCU and remove mains power cable. The power fail alarm (intermittent two tone alarm) should be heard.
N/A
Performance test Performance test
Check general Vibration level
Switched on SCU
Switched on SCU
N/A
N/A
Check cover Visually inspect the Upper and Lower Mattress covers together with the Cable Management clips for signs of damage or deterioration, including staining of the
N/A
white inside cover surface.
Mattress External
Check the umbilical hose cover Visually inspect the Umbilical Hose cover for signs of damage or deterioration. Replace if necessary.
N/A
Mattress External
Check security straps Visually inspect the Mattress cover security straps for signs of damage or deterioration. Replace if necessary.
N/A
Mattress Internal
Open the Top and Bottom cover Un-zip the Upper and Lower Mattress covers and remove the Top and Bottom deck ‘H’ clips (Non-Umbilical Hose side) to allow access to the Top / Bottom deck piping
N/A
Mattress Internal
Check pipes Visually inspect the pipes and connectors for correct fitment, signs of damage or deterioration. Re-fit / replace if necessary.
N/A
Mattress Internal
Replace if necessary
Check air cells Visually inspect the Top and Bottom deck air cells for signs of damage or deterioration. Replace if necessary.
N/A
Mattress Internal
Mattress External
Check general Noise level
Check sleeve Replace if necessary
N/A
Mattress Internal
Check MCM Check MCM pipes, check function. Replace if necessary
Mattress Internal
Check air connector Visually inspect the Umbilical Air Connector for signs of damage i.e. deterioration, cracking or splitting. Replace if necessary.
Label
If is everything correct place the PPM sticker on the SCU cover.
Report
Complete the preventive check report
N/A
N/A
Preventive check report
Periodic Preventative Maintenance protocol SCU serial number:
SCU model number:
Mattress serial number:
Mattress model number:
Customer:
City: Fault Detection
Visual Checks √
Test Function √
External SCU visual Checks / Operation
☐
☐
☐
☐
Internal SCU visual Checks / Operation
☐
☐
☐
☐
External mattress Checks / Operation
☐
☐
☐
☐
Internal mattress Checks / Operation
☐
☐
☐
☐
Performance tests with the SCU open
☐
☐
☐
☐
Performance tests with the SCU closed:
☐
☐
☐
☐
Electrical safety check. (500v insulation)
☐
☐
☐
☐
Check fault log with SDIU
N/A
N/A
N/A
N/A
Check last home & back lash readings with SDIU
N/A
N/A
N/A
N/A
Reset service indicator
N/A
N/A
N/A
N/A
Power switches correctly & green internal lamp comes on
☐
☐
☐
☐
All LED’s flash once
☐
☐
☐
☐
Single alarm Tone
☐
☐
☐
☐
Pressure & cycle check (Alternating / APT Mode)
☐
☐
☐
☐
Pressure check (Static / MAX Mode)
N/A
N/A
N/A
N/A
Pressure check (Evaluation / CLP Mode)
N/A
N/A
N/A
N/A
All buttons on panel
☐
☐
☐
☐
Back rest raise
N/A
N/A
N/A
N/A
Check MCM function
☐
☐
☐
☐
Battery Back-Up
N/A
N/A
N/A
N/A
General noise and vibration level when compressor running
☐
☐
☐
☐
Operation
YES
NO
Electrical Test (measurement according to the norms in the country) measured Insulation test.
allowed
unit
>2.000
MΩ
Replaced spare part number(s) / action made on the mattress:
Results: The mattress passed the service inspection and can be returned to use: ☐
YES
neds to be repared
Date:
Validity expires on: Name
Checked done by
NO
Signature