Quick Guide ATMOS® C 341 Battery This Quick Guide does not replace the operating instructions.
For your safety • Recharge the battery at the latest after 3 months, even if you do not use the device.
• The device may only be used by persons who were medically trained, and were trained in the medical suction.
• Perform a function check after each use. Perform a function check every 4 weeks in case you do not use the device for a longer period.
• Do not operate the device without a bacterial filter.
Device overview 1 4 5 2
6
3
7
1 Control panel 2 Battery compartment cover 3 Device base
8
12
9
11
4 Release button wall and
10 8 Display of battery status 9 Vacuum controller
device support 5 Hose rewind with suction hose 6 Connection charging accessories 7 Guide for the wall and device support
10 Vacuum display 11 On/off button
12 Power indicator
Display of the battery status • The following display values are not valid during battery charging. • The battery status must be supervised visually. • If only the red LED lights up, although the battery has been loaded for a longer period, the battery is defective.
≤ 100 % ca. 70 %
Support Mounting
Removal
ca. 40 % < 10 %
Switching on and operate Switching on
Vacuum adjustment
Switching off
1. Press the on/off button.
2. Turn the vacuum controller for adjusting the required vacuum.
3. Press the on/off button for at least 1 second.
»» The pump starts. »» The LED above the on/off button is illuminated as long as the device is switched on.
»» The vacuum display shows the current suction performance.
Cleaning and disinfection
After each patient / after each suction
Daily*
S
Sterilization
Cleaning
Parts to be reprocessed
D
Manual
When Every 14 days*
How C
5. Perform a function check after each cleaning and disinfection.
Function check
Disinfection
What
4. Clean and disinfect the device after every use.
1. Please check whether the following parts are damaged or torn: -
all hoses
-
Canister system
2. Switch on the device. 3. Check: LEDs and battery status. 4. Connect a fingertip to the suction hose and close the auxiliary air vent.
Surfaces Housing
X
X
X
X
Device base
X
X1
X
X
Hose rewind
X
X
X
X
Wall and device support
X
X
X
Power supply and recharging unit
X
X
X
DDS secretion canister
X
X1
X2
X
X
Outer canister lid (DDS)
X
X
1
X
2
X
X
Inner canister lid (DDS)
X
X1
X2
X
X
Float ball (DDS)
X
X1
X2
X
X
Filter holder
X
X1
X2
X
X
Sealing ring
X
X
X
X
1
5. Close the front opening of the fingertip with your thumb. 6. Select the vacuum -0.5 bar. 7. Please check whether the device reaches the vacuum after approx. 20 seconds: The pump switches off and the vacuum display shows -0.5 bar. 8. In case the device does not reach the vacuum within 20 seconds: Check the device on potential sources of error.
Canister system
Bacterial filter
X
Fingertip Support for canister system Suction hose (DDS)
X
X
X
X
X
X
X
1
X
2
X
X
X
X1
X2
X
X
X
X1
1) Preferred: machine cleaning and disinfection in the washer disinfector (rinse cycle up to 95°C) 2) If required hot-steam sterilization at 134° C with devices acc. to EN285
GA2GB.320101.0 2017-03 Index: 01
1