multipiezo Use and Maintenance Manual Rev 05
Use and Maintenance Manual
254 Pages

Preview
Page 1
G
D
I
N
J
M
F
x3
E
A
P
B O
C
K
H
O-RING 6,5X2 NBR x6
L
3
4
1
6
2
7
5
Copyright © Mectron S.p.A. 2021. All rights reserved. No part of this document may be reproduced, in any form whatsoever, without the prior written consent of the copyright holder.
I
multipiezo CONTENTS
1
Introduction
2
Identification Data
7
3
Delivery
9
4
Installation
13
5
Use
16
6 7 8
Pump/Flush Function Disassembling Parts for Cleaning and Sterilisation Cleaning and Sterilisation
27 28 31
1.1 1.2 1.2.1 1.2.2 1.2.3 1.2.4 1.3 1.4 1.5 2.1 2.2 2.3 3.1 4.1 4.2 4.3 5.1 5.2 5.3 5.4 5.5
8.1 8.1.1 8.1.1.1 8.1.1.2 8.2 8.2.1 8.2.2 8.3 8.3.1 8.3.2 8.3.3 8.4 8.4.1 8.4.1.1 8.4.1.2 8.4.1.3 8.4.1.4 8.4.2 8.4.2.1 8.5 8.5.1
II
Intended Use Description of the Appliance Eligible Patient Group Patient Selection Criteria Indications for Use Users Disclaimer Safety Requirements Symbols Appliance ID Plate Handpiece Identification Data Inserts Identification Data List of Components First Installation Safety Requirements During Installation Connecting the Accessories Power On/Off Description of the Keyboard Safety Requirements Before and During Use Instructions for Use Important information on the Inserts
Preparation Disinfecting the Tank Irrigation Circuit Required Materials Procedure Cleaning the Non-Sterilisable Parts Required Materials Cleaning Method Cleaning the Tank and Cap Preparation Required Materials Cleaning Method Cleaning the Sterilisable Accessories Manual Cleaning Required Materials Scaler Handpiece Inserts Inserts Tightening Wrench Automatic Cleaning Required Materials Cleaning Check Required Materials
1 1 2 2 2 2 3 3 4 6 7 8 8 9
13 13 14
16 17 21 23 26
32 32 32 32 38 38 38 39 39 39 40 41 42 42 43 46 48 50 50 52 52
8.6 8.6.1 8.7 8.7.1 8.7.2 8.7.2.1
Drying and Lubrication Required Materials Sterilisation Preparation Sterilisation Method Special Information
53 53 54 54 55 56
9
Maintenance
56
10 11
Methods and Precautions for Disposal Technical Data
58 58
12
Troubleshooting
67
13
Warranty
73
9.1 11.1 11.1.1 11.1.2 11.1.3 11.1.3.1 11.1.3.2 11.1.3.3 11.1.4 12.1 12.2 12.3 12.4
Replacing the Peristaltic Pump
Electromagnetic Compatibility IEC/EN 60601-1-2 Guide and Manufacturer’s Declaration - Electromagnetic Emissions Accessible Parts of the Casing Guide and the Manufacturer's Declaration - Electromagnetic Immunity Power Connection BC Input Points of Contact with the Patient Parts Accessible to the Input / Output Signals Specifications of the Tests for the Immunity of the Accessible Parts of the Casing to the Wireless RF Communications Device Diagnostic System and Symbols on the Keyboard Quick Troubleshooting Replacing the Fuses Shipping to an Authorised Mectron Service Centre
56 60 60 61 62 62 64 65 66
67 69 71 72
III
Introduction
1
INTRODUCTION
Carefully read this manual before proceeding with installation operations, use, maintenance or other intervention operations on the appliance. This manual must always be available to the operator. Important: To avoid damages to persons or property, carefully read all the "Safety Requirements" in the manual. Depending on the level of severity, the safety requirements are classified with the following indications: WARNING: (Always referred to damage to persons) CAUTION: (Referring to possible damage to property) The aim of this manual is to make the operators aware of the safety regulations,
1.1
installation procedures, instructions for proper use and maintenance of the appliance and its accessories. Use of this manual for aims other than those strictly linked to the installation, use and maintenance of the appliance is prohibited. The information and illustrations in this manual were updated on the edition date shown on the last page. MECTRON is engaged in continuously updating its products with possible changes to the components of the apparatus. In case you encounter discrepancies between the descriptions found in this manual and the equipment in your possession you can: • • •
check for any available updates in the MANUALS section of the MECTRON website1; ask clarifications to Your Dealer;
contact MECTRON After Sales Service.
