Nonin Medical Inc
Model X-100 SenSmart Universal Oximetry System Operators Manual 2015
Operators Manual
103 Pages
Preview
Page 1
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner. Follow Instructions for Use. Nonin® reserves the right to make changes and improvements to this manual and the products it describes at any time, without notice or obligation. Nonin Medical, Inc. 13700 1st Avenue North Plymouth, Minnesota 55441-5443, USA +1 (763) 553-9968 (800) 356-8874 (USA and Canada) Fax: +1 (763) 553-7807 E-mail: [email protected] Nonin Medical B.V. Prins Hendriklaan 26 1075 BD Amsterdam, Netherlands +31 (0)13 - 79 99 040 (Europe) Fax: +31 (0)13 - 79 99 042 E-mail: [email protected] nonin.com
0123
EC
REP
MPS, Medical Product Service GmbH Borngasse 20 D-35619 Braunfels, Germany References to “Nonin” in this manual shall imply Nonin Medical, Inc. Nonin, EQUANOX, and
are registered trademarks of Nonin Medical, Inc.
The Bluetooth word mark and logos are owned by the Bluetooth SIG, Inc. and any use of such marks by Nonin Medical, Inc. is under license. Other trademarks and trade names are those of their respective owners. ©2015 Nonin Medical, Inc. 9544-001-07
Contents Indications for Use ... 1 Contraindications ... 1 Warnings ... 1 Cautions ... 3 Declaration of Conformity with FCC and Canadian Ministry of Health Rules for Electromagnetic Compatibility ... 5 Federal Communications Commission (FCC) Notice ... 5
Guide to Symbols... 6 System Components and Set Up... 12 System Configurations ... 12 Multiple Channels... 12 Single Channel ... 12 X-100M (Monitor) ... 14 X-100H (Hub) ... 17 Connect/Disconnect the Hub...17 X-100HH (Hub Holster) ... 18 Using the Hub Holster ...18 X-100SP (Signal Processor)... 19 Connect a Signal Processor to Hub or Monitor ...19 Disconnect a Signal Processor from Hub or Monitor ...19 Connect/Disconnect a Sensor to the Signal Processor...20 Replacing the Lock on the Signal Processor...20 Cable Clips ... 20 X-100EC (Extension Cable) ... 21 rSO2 and SpO2 Sensors ... 21 Battery... 21
System Operation... 22 Start-up Sequence ... 22 Sensor Application ... 23 Monitoring a Patient ... 23 Operating Screens and Menus ... 24
Monitoring Screen... 25 Monitoring Screen – Description ... 25 Channels ... 25 Event Marks ... 26 Event Mark Table ... 26 Timescale ... 26 Scrolling Cursor and Cursor Values ... 27 Graphs... 27 Monitoring Screen – Procedures... 28 Set All rSO2 Channel Baselines to Current %rSO2 Values... 28 Mark an Event ... 28 View the Event Mark Table ... 28 Change the Timescale ... 28 Scroll Through the Timescale... 28 Graph Set-up... 28 i
Contents (Continued) Settings Menu Screen ... 29 Settings Menu – Description... 29 Sensor Site ... 29 Sensor Type ... 30 Baseline (rSO2 channels only)... 30 Alarm Limits ... 31 Graph Position ... 32 Preset #: ... 32 Settings Menu – Procedures ... 33 Open the Settings Menu ... 33 Assign a Pre-Defined Sensor Site Name... 33 Customize a Sensor Site Name... 33 Clear a Sensor Site Name ... 33 Set Sensor Type ... 34 Set Individual Baseline Values... 34 Set Alarm Limits... 35 Set Graph Position(s)... 35 Review/Select a Preset... 35
Presets Menu Screen... 36 Presets Menu – Description... 36 Presets Menu – Procedures ... 37 Open the Presets Menu ... 37 Activate a Preset... 37 Save Current Settings as a Preset... 37 Delete a Preset ... 38 Rename a Preset ... 38 Lock/Unlock a Preset ... 39
