DP-5/DP-7 Series
Digital Ultrasonic Diagnostic Imaging System
Service Manual
Revision 13.0
Table of Content Table of Content ... i Revision History... I Intellectual Property Statement ... II Applicable for ... II Statement... II Responsibility on the Manufacturer Party ... III Customer Service Department ... III 1
Preface ... 1-1 1.1 1.2
Meaning of Signal Words ... 1-1 Meaning of Symbols ... 1-1 1.2.1 Meaning of Safety Symbols... 1-1 1.2.2 Warning Labels ... 1-2 1.2.3 General Symbols ... 1-2 1.3 Safety Precautions ... 1-3 1.3.1 Electric Safety ... 1-3 1.3.2 Mechanical Safety ... 1-4 1.3.3 Personnel Safety ... 1-5 1.3.4 Others ... 1-5
2
Specifications ... 2-1 2.1
Overview ... 2-1 2.1.1 Intended Use ... 2-1 2.1.2 Introduction of Each Unit ... 2-1 2.1.3 Peripherals Supported ... 2-6 2.2 Specifications ... 2-6 2.2.1 Dimensions and Weight ... 2-6 2.2.2 Electrical Specifications ... 2-6 2.2.3 Environmental Conditions ... 2-7 2.2.4 Monitor Specification ... 2-7
3
System Installation ... 3-1 3.1
Preparations for Installation ... 3-1 3.1.1 Electrical Requirements ... 3-1 3.1.2 Installation Condition ... 3-2 3.2 Unpacking ... 3-3 3.2.1 Unpacking... 3-3 3.2.2 Checking... 3-4 3.3 Installation of Main Unit ... 3-4 3.3.1 Display Adjusting ... 3-4 3.3.2 Connecting the Power Cord ... 3-5 3.3.3 Installing Transducer& Gel Holder ... 3-6 3.3.4 Connecting the Transducer ... 3-6 3.4 Installing Peripherals ... 3-7 3.4.1 Footswitch Installation ... 3-7 3.4.2 Installing a Graph / Text Printer ... 3-7 3.4.3 Video Printer Installation ... 3-7
i
3.5
4
System Configuration ... 3-8 3.5.1 Running the System ... 3-8 3.5.2 Enter Doppler ... 3-8 3.5.3 System Preset ... 3-9 3.5.4 Print Preset ... 3-11 3.5.5 Network Preset ... 3-13 3.5.6 System Information ... 3-15
Hardware Principle ... 4-1 4.1 4.2
General Structure of Hardware System ... 4-1 Main Unit Module ... 4-2 4.2.1 Probe Board ... 4-2 4.2.2 Probe Connecting Board ... 4-4 4.2.3 Main board ... 4-4 4.2.4 IO Interface Board ... 4-7 4.2.5 IO Rear board ... 4-8 4.3 Power supply Module ... 4-9 4.3.1 Power Output of the Power supply module and Supporting Function Distribution ... 4-9 4.3.2 System Power-on Control ... 4-10 4.4 Monitor ... 4-13 4.5 Control Panel ... 4-14 4.6 Ultrasound System Monitoring ... 4-15 4.7 Ultrasound System Indicator ... 4-16
5
Function Checking and Testing ... 5-1 5.1
Preparation... 5-1 5.1.1 Personnel ... 5-1 5.1.2 Tool ... 5-1 5.2 Checking System Status ... 5-1 5.2.1 System Running Status ... 5-1 5.2.2 System Running Status ... 5-1 5.3 General exam ... 5-2 5.3.1 Check Flow ... 5-2 5.3.2 Checking Content ... 5-2 5.4 Function Checks ... 5-4 5.4.1 Check Flow ... 5-4 5.4.2 Checking Content ... 5-5 5.4.3 Checking Content ... 5-5 5.5 Performance Test ... 5-10 5.5.1 Test Process ... 5-10 5.5.2 Test Content ... 5-10
6
Software installation and Maintenance ... 6-1 6.1 6.2
Enter the Maintenance Window ... 6-1 System Software Install/ Restore ... 6-2 6.2.1 Restore the Operating System and Doppler ... 6-2 6.2.2 Doppler Restoration ... 6-5 6.3 Installation of Optional Devices... 6-8 6.4 Data Backup and Storage ... 6-9 6.4.1 Manage Settings ... 6-9 6.4.2 Patient Data Backup and Restore ... 6-10 6.5 Software Maintenance ... 6-11
ii
6.5.1 Product Configuration ... 6-11 6.5.2 Log Maintenance ... 6-12 6.6 Display Parameter Setting ... 6-14 6.7 HDD Partition ... 6-14
7
