Service Manual
906 Pages
Preview
Page 1
P R E L I M I N A R Y
3.MAR.2013 GEHC_FRNT_CVR.FM
GE Healthcare
Vivid S5™ and Vivid S6™ Service Manual Vivid S5™ and Vivid S6™ with software version SW 12.xx (BT12)
Operating Documentation
Part Number: 2421482-100 Revision: 13
P R E L I M I N A R Y GE Healthcare
3.MAR.2013 GEHC_FRNT_CVR.FM
Operating Documentation
Part Number: 2421482-100 Revision: 13
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGAL
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Important Precautions TRANSLATION POLICY
WARNING (EN)
AVERTISSEMENT (FR)
WARNUNG (DE)
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES. • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD. • FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS. CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS. • SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE TRADUIRE. • NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS. • LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES. DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE. • FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN. • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
-
i
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGA
EALTHCARE
DIRECTION 2421482-100, REVISION 13
AVISO (ES)
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN. • NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO. • LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA. ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS. • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO. • NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA. • O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS. ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS. • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO. • NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO PREVIAMENTE ESTE MANUAL. • A INOBSERVÂNCIA DESTE AVISO PODE RESULTAR EM FERIMENTOS NO TÉCNICO DE ASSISTÊNCIA, OPERADOR OU PACIENTE EM CONSEQUÊNCIA DE CHOQUE ELÉCTRICO, PERIGOS DE ORIGEM MECÂNICA, BEM COMO DE OUTROS TIPOS.
AVVERTENZA (IT)
ii
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE. • SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE. • SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO. • NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI. -
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGAL
EALTHCARE
DIRECTION 2421482-100, REVISION 13
HOIATUS (ET)
VAROITUS (FI)
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES. • KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST. • ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST. • KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA, OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE VÕI MUU OHU TAGAJÄRJEL. TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI. • JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA. • ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN. • MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN VAARATILANTEEN VUOKSI. ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ ∆ΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ. • ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙ∆ΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ. • ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ. • ΕΑΝ ∆Ε ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝ∆ΕΧΕΤΑΙ ΝΑ ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK. • EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ, MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
-
iii
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGAL
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS IZLASĪŠANAS UN SAPRAŠANAS. • ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES SNIEDZĒJAM, OPERATORAM VAI PACIENTAM. ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE • NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR KITŲ PAVOJŲ.
ADVARSEL (NO)
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE. • IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT. • MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER. NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU ANGIELSKIM. • JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USŁUGI SERWISOWE WYMAGA UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI, OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TŁUMACZENIA SPOCZYWA NA KLIENCIE. • NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM. • NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOŻE GROZIĆ OBRAŻENIAMI CIAŁA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU ZAGROŻEŃ.
-
v
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGA
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE ETMEYİNİZ. • BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
(JA)
Traditional Chinese
viii
-
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGA
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes. The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
OMISSIONS & ERRORS If there are any omissions, errors or suggestions for improving this documentation, please contact the GE Healthcare Global Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details. Mail the information to: Service Documentation, GE Medical Systems, Israel Ltd. 4 Etgar Street 39120 Tirat Carmel Israel GE Healthcare employees should contact the product's Lead Service Integrator (LSI) to report service documentation issues. DO NOT use TrackWise for this purpose. The LSI will then use the internal problem reporting tool to communicate these issues to the writer.
SERVICE SAFETY CONSIDERATIONS DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury. For a complete review of all safety requirements, see the Chapter 1 Safety Considerations section in the Service Manual.
x
-
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGAL
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS All products and their name brands are trademarks of their respective holders.
COPYRIGHTS All Material Copyright© 2008, 2010, 2011, 2012 and 2013 by General Electric Company Inc. All Rights Reserved.
-
xi
P R E L I M I N A R Y GE H
3/3/13 VS56_PRECAUTIONS_LEGA
EALTHCARE
DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Revision History Revision
Date
Reason for change
0
01. MAY. 2007
For review only - not released.
1
22. OCT. 2007
For review only - not released.
2
04. DEC. 2007
Initial Release.
3
30 JAN. 2008
Chapter 5 complete re-write. Updates throughout (hardware), as applicable.
4
13 APR. 2008
Updates throughout (new mechanics and newly-released parts).
5
30 APR 2009
BT10 Updates. Chapter 8 restructured plus additional procedures. Chapter 9 restructured.
6
16 JUL 2009
Updates to software upgrade procedures, database merge (import/export), diagnostics (InSite ExC). Part numbers updated throughout.
