Installation Instructions
20 Pages
Preview
Page 1
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10550800 Rev. D
CE Notice
symbol on this product indicates compliance with the Medical Device Directive of the European The Community. A "Declaration of Conformity" in accordance with the Directive has been made and is on file. European Communities should contact the Authorized Representative listed below regarding any Medical Device Directive (MDD) inquiries. Direct inquiries may be made to the following addresses:
Porter Regulatory Affairs:
EU Authorized Representative:
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000
Parker Hannifin Ltd Instrumentation Products Division Riverside Road, Pottington Business Park Barnstaple, EX31 1NP, England Phone: +44 (0) 1271 313131 Fax: +44 (0) 1271 373636
MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLYIN ACCORDANCE WITH UL 2601-1, ANDCAN/CSA C22.2 NO. 601.1
Symbol
Description
Alternating Current
Earth (ground)
Attention, consult the accompanying documents
Type B Equipment
SAFETY SUMMARY WARNINGS The Matrx by Parker DIGITAL MDM® mixer is a specialized medical device and should be operated ONLY by a professional trained in the use of nitrous oxide. Nitrous oxide and oxygen mixtures should only be administered by or under the supervision of a medical professional. The Matrx by Parker DIGITAL MDM mixer is designed for use with an oxygen supply assuring 100% medical oxygen concentration. The use of other oxygen sources (i.e., oxygen concentrators) could result in an unsafe condition to the patient. Although this analgesia machine is designed to prevent the interchanging of the two gases, it is the responsibility of the user to determine that only oxygen flows into the oxygen connection and only nitrous oxide flows into the nitrous oxide connection located on the back of the analgesia head. ELECTRICAL CONNECTIONS Do not use an electrical cord with nicks, cuts, or other damage. Do not use the unit with an extension cord as it will be subject to electromagnetic interference (EMI). BAG TEE REQUIREMENTS A bag tee containing an air inlet valve is required for use with this unit. The bag tee is factory mounted on the unit or provided separately for remote mounting. Matrx by Parker Scavenging Nasal Inhalers provide the patient connection. CAUTIONS FACTORY SETTINGS Do not change any internal factory set adjustments. The Matrx DIGITAL MDM mixer is adjusted and calibrated for proper operation prior to shipment. Any change could alter the specified operation and accuracy. FACILITY PIPING The user MUST perform tests independent of the dealer or contractor to verify that all pipelines are connected correctly. If you are unfamiliar with piping systems, order Matrx manual 10311700. The user has the ultimate responsibility for properly connected pipelines essential to patient safety. DIAMETER INDEXED SAFETY SYSTEM Gas connections in the DIGITAL MDM mixer are mechanically indexed using the Diameter Indexed Safety System (DISS). This feature helps to prevent the crossing of pipelines. Under no circumstances should these connections be altered. DISCONNECT POWER There are no user-serviceable parts inside this unit. A shock hazard exists if power is not disconnected prior to any internal servicing or maintenance. Always disconnect the power source before inspection or repair. ALARM TEST The user should test the alarm system for proper operation prior to each clinical use.
i
Table of Contents Notes and Contact Information ...iv Introduction ...1 Intended Use...1 User and Installer Responsibility ...1 System Notes...1 NFPA Requirements and Warnings...1 Section 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9
Installation Instructions ...3 Description ...3 Planning ...3 Pre-Installation ...4 Mixer Head Installation ...4 Bag Tee and Scavenging Control Valve Installation ...6 J Hook Installation...8 Hose Connection ...8 Rubber Goods Installation ...9 Electrical Connection ...9
Section 2 2.1
Specifications...10 Minimum Space Requirements...10
Section 3 3.1 3.2 3.3 3.4
Safety Checks ...10 Final Connection Checklist ...10 Cross Pipeline Tests ...10 Gas Outlet Station Tests...10 Mixer Tests ...11
Section 4 4.1 4.2
Return Information and Warranty ...11 Return Goods Procedure...11 Warranty ...11
Section 5
Replacement Parts ...12
ii
Table of Illustrations Figure 1.
Outside View of a Typical Digital MDM® Cabinet Mount with Flush Mounted Bag Tee/Scavenging Valve ...3
Figure 2.
