NeuroSoft
Neuro-MS-D Magnetic Stimulator Technical Manual TM058.07.001.002 Feb 2018
Technical Manual
99 Pages
Preview
Page 1
Technical Manual
Neuro-MS/D Magnetic Stimulator
ТМ058.07.001.002 (12.02.2018)
Contents Introduction ... 5 Safety Requirements ... 6 Indications for Use ... 6 Contraindications ... 6 Warnings and Precautions ... 7 Patient Safety... 10 Possible Side Effects ... 11 1.
Description ... 15 1.1 Main Specifications ... 15 1.2 Delivery Set ... 17 1.3 Principle of Operation ... 22 1.4 Controls, Connectors and Indicators... 25 1.4.1 Controls, Connectors and Indicators of Main Unit ... 25 1.4.2 Controls, Connectors and Indicators of Cooling Unit... 31 1.4.3 Controls, Connectors and Indicators of Extra Power Supply Unit... 33 1.4.4 Coil Controls, Connectors and Indicators... 35 1.5 Synchronization ... 36 1.6 Labeling... 37
2
Setting-Up Procedures ... 39 2.1 Requirements to Personnel Conducting Stimulator Assembly and Installation ... 39 2.2 Room Selection and Placement ... 39 2.3 Unpacking and Check of Delivery Set ... 42 2.4 Assembly and Connection of Stimulator ... 42 2.4.1 General Requirements to Assembly ... 42 2.4.2 Preparation of Cooling Unit to Operation ... 43 2.4.3 System Assembling ... 44
3
Operating Order ... 51 3.1 Operation in Stand-alone Mode ... 51 3.2 Operation in PC-mode ... 53 3.3 Operation in Control Mode via Trig-in ... 53 3.4 Operation in Syncronization Mode with External Device via Trig-out ... 54 3.5 Acquisition of Evoked Potentials... 54 3.6 Actions in the Emergency ... 55 3.7 Troubleshooting... 55
4
Servicing ... 57 4.1 General Requirements ... 57 4.2 Maintenance Service ... 57 4.3 Cleaning and Disinfection ... 57 4.4 Life Time ... 58 4.5 Service Menu of Main Unit... 59 4.6 System Information ... 60
3
5
Packing and Transportation ... 62
6
Storage Data ... 62
7
Conservation... 62
8
Disposal ... 62
9
Delivery Set and Package Data ... 63
10 Warranty ... 63 11 Reclamation Data ... 64 Annex 1 ... 65 Electromagnetic Emissions and Immunity ... 65 Annex 2 ... 69 Safety Screening Questionnaire for Transcranial Magnetic Stimulation (TMS) ... 69 Annex 3 ... 71 Patient Treatment Record Form ... 71 Annex 4 ... 72 Workflow for Depression Treatment in PC-mode... 72 1. Getting Ready ... 72 2. First Visit of a Patient ... 72 3. Second and Next Visits ... 84 Annex 5 ... 87 Workflow for Depression Treatment in Stand-alone Mode... 87 1. Getting Ready ... 87 2. First Visit of a Patient ... 87 3. Second and Next Visits ... 95 References ... 98
4
Description
Introduction This user manual (hereinafter referred to as “manual”) is the combined document describing operation and servicing of Neuro-MS/D magnetic stimulator (hereinafter referred to as “stimulator”). This manual is the document which certifies the stimulator specifications guaranteed by the manufacturer. Do not start working with stimulator before you have read this document! You can send your responses and recommendations to Neurosoft Company by the following address: P.O. Box 10, Ivanovo, 153000, Russia or by e-mail: [email protected] You can find additional information on Neurosoft products in the Internet: www.neurosoft.com or ask questions by phones: +7 (4932) 59-21-12 (Service Center), +7 (4932) 95-99-99; +7 (4932) 24-04-34. You can also contact SAS Neuromed Company, Authorized European Representative of Neurosoft Company (Mr. Pierre Scholl) by the following address: Chemin du tomple 84330 Le Barroux, France Phone: + 33 490-650-470, +622-748-384 Fax: +33 490-650-470 E-mail: [email protected]
5
Neuro-MS/D Technical Manual
Safety Requirements Indications for Use The Neuro-MS/D stimulator is intended for noninvasive diagnostics (together with digital EMG systems) and therapeutic intervention in brain cortex, spinal cord and peripheral nervous system by brief magnetic pulses. The stimulator can be applied in different fields of medicine including neurology, psychiatry, epileptology, neurosurgery, orthopedy, pediatrics, electrodiagnostics and others at: complex diagnostics, to study the cortical motor pathway integrity at wide range of disease-associated movement disorders; localization of speech center, motor cortical representations, etc. treatment of wide range of disorders including pain syndrome, depression, movement disorders (including Parkinson’s disease), amyotrophic lateral sclerosis, multiple sclerosis, epilepsy, impairment of consciousness, Alzheimer's disease, tinnitus, anxiety disorders, obsessive-compulsive disorders, auditory hallucinations and negative symptoms at schizophrenia, alcohol, nicotine, medicine withdrawal and etc. The stimulator can be used in patient care facilities, diagnostics centers, neurosurgical clinics, experimental laboratories of scientific and research institutes.
