Edi MODULE INSTALLATION INSTRUCTIONS
GENERAL This document describes installation of the Edi Module accessory that is required for NAVA and NIV NAVA on the SERVO ventilator systems. The SERVO ventilator systems are referred to as “the system” in this document.
To enable NAVA/NIV NAVA, the following is required: Edi Module (1) including Edi cable (2) and Edi test plug (3). The Edi Module requires one free slot in the module unit. NAVA/NIV NAVA software option. The NAVA/NIV NAVA
The User's Manual for the SERVO ventilator system is
software option is individually created for each system
referred to as “the User's Manual” in this document.
and can only be installed on this system. Serial number
Refer to the User's Manual for further information regarding NAVA/NIV NAVA, such as: Brief device description and description of intended use. Basic principles.
of the system must be stated when ordering the software. The NAVA/NIV NAVA software option must be ordered separately. Edi catheter (4). Single-use catheter with measuring
Set-up
electrodes. Available in different sizes. The Edi catheter
Operating/non-operating conditions and other technical
must be ordered separately.
data. Cleaning. Only personnel trained and authorized by MAQUET shall be permitted to install, service or repair the system. The Edi Module is a complete unit and must not be disassembled. Disassembling the Edi Module will make it defective. After any installation, maintenance or service intervention in the system, perform a Pre-use check. Refer to instructions in the User's Manual. Special waste: This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
©Maquet Critical Care AB | Röntgenvägen 2, SE-171 54 Solna, Sweden, Phone: +46 8 730 73 00, www.maquet.com
MEMBER OF THE GETINGE GROUP
Page 1 (3) Order No: 66 72 322, Revision 01 September 2013
INSTALLATION Edi Module
NAVA/NIV NAVA software option
Before connecting the Edi Module, check for visible
The NAVA/NIV NAVA software option must be installed on
damages on the module. If damages are found, the module
the system. This software option will change the function
must not be used.
(configuration) of the system to enable use of NAVA/NIV
Note: If the Edi Module is exposed to rough handling, e.g.
NAVA. The upgrade procedure is described in the ‘System
dropped on the floor, an Electrical safety test of the module
SW and Software option – Installation Instructions’ for the
must be performed. Refer to standard procedures regulated
system. The Edi Module must be connected during the
by IEC 60601-1:2005, Class 1, Type CF or corresponding
software upgrade as parts of the software are distributed to
national standards.
the module.
Install the Edi Module:
Note: The NAVA/NIV NAVA software option is installed as
Insert the module into a free slot in the module unit.
default in some systems. Installation of the NAVA/NIV NAVA
Make sure that the locking device snaps in.
software option is not required on these systems.
Depending on system, PC Card or USB memory stick must be used for software option installation. SERVO-i User Interface The SERVO-i User Interface must have a 'Neural access' fixed key. If required: Clean the fixed key area with ethylene alcohol or isopropyl alcohol. Attach the 'Neural access' sticker in appropriate language. Note: 'Neural access' stickers in different languages are delivered with the NAVA/NIV NAVA software option.
Connecting the Edi Module to the system. SERVO-i shown in illustration. Connection of the Edi cable and the Edi catheter is described in the 'User's Manual'.
©Maquet Critical Care AB | Röntgenvägen 2, SE-171 54 Solna, Sweden, Phone: +46 8 730 73 00, www.maquet.com
MEMBER OF THE GETINGE GROUP
Page 2 (3) Order No: 66 72 322, Revision 01 September 2013
Check after installation Check that the Edi Module and the Edi Cable works properly: Connect the Edi Cable to the Edi Module. Set the system to Standby. Connect the Edi Test Plug to the Edi Cable connector. This will start the automatic self-test. After any installation, maintenance or service intervention in the system, perform a Pre-use check. Refer to instructions in the User's Manual.