LMA (Laryngeal Mask Company)

C Trach System

C Trach Best Practice Guide Issue PAK2100009a

Guide

2 Pages

English  Best Practice Training Guide  Authorised EU Representative: LMA Deutschland GmbH, Mildred-Scheel-Strasse 1, 53175 Bonn, Germany Tel +49 (0) 228 1800 180 Fax +49 (0) 228 1800 181 Email: marketing@LMACO.com www.LMACO.com Authorised Representative in USA: LMA North America Inc, 4660 La Jolla Village Drive, Suite 900, San Diego CA. 92122 Tel (800) 788-7999 Fax (858) 622-4130 Email: custserv@LMANA.com www.LMACO.com Issue: PAK2100009a  Pre-use (LMA CTrach™ system) 1. Select LMA CTrach™ Airway (Airway) size using recommended LMA™ size selection. 2. Fully deflate the Airway into the recommended profile. 3. Pre-lubricate the Airway with the lubricated ETT. Do not pass the ETT beyond the Epiglottic Elevator Bar (EEB). 4. Clean the fibreoptics in the magnetic latch connector with an alcoholic wipe. 5. Apply anti-fogging solution (if necessary) to the fibreoptic distal end located in the bowl of the mask e.g. Clear Vue, FRED™. 6. Focus the Airway and LMA CTrach™ Viewer (Viewer) using the pre-focus card. When focusing reduce the ‘light output’ until there is no glare on the screen. NO FURTHER FOCUS ADJUSTMENT IS REQUIRED DURING USE  Pre-insertion (Patient) 1. Achieve sufficient depth of anaesthesia. 2. Suction the patient’s airway if appropriate.
File Type: PDF
File Size: 967 KB
File Name: c_trach_best_practice_guide_issue_pak2100009a.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions