MINDRAY
DP-8800Plus, DP-9900Plus and DP-9900 Service Manual V3.0
Service Manual
118 Pages
Preview
Page 1
DP-8800Plus/DP-9900Plus/DP-9900
Digital Ultrasonic Diagnostic Imaging System
Service Manual
© 2007-2009 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rent, adaptation and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
,
,
,
,
are the registered trademarks or
trademarks owned by Mindray in China and other countries.
All other trademarks that
appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners.
Responsibility on the Manufacturer Party Contents of this manual are subject to changes without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual. Mindray is responsible for safety, reliability and performance of this product only in the condition that: I-1
• all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel;
• the electrical installation of the relevant room complies with the applicable national and local requirements;
• the product is used in accordance with the instructions for use.
WARNING:
It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
Warranty THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty shall not extend to: z
any Mindray product which has been subjected to misuse, negligence or accident;
z
any Mindray product from which Mindray's original serial number tag or product identification markings have been altered or removed;
z
any product of any other manufacturer.
Return Policy Return Procedure In the event that it becomes necessary to return this product or part of this product to Mindray, the following procedure should be followed: 1.
Obtain return authorization: Contact the Mindray Service Department and obtain a Customer Service Authorization (Mindray) number. The Mindray number must appear I-2
on the outside of the shipping container. Returned shipments will not be accepted if the Mindray number is not clearly visible. Please provide the model number, serial number, and a brief description of the reason for return. 2.
Freight policy: The customer is responsible for freight charges when this product is shipped to Mindray for service (this includes customs charges).
3.
Return address: Please send the part(s) or equipment to the address offered by Customer Service department
Company Contact Manufacturer: Address:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, ShenzhenShenZhen518057, P.R.China,518057
Tel:
+86 755 26582479 26582888
Fax:
+86 755 26582934 26582500
EC-Representative: Address:
Shanghai International Holding Corp. GmbH(Europe) Eiffestraβe 80, 20537 Hamburg Germany
Tel:
0049-40-2513175
Fax:
0049-40-255726
I-3
Safety Precautions 1.
Meaning of Signal Words
In this operation manual, the signal words
DANGER,
WARNING,
CAUTION and NOTE are used regarding safety and other important instructions. The signal words and their meanings are defined as follows. Please understand their meanings clearly before reading this manual. Signal word
Meaning
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
NOTE
2.
Indicates a potentially hazardous situation which, if not avoided, may result in property damage.
Meaning of Safety Symbols Symbol
Description Type-BF applied part Note: All ultrasound transducers that can be connected to this system are Type-BF applied part. "Attention" indicates the points requiring attention. Be sure to read the operation manual concerning these points before using the equipment.
S-1
3.
Safety Precautions
Please observe the following precautions to ensure patient and operator safety when using this system.
DANGER:
Do not use flammable gasses such as anesthetic gas, oxygen or hydrogen, or flammable liquids such as ethanol, near the system, because there is danger of explosion.
WARNING: 1. Do connect the plugs of the system and peripherals connected to the system to the wall receptacle meeting the ratings indicated on the rating nameplate. Using adapter or multi-functional receptacle may affect the system’s grounding performance and thus cause leakage current exceeding safety requirement. Be sure to connect the video printer to the designated AC outlet on the system with the provided cable. Using other cables may cause electric shock. 2. Be sure to connect the potential-equalization lead wire before inserting the system power plug into the receptacle and be sure to remove the system power plug from the receptacle before disconnecting the wire, in order to avoid electric shock. 3. Connect the earth conductor only before turning ON the system. Disconnect the grounding cable only after turning OFF the system. Otherwise, electric shock may result. 4. For the connection of power and grounding, follow the appropriate procedures described in the operation manual. Otherwise, there is risk of electric shock. Do not connect the grounding cable to a gas pipe or water pipe, otherwise functional grounding may not be effective or there may be risk of a gas explosion. 5. Before cleaning the system, be sure to disconnect the power cable to avoid electric shock.
