Users Guide
44 Pages
Preview
Page 1
Dome® Ex Display User’s Guide
© 2014 NDS Surgical Imaging. All rights reserved. Information in this document has been carefully checked for accuracy; however, no guarantee is given to the correctness of the contents. This document is subject to change without notice. Go to web site ndssi.com for the latest documentation. NDSsi provides this information as reference only. Reference to products from other vendors does not imply any recommendation or endorsement. This document contains proprietary information protected by copyright. No part of this manual may be reproduced by any mechanical, electronic, or other means, in any form, without prior written permission of NDSsi. NDS Surgical Imaging, NDSsi, the NDSsi logo, Dome, the Dome by NDSsi logo, Dome E2cHB, Dome E3, Dome E3cHB, Dome E5 are either registered or unregistered trademarks of NDS Surgical Imaging. All other trademarks are the property of their respective owners. 020-0932-00 Rev. D
About the Display The Dome® Ex displays are available in 2, 3, and 5 megapixels and contain a TFT LCD panel. The display’s thin film transistors, in a transmissive-type display, use an integrated cold cathode fluorescent tube (CCFT) backlight system. The display is designed for medical imaging in diagnostic settings, in portrait or landscape orientation, and comes fully tuned with gamma correction that complies with the DICOM Part 14 Standard. The alldigital design enables the display to produce the sharp, crisp images critical to softcopy medical viewing. Unwanted analog display image artifacts are eliminated. For display classification and intended use; environmental guidelines; safety precautions; handling, cleaning, and storage tips; and specifications, see “Technical Information” on page 6.
System Requirements • Windows 7 (x86 or x64) or Windows XP Service Pack 3 (x86 or x64) • Intel or AMD multi-core processor • PCI Express x16 lane slot • 50 MB hard disk space • 512 MB RAM • CD-ROM drive • Power supply, 350 watts or greater
Dome Ex display mounted on desk stand
DC power adapter*
Power cords
DVI cable
DisplayPort cable**
* Some Dome Ex displays ship with a DC adapter that operates per power switch. ** For Dome E3cHB unit only.
2 | Dome Ex Display
Display Components Review this illustration of the back panel to identify controls and ports on the display unit. Also see “Connector Ports” on page 21.
Dome E2cHB, Dome E3, and Dome E5
Dome E3cHB
All Dome Ex displays include the following on the connector plate: • Power input. Drives power to the display; 8-pin port on Dome E2cHB, Dome E3, and Dome E5, and 6-pin port on Dome E3cHB. • Video input. Drives the data to the display. All Dome Ex displays have a standard DVI connector. The Dome E3cHB unit also includes a DisplayPort connector port. • Reset button. Restores the display configuration to default setting. • LED lights. Provides information on the status of the display. NOTE: Usage of the USB-B port on Dome E3cHB to be defined at a future time.
About the Display | 3
Installing the Display
Turn your computer off. Leave the power cord plugged into the grounded outlet. Use the Dome Ex display with the power adapter and video cable shipped.
WARNING No modification of this equipment is allowed. In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States only).
Connecting the Video and Power Cables IMPORTANT! For the Dome E3cHB unit only. Install either the DVI or the DisplayPort video cable, not both.
To connect the cables 1 Plug one end of the video cable into the video port on the connector plate. Secure the connection. 2 Plug the power cord into the power input port on the connector plate. 3 Plug the other end of the video cable into the video port on your computer system. Secure the connection. 4 Plug the other end of the power cord into the power supply. 5 Plug the power supply cord(s) into a grounded AC outlet. 6 If you are using a BridgePower adapter, turn on the adapter power switch before you turn on the computer.
4 | Dome Ex Display
HARDWARE INSTALLATION TIPS • Access to ports. For displays mounted on a desk stand, rotate the LCD panel from portrait to landscape for better access to the ports. • Threading cables. Thread the DVI cable and power cord through the stand column. Make sure the cable and cord run through the notches. To reattach the stand cover, align the hooks with the slots on the stand. Press the cover into place. A click sound signals a secure connection.