Intended Use
With the appropriate inserts it is possible to perform the following treatments: •
•
•
Scaling: all the procedures for the removal of deposits of bacterial plaque and supragingival, subgingivali and inter dental stones, and removal of stains; Periodontics: periodontal therapy for scaling and root-planing/debridement, including cleaning and irrigation of the periodontal pocket;
•
•
Cleaning treatment of the implant surface;
•
Endodontics: all treatments for canal preparation, irrigation, filling, guttapercha condensation, endodontic reprocessing and retrograde cavity preparation;
Restorative and prosthetics: cavity preparation and removal of carious tissue, removal of prosthetics and excess restoration materials, amalgam condensation, finishing of prosthetic stump; Extraction techniques.
The Mectron piezoelectric ultrasonic scaler devices can be used on patients of any age and gender requiring a tooth polishing treatment. There are no contraindications for specific population segments.
1
https://manuals.mectron.com/
1
multipiezo WARNING: The appliance must be used in a dental practice or clinic, or in professional oral hygiene and preventive care centres. Do not use the device in environments where the atmosphere is saturated with flammable gases (anaesthetic mixtures, oxygen, etc.).
1.2
WARNING: Qualified and specialised staff. The appliance is intended exclusively for use by a dental surgeon, dentist, or dental hygienist, with suitable medical/health training. WARNING: Intended use. Use the appliance exclusively for the use for which it is intended. Failure to observe this regulation can cause serious injury to the patient, the operator, and damage/failure of the device.
Description of the Appliance
multipiezo is a multi-purpose piezoelectric ultrasonic scaler designed to equip the operator with a product featuring an innovative design and unique technical characteristics, and to ensure utmost patient comfort during treatment. The user interface is optimised to make all functions immediately available, by integrating them into the touch keyboard. The handpiece
1.2.1
is fitted with a 360° adjustable LED light and can be autoclaved at a maximum temperature of 135°C for a maximum period of 20 minutes. The appliance is equipped with an automatic tuning circuit that optimises power and frequency for each insert available, in order to always operate in conditions of maximum efficiency.
Eligible Patient Group
This medical device is designed to be used with the following patient groups: •
Children;
•
Adults;
• •
Adolescents;
This medical device can be used on patients of any age, weight, height, gender, and nationality, where applicable.
Seniors.
1.2.2
Patient Selection Criteria
It is not recommended to use the device in the following cases: 1.
Patients with active implantable medical devices (for example: pacemakers, hearing aids, and/or other electromagnetic prostheses) without prior authorisation from their doctor;
1.2.3
Indications for Use
2.
Patients with clinical conditions not suitable for treatment of the sites (for example: local anaesthesia).
All piezoelectric ultrasonic scaler device models are intended for professional use only. The user is therefore the only person able to decide if and how to treat their patients.
Use of the device is suitable for all eligible patients (see Chapter 1.2.1 on page 2 ) whose doctor has prescribed professional scaling, periodontology, an extraction technique, endodontic cleaning, implant site surface cleaning, a restorative or prosthetic technique, as part of the intended use of the device (see Chapter 1.1 on page 1 ).
2
Introduction 1.2.4
Users
1.3
Disclaimer
The device must be used exclusively by specialised and properly trained personnel, specifically the dental surgeon/dentist or dental hygienist, who must be able-bodied, adult, of any weight, age, height, gender, and nationality.
The manufacturer MECTRON disclaims all liability, express or implied, and cannot be held responsible for direct or indirect personal injury and/or property damage, occurring as a result of incorrect procedures linked to the use of the appliance and its accessories. The manufacturer MECTRON cannot be held liable, expressly or implicitly, for any type of personal injury and/or property damage, inflicted by the user of the product and its accessories and which occurs in the following cases, by way of example and not of limitation: • •
• •
•
•
Misuse or use during procedures other than those specified in the destination of use of the product; The environmental conditions for preservation and storage of the device are not complying with the requirements indicated in Chapter 11 on page 58; The appliance is not used in accordance with all the instructions and requirements described in this manual;
The electrical system of the places where the equipment is used do not comply with the laws in force and the related regulations;
Assembly operations, extensions, readjustments, upgrades and repairs of the device are carried out by personnel not authorised by MECTRON; Misuse, abuse, abnormal use, negligent use, intentional misconduct or use exceeding the limits of the appliance
• •
•
• • • • •
indicated and allowed and/or normal wear or deterioration, ill-treatment and/ or incorrect interventions;
Any attempt to tamper with or modification of the appliance under every circumstance;
Use of non original MECTRON inserts resulting in permanent damage to the thread of the handpiece with impaired functioning and risk of injury to the patient;
Use of non original MECTRON inserts used in accordance with the settings designed and tested on the original MECTRON inserts. Correct use of the settings is only guaranteed with original MECTRON inserts;
Shortage of stock material (handpiece, inserts, wrenches) to be used in case of a standstill due to faults or problems; Incorrect/omitted maintenance compared to what is stated in Chapter 9 on page 56 of this manual; Breach of the requirements and the information contained in Chapter 5.5 on page 26 of this manual; Breach of the requirements and the information contained in Chapter 8 on page 31 of this manual;
Unauthorised repairs in accordance with the indications contained in Chapter 12.4 on page 72 this manual.