Contraindications Although the magnetic stimulation is a safe procedure, each patient must be screened for contraindications before the treatment start.
The only absolute contraindication to TMS/rTMS is the presence of metallic hardware in close contact to the discharging coil (such as cochlear implants, or an Internal Pulse Generator or medication pumps). In such instances there is a risk of inducing malfunctioning of such implanted devices [13].
All the rest contraindications are relative ones. In each individual case the benefit of the treatment/investigation should be weighed against possible adverse effects. This judgment is best left to a skilled clinician familiar with the characteristics of the device. The written informed consent of the patient should be obtained.
6
Description
Relative Contraindications: Conditions of increased or uncertain risk of inducing epileptic seizure are: a. Related to the protocol of stimulation: i. Any ‘‘novel paradigm. ii. Conventional high-frequency rTMS protocol with parameters of stimulation (intensity, frequency, train length or intertrain duration) exceeding the known safety limits reported in the Table 1 – Table 3. b. Related to the disease or patient’s condition: i. Personal history of epilepsy (untreated patients with one or a few past episodes), or treated patients. ii. Vascular, traumatic, tumoral, infectious, or metabolic lesion of the brain, even without history of seizure, and without anticonvulsant medication. iii. Administration of drugs that potentially lower seizure threshold, without concomitant administration of anticonvulsant drugs which potentially protect against seizures occurrence. iv. Sleep deprivation, alcoholism. Conditions of increased or uncertain risk of other events are: c. Related to patient’s condition: i. Implanted brain electrodes (cortical or deep-brain electrodes). ii. Pregnancy. iii. Severe or recent heart disease1.
Warnings and Precautions Read carefully the following warnings and precautions before starting to use NeuroMS/D magnetic stimulator. During rTMC session you must keep watching over the signs of seizure and terminate the treatment session if those signs appear. If patient has taken a medication that may alter the seizure threshold, detect the motor threshold before the treatment session start! The risk of seizure during rTMC session is increased in patients having history of convulsion including of unknown etiology, patients with epilepsy, head injuries, stroke, severe headache, neurological diseases that that may affect a seizure threshold altering and also patients taking tricycling antidepressants, neuroleptic agents and other epileptogenic drugs.
Rossi, 2009 [13].
7
Neuro-MS/D Technical Manual
During the treatment observe carefully the patients for clinical worsening of depression and suicidality. If the signs of worsening of depression and suicidality (behaviour) continue to appear, terminate the treatment with the use of Neuro-MS/D magnetic stimulator.
Do not start working with stimulator before you have read this document! The stimulator must be used by the qualified medical personnel trained to operate on it and knowledgeable of magnetic stimulation application. During the work on the stimulator it is required to observe the working regulations concerning the safety rules while operating on electrical installations.
High voltages (1800V) are presented within the stimulator. To provide safety measures and exclude the hazard of medical staff’s or patient’s electric shock, the medical staff is PROHIBITED: to use the stimulator which was mounted and installed incorrectly, without following this manual instructions; to eliminate faults connected with opening of the components included in the delivery set; to work with the stimulator when the electronic unit box, computer or other devices used together with the stimulator are open. Always check the stimulator and coils for the absence of cracks and other defects. Do not open and do not try to repair the electronic units of magnetic stimulator! Before the stimulation make sure, that there are no patient cables or electrodes connected to a patient and also metal parts contacting with a patient in impact area (within 5 cm and less from the coil) as far as the electromagnetic field can create the electrical current leaking through a patient.