Structure and Assembly/Disassembly ... 7-1 7.1 7.2 7.3
Structure of the Complete System ... 7-1 Field Replaceable Unit ... 7-2 Preparation... 7-19 7.3.1 Tools Required ... 7-19 7.3.2 Engineers Required ... 7-19 7.3.3 Preparations for Installation ... 7-19 7.4 Assembly/Disassembly ... 7-19 7.4.1 Monitor assembly ... 7-20 7.4.2 IO Rear board ... 7-23 7.4.3 Machine Fan ... 7-23 7.4.4 Assembly of the Power Module ... 7-24 7.4.5 Main Unit Box Module ... 7-25 7.4.6 Probe Board Assembly ... 7-29 7.4.7 Control Panel Assembly ... 7-30 7.4.8 DVD Assembly ... 7-34 7.4.9 Other... 7-34
8
System Diagnosis and Support... 8-1 8.1
General Status Indicator ... 8-1 8.1.1 The Status Indicators on the Control Panel ... 8-1 8.1.2 The Status Indicators of the Power on the IO Rear Board ... 8-2 8.2 Starting Process of the Whole System ... 8-3 8.2.1 Start Process of Complete System ... 8-3 8.2.2 The start-up Process of BIOS ... 8-4 8.2.3 The Start-up of Linux ... 8-5 8.2.4 The Start-up of Doppler ... 8-5 8.3 Alarming and Abnormal Information ... 8-8 8.3.1 Turning on the System Configuration File is Abnormal ... 8-8 8.3.2 The voltage of system power is abnormal ... 8-8 8.3.3 Temperature Alarming ... 8-9 8.3.4 Fan Alarming ... 8-9 8.3.5 Battery Alarming ... 8-9 8.3.6 PHV Related Alarming... 8-10
9
Care and Maintenance ... 9-1 9.1
Overview ... 9-1 9.1.1 Tools, Measurement Devices and Consumables ... 9-1 9.1.2 Care and Maintenance Frequency ... 9-2 9.2 Cleaning the System ... 9-3 9.2.1 Flow of Cleaning ... 9-3 9.2.2 Content ... 9-3 9.3 Maintenance Check ... 9-5 9.3.1 System Function ... 9-5 9.3.2 Peripherals and Options Check ... 9-6 9.4 System Maintenance ... 9-7 9.4.1 Electrical Safety Maintenance ... 9-7
iii
9.4.2
Mechanical Safety Maintenance ... 9-7
10 Troubleshooting of Regular Malfunctions ... 10-1 10.1 System cannot be powered on ... 10-1 10.1.1 Module or Board Related ... 10-1 10.1.2 Key Points Supporting Troubleshooting ... 10-1 10.1.3 Troubleshooting ... 10-1 10.2 System Cannot Start up Normally ... 10-2 10.2.1 Module or Board Related ... 10-2 10.2.2 Key Points Supporting Troubleshooting ... 10-2 10.2.3 Troubleshooting ... 10-2 10.3 Image Fault ... 10-3 10.3.1 Module or Board Related ... 10-3 10.3.2 Key Points Supporting Troubleshooting ... 10-4 10.3.3 Troubleshooting ... 10-4 10.4 Probe Socket System Malfunction ... 10-5 10.4.1 Module or Board Related ... 10-5 10.4.2 Key Points Supporting Troubleshooting ... 10-5 10.4.3 Troubleshooting ... 10-5 10.5 IO Interface System ... 10-6 10.5.1 Module or Board Related ... 10-6 10.5.2 Key Points Supporting Troubleshooting ... 10-6 10.5.3 Troubleshooting ... 10-6 10.6 Control Panel ... 10-7 10.6.1 Module or Board Related ... 10-7 10.6.2 Key Points Supporting Troubleshooting ... 10-7 10.6.3 Troubleshooting ... 10-7 10.7 Monitor ... 10-8 10.7.1 Module or Board Related ... 10-8 10.7.2 Key Points Supporting Troubleshooting ... 10-8 10.7.3 Troubleshooting ... 10-8
Appendix A
ELECTRICAL SAFETY INSPECTION ... A-1
Appendix B
Phantom Usage Illustration ... B-1
Appendix C
Requirements of Performance Indices ... C-1
Appendix D
Boot Screen ... D-1
iv
Revision History Mindray may revise this publication from time to time without written notice. Revision
Date
Reason for Change
1.0
2011.10.25
Initial release
2.0
2012.9.24
Update the test items, test methods and limit values of electrical safety inspection.