7
July 2010 (Draft 2)
BT11 updates Chapter 3 restructured (printers and peripherals); Chapter 9 restructured Added ECG/Respiratory Option Additional BT11 updates throughout.
8
August 2010
9
September 2010
10
2011-MAR-22
11
2012-JAN-03
Updated document templates throughout
12
2012-AUG-12
Additional BT12 updates throughout
2013-FEB-26
Additional BT12 updates throughout. Includes the following: Addition of DVD Drive as an option; Software installation (and /or loading latest version User Manuals) via USB Flash Drive; Chapter 3 - general update of system installation checks Chapter 4 - Functional Checks - replaced Image Testing section Chapter 8 - update of Replacement Procedures (incl. addition of AC Distribution Box Type-B) Chapter 9 - update of Replacement Spare Parts list.
Parts and replacement procedures updated Additional BT11 updates throughout (includes additional part numbers and updates to software upgrade procedures) Additional BT11 updates throughout Chapter 9 (Renewal Parts) completely restructured and mini-TOC added BT12 updates
13
List of Effected Pages(LOEP)
xii
Pages
Revision
Pages
Revision
Pages
Revision
Title Page
13
4-1 to 4-48
13
10-1 to 10-36
13
Warnings i to xxxvi
13
5-1 to 5-61
13
Back Cover
N/A
TOC vii to xxxviii
13
6-1 to 6-11
13
1-1 to 1-22
13
7-1 to 7-177
13
2-1 to 2-14
13
8-1 to 8-259
13
3-1 to 3-203
13
9-1 to 9-29
13
-
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Table of Contents CHAPTER 1 Introduction Overview... 1 - 1 Purpose of Chapter 1... 1 - 1 Contents in this Chapter... 1 - 1 Service Manual Overview... Contents in this Service Manual... Typical Users of the Basic Service Manual... Vivid S5™ and Vivid S6™ Models Covered in this Manual... Product Description... Overview of the Vivid S5™/Vivid S6™ Ultrasound Scanner... Purpose of Operator Manual(s)...
1-2 1-2 1-3 1-3 1-4 1-4 1-4
Important Conventions... Conventions Used in this Manual... Model Designations... Icons... Safety Precaution Messages... Standard Hazard Icons...
1-5 1-5 1-5 1-5 1-5 1-6
Safety Considerations... 1 - 7 Introduction... 1 - 7 Human Safety... 1 - 7 Mechanical Safety... 1 - 9 Electrical Safety... 1 - 11 Probes... 1 - 11 Peripherals... 1 - 12 Safety and Environmental Guidelines... 1 - 12 Vivid S5™/Vivid S6™ Battery Safety... 1 - 14 Patient Data Safety... 1 - 15 Dangerous Procedure Warnings... 1 - 16 Lockout/Tagout (LOTO) Requirements... 1 - 17
xiii
Table of Contents
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Product Labels and Icons... Universal Product Labels... Labels on Systems with Type-B Power Supply... Labels on Systems with Type-A Power Supply... Universal Label - Vivid S5™ and Vivid S6™Systems... Label Descriptions...
1 - 18 1 - 18 1 - 18 1 - 19 1 - 20 1 - 21
Returning/Shipping Probes and Repair Parts... 1 - 24 EMC, EMI, and ESD... Electromagnetic Compatibility (EMC)... CE Compliance... Electrostatic Discharge (ESD) Prevention... General Caution...