Case Menu Screen ... 40 Case Menu – Description ... 40 Start New Case... 40 Edit Patient ID ... 40 Case Menu – Procedures ... 41 Open the Case Menu... 41 Start a New Case... 41 Edit a Patient ID ... 41
System Menu Screen... 42 System Menu – Description... 42 Brightness ... 43 Alarm Volume ... 43 rSO2 Low Alarm Mode ... 43 Pulse Tone Volume ... 43 Pulse Tone Source ... 44 Data Output Modes... 44 Clear Memory ... 44 Restore Factory Defaults ... 44 System Information ... 44 Date / Time ... 45 Nurse Call Mode ... 45 ii
Contents (Continued) Bluetooth ... 46 Language ... 46 Patient ID Request ... 47 System Name... 47 Default Preset... 47 Institution Default Limits ... 47 Institution Password ... 48 System Menu – Procedures... 48 Open the System Menu... 48 Adjust the Display Brightness... 48 Adjust the Alarm Volume... 49 Set the rSO2 Low Alarm Mode ... 49 Adjust the Pulse Tone Volume ... 49 Select a Pulse Tone Source... 50 Set Bluetooth and/or RS-232 Data Output Formats... 50 Clear the Memory... 50 Restore Factory Defaults... 51 Review System Information... 51 Set the Date and Time ... 51 Set-up Nurse Call ... 51 Enable/Disable Bluetooth Radio... 52 Change Language... 52 Set Patient ID Request... 52 Assign System Name ... 53 Select Default Preset... 53 Set Institution Default Limits... 54 Change the Institution Password... 54
Alarms ... 55 High Priority Alarms ... 55 Medium Priority Alarms ... 56 Silencing Alarms ... 56 Error Codes... 57
Memory and Data Output Features... 58 Memory ... 58 Patient Data Output... 58 Bluetooth Technology... 59 Bluetooth Connection ... 59 Determine the Bluetooth Address and PIN for the Monitor ...59 Bluetooth Security ... 60 RS-232 Connection to Printer ... 60 Using the Dymo Printer ...60 Connecting the Device into a Medical System ... 61 Data Output Formats... 61 Nonin 1 ... 62 Nonin 2 ... 65 Nonin 3 ... 66 Nonin 4 ... 67 Nonin 5 ... 68 iii
Contents (Continued) Printer ... 69
SenSmart Download Software... 70 System Requirements ... 70 Installing SenSmart Download Software ... 70
Care and Maintenance... 71 Cleaning Instructions ... 71
Parts and Accessories ... 72 Troubleshooting... 73 Service, Support, and Warranty ... 76 Service and Support ... 76 Warranty ... 76
Technical Information... 78 Manufacturer’s Declaration... 78 Equipment Response Time ... 81 Example – SpO2 Exponential Averaging ... 82 Testing Summary... 82 rSO2 Principles of Operation... 82 SpO2 Principles of Operation... 82 rSO2 Accuracy Testing ... 83 SpO2 Accuracy Testing... 83 Pulse Rate Accuracy Testing (Motion and Non-motion) ... 83 Low Perfusion Accuracy Testing... 84 Specifications ... 84 Transmitter... 86
External Monitor Installation Instructions ... 87 Philips Monitors ... 87 Components ... 87 VueLink Components... 87 IntelliBridge Components... 87 Connection Specifications ... 87 Connecting the X-100M Monitor to the Philips Monitor ... 88 X-100M Configuration ... 88 Philips Interface Module Installation and Configuration ... 88 Setting Up the Connection – VueLink ... 88 Setting Up the Connection – IntelliBridge ... 90 Philips Monitor Display Configuration ... 91 Setup Philips Monitor with VueLink Interface Module to Display X-100M Numerics... 91 Setup Philips Monitor with IntelliBridge Interface Module to Display X-100M Numerics... 92 Alerts... 93 Patient Alarms ... 94 Equipment Alarms... 95
iv