Installing the Display | 5
Technical Information The design of the Dome® Ex digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display. Read all instructions before setting up the display. Classification: Shock Protection: Class I. Degree of Protection Against Electric Shock: No applied part. Degree of Protection Against Harmful Ingress of Water: Ordinary equipment (IPX0). Degree of Safety in the Presence of Flammable Anaesthetic Mixture with Air or with Oxygen or Nitrous Oxide: Not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide. Mode of Operation: Continuous. No applied part. Important recycle instruction: HAZARDOUS SUBSTANCE. Lamp(s) inside this product contain
mercury. This product may contain other electronic waste that can be hazardous if not disposed of properly. Recycle or dispose in accordance with local, state, or federal laws. For more information, contact the Telecommunications Industry Association at WWW.ECYCLINGCENTRAL.COM. For lampspecific disposal information, check WWW.LAMPRECYCLE.ORG. DISPOSAL. Do not use household or municipal waste collection services for disposal of electrical and electronic equipment. EU countries require the use of separate recycling collection services. ENVIRONMENT-FRIENDLY USE PERIOD. Hazardous substances are present. The number encircled by the recycling symbol indicates the safe-use period (in years). China requires the use of recycling services at the end of product life.
6 | Dome Ex Display
Symbol explanations READ DOCUMENTATION. Follow operating instructions or consult instructions for use. MANUFACTURER.
DATE OF MANUFACTURE.
SERIAL NUMBER. AUTHORIZED REPRESENTATIVE. European Community. DANGEROUS VOLTAGE. Important precautions about electric shock. Read the accompanying text carefully, to prevent damage to display components and to guard your safety. DIRECT CURRENT. BAROMETRIC PRESSURE. Transport and storage 12,000 meters (39,400 feet), maximum in unpressurized container. RELATIVE HUMIDITY. Transport and storage 5% to 90% (noncondensing). TEMPERATURE. Transport and storage -10° to 60° C for Dome E3, Dome E3cHB, and Dome E5; -20° to 60° C for Dome E2cHB. WARNING.
Technical Information | 7
The Dome Ex display is an AMLCD display designed for viewing medical X-ray images. This unit should not be used near patients and should be kept outside of 1.5 m perimeter and 2.29 m vertical.
Intended use Dome Ex line, model Dome E2cHB is intended for use in the displaying and viewing of medical images for review, analysis, and diagnosis by trained medical practitioners. For the European market, Dome Ex line, model Dome E3 is intended for use in the displaying and viewing of radiography images for review, analysis, and diagnosis by trained medical practitioners. Dome Ex line, model Dome E3cHB is intended for use in the displaying and viewing of medical images for review and analysis by trained medical practitioners. Dome Ex line, model Dome E5 is intended for use in the displaying and viewing of medical images, including radiographs of the breast, for review and analysis by trained medical practitioners. WARNING! In the Dome Ex display series, only the Dome E5 display can be used for primary image diagnosis in mammography. The Dome E2cHB, Dome E3, and Dome E3cHB units must not be used for primary image diagnosis in mammography. Federal law restricts this device to sale by or on the order of a medical practitioner.
Safety precautions External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (IEC 60601-1:2005 and EN 60601-1:2006 series for medical electrical equipment). In addition, all such combinations (systems) must comply with the standard IEC 60601-1:2005, Safety requirements for medical electrical systems. Equipment not complying to IEC 60601-1:2005 must be kept outside the patient environment, as defined in the standard as at least 1.5 meters from the patient or the patient support.
8 | Dome Ex Display
Any person who connects external equipment to signal input, signal output, or other connectors has formed a system and is therefore responsible for the system to comply with the requirements of IEC 60601-1:2005. If in doubt, speak with a qualified technician. The power cord is the only disconnect device that allows you to cut power from the AC main. Position the display in order to operate the disconnect device without difficulty. Never touch the display and the patient at the same time. WARNING! No modification of this equipment is allowed.
Safety tips • Never open the display case, even when the power is off. Dangerous voltage inside may cause electric shock or death. • To avoid damage to the display, use the grounded power supply and video cable supplied by NDSsi, or use certified replacements. • Be sure the display is electrically grounded. You must connect the third grounding pin on the US power cord to a grounded outlet. The European power cord does not have a third grounding pin, but it must be plugged into a grounded outlet. • If you cannot insert the plug into the outlet you plan to use, have a licensed electrician replace the outlet with a properly grounded outlet. If the power cord connects directly into the computer, make sure the computer is grounded. • Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills. • In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States). GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTABLE MARKED “HOSPITAL ONLY” OR “HOSPITAL GRADE.” WARNING! TO AVOID THE RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.