3.0
2013.04.17
Add FRU part number for HDD of different regions/types (CE/FDA/human/vet) In chapter 7.2: 1. Correct part number of dual-encoder board. 2. Correct part number of IO interface board.
4.0
2013.05.07
3. Replace control panel modules of DP-5/DP-7 by corresponding top cover assembly of keyboard (including photos and part numbers). 4. Delete FRU unit of main control panel PCBA.
5.0
2013.6.24
Add “The attentions to the assembly/disassembly, otherwise the hard disk will be damaged” to Chapter 7.4.9.2
6.0
2015.1.14
Section 7.2, add new main board FRU number and compatibility description.
7.0
2015.7
Section 2.1.3, add SONY UP-D898MD and SONY UP-X898MD video printers
8.0
2016.8
Update the labels in 1.2.2.
9.0
2017.11
In section 7.2, add FRUs: DC-DC board (FRU), AC-DC board (115V FRU), and AC-DC board (230V FRU)
10.0
2018.01
In section 7.3.1 and 7.4.5.1, modify the main unit fan disassembling method
11.0
2018.04
In section 7.2, add FRUs: PC module and HDD; and update information about the printers
12.0
2019.3
Add FRUs: Power Module Assembly and AC-DC board
13.0
2019.08
Add BayTrail PC Module and relevant comments
© 2011-2019 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All Rights Reserved.
I
Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may referring to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. ,
,
,
,
,
BeneView,
WATO, BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners.
Applicable for This service manual is applicable for the service engineers, authorized service personnel and service representatives of this ultrasound system.
Statement This service manual describes the product according to the most complete configuration; some of the content may not apply to the product you are responsible for. If you have any questions, please contact Mindray Customer Service Department. Do not attempt to service this equipment unless this service manual has been consulted and is understood. Failure to do so may result in personnel injury or product damage.
II
Responsibility on the Manufacturer Party Mindray is responsible for the effects on safety, reliability and performance of this product, only if:
All installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel;
The electrical installation of the relevant room complies with the applicable national and local requirements;
The product is used in accordance with the instructions for use.
Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty shall not extend to:
Any Mindray product which has been subjected to misuse, negligence or accident;
Any Mindray product from which Mindray's original serial number tag or product identification markings have been altered or removed;
Any products of any other manufacturers.
WARNING:
It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
Customer Service Department Manufacturer:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address:
Mindray Building,Keji 12th Road South,High-tech industrial park,Nanshan,Shenzhen 518057,P.R.China
Website:
www.mindray.com
E-mail Address:
Tel:
+86 755 81888998
Fax:
+86 755 26582680
III
1
Preface
This chapter describes important issues related to safety precautions, as well as the labels and icons on the ultrasound machine.
1.1
Meaning of Signal Words
In this operator’s manual, the signal words DANGER, WARNING, CAUTION and NOTE are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual.
Signal word
Meaning
DANGER
Indicates death or serious injury may occur imminently in this hazardous situation if not avoided.
WARNING
Indicates death or serious injury may occur potentially in this hazardous situation if not avoided.
CAUTION
Indicates minor or moderate injury may occur potentially in this hazardous situation if not avoided.
NOTE
1.2
Indicates property damage may occur potentially in this hazardous situation if not avoided.