S-2
6. No waterproof device is applied to the system. Do not use the system in any place with the possibility of water ingress. There is risk of electric shock if any water is sprayed on or into the system. If you carelessly spray any water onto/into the system, contact the Mindray sales office, customer service department or representative. 7. Use the transducer carefully. In case that the human body contacts the scratched transducer surface, immediately stop using the transducer and contact the Mindray sales office, customer service department or representative. There is risk of electric shock if using the scratched transducer. 8.Be careful not to let the patient contact the ultrasound equipment. If the ultrasound equipment is defective, there is risk of electric shock. 9. Do not use the transducers other than those provided by Mindray. If a transducer other than those provided by Mindray is connected, the equipment and the transducer may be damaged, causing an accident such as a fire in the worst case. 10. Do not subject the transducers to knocks. Use of defective transducers may cause electric shock. 11. Do not open the shell or panel. If open the shell when the equipment is powered on, there may be a short circuit or electric shock. 12. Do not use the equipment at the same time use equipment such as an electric scalpel, high-frequency therapy equipment or a defibrillator, etc. 13. Precautions during transportation: When moving the equipment, hold the handle. If the user holds other sections, the equipment may be subject to unnatural force and may be damaged. Do not move the equipment in the left/right direction. If the equipment is moved in the left/right direction, the equipment may fall. 14. Accessory equipments connected to the analogue and digital interfaces must comply with relevant IEC standards. Furthermore all configurations should comply with the standard IEC60601-1-1. Everybody who connects additional equipments to the signal input part or signal output part configures a medical system, and is therefore
responsible
that
the
system
complies
with
the
requirements of IEC60601-1-1. If in doubt, consult Mindray customer service department or your local distributor.
S-3
15. Prolonged and repeated use of keyboards can result in hand or arm nerve disorders in some individuals. Observe the local institution work safety/health regulations on keyboard use.
CAUTION: 1.Precautions concerning clinical examination techniques (1) This system can only be used by medical personnel fully trained in clinical examination techniques. (2) This manual does not describe clinical examination techniques. Selection of the proper clinical examination technique must be based on specialized training and clinical experience. 2.Malfunctions due to radiowaves (1) Use of radiowave-emitting devices in the proximity of this kind of medical electronic system may interfere with its operation. Do not bring or use devices which generate radio waves, such as cellular telephones, transceivers, and radio controlled toys, in the room where the system is installed. (2) If a user brings a device which generates radio waves near the system, he must be instructed to immediately turn OFF the device. This is necessary to ensure the proper operation of the system. 3. Precautions concerning installation and movement of the system (1) Be sure to install the system on a level floor and lock the casters. Otherwise, the system may move, and injure the patient. (2) Do not push the system from the sides. If the system is pushed from the sides, it may fall down and cause injury. (3) When the system is moved over a sloped surface, it must be moved slowly by two persons. Otherwise, the system may slide unexpectedly and cause a serious injury. (4) Do not sit on the system. The system may move, causing you to lose your balance and fall. (5) Do not place any objects on top of the monitor. They may fall, causing injury.
S-4
(6) Confirm that the peripheral units are secured before you move the system. Otherwise, the peripheral units may fall and cause injury. (7) When the system is moved over a step, exercise caution to protect the system from falling. When holding the system at the bottom to help move it over a step, take special care to prevent hand injuries. 4.Do not vibrate the equipment excessively (when moving the equipment); otherwise the mechanical parts (such as casters) may be damaged. If the system is often moved on a bumpy floor, contact the Mindray sales office, customer service department or representative. 5. Do not connect this system to outlets with the same circuit breakers and fuses that control the current of devices such as life-support systems. If this system malfunctions and generates an overcurrent, or when there is an instantaneous current at power ON, the circuit breakers and fuses of the building’s supply circuit may be tripped. 6. Always keep the machine dry. Avoid transporting this machine quickly from the cold place to the warm place, otherwise condensation or water drops may be formed, which will cause short circuit. 7. If the circuit breaker is tripped or the fuse is blown, it indicates that the machine or the peripheral devices have problems. In these cases, the user cannot repair by him but contact the Mindray sales office, customer service department or representative. 8. There is no risk of high-temperature burns during routine ultrasound examinations, even if, due to environment temperature and exam modes, the surface temperature of the transducer exceeds the body temperature of the patient. To prevent high-temperature burns, do not apply the transducer to the same region on the patient for a long time. Apply the transducer only for as long as required time for diagnosis. 9.After the sterilization or disinfection of accessories, chemicals must be washed out or gases must be discharged thoroughly from the accessories. Remaining residual chemicals or gases will not only result in damage to the accessories but also can be harmful to human bodies.