8
Description
Do not allow the coil to come into close proximity (less than 1 meter) with electronic equipment (monitor, computer, etc.) and also the magnetic carriers (credit cards, floppy disks, CD disks, audio recording tapes, hotel room keys, electronic automotive ignition keys, watches, etc.). The magnetic field emitted by the coil can result in its damage and information loss. It is prohibited to immerse the coil in water, ice or refrigerate to cool it off. Do not use Neuro-MS/D stimulator near water or other liquids. Do not place liquids on/near the case of magnetic stimulator, its cables and coils. It is prohibited to use cooled coils without cooling unit. It can cause coil surface overheating (even after the stimulation stop), patient’s burn and magnetic stimulator damage.
The stimulus evoked by the stimulator generates a loud click that can frighten a patient. Tell a patient about the stimulation start and give her/him the ear plugs.
Hearing protection is also advised for the clinical operator if he/she is located within 3 meters of coil during patient treatment.
The stimulator must not be used in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
The operator must be protected from the continuous impact of the magnetic field. It is recommended to use the holder to fix the coil.
If the coil heats over 41°C, the magnetic stimulator stops the stimulation and indicates the overheating. As far as the coil temperature on its surface still grows even after the stimulation stop owing to the thermal lag, it is required to remove immediately the coil from the patient surface to avoid the patient’s burn and discomfort.
9
Neuro-MS/D Technical Manual
To protect a patient from excessive impact by the electromagnetic field, generate the minimal required number of pulses.
Keep out of the reach of children.
It is recommended to replace the coil with a new one after a year of use.
Patient Safety The general principles of safety regulations are stated in: “Risk and safety of repetitive transcranial magnetic stimulation: report and suggested guidelines from the International Workshop on the safety of repetitive magnetic stimulation, June 5-7. 1996” by Eric M. Wassermann (accepted for publication: 23 May 1997). “Safety of different inter-train intervals for repetitive transcranial magnetic stimulation and recommendations for safe ranges of stimulation parameters” by Robert Chen, Christian Gerloff, Joseph Classen, Eric M. Wassermann, Mark Hallet, Leonardo G. Cohen (accepted for publication: 23 May 1997). “Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice.and research” by Simone Rossi, Mark Hallet, Paolo M. Rossini, Alvaro Pascual-Leone and The Safety of TMS Consensus Group (accepted for publication: 21 August 2009). The main issues to consider are intensity, frequency, train duration and intertrain interval (Table 2, Table 3 and Table 4). a
Table 1. Table of safe train duration (s) of single trains of rTMS based on the NINDS experience Frequency (Hz)
10
Intensity (% of Motor threshold) 100
110
120
130
140
150
160
170
180
190
200
210
220
1
>1800
>1800
360
>50
>50
>50
>50
27
11
11
8
7
6
5
>10
>10
>10
>10
7.6
5.2
3.6
2.6
2.4
1.6
1.4
1.6
1.2
10
>5
>5
4.2
2.9
1.3
0.8
0.9
0.8
0.5
0.6
0.4
0.3
0.3
20
2.05
1.6
1.0
0.55
0.35
0.25
0.25
0.15
0.2
0.25
0.2
0.1
0.1
25
1.28
0.84
0.4
0.24
0.2
0.24
0.2
0.12
0.08
0.12
0.12
0.08
0.08
Description
a) Based on Wassermann et al. (1996), [6] and Rossi et al. (2009), [4]. Table 2. Safety recommendations for inter-train intervals for 10 trains of rTMS at < 20 Hz Inter- train interval (s)
Stimulus intensity (% of MT) 100%
105%
110%
120%
5
Safe
Safe
Safe
Insufficient data
1
Unsafe (3)
Unsafe
a
Unsafe (2)
Unsafe (2)
a
Unsafe (2)
Unsafe (3)
a
0.25
Unsafe
Unsafe
a)
These stimulus parameters are considered unsafe because adverse events occurred with stimulation of lower intensity or longer inter-train interval, but no adverse event was observed with these parameters. Table 3. Parameter safety issues: commonly employed stimulation parameters rTMS frequency
No. of studies
4-9Hz
>10
Variable
Variable
Variable
10 Hz
>50
5-6 pulse-trains for 400-500 ms
3,2 s
250
20-25 Hz
>20
10 pulse-trains for 400-500 ms
17,1 s
80
a)
Average train duration
Average inter-train interval
Average no. of trials
Based on Rossini P.M. et al. (1994) [5].
