Contents
User Manual
Section
Title
1
Operation
Page
1-1
Memorandum of Registration
4
1-2
Important Information
5
1-3
Introduction
7
1-4
Identifying Controls
8
1-5
Rear Panel Layout
9
1-6
Operating Instructions
10
1-7
Routine Checks
14
1-8
Warranty
16
2-1
Unpacking
18
2-2
Technical Specification
19
2-3
Chemical Data
21
2-4
Dimensions
23
2-5
Installation
24
2-6
Hyper terminal set-up
26
2-7
Operating Process
27
2-8
Troubleshooting
29
2-9
Service Record
31
2-10
Decontamination Certificate
32
Installation
2
Ref: UM SI-SA TOC 2 10/08 Rev.C
2
Section 1: Operation
Section 1-1 - Memorandum of Registration
User Manual
THIS MANUAL IS VALID ONLY FOR MEDISAFE SI-SA ULTRASONIC IRRIGATORS MANUFACTURED AFTER MARCH 1ST 2004 (SERIAL NUMBERS COMMENCING 0304) Note to Person Installing this Product This Memorandum of Registration MUST be completed by the person undertaking installation and witnessed by the owner of the product. Note to Owner of this Product Be sure that this Memorandum of Registration is completed by the person undertaking to install the product. In the event of a Warranty Claim the information contained within this Registration is your evidence that the product has been correctly installed. Important Note: Failure to provide this evidence in the event of a warranty claim may result in the claim being rejected (this does not affect your statutory rights as a customer).
Owner’s Name:
Installation Address:
Installation Date:
Serial Number:
Software Version:
Installed by:
Witnessed by:
Service and Maintenance Requirements: The Medisafe SI-SA Sonic Irrigator should be serviced on a regular basis to ensure that the product continues to perform in accordance with its design parameters. See page 31 for further information.
Ref: UM SI-SA MOR 4 10/08 Rev.C
4
Section 1-2 - Important Information
User Manual
Product Identity
Decontamination: This product MUST be decontaminated in accordance with local regulations before transporting. In the event that the product must be returned to the supplier, it is vital that the Decontamination Certificate (Section 2-10) of this manual is completed and returned with the product.
The model and serial number are located on the silver label to the rear of the unit. They can also be viewed on the printer attached to the machine.The model number can also be viewed from the Front Panel LCD Display for a 5 second period when the unit is first switched on.
Intended Purpose: The product is not suitable for use with flammable materials or gases and should not be used for any purpose other than that defined by the manufacturer as the Intended Purpose (Section 1-3).
Electrical The power cord and connector supplied with this machine are wired according to the following colour code: BLUE BROWN GREEN/YELLOW
Chemicals: This product contains a heating device and the use of substances or chemicals other than those specified may be hazardous. Never use toxic, corrosive, or flammable liquids.
Neutral Live Earth
If the supplied connector is removed, be sure to observe the colour code when reconnecting to an appropriate plug or fused spur.
Read the Chemical Safety Information in Section 2-3 of this manual before use.
For details of fuse replacement refer to Section 1-5: Rear Panel Layout.
Using non-approved chemicals may cause damage to the machine and could void both your warranty and your guarantee of performance.
Important Note: To prevent fire or shock hazard, do not expose the unit to rain or moisture.
Manufacturer Information: Medisafe UK Ltd The Snap Factory Twyford Road Bishop’s Stortford Hertfordshire CM23 3LJ England
Dangerously high voltages are present inside the unit. Do not open the cabinet. Refer servicing and repair to qualified personnel only. Installation and Maintenance: Observe the Routine Checks (Section 1-7), Installation Instructions (Section 2-5) and the Servicing Requirements (Section 2-9) outlined within this manual.
Telephone: Fax: E-mail: Web:
Heat: The temperature within the unit can reach 60oC. during and after operation. Take care when removing items on completion of a cycle and allow good time for the items to cool to a safe temperature before handling.