Technical Information | 9
Unpacking and handling tips The Dome Ex display is a precision instrument that requires proper care to maintain product operation and adherence to specification. Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel. • Use both hands to grasp the display case when lifting it from the shipping carton, but avoid touching the screen. • Do not apply pressure to the screen or touch the screen with bare fingers or objects. Pressure can affect image quality. Cosmetics and oils on the skin are both detrimental to the screen and difficult to remove. • Allow the display to warm up to room temperature before turning it on. Avoid sudden temperature changes in the environment, as this may cause condensation, which damages the display. • Secure the display properly onto a standard VESA 100-mm mounting unit if you elect not to use the desk stand. • Do not set up the display near strong light or heat sources. • Do not block the vents on the back of the display or install the display in a built-in enclosure. Blocked vents cause excessive heat to build up inside the display, increasing risk of fire. • When installing components, turn off your computer, but leave it plugged into a grounded outlet. • Do not remove the back cover or disassemble the display. There are no user-serviceable parts inside.
Preventing fire and injury • Replace the power supply or cables if damaged. • Use only the power source indicated in this guide or listed on the display. • Do not plug the power supply into an overloaded AC outlet or extension cord. Overloaded AC outlets and cords can result in electric shock or fire. • Do not drop or push objects into the display case. Internal components contain high voltage. • Unplug the power cord from the wall outlet during thunderstorms. • Do not place magnetic devices, such as motors, near the display.
10 | Dome Ex Display
Cleaning the display Observe the following guidelines to maintain the display and the LCD screen. • Use a clean, lint-free, absorbent cotton cloth to clear off any residual glue from removal of the protective film or to remove surface dust. Apply light pressure to remove the dust. • Dampen a clean cloth with a small amount of isopropyl alcohol to remove glue or dust if the screen is still not clean. Do not saturate the cloth; otherwise, alcohol may seep into the display case and collect in the enclosure. Use a clean, dry cloth to completely remove the alcohol residue. • Do not use chemically treated dust cloths, acetone, toluene, or harsh solvents on the display case or the screen. They can damage the polarizer and the display case. • Do not expose the display to water or excessive moisture. Do not allow water or other stains to stand on the unit. Wipe liquids off immediately to prevent damage to the display case and the screen.
WARNING! DISCONNECT SUPPLY BEFORE SERVICING AVERTISSEMENT COUPER L’ALIMENTATION AVANT L’ENTRETIEN ET LE DEPANNAGE
Shipping/storing the display Keep the display in its shipping container until installation. Return the display to its original container whenever you need to store the unit, move it to another location, or return it for repair. The packaging supplied by the manufacturer protects the display while it is in transit. See environment specifications for more information. Before returning the display to the container, do the following: 1 Swivel the display panel to landscape mode. 2 Push the panel down to the lowest position. 3 Use the stand lock to anchor the panel.
Disposal information Follow your local governing ordinance and recycling policy for proper disposal or recycling of display components. Technical Information | 11
Dome E2cHB Specification In locations where 240V outlets are used, connect the Dome E2cHB display to a centertapped, 240V, single-phase supply only (for Canada and the United States). Category
Characteristic Item
Specification
LCD
Technology
TFT AMLCD Dual Domain IPS, 3 subpixels per pixel
Screen size, diagonal
498 mm (19.6 in.)
Display resolution
1200 x 1600 pixels (portrait) 1600 x 1200 pixels (landscape)
Pixel pitch
249 m, 102 DPI
Viewing angle
170° horizontal/vertical
Response time
25 ms (typical)
Refresh rate
60 Hz
Brightness
800 cd/m2 (typical)
Contrast ratio
700:1 (typical)
DICOM LUT
16-bit architecture with 0.3 JND precision
Colors supported
16.7 M
Pixel arrangement
RGB vertical stripe
Display control - brightness/ contrast
DDC2B+
Optical
Interface
Panel Connectivity
Physical
Power Requirement
Display identification
EDID read using DDC2B+
Display status
Dual-stack tricolor LEDs on back panel
Video input
DVI-D
Pixel data sizes
8-bit, 16-bit, and 24-bit per pixel color
VGA compatibility
640 x 480 to 1280 x 1024
USB interface
Built-in USB 2.0 HUB on stand
Display size (without stand), HxWxD
440 mm x 343 mm x 95 mm (17.3 in. x 13.5 in. x 3.7 in.)
Display weight (without stand)
5.4 kg (12 lb)
Display weight (with stand)
9.1 kg (20 lb)
Mounting options
100 x 100 mm VESA mounting standard. Desktop stand includes height, pivot, swivel, and tilt adjustment.
Adapter
BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.)