Possible Side Effects 1. Increase of hearing threshold. According to the publications, some patients suffered from the transient hearing threshold increase after TMS. However the majority of studies have shown that the negative impact of TMS on hearing is missing when the earplugs are used. Thus it is recommended to: use the approved hearing protection (earplugs or ear muffs) by specialists trained to insert these devices; refer a patient for the auditory examination if she/he complains of hearing loss, tinnitus or aural fullness after TMS procedure; make a decision on TMS performing only in cases of a favorable risk/benefit ratio in patients with pre-existing noise induced hearing loss, under concurrent treatment with ototoxic medications (Amynoglycosides, Cisplatine) and also young children. avoid performing the TMS in patients with cochlear implants. 2. Epileptiform abnormalities in EEG. In some publications it is stated that patients with brain damage, drug treatment or discontinuation of the treatment are more vulnerable to TMS. In such patients the epi-
11
Neuro-MS/D Technical Manual
leptiform abnormalities recorded with EEG during or after repetitive TMS (in an hour) may occur. 3. Seizures. The published data have revealed that the risk of rTMS to induce seizures is very low. Before the publication of the safety guidelines with stimulation parameter limits (Wassermann, 1998) it was reported about seven cases of TMS-induced seizures after high-frequency rTMS. Seven of them occurred in healthy subjects during the studies intended to define the safety limits and one of them is in patient suffering from depression. In the above-mentioned cases the parameters outside the safety limits were used and were later included to safety guidelines. Also other possible factors that might contribute for seizures were also considered (for example, taking antidepressants or antipsychotic drugs that change the cortex excitability). After the recommendations were published, four cases of TMS-induced seizures with parameters outside the safety limits and four cases were occurred with rTMS parameters considered safe. In three cases of four (both with use of permissible stimulation parameters and with parameters exceeded the permissible ones) the TMS-induced seizures were observed in patients taking pro-epileptogenic medications or with a history of sleep deprivation. The detailed analysis of clinical features evidences that the episode was a syncope rather than a seizure in two cases of four with recommended stimulation parameters (it is related to the case of generalized tonic clonic seizure in a patient with tinnitus at rTMS reported by Nowak in 2006) and in one case of four with the use of stimulation parameters exceeding the recommended ones. A recent review have shown that the risk of seizures in epileptic patients at rTMS is 1.4%. No cases of induced status epilepticus was reported in such patients. 4. Syncope Vasodepressor (neurocardiogenic) syncope is a common response to anxiety and psycho-physical discomfort. It is a common reaction that can occur both at magnetic stimulation and other noninvasive and minimally invasive procedures. In spite of the fact that syncope may occur more often in comparison with the induced seizures during the TMS, the risk of syncope occurrence is still rather low. 5. Local pain, headache, discomfort Most patients consider the single pulse stimulation to be painless. The pain intensity at rTMS depends on the individual susceptibility, scalp location, coil design, intensity and frequency of stimulation. In 2% of cases the pain intensity requires to discontinue the treatment. 6. Cognitive changes
12
Description
TMS can produce both positive and negative changes. Most of them do not result in any stable changes. 7. Acute psychiatric changes Treatment-emergent mania Anxiety Agitation Suicidal ideation Insomnia In all cases described in the publications the acute psychiatric changes listed above were transient with the spontaneous resolution or were controlled with the use of pharmacological treatment. 8. Other biological effects
The brain tissue heating at single pulse stimulation is very low and is less than 0.1°C by publication review. The closer is the coil to the stimulated tissue, the higher is the probability of tissue temperature increase. It may be important with infarction area tissues or cysts. However there is no significant tissue temperature change at TMS which is proved by clinical data evidencing the absence of hazardous effect of rTMS.
In spite of the reported possible hazardous effect of rTMS at the structural and cellular level of the brain (Matsumiya et al., 1992) reported microvascular changes in the neuropil portion of cortical layers II-VI in rats stimulated with 2.8 Tesla at 100 stimuli) no other experimental evidence of neuropathological damage of variable magnetic field to nervous tissues is obtained.