Ref: UM SI-SA II 5 10/08 Rev.C
5
+44 (0) 1279 461641 +44 (0) 1279 461643 [email protected] www.medisafe.uk.com
Section 1-2 - Important Information
User Manual
IMPORTANT! Extended Periods of Inactivity: If the unit is to be left unused for periods in excess of one month, then the following precautions must be observed to prevent the CPE pumps and the various control valves from sticking. 1.
Enter unit into Test Mode by pressing and
13.
holding the ‘STANDBY’ key until the LCD
After 5 seconds, press ‘STAND-BY’ key to close drain.
display indicates ‘TEST MODE’ and then
2.
release.
14.
Repeat steps 12 & 13 at least 4 times.
Momentarily press the ‘+’ or ‘-’ keys to
15.
Press and hold the ‘+’ or ‘-’ keys until the
scroll through menu and check that the
LCD display indicates ‘COLD IN’
LCD display indicates ‘HOT IN’. 16. 3.
Momentarily press the ‘START’ key and
Momentarily press ‘START’ key to activate cold valve.
observe the rising water level in the wash tank.
17.
After 5 seconds, press ‘STAND-BY’ key to turn off cold valve.
4.
When the tank is approximately ¼ full, press the ‘STANDBY’ key to stop water
18.
Repeat steps 16 & 17 at least 4 times.
19.
Press and hold the ‘+’ or ‘-’ keys until the
from entering tank. 5.
Connect one end of the hose intended to
LCD display indicates ‘DRAIN’.
supply 3E-Zyme® into the connector
Continue holding until wash tank is empty.
marked ‘3E-ZYME®’. 6.
20.
Turn unit off from rear panel.
21.
After 5 seconds, turn unit back on.
Place other end of hose into detergent bottle below fluid level and secure in position.
7.
Press and hold the ‘+’ or ‘-’ keys until the
Failure to observe this procedure could result
LCD display indicates ‘W-CHEM’
in valves remaining open in use which could, in turn, lead to operational errors and possibly
8.
Press and hold the ‘START’
flooding.
key to turn activate wash chemical pump. The 9.
above
procedure
should
also
be
After 10 seconds, press the ‘STANDBY’ key
observed when the unit is likely to be exposed
to deactivate pump.
to ambient temperatures close to, or below, freezing point.
10.
Repeat steps 8 & 9 at least 4 times.
11.
Press and hold the ‘+’ or ‘-’ keys until the
procedure being ignored will NOT be covered
LCD display indicates ‘DRAIN’
under the standard warranty (Section 1-8).
Breakdowns which are attributed to the above
12.
Momentarily press ‘START’ key to open drain.
Ref: UM SI-SA II 6 10/08 Rev.C
6
Section 1-3 - Introduction
User Manual
Service Contracts and Extended Warranty Packages:
Intended Purpose and Performance The Medisafe SI-SA Sonic Irrigator is principally intended to clean hollow (cannulated) instruments, surgical instruments, and associated products. This is achieved through a combination of ultrasonic activity and low-frequency pulsed fluid re-circulation (Cannulated Pulse Enhancement) to dislodge and flush away unwanted debris from both the outside and the inside (hollow instruments only) of items placed in the wash tank.
Your Medisafe SI-SA Sonic Irrigator is supplied with a conditional 12 month parts and labour warranty (Section 1-8) which is effective from the date of installation. Medisafe offer a range of Service Contracts and Extended Warranty Packages which provide continued maintenance and breakdown cover once the original warranty has expired (UK, USA & Germany only).
Washing is achieved using a unique, specially formulated and clinically validated decontamination fluid (3E-Zyme®) which is introduced into the cycle during the program.
Validation Option (UK only): Medisafe can also provide a Validation service which addresses the essential requirements of HTM2030. The work will be performed by an authorised Medisafe engineer and will ensure that your SI-SA Sonic Irrigator continues to conform to HTM2030 guidelines.