Installation of Demand Valves ...4
Figure 3.
Digital MDM® Mixer Head Configurations...5
Figure 4.
Screw and Tee Nut Adjustment ...5
Figure 5.
Digital MDM® Mixer Head Installation...6
Figure 6.
Flush Mounting the Bag Tee and Scavenging Control Valve ...6
Figure 7.
Surface Mounting the Bag Tee and Scavenging Control Valve ...7
Figure 8.
Mounting the Bag Tee and Scavenging Control Valve Inside the Cabinet...8
Figure 9.
Connection Schematic...9
Figure 10.
Installation of Rubber Goods ...9
Figure 11.
Installation of Power Cord...9
iii
Notes and Contact Information
Unit Serial Number
Date Purchased
Dealer Name
Dealer Phone Number
Notes:
Contact Information:
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000
iv
Introduction CAUTION Federal Law in the U.S.A. and Canada restricts this device to sale by or on the order of a licensed practitioner. Intended Use The cabinet mount installation kit and accessories are intended for use in mounting the Digital MDM® mixer and associated accessories. The Digital MDM mixer is intended for use by trained medical personnel only. User and Installer Responsibility It is the responsibility of both the user and the installer to ensure that the correct gas, at the correct pressure, is supplied to the fittings at the rear of the unit. See Specifications in the Digital MDM Operations Manual. System Notes Matrx by Parker recommends the use of a permanently mounted oxygen analyzer in the patient circuit to monitor the oxygen levels being administered to the patient. Additional monitoring should be employed particularly if the delivered oxygen and nitrous oxide is being used as background to intravenous or inhalant deep sedation and anesthesia. TheAmerican Society of Anesthesiology (ASA) encourages the use of an ECG, a Blood Pressure Monitor, and a Pulse Oximeter in these cases. NFPA Requirements and Warnings
WARNING Medical gas fittings are manufactured according to the Diameter Indexed Safety System (DISS) to prevent the delivery of the wrong gas to the patient. The correct supply line sizes are as follows: oxygen 1/2” OD copper nitrous oxide 3/8” OD copper evacuation 5/8” OD copper The installer must ensure that each gas is connected to the proper stop valve to preserve the integrity of the Diameter Indexed Safety System (DISS). 3. Type K or L seamless, copper tubing or standard weight brass piping is required. 4. Copper-to-copper joints made using phosphorous brazing filler are allowed without flux. 5. During brazing, piping must be continuously purged with dry nitrogen. 6. The piping system must be tested for leaks at 150 psi followed by a 24-hour pressure test at 75 psi. NOTE Do not connect the Digital MDM during these pipeline tests.
General System Requirements The Digital MDM mixer complies with the National Fire Protection Association (NFPA) 99C code requirements. The NFPA is a body which has formulated a set of regulations widely adopted by local building authorities as the basis for their own local building code requirements. Do not change fittings or tubing sizes from those specified in this manual. The fittings and tubing incorporate unique tubing sizes for each gas, an important safety feature to prevent accidental crossing of lines during installation or maintenance. The following basic requirements must be observed:
Manifold Requirements The Digital MDM mixer is intended for Type II facilities consisting of less than 2000 cu.ft. of gas and less than ten use points for the gas. 1. All systems shall be a minimum 2+2 configuration. 2. For a non-remote tank room, manual or automatic switchover is permitted. For a remote tank room, automatic switchover is required. NOTE A tank room is considered remote when it cannot be accessed from within the office.
1. Remote tank rooms must be locked. 2. Ventilation of at least 72 square inches is required in the tank room or the cabinet.
3. A remote tank room requires a remote shutoff valve. 4. Each tank requires independent pressure regulators.
1
5. All systems require both audio and visual alarms for high and low pressure and/or automatic switchover. The audio alarm can be cancellable; however, the visual alarm must remain until the condition is corrected. 6. Line pressure gauges are required for each gas. 7. A pressure relief valve set at 75 psi is required for each gas. 8. All threaded connections must conform to Diameter Index Safety System (DISS) standards to prevent cross connections. The system of different tubing sizes helps to avoid the connection of a flowmeter or outlet to the wrong gas supply line. 9. Mechanical means, such as 3/8” OD piping for nitrous oxide and 1/2” OD piping for oxygen, must be used to ensure correct connections. For more information on the NFPA 99C code requirements, consult your local Matrx by Parker representative or your local dental dealer.