Figures Figure 1. Monitoring Screen Symbols (Four-Channel View)... 8 Figure 2. System Components and Set Up (4 Channels Connected) ... 13 Figure 3. System Set Up (Single Channel) ... 13 Figure 4. X-100M Front View ... 14 Figure 5. X-100H – Hub ... 17 Figure 6. Connect Hub or Signal Processor to Monitor ... 17 Figure 7. X-100HH – Hub Holster ... 18 Figure 8. Insert Hub into Hub Holster ... 18 Figure 9. X-100SP – Signal Processor ... 19 Figure 10. Connect Sensor to Signal Processor... 20 Figure 11. Signal Processor Lock Replacement... 20 Figure 12. X-100EC – Extension Cable ... 21 Figure 13. Start-up Screen ... 22 Figure 14. Select a Preset Screen ... 22 Figure 15. Monitoring Screen with Four Channels Connected (3 rSO2 and 1 SpO2)... 25 Figure 16. Event Marks, Timescale, and Scrolling Cursor... 26 Figure 17. Event Mark Table ... 26 Figure 18. Graphs ... 27 Figure 19. Settings Menu Screen ... 29 Figure 20. Presets Menu Screen ... 37 Figure 21. Case Menu Screen ... 40 Figure 22. System Menu Screen ... 42 Figure 23. System Information Pop-up ... 45 Figure 24. Bluetooth Information Pop-up ... 46 Figure 25. Institution Defaults Pop-up ... 48 Figure 26. RS-232 Serial Data Port ... 58 Figure 27. Sample Label from Dymo Printer... 60 Figure 28. X-100M Connection to Philips Monitor with VueLink ... 89 Figure 29. Philips Indicator on Model X-100M Display ... 89 Figure 30. X-100M Connection to Philips Monitor with IntelliBridge ... 90
v
Tables Table 1. Labeling and Packaging Symbols ... 6 Table 2. X-100M Monitoring Screen Symbols and Indicators ... 8 Table 3. X-100M Features... 14 Table 4. X-100H Features ... 17 Table 5. X-100HH Features... 18 Table 6. X-100SP Features ... 19 Table 7. Commonly-Used Sensor Site Names... 30 Table 8. Alarm Limit Settings... 31 Table 9. Factory Default Alarm Limit Settings ... 44 Table 10. High Priority Alarms... 55 Table 11. Medium Priority Alarms ... 56 Table 12. Error Codes ... 57 Table 13. Nonin 1 Data Output Format ... 62 Table 14. Electromagnetic Emissions ... 78 Table 15. Electromagnetic Immunity ... 79 Table 16. Guidance and Manufacturer’s Declaration-Electromagnetic Immunity ... 80 Table 17. Recommended Separation Distances ... 81 Table 18. Philips Monitor Patient Alarms... 94 Table 19. Philips Monitor Equipment Alarms... 95
vi
Indications for Use
Indications for Use Nonin’s SenSmart™ Model X-100 Universal Oximetry System is a modular system and is indicated for use in simultaneously measuring, displaying, monitoring, and recording up to six (6) channels of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate or cerebral or somatic hemoglobin oxygen saturation (rSO2) of blood underneath the sensor. Patient populations include adult, pediatric, infant, and neonate through the use of SenSmartcompatible sensors. The SenSmart system is intended for use in hospitals, long-term care, medical facilities, sleep laboratories, subacute environments, and Emergency Medical Services (EMS), including patient transport. The X-100 SenSmart system may be used for spot-checking and continuous monitoring with patient alarms. The SenSmart pulse oximetry (SpO2) functionality is suitable for use in both motion and non-motion conditions, including patients who are well or poorly perfused.
Contraindications Do not use this device in an MR environment. Explosion Hazard: Do not use in an explosive atmosphere or in the presence of flammable anesthetics or gases. This system is not intended to be used simultaneously on multiple patients. Refer to the applicable sensor instructions for use for additional contraindications, warnings, and cautions.