If a wash cycle fails to achieve any one of the predetermined parameters then the system reports a failed cycle. The unit registers such a failure on the LCD and the printer, this brings it to the attention of the operator, who is then aware that the cycle has been aborted. The Medisafe SI-SA Sonic Irrigator has been designed to perform within strict and carefully controlled tolerances. A stringent quality control programme ensures that every unit is manufactured and tested under controlled conditions to ensure that all units perform identically. Declaration of Conformity: The Medisafe SI-SA Sonic Irrigator is designed and manufactured in the UK by Medisafe UK Ltd. and complies with the essential requirements of the Medical Devices Directive 93/42/EEC (where appropriate).
Ref: UM SI-SA I 7 10/08 Rev.C
7
Section 1-4 - Identifying Controls
User Manual
Front Panel Controls:
Sonic Irrigator SA
LCD DISPLAY Provides visual indication of operational status
INCREMENT Used to increase values in Set-up menu
MODE Used to access program number and time date setup
DECREMENT Used to decrease values in Set-up menu
ENTER Used to enter changes in Set-up menu
STANDBY Used to reset unit after pausing and to exit Set-up menu
START Used to start cycle or continue after pausing
Ref: UM SI-SA IC 8 10/08 Rev.C
8
Section 1-5 - Rear Panel Layout
User Manual
Rear Panel Layout
SERIAL DATA PORT
FUSES Mains supply fuse (32mm cartridge type) 8AT 230V 12.5AT 110V
RS232 data output to printer or PC
Power Cord
Fuse Replacement: The SI-SA comes fitted with with plug and a 13 amp (13A) fuse. If you need to replace the fuse, use only those rated at 13 amp (13A) and ASTA approved to BS1362. To change the fuse follow the steps indicated below, 1. 2. 3. 4. 5.
Ensure that power supply to machine is disconnected. Using a screwdriver, push the centre section inwards whilst turning anti-clockwise. Remove centre section complete with fuse. Substitute damaged fuse with one of identical type and rating as stated above. Replacement is a reversal of the above procedure.
Ref: UM SI-SA RPL 9 10/08 Rev.C
9
Section 1-6 - Operating Instructions
User Manual
Operating a Wash Cycle
Changing Programme Parameters
Although your Medisafe SI-SA Sonic Irrigator is simple to operate, it is advisable to read the following information carefully before attempting to operate the product. Please pay particular attention to the following points:
Various parameters relating to the wash cycle may be programmed by the user. Details of these parameters and how to adjust them can be found on pages 12 & 13 of this manual. Note: The performance of this product is validated on the factory default settings at the time of installation.
It is essential that the checks and procedures detailed in Section 1-7, are performed before EVERY operation. Your own personal safety is important and it must be remembered that contaminated surgical instruments are being handled when loading the unit. Follow the safe-handling recommendations applicable to your facility and the advice offered here on loading a basket.
Failed Cycles In the event of a cycle failing or being aborted, the unit may be reset by momentarily pressing the ‘STANDBY’ key or by removing power from the machine.
Aborting a Cycle
STAND-BY
Occasionally it may be necessary to terminate a wash cycle before the cycle has been completed. This can be achieved by observing the procedure on page 12.
Note: All paused or aborted cycles will be recorded by the printer.
Ref: UM SI-SA OI 10 10/08 Rev.C
10
Section 1-6 - Operating Instructions
User Manual
Operating a Wash Cycle 1.
Load instruments into basket and place basket into wash tank. Ensure that the basket feet (flush ports) are correctly located over docking ports on base of wash tank. Slide basket side ways to lock in position.
2.
Apply power to the product.
3.
Close lid.
4.
Check that the following message is displayed on the LCD DISPLAY:
SA ready... Press START
Important Information on Loading Instruments The main basket is capable of cleaning up to 12 hollow instruments in one session.
5.