Meaning of Symbols
The meaning and location of the safety symbols and warning labels on the ultrasound machine are described in the following tables, please read them carefully before using the system.
1.2.1
Meaning of Safety Symbols
Symbol
Meaning Type-BF applied part The ultrasound transducers connected to this system are type-BF applied parts. The ECG module connected to this system is Type-BF applied part.
Location Below the IO panel
Caution.
Preface 1-1
1.2.2
Warning Labels
No. 1.
Warning Labels
a
Meaning a. Do not place the device with the mobile trolley on a sloped surface. Otherwise the device may slide, resulting in personal injury or the device malfunction. Two persons are d required to move the device over a sloped surface. b. Do not sit on the device. c. DO NOT push the device when the casters are locked. c d. Caution! please carefully read this manual before use system
b
1.2.3
General Symbols
This system uses the symbols listed in the following table, and their meanings are explained as well. No.
Symbol
Description
Location
1
Equipotentiality
On the power supply panel
2
Power button
On the control panel
3
Network port
4
USB ports
On the IO panel
5
Video
Video output
6
S-Video
S-Video output
7
Remote control port
8
Reserved,Audio output
9
VGA signal output
10
Display brightness adjusting
11
Display contrast adjusting
12
AC Status Indicator
13
Battery Status Indicator
1-2 Preface
On the IO panel
On the nether right corner of the display
On the upper right corner of the control panel
No.
Symbol
Description
14
Standby indicator
15
Hard disk indicator
16 17 18
Probe port A
A
Probe port B
B
Below the probe socket
Probe port C
C
19
Product serial number
20
Manufacture date
21
This product is provided with a CE marking in accordance with the regulations stated in Council Directive 93 / 42 / EEC concerning Medical Devices. The number adjacent to the CE marking (0123) is the number of the EU-notified body certified for meeting the requirements of the Directive.
1.3
Location
Product label
Safety Precautions
Please read the following precautions carefully to ensure the safety of the patient and the operator when using the probes.
DANGER
1.3.1
Do not operate this system in an atmosphere containing flammable or explosive gases such as anesthetic gases, oxygen, and hydrogen because an explosion may occur.
Electric Safety
WARNING:
1. Connect the adapter power plug of this system and power plugs of the peripherals to wall receptacles that meet the ratings indicated on the rating nameplate. Using a multifunctional receptacle may affect the system grounding performance, and cause the leakage current to exceed safety requirements. Use the power cable provided with this system. 2. Before cleaning or disassembling the system, disconnect the power cord from the socket. Otherwise, it may cause electric shock. 3. Make sure all connections are correct before connecting the power supply cable during system maintenance. Otherwise damage may result by hot plug. Preface 1-3
4. Do not use this system simultaneously with equipment such as an electrosurgical unit, high-frequency therapy equipment, or a defibrillator, etc.; otherwise electric shock may result. 5. This system is not water-proof. If any water is sprayed on or into the system, electric shock may result.
CAUTION:
1. DO NOT connect or disconnect the system’s power cord or its accessories (e.g., a printer or a recorder) without turning OFF the power first. This may damage the system and its accessories or cause electric shock. 2. Avoid electromagnetic radiation when perform performance test on the ultrasound system. 3. In an electrostatic sensitive environment, don’t touch the device directly. Please wear electrostatic protecting gloves if necessary, 4. You should use the ECG lead wires provided with the ECG module. Otherwise it may result in electric shock.
1.3.2
Mechanical Safety
WARNING:
1. When this system is moved, please hold the handle. If other parts of the system are held, it may cause damage due to the abnormal force. Do not push the system from the left/right side; otherwise, it may be toppled over. 2. Do not subject the transducers to knocks or drops. Use of a defective transducer may cause an electric shock.
CAUTION:
1. When move the system on the steps, fasten and fully secure any peripheral device before moving the system, please take care to prevent the system from toppling. 2. Do not expose the system to excessive vibration (during the transportation) to avoid device dropping, collision, or mechanical damage. 3. Please install the system on a flat plane with the four casters locked. Otherwise, damage may be resulted by accidental moving. 4. Move the system only when the system is turned off or in standby mode, otherwise damage may result to the hard disk.
