ENDOCAPSULE 10 SYSTEM Preparation and settings for capsule endoscopy This quick reference guide is only a summary of the ENDOCAPSULE instructions for use. For detailed operating instructions, be sure to follow the ENDOCAPSULE instruction manual that was included with your device when purchased. Before the examination
1
. Log in to workstation. Register 2 patient information and initialize recorder.
2
Refer to the back page Note: Check that the battery is fully charged (green light is lit).
1. Place recorder in the cradle.
4 3
3
3. Insert antenna sheet into the cover. Orange colored lines printed on the surface should face front. The cable always stays on the patient’s left side.
5
4. Fasten the antenna belt around the patient’s body. The alignment hole should fit the umbilical area.
8 3
7
7. Insert the capsule activator into the capsule container. The LED should start blinking.
6
5. Attach the recorder holder on the left side of the patient’s body. Connect the recorder and the antenna unit. Slide the connection part until it clicks.
6. Press “POWER” for at least 1 sec. Patient ID and name are displayed on the upper left screen.
IMPORTANT NOTICE TO PATIENTS
8. Press “MENU” and select “Real Time View”. Confirm that images are received properly.
When removing the capsule from the container, be careful NOT to touch the top cover of the capsule. Otherwise, clear images may not be captured.
After the capsule ingestion 9. Press “MENU” and select “Real Time View”. Confirm that images from the body are being received properly. If necessary, images can be played back or captured.
9
10
10. Lock the button operation. Go to “Patient Message” and press “MENU” for at least 2 sec. When patient guidance is activated, the patient does not necessarily need to touch the recorder. Beep or vibration will stop after 30 sec.
After the examination 11
12
13
11. Unlock the recorder. Press “MENU” for more than 2 sec. Open “Real Time View” to confirm that the examination is complete. 12. Place recorder in the cradle. 13. A message will be displayed on the screen. Click “OK” to start downloading image data. When completed, a check mark will be displayed on the main screen.
ENDOCAPSULE 10 SYSTEM
How to register patient information
New
Schedule Exam
Save & New
Save & Close
Enter required patient information. Minimum requirements are indicated by red * marks and red numbers. Click “Save & New” to continue registering. Click “Save & Close” to return to main screen. Note: In order to send patient messages, register from Patient Guidance tab.
How to initialize the recorder
Initialize recorder
OK
Q&A What is the recommendation for patient clothing?
How can I recognize when the battery needs to be replaced?
Wear thin clothing without any metal. Wearing a top or bottom with buttons or zippers is not recommended to avoid un-preferable effects for the antenna.
A message “Need new battery soon” will be displayed on the recorder. Be sure to replace it before the message “Replace Battery” pops up. Why does the display on the recorder disappear?
If the process is canceled, place the recorder on the cradle again and restart the workstation. The data transfer may be interrupted, but the data will be protected. How long does it take to recharge the battery? It takes about two hours. The green light will tell you when the battery has finished charging.
The display light will turn off to save battery power after 30 seconds without operation. How do I clean the equipment? The antenna unit holder and recorder holder can be machine-washed. Be sure to let them dry naturally and completely. Other equipment such as the antenna unit, recorder, cradle and capsule activator should have their surfaces wiped immediately after use with 70% ethyl or isopropyl alcohol.
Specifications, design, and accessories are subject to change without any notice or obligation on the part of the manufacturer.
Postbox 10 49 08, 20034 Hamburg, Germany Wendenstrasse 14–18, 20097 Hamburg, Germany Phone: +49 40 23773-0, Fax: +49 40 237765 www.olympus-europa.com
E0429376 · 300 · 02/14 · PR
If the recorder is removed from the cradle accidentally during initialization or downloading, will the data be affected?