3
multipiezo 1.4
Safety Requirements
WARNING: Contraindications. Do not use the appliance on patients with Pacemakers or other implantable electronic devices. This regulation also stands for the operator. WARNING: Contraindications. Do not perform scaling treatments without water spray in order to avoid an overheating of the insert that can cause damage to the tooth. Treatments without water spray can only be those performed with the "Dry Work" inserts without water passage. CAUTION: Contraindications. Ultrasonic scaler. Do not perform treatments on metal or ceramic prosthetics. Ultrasonic vibrations could lead to decementation of the dentures. WARNING: Contraindications. Interference from other equipment. An electrosurgical scalpel or other electrosurgical units positioned near the multipiezo appliance may interfere with the correct operation of the appliance itself. WARNING: Contraindications. Interference with other equipment. Even if compliant with standard IEC 60601-1-2, multipiezo may interfere with other devices in the vicinity. multipiezo must not be used near or stacked with other appliances. However, if this should become necessary, it is necessary to check and monitor the correct functioning of the appliance in that configuration. WARNING: Risk of explosion. The appliance cannot operate in environments where the atmosphere is saturated with flammable gases (aesthetic mixtures, oxygen, etc.).
4
CAUTION: In case the final user, operating in their own medical room or surgery, in order to comply with mandatory requirements, must periodically inspect the equipment present in the surgery, the test procedures to apply to medical electrical equipment and medical electrical systems for the safety assessment must be carried out following the standard EN 62353 'Medical electrical equipment - Periodic inspections and tests to be carried out after repair of medical electrical equipment'. The frequency of periodic inspections in the intended conditions of use described in this “Use and Maintenance” manual is once per year or every 2000 hours of use, whichever condition is satisfied first. WARNING: Checking the condition of the device before treatment. Always check that there is no water under the appliance. Before every treatment always check the perfect operation of the appliance and efficiency of the accessories. If operation anomalies are observed do not carry out the treatment. If the faults concern the appliance, contact an Authorised MECTRON Service Centre. CAUTION: The electrical system of the premises in which the appliance is installed and used must comply with the rules in force and the relevant requirements of electrical safety. CAUTION: To avoid the risk of electrical shocks, this appliance has to be connected exclusively to a power supply with an earth safety device. WARNING: Cleaning and sterilising new or repaired instruments. All the new or repaired appliance accessories are not sterile. At first use and after each treatment they must be cleaned and sterilised by carefully following the instructions in Chapter 8 on page 31.
Introduction WARNING: Infection control. For maximum patient and operator safety, before using all the parts and reusable accessories, make sure they have been previously cleaned and sterilised by following the instructions in Chapter 8 on page 31. CAUTION: Contraindications. After having autoclaved the handpiece, inserts, torque wrench or any other sterilisable accessory, wait for them to completely cool down before reusing them. WARNING: Insert Breakage and Wear. High frequency oscillations and wear can, in rare cases, lead to the breakage of the insert. Deformed or damaged inserts are susceptible to breakage during use. Broken or worn inserts must not ever be used. In the event of breakage check that no fragments remain in the treated part and at the same time use suction effectively to remove them. It is necessary to instruct the patient to breathe through the nose during the treatment, or use a dental dam, so as to avoid to ingest fragments of broken inserts. Check the state of wear of the insert and its integrity before and during each use. If a drop in performance occurs, see to its replacement. The state of wear of the most common inserts (S1, S1-S, S2, S5, P2, P4, P10) can be checked using the supplied INSERT-CARD. To use the INSERT-CARD correctly: •
Position the insert on the INSERT-CARD so that the profile corresponds to the one printed on the card. The printed profile on the card has a red line
•
indicating the limit of wear;
If the insert is shorter than the limit of wear, its performance will be significantly inferior compared to that of a new insert, and should therefore be replaced.