Warnings This device is intended only as an adjunct device in patient assessment. It should not be used as the sole basis for diagnosis or therapy decisions. It must be used in conjunction with other methods of assessing clinical signs and symptoms. Use only Nonin-branded SenSmart oximeter signal processors, sensors, and accessories, otherwise patient injury can result. These sensors are manufactured to meet the accuracy specifications for this device. Using other manufacturers’ sensors can result in improper oximeter performance. Inspect the sensor application sites in accordance with the sensor instructions for use to ensure correct sensor alignment and skin integrity. Patient sensitivity to the sensor may vary due to medical status or skin condition. Avoid excessive pressure to the sensor application site(s) as this may cause damage to the skin beneath the sensor. Always inspect the device before use. Do not use a damaged device or sensor. Before using any sensor, carefully read the sensor instructions for use, which contains sensor application information for each sensor. To prevent improper performance and/or patient injury, verify compatibility of the monitor, sensor(s), and accessories before use. No modifications to this device are allowed as it may affect device performance. Verify all alarm settings and limits during system startup to ensure that they are set as intended. Protect from exposure to water or any other liquid, with or without AC power. Use the device only with Nonin-specified power supplies.
1
Indications for Use
Warnings (Continued) As with all medical equipment, carefully route patient cables and connections to reduce the possibility of entanglement, strangulation, or tripping. For remote monitoring, use the X-100M monitor only within its designated range (approximately 100 meters (300 feet) spherical radius from monitor to remote location). Moving outside this range may cause missing or lost data at the remote monitoring location. Memory is cleared if error code E06 appears on the display screen. The device presets are deleted if error code E09 appears on the display screen. The device configuration is deleted if error code E10 appears on the display screen. This device turns off after approximately 30 minutes when in low battery alarm condition. If this device is used adjacent to or stacked with other equipment, the device should be observed carefully to verify normal operation. The battery pack must be installed at all times while the device is operating-even when operating on AC power. Do NOT use the device without the battery pack installed. The use of signal processors, sensors, accessories, and cables other than those listed in the Parts and Accessories List may result in increased electromagnetic emission and/or decreased immunity of this device. The device’s Nurse Call and Bluetooth features should not be used as the primary source of alarm notification. The user must verify the device Bluetooth pairing to ensure the correct patient is remotely monitored. Ensure all alarm volumes are set appropriately and are audible in all situations. Do not cover or otherwise block any speaker openings. This device is a precision electronic instrument and must be repaired by qualified technical professionals. Field repair of the device is not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty. The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is hindering the pulse measurement before relying on the SpO2 measurement. Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results for SpO2. Readings of this device may be affected by the use of an electrosurgical unit (ESU). Keep electrosurgical/ electrocautery instruments away from the sensors and signal processors, as they may cause damage or result in erroneous readings. When defibrillation voltages are applied to the patient, the X-100M monitor may recover to a menu screen. The X-100M will return to the monitoring screen after the menu screen times out (2 minutes) or with operator interaction (press Menu button). It is the user’s responsibility to implement the interface between the nurse call system and the Model X-100M, and to adequately test the interface between the Model X-100M and the nurse call system to ensure operation. The X-100M monitor has not been evaluated with specific nurse call systems. Portable and mobile RF communications equipment can affect medical electrical equipment. Radios and cell phones or similar devices can affect the equipment and must be kept at least 2 meters (6.5 feet) away from the equipment. A hazard can exist if different presets are used on multiple X-100M monitors in one care area. Refer to the applicable sensor instructions for use for additional contraindications, warnings, and cautions.
2
Indications for Use
!