U.S. Gas Color Coding All products in this manual are manufactured according to the standard ANSI (USA) color coding. For products manufactured with other gas color-coding standards, call Matrx by Parker Customer Service at 1-888-723-4001 for assistance. Gas Oxygen Nitrous
Color Green OxideBlue
WARNING Route all hoses and cords so that there are no obstructions, pinched or kinked hoses, or potential abrasions of hoses or cords when the installation is completed. WARNING Although the Digital MDM® mixer is designed to restrict the interchanging of the two gases, it is the responsibility of both the user and the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located on the back of the mixer head.
2
Section 1
Installation Instructions
1.1 Description These instructions describe the installation of the Digital MDM® mixer and a bag tee in a cabinet’s vertical surface. This manual includes installation instructions for an optional Matrx by Parker Scavenging System, as well as an optional gas outlet station. Figure 1 illustrates a typical installation.
WARNING These instructions fully comply with the National Fire Protection Association (NFPA) guideline No. 99C; see the Introduction for details. All equipment must be installed in accordance with these guidelines, as well as all local codes and ordinances. Matrx by Parker makes no warranty for cabinets and equipment that have been installed improperly. Planning should include the following: 1. The cabinet must be manufactured with the proper materials. • The mounting brackets are designed for mounting in wood, and wood and laminate, approximately 3/4” thick. • Due to the nature of the mounting and the need for equipment stability during use, plaster board and other soft wall materials are not suitable. 2. Determine the location of the mixer head. • Typically, the mixer head is located at the approximate eye level of a seated person.
Figure 1. Outside View of a Typical MDM® Digital Cabinet Mount with Flush Mounted Bag Tee/Scavenging Control Valve
• Consult with the practitioner to determine the desired location. 3. Determine the location of the bag tee.
IMPORTANT A bag tee containing an air inlet valve is required for use with this unit. The bag tee is provided separately for remote mounting. A standard Scavenging Nasal Inhaler or a set of rubber goods provides the patient connection.
• The bag tee is either flush mounted in the cabinet surface (hoses are not externally visible), surface mounted to the cabinet surface, or mounted to a bracket inside the cabinet. • If a scavenging system is to be installed, it is mounted next to the bag tee.
To reduce trace N2O in the operator’s environment, it is strongly recommended, though not required, that a Scavenging System be included with this installation. The scavenging system removes unwanted gases from the exhalation of the patient and the room.
• The bag tee/scavenging control valve is supplied assembled to the appropriate mounting hardware.
• Consult with the practitioner to determine the 1.2 Planning desired location. Planning is essential to the proper installation of the equipment. Without proper planning, minor changes can 4. If necessary, determine the location of the gas outlet quickly become major problems. Read the entire installation station. instructions before attempting any and every installation. • If the outlet station delivers only oxygen, it can be located next to the mixer head. If this is the case, the oxygen outlet is supplied assembled to the mixer head bezel. 3
• The outlet station can be surface mounted to the cabinet surface, flush mounted in the cabinet surface, or mounted inside the cabinet. See the Outlet Stations installation and Parts Manual.
4. Solder a Matrx shutoff valve to each gas line that will connect to the mixer head and the scavenging control valve; see Figures 2 and 9. For soldering requirements, see the NFPA 99C guidelines in the Introduction.
• Consult with the practitioner to determine the desired location.
NOTE Each mixer must have its own set of shutoff valves for its gases.
5. The cabinet must be properly designed. 5. Install the cabinet according to the manufacturer’s instructions.
• The cabinet must be large enough to contain the mixer head, the bag tee and scavenging system (if mounted inside), the gas outlet station, the gas shutoff valves, and all the connecting hoses andfittings.
1.4 Mixer Head Installation There are 3 possible configurations for the mixer head (see Figure 3):
• NFPA 99C specifically prohibits hoses and flexible connectors from being installed in walls, floors, ceilings, and partitions.
• small - use when there is no preexisting hole in the cabinet and no outlet station is to be mounted next to the mixer head
• The minimum depth of the cabinet, where the mixer head mounts, is 6-1/2”. See Section 2, Specifications.