If the layer of titanium nitride (gold-plated surface), where present, is visibly worn, the insert must be replaced. The use of a worn insert reduces its efficiency. Diamond Inserts: diamond inserts must be replaced when the layer of titanium nitride is visibly worn and in any case after a maximum of 10 treatments. When the nitriding is consumed, the cutting edge loses effectiveness; a possible regrinding harms the insert, therefore it is prohibited. Verify that the insert is not worn. During the operation frequently check that the insert is intact, especially in the apical part. During the operation avoid prolonged contact with the retractor or with metal instruments in use. Do not exert excessive pressure on the inserts during use. WARNING: Exclusively use original MECTRON inserts, accessories and spare parts. CAUTION: No modification is allowed to this appliance. WARNING: In case of an adverse event and/or accident attributable to the device during correct use and in accordance with the intended use, a report must be made to the Competent Authority and manufacturer indicated on the product label.
5
multipiezo 1.5
Symbols Symbol
0051
Description Device compliant with Regulation (EU) 2017/745. Notified body: IMQ S.p.A.
Symbol
Nemko brand Compliance with UL CSA regulations
Medical device
Caution! Read the instructions for use
Operating instructions
Manufacturer
Date of manufacture
Serial number
Batch number
Product code
Not sterile
The sterilisable materials must be autoclaved and can resist up to a maximum temperature of 135°C
“B” Part applied according to regulation EN 60601-1
Alternating current
Connection of the control pedal I
6
Description
Power switch set to “on”
Generic warning signal a)
0
Power switch set to “off”
The appliance and its accessories must not be disposed of or treated as municipal solid waste
Temperature limits for transport and storage
Moisture limits for transport and storage
Atmospheric pressure limits for transport and storage
Identification Data Symbol
Description
Symbol
QTY.1
Quantity in package: 1
Rx only
Table 1 – Symbols a) The symbol is represented by a yellow triangle and black graphic symbol.
NOTE: For other symbols, refer to Chapter 12.1 on page 67.
IDENTIFICATION DATA
A correct description of the model and of the serial number of the appliance will allow the After Sales Service to provide fast and
2.1
CAUTION: US federal law restricts sale to orders by licensed dentists or dental hygienists only.
effective answers. Always provide this information every time that you contact MECTRON After Sales Service.
Appliance ID Plate
Each appliance is equipped with an identification plate indicating the main technical characteristics and serial number. The identification plate is located under the appliance. The complete technical specifications are provided in Chapter 11 on page 58. NOTE: The complete list of symbols is provided in Chapter 1.5 on page 6.
manufacturer Mectron S.p.A. Via Loreto 15/A 16042 Carasco -GE- Italy
multipiezo 100-240 V~ 90 VA - 50/60 Hz.
00000000
106000000
0051
YYYY-MM-DD made in Italy
Bar Code HIBC 128
Rx Only
2
Description For US market only
Additional symbols and characteristics of the device are provided on a separate plate. This identification plate is located under the appliance. NOTE: The complete list of symbols is provided in Chapter 1.5 on page 6.
multipiezo Device for intermittent operation: 55 s ON - 30 s OFF with irrigation 30 s ON - 120 s OFF without irrigation
7
multipiezo Handpiece Identification Data
The handpiece name, MECTRON logo, serial number, product code, HIBC matrix code, and a series of symbols are laser engraved on the scaler handpiece.
MANIPOLO ABLATORE SLIM
114000000
0476 0051
YYYY-MM
03120000
MANIPOLO ABLATORE SLIM
NOTE: The complete list of symbols is provided in Chapter 1.5 on page 6.
114000000 03120238
107000000
YYYY-MM
03120238
0476 YYYY-MM 0051
0476 0051
2.2
MANIPOLO ABLATORE SLIM
114000000 03120238 YYYY-MM
Data Matrix HIBC 128
2.3
Inserts Identification Data
The name of the insert itself (Reference 1), MECTRON logo (Reference 2), and insert batch number (Reference 3) are laser engraved on each insert.