The effect of TMS on hormonal state is still being studied. Thus, some researchers (Bridgers, Delaney, 1989; Wassermann et al., 1996) reported a decline of prolactine in blood serum and no change in concentration of thyroid-stimulating hormone, follicle-stimulating hormone and cortisol after rTMS. Other studies (George, 1996; Szuba, 2001) have shown the transient increase of thyroid-stimulating hormone with other hormones of hypothalamopituitary system unaffected. Evers (2001) performed the placebo-controlled cross-over trial. The left dorsolateral prefrontal cortex was stimulated using rTMS stimuli of suprathreshold and infrathreshold intensity by 10 Hz and 20 Hz protocols. He reported the reliable moderate decrease of cortisol level and thyroid-stimulating hormone level only after the infrathreshold stimulation, the concentration of prolactine and follicle-stimulating hormone remained unaffected. Considering that cortisol and thyroid-stimulating hormone are released in stressful situations, the decrease of these hormones suggests a relaxing effect by this type of rTMS in healthy subjects. According to Padberg
13
Neuro-MS/D Technical Manual
(2002) the plasma concentrations of neuroactive steroids were not affected by rTMS when using depression treatment protocol approved by FDA.
14
The effect of rTMS on neurotransmitters is described in some studies. Keck (2000) and Strafella (2001) reported on increased dopamine release in some brain areas including the hippocampus. Theoretically, such effect should induce the beneficial effect on parkinsonian symptoms, mania and other abnormalities. The stimulation of the left dorsolateral prefrontal cortex (20 Hz, 20 min per day) may change cortical glutamate/glutamine levels not only close to the stimulation site but also in remote cortex areas (Michael, 2003). 10 Hz rTMS of left dorsolateral prefrontal cortex can modulate the tryptophan and 5-hydroxytryptamine metabolism in limbic areas in normal subjects without inducing behavioral changes (Sibon, 2007).
The effect on the immune system is still studied. In few publications related to the after-effects of rTMS on immune system there are no data that rTMS may impact negatively to immune system.
The effect of rTMS on antonomic functions was reviewed in some studies. Thus, rTMS (500 ms, 20 Hz, 70-90% with maximum intensity) induced a short lasting increase in heart rate and blood pressure (Foerster, 1997). Udupa (2007) concluded the improvement of sympathovagal balance in patients with depression. Vernieri (2009) noted that rTMS may decrease the vaso-motorreactivity that should be taken into account when using rTMS to treat acute stroke patients.
Description
1.
Description 1.1 Main Specifications Table 4 contains the main specifications of the stimulator. Table 4. Main Specifications
Parameters
Values
Stimulation Parameters Pulse duration: standard biphasic
250–330 µs
Maximum frequency at maximum power without extra power unit with extra power supply unit
2
5 Hz 2 20 Hz Power Supply
Supply voltage
(22022)/(23023) V 50 or 60 Hz
Maximum power consumption during the stimulation: main unit cooling unit extra power supply unit
≤1000 V·A ≤1000 V·A ≤3000 V·A
Trig-in/out Trig-in
TTL/CMOS
Trig-out
TTL/CMOS
Trig-in pulse width
≥ 200 µs
Stimulus delay relative to trig-in signal
100...200 µs
Trig-out pulse width (user-defined)
10 µs…50 ms
General Specifications and Parameters Interface
USB
Safety
I
Work parts Cooling unit tank volume while filling up to the maximal level Type of cooling agent Ingress protection
BF type 3.5 l Silicone oil 5 cSt IP20
Operation temperature
(10–35)º C
Relative air humidity
≤80%
Dimensions: main unit cooling unit extra power supply unit
460215520 mm 470215520 mm 360170520 mm
depends on coil type. The data is given for RC-02-150 coil.