12 | Dome Ex Display
Power Supply WARNING: Use only the power adapter supplied with the Dome E2cHB display unit: the BridgePower BPM130S12F02 (130W) model. Category
Characteristic Item
Specification
Power input requirements
Voltage selection
Auto-ranging
Voltage
100–240V AC
Current
3.0 A (3.0–1.5 A)
Power output requirements
Frequency
50–60 Hz
Voltage
12 V DC ±5%
Current
10.8 A (130W)
Power consumption
Wattage
60 W (typical)
Physical
Size
228.6 mm x 76.2 mm x 50.8 mm (9 in x 3 in x 2 in)
Weight
1.3 kg (2.75 lb)
Reliability Characteristic item
Specification
Display
MTBF 50,000 hours
Backlight
MTBF 50,000 hours to 50% brightness with backlight on continuously
Environment Characteristic item
Specification
EMI shielding Temperature Humidity
Altitude
No emission of low-level radiation Operating range
10° C ~ 40° C
Storage range
-20° C ~ 60° C
Operating range
30% ~ 75% Relative Humidity (noncondensing)
Storage range
5% ~ 90% Relative Humidity (noncondensing)
Operating
2,000 m (6,560 ft) maximum
Transport/storage
12,000 m (39,400 ft) maximum in unpressurized container Technical Information | 13
Dome E3 Specification In locations where 240V outlets are used, connect the Dome E3 display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category
Characteristic Item
Specification
LCD
Technology
TFT AMLCD Dual Domain IPS, 3 subpixels per pixel
Screen size, diagonal
528 mm (20.8 in.)
Display resolution
1536 x 2048 pixels (portrait) 2048 x 1536 pixels (landscape)
Pixel pitch
206 m, 123 DPI
Viewing angle
170° horizontal/vertical
Response time
25 ms (typical)
Refresh rate
60 Hz
Brightness
1000 cd/m2 (typical)
Contrast ratio
900:1 (typical)
DICOM LUT
16-bit architecture with 0.2 JND precision
Colors supported
Up to 1024 shades of gray, programmable gamma from palette of up to 3061 shades
Pixel arrangement
Subpixel vertical stripe
Display control -brightness/ contrast
DDC2B+
Display identification
EDID read using DDC2B+
Optical
Interface
Panel Connectivity
Physical
Power Requirement
Display status
Dual-stack tricolor LEDs on back panel
Video input
Dual Link DVI-D
Pixel data sizes
8-bit and 10-bit per pixel grayscale, 24-bit color mapped to luminance
VGA compatibility
640 x 480 to 1280 x 1024
USB interface
Built-in USB 2.0 HUB on stand
Display size (without stand), HxWxD
475 mm x 368 mm x 102 mm (18.7 in. x 14.5 in. x 4.0 in.)
Display weight (without stand)
6.4 kg (14 lb)
Display weight (with stand)
10.0 kg (22 lb)
Mounting options
100 x 100 mm VESA mounting standard. Desktop stand includes height, pivot, swivel, and tilt adjustment.
Adapter
BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.)
14 | Dome Ex Display
Power Supply WARNING: Use only the power adapter supplied with the Dome E3 display unit: the BridgePower BPM130S12F02 (130W) model. Category
Characteristic Item
Specification
Power input requirements
Voltage selection
Auto-ranging
Voltage
100–240V AC
Current
3.0 A (3.0–1.5 A)
Frequency
50–60 Hz
Power output requirements
Voltage
12 V DC ±5%
Current
10.8 A (130W)
Power consumption
Wattage
60 W (typical)
Physical
Size
228.6 mm x 76.2 mm x 50.8 mm (9 in x 3 in x 2 in)
Weight
1.3 kg (2.75 lb)
Reliability Characteristic item
Specification
Display
MTBF >50,000 hours
Backlight
MTBF 50,000 hours to 50% brightness with backlight on continuously
Environment Characteristic item
Specification
EMI shielding
No emission of low-level radiation
Temperature Humidity Altitude
Operating range
10° C ~ 40° C
Storage range
-10° C ~ 60° C
Operating range
30% ~ 75% Relative Humidity (noncondensing)
Storage range
5% ~ 90% Relative Humidity (noncondensing)
Operating Transport/storage
2,000 m (6,560 ft) maximum 12,000 m (39,400 ft) maximum in unpressurized container
Technical Information | 15
Dome E3cHB Specification In locations where 240V outlets are used, connect the Dome E3cHB unit to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category
Characteristic Item
Specification
LCD
Technology
TFT AMLCD Dual Domain IPS, 3 subpixels per pixel
Screen size, diagonal
541 mm (21.3 in.)