When loading instruments into basket be sure to observe the following precautions:
Observe the messages shown on the LCD DISPLAY and follow this procedure: START
Do not drop items into the empty wash tank as this could damage the ultrasonic transducers.
The wash cycle will now commence and proceed automatically as detailed in ‘Operating Process’ in Section 2-7 of this manual.
Do not place items to be cleaned directly into the wash tank or allow them to make contact with the sides of the tank. Do not load basket with instruments pointing at the lid or the level sensors.
6.
Do not overload basket. Always attach tubing to a Distal Connector of suitable size for the instrument being cleaned or to a quick connect port if the instrument has one.
Cycle complete...
Note: The prevailing wash cycle will be aborted if the lid is opened at any time during the cycle.
Always open box jointed instruments. Always ensure that instruments are open.
taps
on
Always dissemble instruments designed to be taken apart
cannulated
which
are
Always connect instruments of similar bore size to the same manifold. Always attach at least 1 leur pipe to each manifold if cannulated instruments are not being cleaned as this will aid water circulation.
Ref: UM SI-SA OI 11 10/08 Rev.C
When the cycle is completed, the following message will be displayed. Open lid and remove load.
11
Section 1-6 - Operating Instructions
User Manual
Changing Programmes
Aborting a Wash Cycle
To select an alternative wash program, or to amend the wash temperature, follow this procedure (see page 13 for parameter ranges).
Press and hold ‘STANDBY’ button
STAND-BY
Press and hold ‘MODE’ button once The following message will be displayed on the LCD DISPLAY
MODE
The parameter to be amended will be displayed on the LCD DISPLAY
Paused: RESET or reSTART
Wash TIME 10 mins
To abort cycle, press ‘STANDBY’ button once
Press and hold INCREMENT or DECREMENT buttons to amend value
STAND-BY
+ To increase number To decrease number
Press ‘ENTER’ button to store amended value ENTER
Press and hold ‘MODE’ button to select next parameter MODE
Press ‘STAND-BY’ button to save changes and exit set-up menu
STAND-BY
Ref: UM SI-SA OI 12 10/08 Rev.C
12
Section 1-6 - Operating Instructions
User Manual
The following parameters are adjustable within the range indicated below:
#
Parameter
Increment
Min
Max
1
Wash Time
minutes
5
25
2
Temperature
Degrees C.
20
60
3
Set Time (hours)
n/a
n/a
4
Set Time (minutes)
n/a
n/a
5
Set Date (day)
n/a
n/a
6
Set Date (month)
n/a
n/a
7
Set Date (year)
n/a
n/a
Ref: UM SI-SA OI 13 10/08 Rev.C
‘INCREMENT’ key to increase ‘DECREMENT’ key to decrease ‘MODE’ key to select next parameter ‘ENTER’ key to save changes
13
Section 1-7 - Routine Checks
User Manual
Routine Checks
Daily Checks
To ensure that your Medisafe SI-SA Ultrasonic Irrigator continues to operate within its design parameters it is recommended that the following checks are performed on a regular basis. Consulting the troubleshooting guide on page 27 of this manual may prove useful when conducting the checks.
The following checks should be performed daily • Check cleanliness of all external surfaces The outside of the unit (including lid seal) may be cleaned by wiping with a decontaminant wash.
Note: The completion of these checks is not a substitute for regular servicing and maintenance by a suitably qualified engineer.
• Check lid and lid seal Inspect the lid seal for any signs of damage. Ensure that lid can be opened and closed without restriction.
Before every cycle Perform the following procedures before each wash cycle:
• Check that the wash tank is clean and free from any deposit The wash tank may be cleaned by wiping with a decontaminant wash or suitable disinfectant.
• Replace Disposable filter in wash tank Carefully remove the small disposable filter located within the drain inside the wash tank. Replace filter with a new item.
• Check condition of hoses connecting basket to ports in wash tank Ensure that hoses are not kinked or ruptured. • Check condition of basket docking ports Check that the two docking ports located on the bottom of the wash tank are undamaged and free of debris.