1 - 25 1 - 25 1 - 25 1 - 25 1 - 26
Customer Assistance... 1 - 27 Contact Information... 1 - 27
xiv
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
CHAPTER 2 Site Preparations Overview...2 - 1 Purpose of Chapter 2... 2 - 1 Console Requirements...2 - 2 Unit Environmental Requirements... 2 - 2 Cooling Requirements... 2 - 2 Lighting Requirements... 2 - 2 Time and Manpower Requirements... 2 - 2 Electrical Requirements... 2 - 3 Vivid S5™ and Vivid S6™ Power Requirements... 2 - 3 Inrush Current... 2 - 3 Site Power Outlets... 2 - 3 Mains Power Plug... 2 - 3 Power Stability Requirements... 2 - 4 EMI Limitations... 2 - 5 Probe Environmental Requirements... 2 - 6 Facility Needs...2 - 7 Purchaser Responsibilities... 2 - 7 Mandatory Site Requirements... 2 - 8 Site Recommendations... 2 - 8 Recommended Ultrasound Room Layout... 2 - 9 Networking Pre-Installation Requirements... 2 - 10 Stand-alone Unit (without Network Connection)... 2 - 10 Unit Connected to Hospital’s Network... 2 - 10 Purpose of the DICOM Network Function... 2 - 10 DICOM Option Pre-Installation Requirements... 2 - 10 Connectivity Installation Worksheet...2 - 11
xv
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
CHAPTER 3 System Setup Overview... 3 - 1 Purpose of Chapter 3... 3 - 1 Installation Reminders... 3 - 2 Average Installation Time... 3 - 2 Installation Warnings... 3 - 2 System Acclimation Time... 3 - 2 Safety Reminders... 3 - 3 Receiving and Unpacking the Equipment... 3 - 4 Unpacking the Cardboard Shipping Carton... 3 - 8 Unpacking the Vivid S5™ or Vivid S6™ with Peripherals... 3 - 8 Unpacking the Vivid S5™ or Vivid S6™... 3 - 8 Verifying the Shipping Carton Contents... 3 - 13 Physical Inspection... 3 - 13 System Voltage Settings... 3 - 13 EMI Protection... 3 - 13 Preparing for Installation... Confirming Customer Order... Regional Language Support Kit... Keypad Replacement and Language Change... Probes... Component Inspection... Damage Inspection Checklist... Front View of the Vivid S5™ or Vivid S6™ Ultrasound Unit... Rear View of the Vivid S5™ or Vivid S6™ Ultrasound Unit... Rear Panel of the Vivid S5™ or Vivid S6™ Ultrasound Unit... System Voltage Confirmation... System Voltage Settings... Ensuring Protection from EMI...
3 - 14 3 - 14 3 - 14 3 - 15 3 - 18 3 - 19 3 - 19 3 - 21 3 - 23 3 - 24 3 - 25 3 - 25 3 - 26
Completing the Hardware Installation... 3 - 27 Connecting Printers... 3 - 28 Printer Connection and Configuration - Overview... 3 - 28 Printer Categories and Outline of Procedure Instructions... 3 - 28 Configuring Printing Buttons P1 and P2... 3 - 28 Setting Printing Orientation (Portrait or Landscape) and Paper Size . 3 - 34 HP470 /HP100 Color Printer... 3 - 36 xvi
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
HP 5400K Deskjet Color Printer... 3 - 37 Configuring the HP 5400K Deskjet Color Printer for Connection via USB Port... 3 - 38 Setting the HP 5400K Deskjet Color Printer as the Default Report Printer... 3 - 41 HP Officejet Pro 8000 Color Printer... 3 - 43 Connecting the HP Officejet Pro 8000 Directly to the Vivid S5™/ Vivid S6™ System... 3 - 43 Connecting the HP Officejet Pro 8000 to a Local Area Network (LAN)... 3 - 46 HP Laserjet CP2025/M451 Color Printer... 3 - 53 Peer-to-Peer Configuration... 3 - 54 Installing the HP Laserjet CP2025 Color Printer on a Network . . 3 - 60 Sony UP-D897 B/W Video Printer... 3 - 60 Configuring the Sony UP-D897 B/W Video Printer... 3 - 62 Sony UP-D25MD Color Video Printer... 3 - 65 Connecting the Color Video Printer... 3 - 66 Connecting Peripherals... 3 - 67 Wireless Network Adaptor... 3 - 67 MOD Drive... 3 - 68 External USB Hard Disk... 3 - 68 USB Memory Stick... 3 - 69 Footswitch... 3 - 69 Connecting Probes... 3 - 72 Available Probes... 3 - 74 Connecting the ECG... 3 - 76 Connecting the Unit to a Power Source... 3 - 78 Voltage Level Checks... 3 - 78 Connecting the Ultrasound Unit to the Electrical Outlet... 3 - 78 Disconnecting the Ultrasound Unit from the Electrical Outlet... 3 - 81 Switching the System ON/OFF... 3 - 82 Switching the System ON... 3 - 82 Switching the System OFF... 3 - 82 Service Dongle - Technical Support Mode... 3 - 83 Entering Technical Support Mode During Boot-up... 3 - 83 Entering Technical Support Mode when System is Powered ON - Procedure 1 3 - 84 Entering Technical Support Mode when System is Powered ON - Procedure 2 3 - 84 Configuration...3 - 85 Overview... 3 - 85 Vivid S5™ and Vivid S6™ Configuration... 3 - 86 Enter Location... 3 - 86 Date and Time Adjustments... 3 - 87 Language Selection... 3 - 88 Units of Measure... 3 - 90 xvii