S-5
10.Before examining a new patient, press 『Patient』 to delete the patient information and data recorded in the memory for the previous patient. Otherwise, the new data may be confused with the data of the previous patient. 11. Do not pull out plugs of the system and its accessories without turning OFF the power. Otherwise it may result in equipment damage or even electric shock. 12. Do not turn OFF the power supply of the system during printing, saving, or invoking. Otherwise it may result in abnormality of these processes.
NOTE:
1. Do not use the system in the vicinity of strong electromagnetic field (such as the transformer), which may affect the performance of the monitor. 2. Do not use the system in the vicinity of high-frequency radiation source (such as the cellular phone), which may affect the performance of the machine or even lead to failure. 3. To avoid damaging the system, do not use the system in following environment: (1) Locations exposed to direct sunlight (2) Locations subject to sudden changes in temperature (3) Dusty locations (4) Locations subject to vibration (5) Locations near heat generators (6) Locations with high humidity 4. Turn ON the system only after the power has been OFF for more than 10 seconds. If the system is turned ON immediately after being turned OFF, it may result in malfunction of the system. 5. Turn OFF the auxiliary power switch or stop transmission through the [Freeze] key before connecting or disconnecting a transducer. If a transducer is connected or disconnected with an image displayed, the system and/or the transducer may malfunction. 6. After using the transducer, remove the ultrasound gel on it and place the transducer on the transducer holder. Otherwise, water in the gel may enter the acoustic lens, thus adversely affecting the performance and safety of the transducer.
S-6
7. You can record the registration data (including the hospital data and patient data). To ensure the security of the data, be sure to back up the data on external storage media. Data stored in the system may be lost due to improper operation or an accident. 8. Do not apply external force to the control panel (e.g. leaning against it). Otherwise it may damage the system. 9. If the system is used in a small room, the room temperature may rise. Therefore, proper ventilation shall be provided. 10. The fuse, which is inside the machine, can be replaced only by the Mindray service engineer or the technician specified by Mindray. 11. When disposing of this system or any part of the system, contact Mindray sales office, customer service department or representative. Do not dispose of this system without consulting Mindray sales office, customer service department or representative first. Mindray does not assume any responsibility for damage resulting from disposal of this system without consulting Mindray. 12. Deterioration of electrical and mechanical safety characteristics (such as generation of a leakage current or deformation/abrasion of mechanical parts) and of image sensitivity and resolution may occur over a period of time. To ensure system performance, signing a maintenance and service contract to avoid accidents and misdiagnose is recommended. 13. The output power outlets on the main unit are used for providing power to recommended external optional devices. Do not connect other devices to the outlets, otherwise the output power may be exceeded and the system may malfunction.
4. Warning Labels Various WARNING labels are attached to this system in order to call the user's attention to potential hazards. The symbol
on the WARNING labels attached to the system indicates safety
precautions. The name, appearance, and the indication of each WARNING label are described in the Operation Manual. Read them carefully and understand them.
S-7
Contents Chapter 1 1.1.
System Introduction ... 1-1
Introduction of DP-8800Plus ... 1-1
1.1.1.
Appearance ... 1-1
1.1.2.
Rear Panel ... 1-3
1.1.3.
Control Panel ... 1-4
1.2.
Introduction of DP-9900Plus/DP-9900 ... 1-6
1.2.1.
Appearance ... 1-6
1.2.2.
Rear Panel ... 1-8
1.2.3.
Control Panel ... 1-10
Chapter 2
Hardware Introduction... 2-1
2.1.
Hardware System ... 2-1
2.2.
Boards ... 2-1
2.2.1.
Main Board ... 2-1
2.2.2.
Transducer Board ... 2-2
2.2.3.
Connection Board... 2-3
2.2.4.
Output Board ... 2-9
2.2.5.
I/O Board ... 2-12
2.3.
Power Supply Board ... 2-12
2.3.1.
Introduction ... 2-12
2.3.2.
Working Procedure ... 2-13
2.3.3.
Circuit Description ... 2-13
2.3.4.
Protection Circuit ... 2-14
2.3.5.
Maintenance ... 2-15
2.4.
Principle of the LCD ... 2-18
2.4.1.
Operating Principle of the Inverter Module ... 2-18
2.4.2.
Working principle of the regulating board ... 2-20
Chapter 3
Disassembly of DP-9900Plus/DP-9900 ... 3-1
3.1.