Cautions
The value of rSO2 data from the system has not been demonstrated in specific disease states, under conditions of hemoglobinopathies or clinical conditions that may affect blood volume, or under hypocapnic and hypercapnic conditions. When using this device in an operating room, it must remain outside the sterile field. When mounting the monitor to a mobile pole, mounting the monitor higher than 1.5 meters or mounting more than 2 kilograms of equipment onto the pole may result in tipping, damage to the equipment, or injury. This equipment complies with IEC 60601-1-2 for electromagnetic compatibility (EMC) for medical electrical equipment and/or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual. Exposure to Radio Frequency Radiation. The radiated output power of the display device is far below FCC radio frequency exposure limits. Nevertheless, the device must be used in such a way that the potential for human contact during normal operation is minimized. To avoid the possibility of exceeding FCC radio frequency exposure limits, remain at least 20 cm (8 in.) away from the display unit’s internal antenna during normal operation. The monitor has been tested and meets allowed limits for exposure. This device is designed to determine regional hemoglobin oxygen saturation of blood underneath the sensor when used with SenSmart regional oximetry sensors. Factors that may degrade performance or affect the accuracy of the measurement include the following: - excessive ambient light or direct sunlight - excessive motion - electrosurgical interference - metal plate or other foreign object in sensor path - moisture on skin - improperly applied sensor
- placement over bony prominence - incorrect sensor type - skin barriers used between sensor and patient skin - anemia or low hemoglobin concentrations - cardiogreen or other intravascular or tissue dyes
- carboxyhemoglobin and other dyshemoglobins - hemoglobinopathies - billirubinemia and/or icterus (jaundice) - non-normocapnic conditions or other conditions that affect blood volume
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin when used with SenSmart pulse oximetry sensors. Factors that may degrade pulse oximeter performance include the following: - excessive ambient light - excessive motion - electrosurgical interference - moisture in the sensor - improperly applied sensor - carboxyhemoglobin - methemoglobin
- blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.) - incorrect sensor type - poor pulse quality - venous pulsations
- anemia or low hemoglobin concentrations - cardiovascular dyes - dysfunctional hemoglobin - artificial nails or fingernail polish - residue (e.g., dried blood, dirt, grease, oil) in the light path
Batteries are a fire hazard if damaged. Do not damage, mishandle, disassemble, service, or replace with nonspecified components. Do not charge Li-Ion batteries at a temperature of 0 °C (32 °F) or less as this may result in significantly reduced battery life. Do not apply sensor over open wound, incision, or compromised skin. Inspect the sensor site(s) prior to applying the sensor(s). Inspect the sensor application sites in accordance with the sensor instructions for use to ensure correct sensor alignment and skin integrity. Patient sensitivity to the sensor may vary due to medical status or skin condition. Do not autoclave, sterilize, immerse, or spray this device with liquid or use caustic or abrasive cleaning agents. Do not use cleaning agents or cleaning products that contain ammonium chloride. 3
Indications for Use
!
Cautions (Continued)
Follow local, state and national governing ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries. Use only Nonin-approved battery packs. In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information. Data is written continuously to memory when the device is on. If the entire memory is filled, portions of the oldest record will be overwritten when new data is written. Setting alarm limits to extremes can render the alarm system useless. A preset cannot be saved with %rSO2 or %SpO2 low alarm limit settings that are lower than the institution default settings. %rSO2 and %SpO2 low alarm limits that are lower than the institution defaults will be replaced in the preset with the institution default setting(s). A 2-minute alarm silence is automatically engaged at startup. A functional tester cannot be used to assess the accuracy of the oximeter monitor or sensor. If this device fails to respond as described, discontinue use until the situation is corrected by qualified technical professionals. Between patients, turn the X-100M monitor off (Standby mode) or start a new case (Case Menu). Failure to do so could result in inaccurate baseline values for the new patient. When the device is turned on or a new case is started, the monitor clears the baseline values, resets the limits to the default values, and begins a new patient record in data memory. For SpO2 monitoring, the device may not work when circulation is reduced. Warm or rub the finger, or reposition the device. In some circumstances, the device may interpret motion as good pulse quality during SpO2 monitoring. Minimize patient motion as much as possible. Do not connect more than one hub in the system. The system will not function. Do not use an extension cable between the monitor and the hub. The system will not function. Duplicate signal processors cannot be used simultaneously and will result in an error message. Do not connect multiple extension cables between the monitor and a signal processor or between the hub and a signal processor. Do not simultaneously touch the accessible connector pins and the patient. The device has been designed for use within the specified ranges. Use outside of these ranges has not been tested and may result in improper oximeter performance. All parts and accessories connected to the serial port of this device must be certified according to at least IEC Standard EN 60950, IEC 62368-1, or UL 1950 for data processing equipment. Failure of a network data coupling (serial cable/connectors/wireless connections) will result in loss of data transfer. Refer to the applicable sensor instructions for use for additional contraindications, warnings, and cautions.