15
Neuro-MS/D Technical Manual
Table 5. Continued
Parameters
Values
Weight: main unit cooling unit filled with cooling agent extra power supply unit
not more than 24 kg not more than 27 kg not more than 12 kg
Cooling agent pressure in coil while cooling unit operating
not more than 0.15 MPa
Table 5. Range of Controlled Parameters and Frequency Step Parameter
Range of Permissible Values
Frequency Step
Stimulation Parameters Stimulation frequency in repetitive stimulation mode
(0.1–30) Hz (Stand-alone mode) (0.1 – 100) Hz (PC mode)
0.1 Hz (0.1 -10 Hz)
50 – 100 Hz (PC mode)
1 Hz
Pulse train duration range
(0.5 – 100) s
0.5s (0.5-9.5 s) 5s (10-100 s)
Inter-train interval range
(0 – 300) s
0.5s (0-29.5s ) 1s (30-100 s)
Session
(0.5-40) min
0.5 min
Stimulation frequency in burst* mode
1 Hz (10-100 Hz)
Capacitor Charge Control Delay for the capacitor charge start (analysis epoch)**
(0–9.90) s
0.01 s (0–0.99 s), 0.1 s (1.00–9.90 s)
10 µs, 1 ms–50 ms
1 ms
Trig-out Trig-out pulse width
*The burst mode is available only if you use Neuro-MS.NET software. **To avoid the noise impact from the power supply while joint operation with EMG equipment during the response acquisition, the delay of capacitor charge start is provided in single pulse mode. Table 6. Dependence of Maximal Stimulus Amplitude on Stimulation Frequency While Working with/without Extra Power Supply Unit with Use of RC-02-150 Big Ring Coil Stimulation Frequency, Hz
16
Maximum Power Without Extra Power Supply Unit
With Extra Power Supply Unit
5
100%
100%
10
69%
100%
15
53%
100%
20
43%
100%
25
37%
89%
30
30%
79%
Description
Safety and Electromagnetic Compatibility Electromagnetic compatibility (EMC) is ensured by conformance to International standard IEC 60601-1-2:2007 requirements. The magnetic stimulator is intended for operation in electromagnetic environment, as specified in the Annex 0. As for safety, magnetic stimulator complies with IEC 60601-1:2005 requirements, it is referred to class I and has BF type work parts complying with IEC 60601-1:2005 requirements.
1.2 Delivery Set The delivery set includes the electronic units with the set of coils and cables. The delivery set of the stimulator corresponds to Table 7 and Table 8. The figures in “Quantity, pcs.” column of Table 7 and Table 8 indicate: 1 - Neuro-MS/D magnetic stimulator (configuration 1 – diagnostic). 2 - Neuro-MS/D magnetic stimulator (configuration 2 – therapeutic). 3 - Neuro-MS/D magnetic stimulator (configuration 3 – advanced therapeutic).
17
Neuro-MS/D Technical Manual
The assembled stimulator (configuration 3) is shown in Fig. 1.
Fig. 1. Assembled magnetic stimulator.
18
1
Cooled angulated figure-of-eight coil 100 mm AFEC-02-100-C (the cooled coil is used together with the cooling unit);
2
Flexible arm for coil positioning К-3;
3
Trolley with custors for magnetic stimulator (4 shelves);
4
Touchscreen computer (is not included in the delivery set);
5
Main unit;
6
Cooling unit;
7
Extra power supply unit;
8
Coil holder (trolley/wall mounted) K-8.
Description Table 7. Base Delivery Set
Order code or main specifications
Name
Quantity, pcs. 1
2
3
Main unit
NS058201.014-020
1
1
1
Cooling unit
NS058201.017-020
–
1
1
Extra power supply unit
NS058201.016-020
–
–
1
NS058105.002
1
1
1
NS058213.001 NS058213.001-01
– –
10 10
10 10
NS058998.005
–
1
1
Big ring coil 150 mm RC-02-150
NS058304.002
1
–
–
Small ring coil 100 mm RC-02-100
NS058304.001
1
–
–
Figure-of-eight coil 100 mm FEC-02-100
NS058304.013
1
–
–
Cooled angulated figure-of-eight coil 100 mm AFEC-02-100-C
NS058304.012
–
1
1
Hydraulic system adapter
NS058214.001
–
2
2
Control cable for Neuro-MS/D cooling unit
NS058103.070
–
1
1
HV (high-voltage) cable for extra power supply unit connection to Neuro-MS/D
NS058103.102
–
–
1
Equipotential cable
NS058103.032-050
–
1
2
IEC C13 – IEC C14 mains supply cable (to connect cooling unit)