1
8
2
3
0476 0051
MANIPOLO ABLATORE SLIM
Delivery
3
DELIVERY
3.1
List of Components
See the Figure on the inside cover. multipiezo is supplied with the standard equipment (see Table 2 on page 9), a set of variable accessories depending on the configuration and client requests (see Table
3 on page 11), and accessories that can be ordered separately (see Table 4 on page 12). NOTE: Both the items included in the standard supply and all accessories can be ordered separately by the client.
Standard supply Item Device core unit
Code
Description
Ref.
04000100
A
Irrigation tank complete with 03150102 cap
B
Pedal FS-05 with cord and plug
C
02900114 01950014
Grey safety cap
01950015
Blue safety cap
03020082
Grey tank cap
03020081
Blue tank cap
Scaler Handpiece
03120238
MANIPOLO ABLATORE SLIM (SLIM SCALER HANDPIECE)
multipiezo O-ring kit
03020175
Tank safety cap Tank cap
D E F H
Table 2 – Standard supply.
9
multipiezo Accessories available for order with standard supply Item
Torque wrench
Inserts
Code
Ref.
02900137
K10 torque wrench b)
02900074
K6 torque wrench d) b)
02900081
K7 torque wrench e) b)
0296xxxx
"S” series reusable scaler inserts b)
0308xxxx 0305xxxx 0345xxxx 0235xxxx 0299xxxx
10
Description
I
“PE” series reusable scaler inserts b)
“R” series reusable scaler inserts b) “ER” series reusable scaler inserts b)
“E” series reusable scaler inserts b) “D” series reusable scaler inserts b)
0219xxxx
“CM” series inserts b)
reusable
scaler
0313xxxx
“ME” series inserts b)
reusable
scaler
03570004
“ICS” series reusable base insert b)
03590009
"IC1” series tip b)
02900112
IC1 tips kit (5 pcs.) b)
03570001
DB1 reusable base support for inserts b)
03590001
TA12D60 diamond insert b)
03590004
TA12D90 diamond insert b)
03590002
TA14D60 diamond insert b)
03590005
TA14D90 diamond insert b)
J
Delivery Accessories available for order with standard supply Item
Code
Use and Maintenance Manual Information: Documentation
Power supply cable
Online
Description
Ref.
03590007
TA14D120 diamond insert b)
03590003
TA16D60 diamond insert b)
03590006
TA16D90 diamond insert b)
03590008
TA16D120 diamond insert b)
03590010
TF12D60 diamond insert b)
03590011
TF16D60 diamond insert b)
03590012
TF12D90 diamond insert b)
03590013
TF16D90 diamond insert b)
02150273
IT/DE/SV version b)
02150274
EN/FR/ES version b)
K
00050030
Black power supply cable 3X0.75 2m IT c)
L
00050031
Black power supply cable 3X0.75 2m D c)
00050034
Black power supply cable 3X0.75 2m UK c)
00050010
Black power supply cable 3X1 2.5m AU c)
00050032
Black power supply cable 3X0.75 2m CH c)
00050029
Black power supply cable 3X0.75 HG 2m JP c)
00050028
Black power supply cable 3X0.75 HG 2m USA c)
02150650
Table 3 – Accessories available for order with standard supply.
11
multipiezo Accessories available for separate order Item
Code
Description
Ref.
Scaler Handpiece
03120142
MANIPOLO ABLATORE LED (LED SCALER HANDPIECE) b)
M
Front cone for LED handpiece
03020158
Front cone for LED SCALER HANDPIECE b)
N
Light guide kit
02900146
Front cone, light guide and decorative ring for SLIM scaler handpiece b)
P
Front cone without light
03020171
Front cone for LED SCALER HANDPIECE b)
G
Peristaltic pump rotor assembly
03020226001
c)
O
Table 4 – Accessories available for separate order. b) Produced by Mectron.
d) To be used only for ME type inserts.
c) Distributed by Mectron.
e) To be used only for the "Crown Preparation” kit.
The packaging of the appliance cannot undergo strong impacts as contains electronic components, therefore the transport and the storage must be carried out with particular care. Do not stack various boxes in order to avoid pinching the underlying packaging. All the material sent by MECTRON was controlled at the time of dispatch. The appliance is shipped appropriately protected and packaged. Upon receipt of the appliance, check for any
12
possible damage caused during transport and in case any damage and/or defects is found, complain to the transporter. Keep the packaging in case any items need to be sent to an Authorised MECTRON Service Centre and to store the appliance during long periods of non-use. WARNING: Before starting the treatment, always make sure to have a stock of material (handpiece, inserts, wrenches) to use in the event of failures or drawbacks.