1-4 Preface
1.3.3 NOTE:
Personnel Safety 1. 2. 3.
The user is not allowed to open the covers and panel of the system, neither device disassemble is allowed. To ensure the system performance and safety, only Mindray engineers or engineers authorized by Mindray can perform maintenance. Only technical professionals from Mindray or engineers authorized by Mindray after training can perform system maintenance.
1.3.4
Others
NOTE:
For detailed operation and other information about the ultrasound system, please refer to the operator’s manual.
Preface 1-5
2
Specifications
2.1
Overview
2.1.1
Intended Use
The DC-5 Series/DC-7 Series diagnostic ultrasound system is intended for use in clinical ultrasonic diagnosis. DP-5 Series model:DP-5、DP-5T, DP-5E, DP-5S, DP-5B DP-7 Series model:DP-7、DP-7T
2.1.2
Introduction of Each Unit
No.
Name
Function
1.
LCD Display
Displays the image and parameters during scanning (tilt angle adjustable)
2.
Control Panel
Refer to the 2.1.2.3 Control Panel.
3.
Support arm
Used to support the control panel and the display(Only DP-7series support)
Specifications 2-1
No.
Name
4.
Probe and gel holder
Used to place the probe or gel
5.
Handle
Used for pushing when moving the system.
6.
Support arm control handle
Used to adjust the height of support arm (Only DP-7 series support)
7.
DVD-RW
Read and Write DVD drive
8.
Probe ports
Used to connect the probe
9.
Placing table
Used to place objects or video printer
10.
Display support arm
Used to support and adjust the position of display
11.
IO Panel
Interface panel used for inputting and outputting signals, refer to 2.1.2.1 IO Panel.
12.
Power supply panel
Electrical port panel, refer to 2.1.2.2 Power Supply Panel.
13.
Caster
Used to secure or move the system.
14.
Probe cable hook
Used to place and fix the probe cable.
2.1.2.1
IO Panel
No.
Symbol
Function
Function
1
USB ports
2
Separate video output, connecting video printer
3
Composite video signal output, connecting video printer
4
Remote control port
5
Audio output (left channel)
6
Audio output (right channel)
7
VGA signal output
8
Network port
9
1、2、3
2-2 Specifications
Power indicators for 12V, 5V and 3.3V
2.1.2.2
Power Supply Panel <3>
<2>
<1>
No.
Name
Function
1
Equipotential terminal
Used for equipotential connection, that balances the protective earth potentials between the system and other electrical equipment.
2
Auxiliary outlet
Supply power for optional peripheral devices
3
Power inlet
2.1.2.3
Power
AC power inlet
Control Panel <2><3><4><5>
<6>
<7>
<11>
<8> <9> <10>
<12> <13> <14>
<1>
<20> <15>
<18>
<16> <17> <23>
<25>
<27>
<19> <21>
<24>
1> <3
<26>
<28>
<32>
5> <3
<33>
<36>
<3 4>
<22>
<3 7>
<39>
<40>
<38>
<30>
<29>
<41>
<43>
<44>
<45>
<42>
Specifications 2-3
No.
Name
Description
Function Off: when system is turned off;
1
/
Power button
Green: when system is turned on by pressing this button; Orange: when system is in standby.
2
/
Indicator 1
AC power supply indicator
3
/
Indicator 2
Battery status indicator
4
/
Indicator 3
Standby indicator HDD status indicator
Indicator 4
NOTE: DO NOT move the machine when the indicator blinking in green. Otherwise the HDD may be damaged by sudden shake.
Exit
Press to exit the current status to the previous status.
/
Press to open or close the accompanying help documents.
Review
/
Press to review the stored images.
9
Report
/
Press to open or close the diagnosis reports.
10
iStation
/
Press to enter or exit the patient information management system.
11
F1~F4
User-defined key
You can assign a function to the key.
12
Biopsy
/
Press to show or hide the biopsy guide line.
13
Setup
/
Press to open/close the setup menu.
14
Del
/
Press to delete the comment, etc.