Display resolution
1536 x 2048 pixels (portrait) 2048 x 1536 pixels (landscape)
Pixel pitch
211.5 m, 120 DPI
Viewing angle
176° horizontal/vertical
Response time
14 ms (typical)
Refresh rate
60 Hz
Brightness
800 cd/m2 (typical)
Contrast ratio
750:1 (typical)
DICOM LUT
16-bit architecture with 0.3 JND precision
Colors supported
Up to 1.06 billion colors; up to 1024 shades of gray, programmable gamma from palette of up to 2041 shades
Optical
Interface
Panel Connectivity
Physical
Power Requirement *
Pixel arrangement
R,G,B vertical stripe
Display control -brightness/ contrast
DDC2B+
Display identification
EDID read using DDC2B+
Display status
Dual-stack tricolor LEDs on back panel
Video input
DisplayPort 1.1 and Dual Link DVI-D
Pixel data sizes
8-bit, 16-bit, 24-bit, and 30-bit per pixel color*
VGA compatibility
640 x 480 to 1280 x 1024
USB interface
Built-in USB 2.0 HUB on stand, USB B upstream port
Display size (without stand), HxWxD
475 mm x 369 mm x 102 mm (18.7 in. x 14.5 in. x 4.0 in.)
Display weight (without stand)
6.6 kg (15 lb)
Display weight (with stand)
10.2 kg (23 lb)
Mounting options
100 x 100 mm VESA mounting standard. Desktop stand includes height, pivot, swivel, and tilt adjustment.
Adapter
BridgePower BPM150S24F05 (CAUTION: Use only the adapter supplied with unit.)
30-bit color is supported on DisplayPort only.
16 | Dome Ex Display
Power Supply WARNING: Use only the power adapter supplied with the Dome E3cHB unit: the BridgePower BPM150S24F05 (150W) model. Category
Characteristic Item
Specification
Power input requirements
Voltage selection
Auto-ranging
Voltage
100–240V AC
Current
2.5 A (2.5–1.5 A)
Frequency
50–60 Hz
Power output requirements
Voltage
24 V DC ±5%
Current
6.25 A (150 W)
Power consumption
Wattage
81 W (typical)
Physical
Size
223.5 mm x 90.0 mm x 56.6 mm (8.8 in x 3.5 in x 2.2 in)
Weight
1.4 kg (3 lb)
Reliability Characteristic item
Specification
Display
MTBF 50,000 hours
Backlight
MTBF 50,000 hours to 50% brightness with backlight on continuously
Environment Characteristic item
Specification
EMI shielding
No emission of low-level radiation
Temperature Humidity
Altitude
Operating range
10° C ~ 40° C
Storage range
-10° C ~ 60° C
Operating range
30% ~ 75% Relative Humidity (noncondensing)
Storage range
5% ~ 90% Relative Humidity (noncondensing)
Operating
2,000 m (6,560 ft) maximum
Transport/storage
12,000 m (39,400 ft) maximum in unpressurized container
Technical Information | 17
Dome E5 Specification In locations where 240V outlets are used, connect the Dome E5 display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category
Characteristic Item
Specification
LCD
Technology
TFT AMLCD Dual Domain IPS, 3 subpixels per pixel
Screen size, diagonal
541 mm (21.3 in.)
Display resolution
2048 x 2560 pixels (portrait) 2560 x 2048 pixels (landscape)
Pixel pitch
165 m, 154 DPI
Viewing angle
170° horizontal/vertical
Response time
18 ms (typical)
Refresh rate
50 Hz
Brightness
1100 cd/m2 (typical)
Contrast ratio
850:1 (typical)
DICOM LUT
16-bit architecture with 0.2 JND precision
Colors supported
Up to 1024 shades of gray, programmable gamma from palette of up to 3061 shades
Pixel arrangement
LCR vertical stripe
Display control -brightness/ contrast
DDC2B+
Display identification
EDID read using DDC2B+
Display status
Dual-stack tricolor LEDs on back panel
Video input
Dual Link DVI-D
Pixel data sizes
8-bit and 10-bit per pixel grayscale, 24-bit color mapped to luminance
VGA compatibility
640 x 480 to 1280 x 1024
USB interface
Built-in USB 2.0 HUB on stand
Display size (without stand), HxWxD
470 mm x 387 mm x 103 mm (18.5 in. x 15.2 in. x 4.1 in.)
Display weight (without stand)
6.4 kg (14 lb)
Optical
Interface
Panel Connectivity
Physical
Power Requirement
Display weight (with stand)
10.0 kg (22 lb)
Mounting options
100 x 100 mm VESA mounting standard. Desktop stand includes height, pivot, swivel, and tilt adjustment.
Adapter
BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.)
18 | Dome Ex Display