• large (centered) - use when there is a preexisting hole in the cabinet and no outlet station is to be mounted next to the mixer head
• Check your local building codes regarding electrical, gas connection, and certification requirements. Typically, gas and electrical connections must be accessible after installation.
• large (with integral outlet station) - use when an oxygen outlet station is to be mounted next to the mixer head Associated with each configuration is an installation template, a mounting shelf, and a bezel.
6. Determine the location of the cabinet. • Consult with the practitioner to determine the desired location.
NOTE The equipment is supplied with one installation template, one bezel (assembled to the mixer head), and one mounting shelf only.
• Consider the necessary connections to the central gas supply. • Consider the availability of an electrical outlet. The mixer head requires an electrical connection and is supplied with a 10’ power cord. It may be necessary to install an electrical outlet. 1.3 Pre-Installation 1. Install the necessary gas plumbing from the central gas supply. 2. If separate gas outlet stations are to be installed, install the necessary gas plumbing. See the Outlet Stations Installation and Parts Manual. NOTE All piping for medical gases must conform to the NFPA 99C guidelines (see Introduction) and local fire and building codes. 3. If necessary, install the electrical outlet and connection.
Figure 2. Installation of Demand Valves. 4
Figure 3. Digital MDM® Mixer Head Configurations. If an oxygen outlet station is to be installed next to the mixer head, it will be supplied assembled to the bezel. 1. Position the installation template at the desired location on the cabinet surface. 2. Mark the (4) mounting holes and the cutout area on the cabinet surface. 3. Drill (4) 1/16” pilot holes for the cutout area, and cut out the area for the mixer head. 4. Using a #20 drill, drill (4) 0.161” mounting holes. 5. From inside the cabinet, insert the (4) tee nuts into the (4) mounting holes. See Figure 4. 6. While supporting the front of the cabinet, tap in the tee Figure 4. Screw and Tee Nut Attachment. nuts with a hammer. Alternately, a C clamp can be used to squeeze the tee nuts into the cabinet surface. Be 8. Align the mounting shelf and bezel assembly with the certain to protect the front surface of the cabinet with a mounting holes, and slide it into the cutout area. See clean piece of wood between the C clamp and the Figure 5. finished surface. 9. Connect the gas hoses and the power cord to the rear of the mixer head. See Sections 1.7 and 1.9.
7. Align and slide the mixer head and bezel assembly into the mounting shelf.
10. Slide a screw cap onto each of the (4) screws. See Figure 4.
5
Figure 5. Digital MDM® Mixer Head Installation 11. With the (4) screws, secure the mixer head, the bezel, and mounting shelf to the cabinet surface. See Figures 4 and 5. 12. Tighten the (4) screws and snap the screw caps over the screw heads. 1.5 Bag Tee and Scavenging Control Valve Installation If a scavenging control valve is to be installed, it is installed next to the bag tee. See Figure 6. The bag tee/scavenging control valve is supplied assembled to the appropriate mounting hardware.
Figure 6. Flush Mounting the Bag Tee and Scavenging Control Valve.
CAUTION During daily operation, the bag tee/scavenging control valve mount is occasionally stressed, for example, when the breathing bag is removed or installed. The attachment to the cabinet surface must be sufficiently rigid to withstand the stress.
1.5.1
The bag tee/scavenging control valve assembly may be installed on plywood attached to studs or on a wood cabinet surface. It must not be installed onto plaster board or other soft wall materials.
Flush Mounting the Bag Tee/Scavenging Control Valve For flush mounting, the bag tee/scavenging control valve is supplied assembled to a mounting plate; see Figure 6.
1. Position the bag tee/scavenging control valve template at the desired location on the cabinet surface. 2. Mark the (4) mounting holes and the (2) holes for the gas hoses.
All hoses and gas connections must be accessible after the installation is complete.
3. Using a #20 drill, drill (4) 0.161” mounting holes.
6
7. Orient the hose barbs (they turn) parallel to the mounting plate.
4. Drill the (2) 3/4” holes for the gas hoses. 5. From inside the cabinet, insert the (4) tee nuts into the (4) mounting holes. See Figure 4.