15
/
Alphanumeric keys
Same as on PC
16
Menu
Main menu
Press to display or hide a mode-specific parameter menu.
17
Arrow
/
Press to enter or exit the arrow comment status.
18
Cine
Cine Review
Press to enter or exit the Cine Review status.
19
/
Direction key
To adjust LCD brightness or contrast when pressing with <Fn> key.
20
TGC
/
Move to adjust time gain compensation.
5
/
6
Esc
7
Help
8
21 22
Focus Freq./THI Depth Zoom
/ /
Press: to switch between Focus and Freq./THI; Rotate: to adjust corresponding parameter Press: to switch between Depth and Zoom; Rotate: to adjust corresponding parameter
23
Patient
Patient Information
Press to open/ exit patient information screen.
24
Body Mark
/
Press to enter or exit the Body Mark status.
25
Probe
Probe switch
Press to switch Probe and Exam Type
2-4 Specifications
No.
Name
26
Comment
27
End Exam
28
Clear
29 30
31
Description
Function
/
Press to enter or exit the character comment status.
/
Press to end an exam.
/
Press to clear the comments or measurement calipers on the screen.
Cursor
/
Press to show the cursor.
Nav.Rot
/
Multifunction knob
Quad-split screen
Press to enter Quad mode from non-Quad mode;
Quad
Press to switch between windows in Quad mode. Press to enter Dual mode from non-Dual mode;
32
Dual
Dual-split screen
Press to switch between windows in Dual mode.
33
M
/
Press to enter the M mode.
34
B
/
Press to enter the B mode
35
Measure
/
Press to enter/ exit Application Measurement Measurement status: press to switch between the fixed and active end of the caliper;
36
Update
/
37
Caliper
/
38
Gain/ iTouch
/
39
Save 1
/
Press to save, user-defined key
40
Save 2
/
Press to save, user-defined key
41
/
/
Press to confirm an operation
42
/
Trackball
Roll the trackball to change the cursor position.
43
/
/
Press to confirm an operation
44
Freeze
/
Press to freeze or unfreeze the image.
45
/
Press to print: user-defined key.
Multi-imaging mode: press to change the currently active window. Press to enter/ exit general Measurement Rotate: to adjust B or M gain Press: to enter/ exit iTouch
Specifications 2-5
2.1.3
Peripherals Supported Name
Model 971-SWNOM (2-pedal, USB port) 971-SWNOM (3-pedal, USB port)
Footswitch
SONY UP-897MD Black / white video printer
Analog
SONY UP-X898MD Digital
Analog Color video printer
MITSUBISHI P93W-Z
SONY UP-D898MD
SONY UP-20 MITSUBISHI CP910E
NOTE:
If the ultrasound system cannot recognize the SONY UP-X898MD and SONY UP-D898MD printers automatically, you may need to change the settings on the printer: push <PUSH ENTER> to enter the main menu and select [DIGITAL]->[DRIVER], and select [897].
2.2
Specifications
2.2.1
Dimensions and Weight
2.2.1.1
DP-5 Series
Dimension: 1315mm (H) ×670mm (D) ×520mm (W) Net weight: about 55Kg (including battery)
2.2.1.2
DP-7 Series
Dimension: 1265~1415mm (H) ×670mm (D) ×520mm (W) Net weight: about 55Kg (including battery)
2.2.2
Electrical Specifications
2.2.2.1
AC Input
Voltage
220-240V~,100-127V~
Frequency
50/60Hz
Power consumption
420VA
2.2.2.2
Battery (optional)
Voltage
2-6 Specifications
14.8V
Current
2.2.3
6600mAh
Environmental Conditions Operating conditions
Storage and transportation conditions
Ambient temperature
0℃~40℃
-20℃~55℃
Relative humidity
30%~85% (no condensation)
30%~95% (no condensation)
Atmospheric pressure
700hPa~1060hPa
700hPa~1060hPa
Warning :
Do not use this system in the conditions other than those specified
2.2.4
Monitor Specification
2.2.4.1
Main Monitor
Voltage
12V
Dimension
15 inch
Resolution
1024×768
Visible angle
≥170 degree
Specifications 2-7