• Check Metal Filter(drain) in wash tank Clean metal filter on drain if required • Wipe ALL water level sensors Using a soft nylon brush or non-greasy alcohol wipes, clean the three plastic water level sensors located in the wash tank to remove all traces of debris, and then dry thoroughly.
Disposable Filter Metal Filter (Drain)
• Check level of detergent (3E-Zyme®) Visually inspect fluid bottles. If low, then replenish as necessary.
Docking ports
Ref: UM SI-SA RC 14 10/08 Rev.C
14
Section 1-7 - Routine Checks
User Manual
Weekly Checks The following checks should be performed every week: •
Check cleaning efficiency of unit.
•
Place 5 TOSI (or equivalent) test strips in the wash basket in the following positions. Secure in place using an appropriate method, i.e., nylon cable ties.
The following table indicates the target wash parameters which must be met to ensure efficient operation of the unit. The accuracy of these parameters should be monitored using a recent print-out from the Medisafe Thermal Printer (or by viewing the information directly from the Front Panel LCD as the cycle progresses) to compare the data with the target parameters listed below.
Run a wash cycle (refer to p.10-13). Check that test strip is completely cleaned with no red or clear residue remaining. •
Check water supply hoses to unit. Ensure that hoses connecting the unit to the water supplies are not trapped or kinked.
Initial Drain duration Pre-Wash duration Dispensing Chemical duration* De-gas duration Main Wash duration* Main Wash temperature* (default 43oC) Rinse duration Repeat Wash duration* Repeat Wash temperature* Repeat Rinse duration Total cycle time (approx.)
Ensure that hose connectors are not leaking. •
Check accuracy of electronic control systems.
Check mains power cord. Ensure that the mains power cord is not trapped or damaged. Check that connector to the mains supply is secure.
Ref: UM SI-SA RC 15 10/08 Rev.C
20s (min) 3m 5-7s 5m 5-25m 20-60oC 2m 5-25 mins 20-60oC 2m *30-35m
* dependent upon user programming
15
Section 1-8 - Warranty
User Manual
Fair wear and tear of any part, normal maintenance service items, and the parts and any material used in connection with the new Sonic Irrigator.
PRODUCT WARRANTY To ensure continued efficiency and long operational life, your new Sonic Irrigator is covered by a one year warranty/5,000 cycle warranty which is arranged and administered by Medisafe (UK) Ltd. Details of the warranty are explained in this section.
Normal deterioration or damage to plated parts, paint coat, and rubber parts caused by daily wear and exposure to atmosphere and environment. Minor irregularities not affecting quality, performance or function of the new Sonic Irrigator parts thereof (e.g. slight noise or vibration) or which appear only under particular abnormal conditions.
Subject to the following terms and conditions, Medisafe (UK) Ltd. warrants that any item of THE BASIC WARRANTY ITEMS of the unit (hereinafter called “new Sonic Irrigator”) manufactured by Medisafe (UK) Ltd. shall be free from any defects in material and workmanship under normal use and maintenance.
Damage caused by insufficient or improper maintenance (e.g. negligence of the daily, weekly, and monthly periodic inspections and any servicing requirements as described and specified within this manual).
Any item of THE BASIC WARRANTY ITEMS covered by the limited warranty will be repaired or replaced free of charge by Medisafe (UK) Ltd. or any distributor authorised by Medisafe (UK) Ltd. to act as their agent (hereinafter called “authorised Medisafe dealer or distributor”).
Damage caused by using a part or lubricants not approved by an authorised Medisafe dealer or distributor or being repaired or replaced by a person who is not an authorised Medisafe dealer or distributor.
THE BASIC WARRANTY ITEMS
Damage caused by accident, abuse, or negligence of the proper handling as set out in this Owner’s Manual, misuse of the new Sonic Irrigator, use of the new Sonic Irrigator under unusual conditions, and modifications of or to the new Sonic Irrigator or any component/parts thereof not recommended or approved by an authorised Medisafe dealer or distributor.
