ENDOSCOPE CLEANING GUIDE
All steps of manual cleaning for endoscope are done in detergent solution or water completely, except for flushing with air and drying the external surfaces at the end of cleaning.
EVIS EUS ULTRASOUND BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-UC190F/290F WARNING WARNING This workflow is a summary of the cleaning steps necessary to properly reprocess your BF-UC190F/290F and does not include disinfecting/sterilizing steps. Follow the detailed steps instructed in the latest ENDOSCOPE REPROCESSING MANUAL for BF-UC190F/290F.
Precleaning at Bedside
1
Leakage Testing
2
Remove the balloon, and wipe the insertion section
Aspirate water and air through the instrument/suction channel
Flush water and air through the balloon channel
Suction valve
3 Detach the suction and biopsy valve, and attach the US connector cap (MAJ-2295)
4
Confirm that air bubbles do not emerge from endoscope 30 sec
water : 30 ml ×3 air : 30 ml ×3
water : 10 sec air : 10 sec
Syringe Irrigation port
Manual Cleaning
5
Prepare the following brushes Cleaning brush
Thoroughly wipe all external surfaces in the detergent solution
6
Brush from the suction cylinder to the distal end
or BW-411B
7
Brush from the instrument channel port to the distal end
Brush the suction cylinder
BW-411B or BW-15B
BW-411B or BW-15B
BW-15B
8
BW-411B or MH-507
or BW-7B
BW-400B
Channel opening cleaning brush or BW-411B
MH-507
Manual Cleaning
9 Brush the instrument
10 Brush from the irrigation port
channel port
11 Brush the irrigation port (balloon channel)
(balloon channel) to the distal end
BW-411B or MH-507
BW-400B or BW-7B
BW-411B or MH-507
12 Brush the balloon grooves
BW-411B or BW-15B
BW-411B or MH-507
Manual Cleaning
13 Aspirate the detergent solution (with attaching MAJ-222)
14 Completely immerse the endoscope and the suction cleaning adapter (MAJ-222) in the detergent solution
MAJ-222
15 Aspirate the detergent solution
16 Flush the detergent solution (through the balloon channel)
30 ml ×1
30 ml ×1
Suction pump
30 sec Detergent solution
Manual Cleaning
Rinsing
17 Wipe all external surfaces and
18 Completely immerse and gently
leave in the detergent solution
rinse the endoscope and MAJ-222 in the water
19 Aspirate water, air
water : 30 sec air : 20 sec
20 Dry the external surfaces by wiping with clean cloths
Suction pump
Flush water, air water : 30 ml ×5 air : 30 ml ×5