System Structure ... 3-1
3.2.
Disassembly of Monitor... 3-2
3.3.
Disassembly of Screen ... 3-3
3.4.
Disassembly of Keyboard ... 3-3 C-1
Contents
3.4.1.
Disassembly of Keyboard Cover ... 3-3
3.4.2.
Disassembly of Trackball... 3-4
3.4.3.
Disassembly of Keyboard Board ... 3-5
3.4.4.
Disassembly of TGC Board ... 3-5
3.5.
Disassembly of CD-ROM ... 3-6
3.5.1.
Locking Casters... 3-6
3.5.2.
Disassembly of Drawer and Transducer Cable Hanger... 3-6
3.5.3.
Disassembly of Rear Cover of Neck... 3-7
3.5.4.
Disassembly of Top Cover ... 3-7
3.5.5.
Disassembly of Left Cover ... 3-8
3.5.6.
Disassembly of Right Cover ... 3-9
3.5.7.
Disassembly of Front Cover ... 3-9
3.5.8.
Disassembly of CD-ROM ... 3-10
3.6.
Disassembly of I/O Board ... 3-11
3.7.
Disassembly of Air Outlet Fan Assembly ... 3-12
3.8.
Disassembly of Hard Disk ... 3-12
3.9.
Disassembly of Power Conversion Board ... 3-13
3.10. Disassembly of Air Intake Fan Assembly ... 3-14 3.11. Disassembly of Output Board, Main Board and Power Supply Board ... 3-15 3.12. Disassembly of Transducer Board ... 3-16 3.13. Disassembly of Cabinet ... 3-16 3.14. Diagram ... 3-17 3.15. Disassembly of the LCD Monitor Assembly ... 3-18 3.16. Replacing the OSD Board of the LCD Monitor... 3-19 3.17. Replacing the inverter board and control board of the LCD ... 3-20 3.18. Replacing the LCD screen of the LCD Monitor ... 3-21 3.19. Disassembly of the LCD support arm assembly ... 3-23
Chapter 4
Disassembly of DP-8800Plus ... 4-1
4.1.
System Structure ... 4-1
4.2.
Disassembly of Monitor... 4-2
4.3.
Disassembly of Screen ... 4-3
4.4.
Disassembly of Printer ... 4-4
4.5.
Disassembly of Keyboard ... 4-4
4.6.
Disassembly of Trackball ... 4-5
4.7.
Disassembly of TGC Board ... 4-5 C-2
Contents
4.8.
Disassembly of Transducer Cable Hanger... 4-6
4.9.
Disassembly of Handle ... 4-7
Chapter 5
Maintenance Requirements ... 5-1
5.1.
Tools and Consumables... 5-1
5.2.
Maintenance Personnel ... 5-2
Chapter 6 6.1.
System Maintenance ... 6-1
Cleaning... 6-1
6.1.1.
Procedure ... 6-1
6.1.2.
Details ... 6-1
6.2.
Software Maintenance ... 6-2
6.2.1.
Software Upgrade ... 6-4
6.2.2.
Installation/Uninstallation of DICOM ... 6-5
6.2.3.
System Self-check... 6-6
6.2.4.
LCD test ... 6-6
Chapter 7
Troubleshooting ... 7-1
7.1.
Power Supply ... 7-1
7.2.
Monitor ... 7-1
7.3.
Troubleshooting of LCD Blank Screen ... 7-2
7.4.
Troubleshooting of Other Malfunctions of the LCD ... 7-2
7.5.
Startup ... 7-3
7.6.
Use ... 7-4
7.7.
Software... 7-5
7.8.
Others ... 7-5
Appendix A Electrical Safety Inspection ... A-1
C-3
Chapter 1
System Introduction
1.1. Introduction of DP-8800Plus 1.1.1. Appearance
<1>
<2> <3>
<5> <6> <7>
<4>
<8> <5>
<9>
<6>
<7>
<10>
Front View
1-1
System Introduction
<11> <12> <13>
<14>
Left View No.
Part
Introduction
<1>
Monitor
Display images and parameters, etc.