8. Align the mounting plate with the (4) holes, and secure with the (4) screws.
6. While supporting the front of the cabinet, tap in the tee nuts with a hammer. Alternately, a C clamp can be used 9. Tighten the (4) screws and snap the screw caps over the screw heads. to squeeze the tee nuts into the cabinet surface. Be certain to protect the front surface of the cabinet with a 10. Connect the gas hoses to the hose barbs. See Section clean piece of wood between the C clamp and the 1.7. finished surface. Mounting the Bag Tee/Scavenging Control Valve Inside the Cabinet For mounting inside the cabinet, the bag tee/scavenging control valve is supplied with a mounting bracket. Figure 8 illustrates the bag tee/scavenging control valve installed on the mounting bracket.
7. Slide a screw cap onto each of the (4) screws. See Figure 4. 8. Orient the hose barbs (they turn) perpendicular to the mounting plate. 9. Align the mounting plate with the (4) holes. The hose barbs should extend into the cabinet surface.
Install the mounting bracket on one of the inside surfaces (right or left), toward the front of the cabinet. During use, the bag tee/scavenging control valve is turned toward the front of the cabinet. For storage, the bag tee/scavenging control valve is turned toward the rear of the cabinet.
10. Connect the gas hoses to the hose barbs. See Section 1.7. 11. Secure the mounting plate with the (4) screws.
CAUTION The mounting bracket must be secured to the cabinet surface with bolts and nuts. A washer must be inside each nut. If necessary, strengthen the cabinet surface with a furring strip running from front to back.
12. Tighten the (4) screws and snap the screw caps over the screw heads. 1.5.2
Surface Mounting the Bag Tee/Scavenging Control Valve
Do not install the mounting bracket to the inside of the cabinet door. Daily use will put too much stress on the door hinges.
For surface mounting, the bag tee/scavenging control valve is supplied assembled to a mounting plate; see Figure 7. The hose barbs are oriented so that, when installed, the hoses attach from below.
1. Position the bracket mounting template at the desired location inside the cabinet surface.
1. Position the bag tee/scavenging control valve template at the desired location on the cabinet surface.
2. Mark the (2) mounting holes.
2. Mark the (4) mounting holes. 3. Using a #20 drill, drill (4) 0.161” mounting holes. 4. From inside the cabinet, insert the (4) tee nuts into the (4) mounting holes. See Figure 4. 5. While supporting the front of the cabinet, tap in the tee nuts with a hammer. Alternately, a C clamp can be used to squeeze the tee nuts into the cabinet surface. Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the finished surface. 6. Slide a screw cap onto each of the (4) screws. See Figure 4.
Figure 7. Surface Mounting the Bag Tee and Scavenging Control Valve
7
3. Drill the (2) 3/8” mounting holes.
1.6 J Hook Installation The Digital MDM® Flush Cabinet Mount Installation Kit includes a J hook for rubber goods storage. Mount the J hook in a convenient location; consult with the practitioner.
4. Slide a screw cap onto each of the (4) short screws, as in Figure 4.
5. If the practitioner wants the hoses to attach from below, 1.7 Hose Connections orient the hose barbs (they turn) parallel to the mounting plate. If the practitioner wants the hoses to WARNING attach from behind, orient the hose barbs perpendicular The mixed gas outlet from the rear of the mixer head to the mounting plate. must be connected to a single bag tee only. 6. Align the mounting plate of the bag tee/scavenging control valve assembly to the front of the mounting bracket. Secure with the (4) screws.
For gas compatibility, all hoses must be supplied by Matrx by Parker. These hoses are braided to prevent occlusion when bent.
7. Tighten the (4) screws and snap the screw caps over the screw heads.
All hoses should be as short as possible and installed in a manner that prevents kinking, crushing, and excessive movement. Properly installed, short hoses minimize back pressure on the system.
8. Align the back of the mounting bracket with the (2) holes on the cabinet surface, and secure with the (2) 21/4” bolts.
All hoses must be fully seated on the hose barbs. This prevents gas leakage, which would result in the mixer’s 9. From the other side of the cabinet, install a flat washer, incorrectly displaying the volume of gas being a lock washer, and a nut on each bolt. Tighten the nuts. delivered to the patient. 10. Connect the gas hoses to the hose barbs. See Section 1.7.