THE BASIC WARRANTY ITEMS are any items (except THE SPECIFIC EXCLUDED ITEMS) of this new Sonic Irrigator which are warranted by an authorised Medisafe dealer or distributor within 12 months or 5,000 cycles (whichever occurs first) of the WARRANTY START DATE when this new Sonic Irrigator is first installed by or for you (the original purchaser) as set forth on the Memorandum of Registration (Section 1-1, p.3).
Damage caused by external influences such as chemical spillage, accident, fire, and disasters due to human fault or negligence or Act of God.
THE BASIC WARRANTY ITEMS listed below are warranted within 12 months or 5,000 cycles (whichever occurs first) from the WARRANTY START DATE against defects in manufacture.
Damage to the external casing, wash chamber, load carriers, or any part of the new Sonic Irrigator which may come into contact with the load carrier during the process of loading/unloading the new Sonic Irrigator.
- solenoid valves - electric motors and pumps - switches - electronic control circuits - water heating elements - ultrasonic transducers THE SPECIFIC EXCLUDED ITEMS
All charges incidental to breakdowns, such as loss of use of the new Sonic Irrigator, loss of time, inconvenience, telephone, transportation, loss or damage to personal property or loss of revenues, or other matter not specifically included.
THE SPECIFIC EXCLUDED ITEMS are any items of this new Sonic Irrigator which are not included in THE BASIC WARRANTY ITEMS and are therefore not subject to the terms and conditions of the PRODUCT WARRANTY.
Damage caused by incorrect installation of the new Sonic Irrigator which could cause operational failure or impairment either immediately after such installation or at any time within the period of this PRODUCT WARRANTY.
Expendable items which will deteriorate and require replacement with normal use of the Washer (for example: lid seal, water inlet filter element, various types of gaskets and hoses, lamp bulbs (if fitted), and other expendable parts not specifically listed)
Medisafe (UK) Ltd. reserves the right of final decision on all manufacturers’ warranty claims.
Ref: UM SI-SA W 16 10/08 Rev.C
Nothing herein affects the statutory rights of a consumer.
16
Section 2: Installation
Section 2-1 - Unpacking
User Manual
Unpacking
Replacement Parts and Consumables
Carefully unpack your Medisafe SI-SA Ultrasonic Irrigator and retain the carton and all packaging materials. If the unit needs to be returned to the supplier for any reason then the original carton and packaging materials should be re-used **.
A range of replacement parts and consumable items approved by Medisafe for use with the SI-SA Ultrasonic Irrigator may be purchased from your supplier. The use of these compatible Medisafe approved products will ensure the continued and reliable operation of your washer.
The unit is supplied with a range of accessories to enable easy installation and efficient operation. Ensure that all these items are present before attempting to install or operate the unit. Complete the following Enclosures Check List and notify your supplier of any shortages or damaged components as soon as possible.
3E-Zyme® Detergent (1 x 4 litre bottle) Part# MED8035
Enclosures Check List
Thermal Printer Paper (1 x 100m roll) Part# MED1014.1
• • • • • • • •
Additional detergent (3E-Zyme®) fluid and thermal printer paper may be purchased from your supplier using the following order codes:
Disposable Filters (pack 24) Part# MED3079
1 Grey drain hose 1 Blue cold water hose 1 Stainless hot water hose 1 Instrument organiser 1 Sample bottle 3E-Zyme® 1 Chemical intake tube 1 Universal basket 1 Owner’s Manual
Using non-approved chemicals, accessories, and replacement parts may cause damage to the machine and could void both your warranty and your guarantee of performance.