<2>
Transducer hanger
<3>
Transducer holder
Place the transducer provisionally
<4>
Printer area
Place the video printer
<5>
Video out
Connect the video printer
<6>
AC out
Connect the AC power cable for the video printer
<7>
Remote
Connect the remote cable for the video printer
<8>
Transducer socket
Connect or disconnect the transducer with the main unit
<9>
Footswitch socket
switch Connect or disconnect footswitch
<10>
Casters
Lock or move the system
<11>
Control panel
User interface
<12>
Handle
Move the system
<13>
CD-RW
Backup data to a CD or read data from a CD
<14>
Power switch
Power on / off
cable Hook for the transducer cable
1-2
System Introduction
1.1.2. Rear Panel
Rear View
<20>
<19>
<17>
<15>
<21>
<20>
<18>
<16>
I/O Panel
No. <15>
Name Video out
Introduction Connect the video printer or other video device
1-3
System Introduction
No.
Name
Introduction
<16>
Graph/text port
printer Connects the graph/text printer
<17>
S-Video port
Video output
<18>
RS-232
Rreserved
<19>
VGA port
VGA output
<20>
USB port
Connects the USB storage device or graph/text printer, etc.
<21>
Ethernet port
Connects network
<22>
AC input
Connects the power cable for the system
<23>
Equipotential terminal
Equipotential connecting
1.1.3. Control Panel < 2>
< 1>
< 3>
< 4>
< 6>
< 5>
< 8>
< 7>
< 9>
< 18 >
< 10 >
< 11 >
< 12 >
IP < 13 >
< 19 >
< 17 >
< 23 >
< 16 >
< 28 >
< 22 >
< 20 >
+
< 24 >
+
ABC
< 25 >
<29>
< 30 >
< 31 >
< 26 >
< 15 > < 21 >
< 14 >
< 32 >
< 27 >
< 34 >
< 33 >
Control Panel
1-4
System Introduction
NO. <1>
Key
Function
Patient
Delete the data for the previous patient, including the ID and measured values, and begin with a new patient.
<2>
Info
The patient information screen appears.
<3>
Preset
Invoke the registered initial settings (presets).
<4>
File
Activate storage or load files system
<5>
Probe
Switch transducers
<6>
ABD/GYN/OB/SML Select exam mode
<7>
Character Key
Used to enter characters and symbols When backlight of SHIFT key is on, the symbol in the upper row on keys can be entered. When pressing CAPS key, the capital letter can be entered.
<8>
TGC
Adjust the ultrasound echo reception sensitivity according to the depth from the body surface.
<9>
Puncture
Enter needle guide menu
<10> F.Posi
Enter the focus position adjusting mode.
<11>
Enter the focus number adjusting mode.
F.Num
<12> A.power
Enter the acoustic power adjusting mode.
<13> IP
Adjust the image quality according to the registered image conditions.
<14> Parameter knob <15> Clear
adjust Adjust some value of Puncture, F.Posi, F.Num, A.Power and IP with combination of corresponding key. Clear comments, body marks, measurements and results on screen
<16> Exit
Exit the current work mode
<17> Arrow
Enter the arrow comment mode
<18> Set
Determine the cursor position for measurement, and confirm the selected items, and adjust the value or items in menus etc.
<19> Change
Change the fixed end or active end in measurement.
<20> Back
Delete the comment or the previous operations, and adjust the value or items in menus etc.
<21> Trackball
Move the cursor or mark during image movement or measurement.
<22> Menu
Open or close the menu on screen.
<23> Measure
Go into measurement mode
<24> BodyMark
Enter body marks mode.
<25> Comment
Go into comment mode
<26> M-Mark
Enter M-Mark mode 1-5
System Introduction
NO.
Key
<27> Depth/Zoom/ Rotation <28> Cine
Function Adjust the viewing depth for display, the zoom mode, and the arrow rotation of the ultrasound images. Switch Manual/Auto CINE review. When the 『Cine』 lamp lights on, the system enter Manual Review mode; when the 『Cine』 lamp is off the system exits Manual Review mode.
<29> Zoom
Enter the mode of amplifying the image
<30> B/M/M+B/B+B
Select image mode
<31> HRev / VRev
Reversal the image horizontally or vertically
<32> Gain
Adjust the sensitivity of black/white images.
<33> Print
Activate a printing function for a video printer connected.
<34> Freeze
Freeze and unfreezes the image. If image is frozen, Output of acoustic power is stopped.
1.2. Introduction of DP-9900Plus/DP-9900 1.2.1. Appearance
Front View
1-6