Installation
4
INSTALLATION
4.1
First Installation
The appliance must be installed in a suitable place that is convenient for its use. WARNING: The place where the device is installed must meet the requirements set out in Chapter 4.2 on page 13. multipiezo can be purchased ready for use
4.2
or may need to be enabled by typing in an activation key. In case your appliance requires an activation key, the procedures to be followed may vary from country to country. Always refer to your dealer for details.
Safety Requirements During Installation
WARNING: Contraindications. Interference with other equipment. Even if compliant with standard IEC 60601-1-2, multipiezo may interfere with other devices in the vicinity. multipiezo must not be used near or stacked with other appliances. However, if this should become necessary, it is necessary to check and monitor the correct functioning of the appliance in that configuration. WARNING: Contraindications. Interference from other equipment. An electrosurgical scalpel or other electrosurgical units positioned near the multipiezo appliance may interfere with the correct operation of the appliance itself. CAUTION: The electrical system of the places where the appliance will be installed and used must comply with the laws in force and the relative electrical safety regulations. CAUTION: To avoid the risk of electrical shocks, this appliance has to be connected exclusively to a power supply with an earth safety device. CAUTION: Place the appliance in such a way as to always have the power plug easily accessible as it is considered to be a disconnection device.
WARNING: Risk of explosion. The appliance cannot operate in environments where there are saturated atmospheres of flammable gases (aesthetic mixtures, oxygen, etc.). WARNING: Install the appliance in a safe place protected from impact or accidental water or liquid spray. WARNING: Do not install the appliance above or near sources of heat. Arrange during installation for a suitable circulation of air around the appliance. CAUTION: Do not expose the appliance to direct sunlight or sources of UV light. CAUTION: The appliance can be transported but must be handled with care when moved. Position the pedal on the floor so that it can only be activated intentionally by the operator. CAUTION: Before connecting the handpiece to its cord, check that the electrical contacts are perfectly dry, on both parts. If necessary dry them with compressed air. CAUTION: Each tank can hold liquids up to a maximum of 500 ml.
13
multipiezo 4.3
Connecting the Accessories
Connect the pedal to the back of the appliance, on the right-hand side, in the socket marked with the symbol by inserting the pedal cord plug until you hear a ‘click’.
1
CAUTION: Pay attention to the positioning of the pedal, which must be such that the pedal is only activated intentionally by the operator. Insert the supply cable in its connection situated on the back of the appliance. Connect it to the socket in the wall;
2
CAUTION: Position the device in a way that the power plug is easily accessible at all times; this is considered a means of isolation.
Unscrew the cap of the tank and fill it with the desired liquid; CAUTION: Each tank can hold liquids up to a maximum of 500 ml.
14
3
Installation Check that the small tube inside the cap is properly installed, then screw the cap onto the tank;
4
CAUTION: Check that the female coupling of the tank cap is clean and free from obstructions.
CAUTION: Check that the male coupling on the appliance core unit is clean and that its O-rings are not worn. CAUTION: Use the support only to install the Mectron 500 ml tank and to stow the handpiece. Do not use the support for other purposes.
Holding the tank in a vertical position, insert it on the machine core unit of the device until it comes into contact;
5
CAUTION: Do not overturn the tank, the cap is not watertight. The outflow of physiological solution or aggressive liquids may cause damage to the surfaces.
Correctly insert the scaler handpiece onto its cord by matching the alignment notch on the handpiece connector with the alignment key on the cord connector. Check that the electrical contacts of both are perfectly dry and if necessary dry them by blowing with compressed air;
6
15
multipiezo Place the handpiece on its respective support.
7
CAUTION: Use the support only to install the Mectron 500 ml tank and to stow the handpiece. Do not use the support for other purposes.
5
USE
5.1
Power On/Off
Switching the appliance on Switching the appliance on The switch is located on the left-hand side of the machine core unit. Flick the switch into the “I” position, taking care not to press the pedal. On start-up, 4 symbols appear (Ref. 7 inside cover) which will progressively turn off. The system will then show the preset configuration and be ready for use. Switching the appliance off The switch is located on the left-hand side of the machine core unit. Flick the switch into the “O” position, taking care not to press the pedal. The system will turn off. CAUTION: Position the device in a way that the power plug is easily accessible at all times; this is considered a means of isolation.
NOTE: each time the appliance is started up, the default setting is applied: •
“power”: 1;
•
“function”: ENDO;
• •
“light”: AUTO;
“irrigation”: 3.
NOTE: When the appliance is turned on, the bottle lights up.
16