Following the schematic in Figure 9, install all hoses at the proper connections. Be aware that your installation may not include all the components shown in Figure 9.
NOTE If desired, the installer may reverse the order of installation, first installing the mounting bracket to the cabinet surface (steps 8 – 9) and then installing the bag tee/scavenging control valve to the mounting bracket (steps 4 – 7).
Figure 8. Mounting the Bag Tee and Scavenging Control Valve Inside the Cabinet
8
Figure 9. Connections Schematic (some components may not be part of your installation).
Figure 11. Installation of Power Cord. 1.8 Rubber Goods Installation Install the rubber goods. See Figure 10. 1.9 Electrical Connection Install the power cord at the rear of the mixer head. See Figure 11. Connect the power cord to the electrical outlet.
Figure 10. Installation of Rubber Goods.
9
Section 2 2.1.
Specifications
Minimum Space Requirements Installation
W×W×D
Digital MDM® mixer head (small bezel)
4-1/4” × 9-3/4” × 6-1/2”
Digital MDM® mixer head (large bezel)
4-1/4” × 12-5/8” × 6-1/2”
Digital MDM® mixer head (large bezel with outlet)
4-1/4” × 12-5/8” × 6-1/2”
Bag tee/scavenger system (flush mount)
18” × 3-1/2” × 4-1/4” *
Bag tee/scavenger system (mounted inside the cabinet on swing arm)
18” × 3-1/2” × 8-1/2” *
* Includes reservoir breathing bag
Section 3 3.1 Final Connection Checklist When all items have been installed, all hose connections must be checked and tested. The installer must ensure that all gas connections are correctly made and leak free. The installation must conform to NFPA 99C guidelines, as well as all local codes and ordinances. For proper United States color coding, see page 2. WARNING Route all hoses and cords so that there are no obstructions, pinched or kinked hoses, or potential abrasions of hoses and cords when the installations complete.
Safety Checks 1. Close the gas shutoff valves on the main supply cylinders. 2. Exhaust the gas from the entire manifold and pipeline system by sequentially connecting an indexed gas hose to each outlet station or connection point. WARNING To prevent injury from hose whip, always hold the free end of the hose. 3. Open the oxygen shutoff valve on the main cylinder only. 4. Verify: • that oxygen gas flows at each oxygen outlet station
NOTE When all connections are complete, all gas stop valves and hoses must be accessible for future maintenance of the mixer and its components. 3.2
Cross Pipeline Tests
• that no gas flows at each nitrous oxide outlet station NOTE Use only oxygen or nitrous oxide when testing gas lines.
WARNING Although this analgesia machine is designed to restrict 5. With the oxygen still pressurized, open the nitrous oxide shutoff valve on the main cylinder. the interchanging of the two gases, it is the responsibility of both the user and the installer to determine that only oxygen flows into the oxygen fitting 6. Verify that nitrous oxide gas flows at each nitrous oxide outlet station and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head. 3.3 Gas Outlet Station Tests If additional gas outlet stations are installed, complete the If the gas pipeline test fails at any step, the pipeline test procedures listed in the Matrx by Parker Outlet Stations must be repaired and tested again, starting at step 1. Installation and Parts Manual. Before repair or testing, the entire system must be purged of any residual gas. If repair is required, follow 3.4 Mixer Tests the NFPA 99C code for brazing requirements; see the Complete the performance checks listed in section 3.1 of Introduction. the Digital MDM® Operations Manual (Part No. 10545100).