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Service Screen Set-up... 3 - 91 Overview... 3 - 91 Open Service Screen... 3 - 91 Adjust Keyboard Backlight Intensity... 3 - 92 LCD Display Adjustment and Calibration... 3 - 92 Keyboard Setup... 3 - 93 Regional Options... 3 - 94 Languages... 3 - 95 Advanced... 3 - 96 Add Printer... 3 - 97 Configuring Peripherals... 3 - 98 Approved Peripherals... 3 - 98 Network Peripherals (Optional)... 3 - 98 Virtual Printing (Optional)... 3 - 99 Software Options Configuration... 3 - 101 Software Option Installation... 3 - 101 On-site Configuration... 3 - 101 Remote Check and Configuration... 3 - 102 Configuration for Network InSite Checkout... 3 - 102 Troubleshooting... 3 - 107 Basic Connectivity... 3 - 107 Prerequisites... 3 - 107 Flow Chart... 3 - 108 Connectivity Setup... 3 - 109 Connectivity Introduction... 3 - 109 The Dataflow Concept... 3 - 109 Dataflow Naming Convention... 3 - 113 Dataflow Examples... 3 - 114 Stand-alone Vivid S5™ or Vivid S6™... 3 - 114 “Sneaker Net” Environment... 3 - 114 Direct Connection from Vivid S5™ or Vivid S6™ to an EchoPAC PC Workstation... 3 - 115 Vivid S5™ or Vivid S6™ and a DICOM Server in a Network... 3 - 115 Physical Connection... 3 - 116 Ethernet Switch Connections... 3 - 116 Local Network Connection to EchoPAC PC Workstation... 3 - 116 Hospital Network Connection to EchoPAC PC Workstation... 3 - 117 Wireless Network Configuration... 3 - 118 Configuring the Wireless Network Adaptor... 3 - 118 Stand-alone Vivid S5™ or Vivid S6™... 3 - 121 “Sneaker Net” Environment... 3 - 121 Connection from Vivid S5™ or Vivid S6™ to an EchoPAC PC Workstation... 3 - 121 Connection between a Vivid S5™ or Vivid S6™ and a DICOM Server on a Network... 3 - 121 xviii
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Connectivity Configuration... 3 - 122 Introduction... 3 - 122 Select TCP/IP Set-up Screen... 3 - 122 Set the Scanner’s Network Information... 3 - 123 Set the Remote Archive’s Network Information... 3 - 123 Vivid S5™/ Vivid S6™ Remote Path Configuration... 3 - 124 Configuring the Vivid S5™ / Vivid S6™ Remote Path... 3 - 124 Configuring the Repository Remote Path... 3 - 125 Configuring the Vivid S5™ or Vivid S6™ and Network PC for a Peer-to-Peer Connection... 3 - 125 Configuring the Vivid S5™ or Vivid S6™ and Network PC for a LAN Connection... 3 - 128 Set Up Connection to a DICOM Server in a Network... 3 - 133 Overview... 3 - 133 DICOM Server IP Address Setting on the Scanner... 3 - 133 Verify the Connection to a Device... 3 - 137 Setup Connection to HL7 Server in a Network... 3 - 138 Introduction... 3 - 138 Preparations... 3 - 138 Configuration of Modality Worklist Services... 3 - 138 Query/Retrieve (Q/R) Setup... 3 - 142 Overview... 3 - 142 Query/Retrieve Setup on the Vivid S5™ or Vivid S6™... 3 - 142 Query/Retrieve Verification... 3 - 146 MPEGVue Export Configuration... 3 - 149 Introduction... 3 - 149 Preparations... 3 - 149 Configuring the Vivid S5™ or Vivid S6™ for MPEGVue Export... 3 - 149 Exporting to MPEGVue... 3 - 152 eVue Dataflow Configuration... 3 - 154 Introduction... 3 - 154 Preparations... 3 - 154 Configuring the Vivid S5™ or Vivid S6™ for eVue Dataflow... 3 - 154 Using eVue... 3 - 156 Using MPEGVue/eVue on a Remote PC... 3 - 158 Installing MPEGVue on a Remote PC... 3 - 158 Manually Setting the Network PC for eVue Import... 3 - 159 Manually Stopping the Firewall on the Network PC... 3 - 166 Un-installing MPEGVue on a Remote PC... 3 - 171 Using MPEGVue Viewer on a Remote PC... 3 - 174 Reading an MPEGVue Exam... 3 - 174 MPEGVue Patient List... 3 - 174 Selecting an Image... 3 - 175 Stopping & Running the Cine-Loop... 3 - 176 Selecting a Different Patient... 3 - 176 Exporting Images or Loops from MPEGvue... 3 - 176 xix
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Performing Measurements on MPEGVue Images... Saved Worksheets... Reports (.CHM files)... Using MPEGVue over E-mail... Managing and Viewing Exams Received... Get New Exams by E-mail... Using MPEGVue Player...