1 Plumbing kit comprising: • • • • • •
12 short hose and leur assys 2 long hose and leur assy 6 ‘Y’ connectors 1 hose adaptor 1 Metal Filter (Drain) Disposable 24 filters
** Before transporting a used Medisafe SI-SA it is vital that all fluids are thoroughly drained. If the unit is being returned to your supplier or Medisafe UK Ltd. then the Decontamination Certificate MUST be completed and returned with the unit. The Certificate can be found in Section 2-10 of this manual.
OPTIONS • Thermal printer • Printer power supply • Printer connection lead
Ref: UM SI-SA U 18 10/08 Rev.C
18
Section 2-2 - Technical Specification
User Manual
Water Supply: Supply Connection (hot): ¾” BSP Male Union Supply Connection (cold): ¾” BSP Male Union Optimum pressure: 1.5 BAR (22psi) Acceptable Pressure Range: 1.5-2.5 bar (21-35psi) hot & cold water Effect of Lower Pressure: increased filling time leading to extended total cycle time Acceptable Flow Rate: not less than 2 litres/min Effect of Reduced Flow: increased filling time leading to extended total cycle time Volume Used per Cycle: Hot:approx 15 litres Cold:approx 50 litres Acceptable Temp. Range: Hot: 55-75oC (with cold tank) * Cold: 5 – 25 * Hardness: not more than 210 ppm
General: Medisafe Part No: Model: Description:
MED11150 230V 50Hz MED11160 110V 50Hz 110V 60Hz - special order only Medisafe SI-SA Ultrasonic Irrigator Medical Device Class I
Intended Purpose: Operating Process:
Internal and external cleaning of surgical instruments, hollow-ware, and associated products Ultrasonic activity
Applied Standards: BS2745: part 1: 1993 BS2745: part 3: 1993 EN60101-1:2001 (2nd Edition) EN61010-2-045 (2001) EN61326: 1997 A1: 1998 EN980 Wash Cycle Parameters: Min. Temp. During Cycle:
Note: * Temperature sensors do not read below 15oC ** Water temperature at start of cycle should be 15oC above the temperature programmed for the wash cycle (wash temp 40oC. Hot water in: 55oC).
determined by temperature of water supply
Max. Temp. During Cycle: ** See Note (Ultrasonic activity also has a heating effect of one to three degrees C per 15 minutes)
Drain: Discharge Orifice Type: 18mm outside dia. unthreaded hose tail ** Drain position: ** Drain entry should be below level of discharge orifice with no part of the drain hose above the orificie (i.e. no loops in the hose)
Vol. of Dispensed Chemicals: 160ml (approx) 3E-Zyme® detergent: Required Services: Electricity Supply: Supply Voltage: single phase 230V or 110V (nominal) AC 50Hz Power Consumption: 0.8KVA max. Termination: 3 pin type earthed plug (fused @ 13A if UK) Lead Length: 1.83m (6 feet)
Ref: UM SI-SA TS 19 10/08 Rev.C
Max. Flow to Drain: Max. Temp. of Effluent:
15 litres/min 50oC
Ventilation: Max. Heat Output of Load: 50oC
19
Section 2-2 - Technical Specification
User Manual
Chemicals: Size and Nature of Containers: 3E-Zyme® Fluid: 4 litre plastic bottle Storage Instructions: 3E-Zyme® Fluid: Refer to Section 2-3 Chemical Data. Safe Handling Instructions: 3E-Zyme® Fluid: Refer to Section 2-3 Chemical Data. Weight: Operational weight (dry): Shipping weight:
51kg 30kg
Dimensions: Wash Chamber Volume:
28,275 cm3
Outer Cabinet Dimensions: 720mm (W) 400mm (H) 380mm (D) with overflow 455mm(D)
Carton Dimensions:
850mm (W) 520mm (H) 720mm (D)
Installed Clearances:
See page 21
Important Note: Dangerously high voltages are present inside the unit. Do not open the cabinet. Refer servicing and repair to qualified personnel only
Ref: UM SI-SA TS 20 10/08 Rev.C
20