4
Indications for Use
Declaration of Conformity with FCC and Canadian Ministry of Health Rules for Electromagnetic Compatibility • Nonin Medical, Inc., of 13700 1st Avenue North, Plymouth, Minnesota, 55441, declares under its sole responsibility that the Model X-100M, to which this declaration relates, comply with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. • Ministry of Health (Canada), Safety Code 6: standards include a substantial safety margin designed to ensure the safety of all persons, regardless of age and health. The exposure standard for wireless mobile phones employs a unit of measurement known as the Specific Absorption Rate, or SAR. The SAR limit set by the FCC is 1.6 W/kg.
Federal Communications Commission (FCC) Notice This equipment has been tested and found to comply with the limits for a class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If not installed and used in accordance with the instructions, it may cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on. The user is encouraged to try to correct the interference by one or more of the following measures: (1) Reorient or relocate the receiving antenna, (2) Increase the distance between the equipment and the receiver, (3) Connect the equipment to an outlet on a circuit different from the outlet where the receiver is connected, or (4) Consult the dealer or an experienced radio/TV technician for assistance. The Model X-100M is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the Federal Communications Commission of the U.S. Government. These limits are part of comprehensive guidelines and establish permitted levels of RF energy for the general population. The guidelines are based on the safety standards previously set by both U.S. and international standards bodies. This EUT has been shown to be capable of compliance for localized specific absorption rate (SAR) for uncontrolled environment/general population exposure limits specified in ANSI/IEEE Std. C95.1-2005. The FCC requires the user to be notified that any changes or modifications to this device that are not expressly approved by Nonin Medical, Inc. may void the user’s authority to operate the equipment. NOTE: No modifications to this device are allowed that in any way affect or alter its antenna or antenna configuration.
5
Guide to Symbols
Guide to Symbols This chapter describes the symbols that are found on the Model X-100 system components and packaging. Detailed information about functional symbols can be found in "System Components and Set Up" on page 12. Table 1. Labeling and Packaging Symbols Symbol
! EC
REP
Description CAUTION! Authorized Representative in the European Community. Consult instructions for use. Follow instructions for use. UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical hazards in accordance with: • ANSI/AAMI ES60601-1 (2005, 3rd Ed.), CAN/CSA C22.2 No. 60601-1 (2008, 3rd Ed.), ISO 80601-2-61:2011, IEC 60601-1-8:2006 • UL 60601-1 and CAN/CSA C22.2 No. 601.1
0123
CE Marking indicating conformance to all applicable directives, including EC Directive No. 93/42/EEC concerning medical devices. Class II, double insulated
SN
Serial number
IP32
Protected against vertically falling water drops when enclosure is tilted up to 15 degrees and ingress of solid foreign objects greater than or equal to 2.5 mm in diameter, per IEC 60529. Direct current Type BF Applied Part (X-100M, X-100H) Indicates separate collection for waste electrical and electronic equipment (WEEE). Non-ionizing electromagnetic radiation. Equipment includes RF transmitters. Interference may occur in the vicinity of equipment marked with this symbol. Defibrillation-Proof Type BF Applied Part (patient isolation from electrical shock). (X-100SP)
6
Guide to Symbols
Table 1. Labeling and Packaging Symbols (Continued) Symbol
Description Do not discard. Do not pull on cable. Retract connector and remove. Lot number
REF
Catalogue number
QTY
Quantity Date of manufacture Manufacturer Non-sterile
NON STERILE
Storage/shipping temperature range Handle with care. Keep dry. Medical prescription required. Use by date
7
Guide to Symbols
1
2
5
6
7
3
4
8
15 19
14
16
10
17 18
9 11
12
13
Figure 1. Monitoring Screen Symbols (Four-Channel View) Table 2. X-100M Monitoring Screen Symbols and Indicators No.