220 V, 10 A, 1,8 m (3G×0.75 sq.mm.)
–
–
1
СЕЕ 7/7 – IEC C13 mains supply cable
220 V, 10 A, 1,8 m (3G×0.75 sq.mm.)
1
2
2
NS007103.005
1
1
1
End cap Patient cap: size 42-54 size 54-66 Silicone oil Coils:
Cables:
USB cable (A→B)
19
Neuro-MS/D Technical Manual
Table 9. Continued
Order code or main specifications
Name
Quantity, pcs. 1
2
3
Holders: 1)
Flexible arm for coil positioning K-3
NS016998.012
–
1
1
2)
Coil holder K-5
NS058998.003
1
–
–
Trolley mount for extra coil H-8
NS059998.004
1
1
1
NS059998.009
–
1
1
NS 4D42006 3/8’’
–
1
1
without additional modules
–
1
1
Trolleys: Trolley with custors for magnetic stimulator T-4/A3)
Spare Parts and Accessories Quick connect coupling Software License for the use of Neuro-MS.NET software
Operational Documentation: Neuro-MS/D technical manual
TM058.07.001.000
1
1
1
“Coils for Neuro-MS/D and Neuro-MS Magnetic Stimulators” technical manual
TM058.02.003.000
1
1
1
Neuro-MS.NET user manual
UM058.02.003.000
–
1
1
Ivanovo, 2012
–
1
1
–
1
1
1
Main unit package
NS058201.014
1
1
1
Cooling unit package
NS058201.013
–
1
1
Extra power supply unit package
NS058201.016
–
–
1
“Transcranial Magnetic Stimulation” handbook by Moacyr Alexandro Rosa and Marina Odebrecht Rosa Package: Cardboard package (set)
Note: 1) Description of flexible arm for coil positioning K-3 is given in “Flexible arm for coil positioning К-3” technical manual 2) Description of coil holder is given in “Coil holder (trolley/wall mounted) K-5” technical manual. 3) Description of trolley with custors for magnetic stimulator T-4/A is given in “Trolley for magnetic stimulator T-4/A” technical manual.
20
Description Table 8. Optional Equipment, Accessories and Software
Order code or main specifications
Name
Quantity, pcs. 1
2
3
1
1
1
1
1
1
NS028353.004
1
1
1
Cooled big ring coil 150 mm RC-02-150-C
NS058304.003
–
1
1
Big ring coil 150 mm RC-02-150
NS058304.002
–
1
1
Small ring coil 100 mm RC-02-100
NS058304.001
–
1
1
Figure-of-eight coil 100 mm FEC-02-100
NS058304.013
–
1
1
Angulated figure-of-eight coil 100 mm 1) AFEC-02-100
NS058304.014
1
1
1
Small figure-of-eight coil 50 mm FEC-02-50
NS058304.005
1
1
1
Angulated figure-of-eight coil 100 mm (placebo) AFEC-02-100-P
NS058304.009
1
1
1
Figure-of-eight coil 100 mm (placebo) FEC-02-100-P
NS058304.004
1
1
1
Cooled figure-of-eight coil 100 mm FEC-02-100-C
NS058304.010
–
1
1
Cooled angulated figure-of-eight coil 100 mm (placebo) AFEC-02-100-C-P
NS058304.011
–
1
1
Angulated figure-of-eight coil, 125 mm AFEC-02-125
NS058304.017
–
1
1
Hydraulic system adapter
NS058214.001
–
2
2
Trig-in cable for synchronization via electrical stimulator: DIN8DIN5 (240°)
NS014103.032
1
1
1
Trig-in cable for synchronization via electrical stimulator: DIN8DIN5 (180°)
NS014103.033
1
1
1
Trig-in cable: DIN8BNC
NS014103.039
1
1
1
Trig-out cable: DIN8BNC
NS014103.047
1
1
1
Coil holder (trolley-wall mounted) K-8
NS059998.004
1
–
–
Flexible arm for coil positioning K-3
NS016998.012
1
–
–
without additional modules
1
–
–
1
1
1
Trigger unit
NS990201.007
Patient button
NS028201.006 NS028201.007
Footswitch Coils:
Cables:
Holders:
Software: License for the use of Neuro-MS.NET software
Operational Documentation: “Trigger unit TU-2” technical manual
TM990.01.001.000
21