10
Section 4 Return Information and Warranty All repairs, unless otherwise specified, should be performed 4.2 Warranty This warranty is given in lieu of all other warranties, by an authorized Matrx by Parker service representative. expressed or implied, of merchantability, fitness for a Call 1-888-723-4001 for Customer Service. particular purpose, or otherwise. 4.1 Returned Goods Procedure No statement or claim about the product by any employee, agent, representative, or dealer of Matrx by Parker shall All returns must be made through an authorized Matrx constitute a warranty by Matrx by Parker or give rise to any dealer. Units for repair should be sent to Matrx by Parker and packaged in the original shipping container if possible. liability or obligation of Matrx by Parker. Enclose a letter detailing the difficulties experienced and the Subject to the next sentences, Matrx by Parker warrants repairs you feel are necessary. Please contact our that each product be free from defects in workmanship and Customer Service Department prior to shipping the unit materials, under normal use and with appropriate prepaid. maintenance, for the time listed below, commencing from the date of delivery to the customer. Ship to: Parker Hannifin Corporation Digital MDM® Cabinet Mount: 1 year Porter Instrument Division 245 Township Line Road Matrx by Parker’s obligations for breach of this warranty, or Hatfield, PA 19440 USA for negligence, or otherwise shall be strictly and exclusively Tel 215-723-4000 limited to repair or replacement of the product. This warranty shall be void on a product on which the serial number has been altered, defaced, or removed. For international returns, contact your local Matrx by Parker Matrx by Parker shall not be liable for any damage, injury, dealer. or loss arising out of the use of the product, whether as a result of a defect of the product or otherwise, if prior to such damage, injury, or loss, the product was (1) damaged or misused; (2) repaired, altered or modified by persons other than Matrx by Parker. Under no circumstances shall Matrx by Parker be liable for incidental or consequential damages as those terms are defined in the Uniform Commercial Code.
11
Section 5
Replacement Parts
The following replacement parts may be obtained from Digital MDM®: Matrx Part No. 20183200 10545100 63882537 91515087 91515088 91515089 62939600 91525136 91505112 91316043
Description Power Cord (USA) Instruction Manual Fuse: 2.5 amp, 250 volt, slo-blo (time delay) 5 mm by 20 mm (Note: international 220-240 volt units require two fuses) Autoclavable nitrous scavenging assembly Adult Autoclavable nitrous scavenging assembly Medium Autoclavable nitrous scavenging assembly Pediatric Vacuum gauge (scavenger control valve) Swing arm assembly, with mounting hardware Pre-installation kit (shutoff valves) 3 liter breathing bag
Mounting Fasteners 91525141 Hardware Kit, 1” screws, tee nuts, (4) black caps 91525169 Hardware Kit, swing arm “wall side” 91525137 Hardware Kit, 1/2” screws, tee nuts, (4) black caps 91525142 Hardware Kit, 1” screws, tee nuts, (4) pebble grey caps
Installation Templates 63936200 10” bezel 63936300 12-5/8” bezel, unit centered 63938200 12-5/8” bezel, with O2 Quick Connect 63940000 Bag tee, scavenging control valve 63939900 Swing arm (wall side template) 63940100 Flush swivel bag tee (alone) template 63940400 Undercounter hinge
12
Matrx Part No.
Description
Hoses 91525143 91525144 92305126 92305127 92515063 92305110 92305111 92305248 92305288 92305289 92515064 92305118 92305119 92305353
Black Scavenging hose, 1/4” I.D., 10’ long Bag tee hose, 5/16” I.D., 10’ long Ohio Male Quick Connect Ohio Quick Connect male X DISS female Ohio Quick Connect male X DISS female Ohio Quick Connect male X 1/4 NPT DISS (Diameter Indexed Safety System) DISS female X DISS female DISS female X DISS female DISS female X DISS female DISS Hand-I-Twist DISS Hand-I-Twist female X DISS female DISS Hand-I-Twist female X DISS female DISS Hand-I-Twist female X 1/4 NPT DISS X Ohio Female Quick Connect DISS female X Ohio QC female DISS female X Ohio QC female DISS female X Ohio QC female
13
Oxygen Nitrous Oxide Vacuum Oxygen Nitrous oxide Vacuum Oxygen Nitrous oxide Vacuum Oxygen Nitrous oxide Vacuum
REVISION HISTORY Rev. C
Date 2/18/10
Ecn # 10-0052
Description of Changes Removed PO Box info throughout document. Changed UL number from 82AN to 34S7. Eu Authorized Rep has changed. Sheet 11: Tel was TelL.
Dr MAM
D
8/19/10
10-0235
Updated CE Mark
ANB
DATE: 2/18/10
ECN: 10-0052
DR: MAM
CH:
Ch RAR
AP:
Digital MDM Cabinet
NUMBER: 10550800-D
REV.
Mount Installation Instructions
10550800
D
Appr SDL