Symbol
Description
1
example: Event Marks Located at the top of the monitoring screen, event marks (A, B, C, D, A B etc.) display when the Event Mark button is pressed.
2
example: Timescale 30MIN Located below the event marks, the timescale shows the amount of data time displaying on the screen.
3
Scrolling Cursor Located below the timescale, the yellow scrolling cursor allows the user to view a channel’s rSO2 or SpO2 reading at a specific time on the trendline. The scrolling cursor does not display until the Left navigation button has been pressed.
4 97 65
Cursor Values When the scrolling cursor is active, cursor oximetry (rSO2 or SpO2) values display on the left side of the monitor screen in a yellow box. NOTE: Pulse rate values are not displayed in the cursor values.
8
Guide to Symbols
Table 2. X-100M Monitoring Screen Symbols and Indicators (Continued) No.
Symbol
Description
5
Ch
Channel Located at the top of each channel, this indicator shows the channel’s number (e.g., Ch 1, Ch 2, etc.). If set, the sensor site name displays to the right of the channel indicator.
6
%rSO2 or %rSO2-T
Regional Hemoglobin Oxygen Saturation NOTE: %rSO2 displays when an absolute regional sensor is attached to a signal processor. %rSO2-T displays when a trending regional sensor is attached to a signal processor. %rSO2 displays from 0 to 100% when a signal processor receives an adequate signal from an attached regional sensor. The channel background display flashes: • Yellow during medium priority alarm conditions (equipment alarms and rSO2 values that are 5% or less above the rSO2 low alarm limit). • Red during high priority rSO2 alarm conditions (set by the high and low rSO2 alarm limits).
7
BL
Baseline When the monitor is turned on, the BL display shows dashes until the user sets the baseline values. The user must set the baselines for each new patient. • For instructions on setting the baseline values to the current rSO2 values, see "Set All rSO2 Channel Baselines to Current %rSO2 Values" on page 28. • For instructions on setting the baseline values separately or to make finite adjustments to the baseline value, see "Set Individual Baseline Values" on page 34.
8
AUC
Area Under the Curve (cumulative saturation below low alarm limit) For each channel, the rSO2 values below the low alarm limit are integrated together and displayed as the cumulative saturation below low alarm limit, also known as AUC (Area Under the Curve). The value is expressed in units of % minutes (%Min). When a baseline value is changed, the AUC recalculates from the beginning of the current record. The AUC will not calculate if a channel’s %rSO2 Low setting is OFF. NOTE: In order for the AUC display to match the Society of Thoracic Surgeons (STS) database definition, the low alarm limit value for each channel must be set to 25% below the patient's baseline.
9
Guide to Symbols
Table 2. X-100M Monitoring Screen Symbols and Indicators (Continued) No.
Symbol
Description
9
example: Low Alarm Limit
The low alarm limit is determined by the %rSO2 Low setting and displays as the following:
46 ...
• Numeric value – This value displays to the right of a trendline graph. The color of the value matches the color of the associated channel. • White, dotted line – This line only displays in a graph when the graph shows a single rSO2 trendline. The dotted line does not display when multiple trendlines are set to display in one graph. NOTE: A low alarm limit does not display on the monitoring screen if the channel’s %rSO2 Low setting is OFF.
10
%SpO2
Percent Functional Hemoglobin Oxygen Saturation %SpO2 data displays from 0 to 100% when a signal processor receives an adequate signal from an attached pulse oximetry sensor. The background of the SpO2 portion of the channel display flashes red during high priority SpO2 alarm conditions (set by the high and low SpO2 alarm limits) and low perfusion alarm conditions.