3 - 176 3 - 177 3 - 177 3 - 177 3 - 182 3 - 184 3 - 195
Storing and Transporting the Unit... 3 - 199 Safety Precautions for Moving the Vivid S5™ or Vivid S6™ Unit... 3 - 199 Cardboard Shipping Carton and Packaging Materials... 3 - 204 Completing the Installation Paperwork... Installation Acceptance Test Criteria... System Installation Details... Product Locator Installation... User Manual(s)...
xx
3 - 205 3 - 205 3 - 205 3 - 205 3 - 206
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
CHAPTER 4 Functional Checks Overview...4 - 1 Purpose of Chapter 4... 4 - 1 Specific Equipment Required... 4 - 1 General Procedures...4 - 2 Overview... 4 - 2 Service Dongle (HASP)... 4 - 2 Exit to Windows from the Vivid S5™ or Vivid S6™ Application... 4 - 4 Power ON/Boot-up... 4 - 4 Connecting Mains Power to the Vivid S5™ or Vivid S6™ Ultrasound Unit 4 - 4 Turning the System ON... 4 - 7 Power Shut Down... 4 - 9 Standby Mode... 4 - 10 Complete Power Down... 4 - 11 Log On to the System as ADM... 4 - 12 Using Removable Media... 4 - 13 Introduction... 4 - 13 Install MO Disk in MO Drive... 4 - 14 Eject and Remove MO Disk from MO Drive... 4 - 14 Install CD-R Disk in CD Drive... 4 - 14 Eject and Remove CD-R Disk from CD Drive... 4 - 15 Labeling Removable Media... 4 - 15 Formatting Removable Media... 4 - 15 DICOM Repair of Removable Media... 4 - 16 Archiving and Restoring System Configuration... 4 - 16 Backing Up System Configuration to Removable Media... 4 - 16 Restoring System Configuration from an MO Disk... 4 - 17 Functional Checks...4 - 18 Basic Controls... 4 - 18 Alphanumeric Keyboard Test... 4 - 18 Extended Keyboard and Trackball Test... 4 - 18 System SN Test... 4 - 21 Speakers Tests... 4 - 21 Light Detector Test for Vivid S6™ only... 4 - 22 Monitor Test... 4 - 22 ECG/Respiratory Test... 4 - 23 Front End Tests... 4 - 26 Visual Test... 4 - 32 xxi
P R E L I M I N A R Y GE HEALTHCARE DIRECTION 2421482-100, REVISION 13
VIVID S5™ AND VIVID S6™ SERVICE MANUAL
Peripherals... B/W and Color Printer Test... Color Printer Test... External USB Hard Disk Test... Mechanical Functional Tests... Arm Movement Test... Operator Panel Movement Test... Monitor Movement Test... Front Wheel Function Test... Rear Wheel Function Test... Back End Processor Tests... USB Test... Standby Test...
4 - 33 4 - 33 4 - 33 4 - 34 4 - 35 4 - 35 4 - 35 4 - 36 4 - 36 4 - 36 4 - 37 4 - 37 4 - 37
Image Testing... 4 - 38 2D Mode (B Mode) Checks... 4 - 38 Introduction... 4 - 38 Preparations... 4 - 38 Adjust the 2D Mode Controls... 4 - 39 M Mode Checks... 4 - 39 Preparations... 4 - 40 Adjust the M Mode Controls... 4 - 40 Color Mode Checks... 4 - 40 Introduction... 4 - 40 Preparations... 4 - 40 Select Color 2D Mode... 4 - 40 Adjust the Color 2D Mode Controls... 4 - 40 Select Color M Mode... 4 - 41 Adjust the Color M Mode Controls... 4 - 41 PW/CW Doppler Mode Checks... 4 - 42 Introduction... 4 - 42 Preparations... 4 - 42 Adjust the PW/CW Doppler Mode Controls... 4 - 42 Tissue Velocity Imaging (TVI) Checks... 4 - 43 Introduction... 4 - 43 Preparations... 4 - 43 Adjust the TVI Controls... 4 - 43 Probe/Connectors Check... 4 - 44 System Turnover Checklist... 4 - 45 Software Configuration Checks... 4 - 45 Site Log... 4 - 46
xxii