11
PR
Pulse Rate and Pulse Rate Bar Graph Pulse rate data displays along with the %SpO2 display when a channel is set up to measure pulse oximetry. The bar graph indicates pulse strength as determined by the oximeter. The height of the bar graph is proportional to the pulse amplitude. The background of the pulse rate portion of the channel display flashes red during high priority pulse rate alarm conditions (set by the high and low pulse rate alarm limits) and low perfusion alarm conditions.
12
Sensor Fault This yellow indicator flashes when a sensor is disconnected, has failed, has not received any usable data in the last 90 seconds, or is not compatible with the monitor.
13
Poor Signal This yellow indicator flashes when there has been a sustained period of poor patient signals from the sensor. Check the sensor site and reposition or replace the sensor if necessary.
14
Signal Processor Communication Error This yellow indicator flashes and the message X-100SP not connected displays when the respective signal processor has stopped communicating with the display. Check the signal processor connections or replace the signal processor to correct the issue. If the message appears in each channel, check the hub’s connection to the monitor.
10
Guide to Symbols
Table 2. X-100M Monitoring Screen Symbols and Indicators (Continued) No.
Symbol
15
Description Alarm Silence This yellow indicator flashes once every 2 seconds when the audible alarm is silenced for 2 minutes. If the alarm volume is at step 4 or lower (less than 45 decibels), the Alarm Silence indicator is solidly lit.
16
Bluetooth The Bluetooth indicator is green when Bluetooth is connected to a host, white when it is enabled but not connected, and gray when it is disabled. See "Enable/Disable Bluetooth Radio" on page 52 for more information.
17 Full
Empty l
Battery The battery indicator shows the approximate percentage of battery life remaining. When AC power is connected, the battery indicator fills up repeatedly to indicate the battery is charging. The indicator stops filling when the battery is fully charged. • Low – battery indicator flashes yellow • Critical – battery indicator flashes red NOTE: When the monitor reaches a low or critical battery condition, an audible alarm sounds. To clear the alarm, connect the monitor to the AC power adapter.
18
19
example: Date and Time The date and time display in 24-hour clock format. To set the date and/ 2013 or time, see "Set the Date and Time" on page 51. 09-25 14:27 rSO2 Axis Scale Fixed scale with a range from 0 to 100%.
11
System Components and Set Up
System Components and Set Up NOTES: • Before using the SenSmart system, please review all contraindications, warnings, and cautions. • Before using the Model X-100M for the first time, the battery should be charged for 4 hours. • Additional, but recommended, set-up tasks include: setting the clock, setting the institution defaults, changing the institution password, and setting up presets. Carefully remove the monitor and accessories from the shipping carton. Save the packaging materials in case the monitor or accessories must be returned. Compare the packing list with the accessories received to make sure the shipment is complete. The standard system configuration includes these non-sterile components: • X-100M, SenSmart universal oximetry monitor • X-100H, hub • X-100HH, hub holster • X-100SP-1 and X-100SP-2, oximetry signal processors for channels 1 and 2 (a garment clip is shipped with each signal processor) • X-100EC-1, 1-meter extension cable, quantity 2 • Operator’s manual/parts and accessories list (CD) • Power supply and cord • SenSmart download software (CD) For a list of compatible sensors and other accessories, see the Parts and Accessories List on the operator’s manual CD.
System Configurations Multiple Channels When using the X-100H hub (figure 2), up to six signal processors can be connected to the hub. If needed, an extension cable may be used between the hub and a signal processor.
Single Channel When using a single channel (figure 3), the signal processor can be connected directly to the monitor. If needed, an extension cable can be used between the monitor and the signal processor.
12
System Components and Set Up
X-100M 100 50 0 100 50 0 100 50 0
X-100H
X-100SP
Sensors
Figure 2. System Components and Set Up (4 Channels Connected)
X-100M 66
X-100SP
Sensor
Figure 3. System Set Up (Single Channel) 13