INSTRUCTIONS
GRASPING FORCEPS
FG-4L-1 FG-14P-1 FG-20P-1 FG-21L-1 FG-25C-1 FG-26C-1 FG-32C/L-1 FG-38SX-1
AUTOCLAVABLE USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Contents
Contents Symbols...
1
Important Information - Please Read Before Use 2 Intended Use ... Instruction Manual ... User Qualifications... Instrument Compatibility ... Reprocessing and Storage ... Repair and Modification ... Signal Words ...
Chapter 1 1.1
Checking the Package Contents...
5
Checking the Package Contents ...
5
Chapter 2 2.1 2.2
Instrument Nomenclature and Specifications ...
6
Nomenclature and Functions... Specifications...
6 7
Chapter 3 3.1 3.2 3.3
17 18 21
Reprocessing... 29
General Policy ... Required Reprocessing Equipment ... Cleaning... Lubrication ... Sterilization ...
Chapter 5 5.1 5.2
Preparation, Inspection and Operation ... 15
Preparation ... Inspection ... Operation ...
Chapter 4 4.1 4.2 4.3 4.4 4.5
2 3 3 3 4 4 4
29 33 35 36 37
Storage ... 40
Inspection Before Storage ... Storage ...
GRASPING FORCEPS
40 41
i
Contents
ii
GRASPING FORCEPS
Symbols
Symbols The meaning(s) of the symbol(s) shown on the package, the back cover of this instruction manual and/or this instrument are as follows: Refer to instructions. Manufacturer Authorized representative in the European Community Lot Number Keep away from sunlight
Keep dry
Not Made with Natural Rubber Latex 70C -40C
Temperature limitation
GRASPING FORCEPS
1
Important Information - Please Read Before Use
Important Information - Please Read Before Use Intended Use Do not use these instrument for any purpose other than their intended use.
FG-4L-1, FG-14P-1, FG-38SX-1 This instrument has been designed to be used with Olympus endoscopes in retrieving foreign bodies or tissue specimens from within the digestive tract, upper airways, tracheobronchial tree, female reproductive and urinary organs.
FG-25C-1, FG-26C-1, FG-32C/L-1 This instrument has been designed to be used with Olympus endoscopes in retrieving foreign bodies or tissue specimens from within the digestive tract and respiratory organs.
FG-20P-1 This instrument has been designed to be used with Olympus endoscopes in retrieving foreign bodies from within the digestive tract, upper airways, tracheobronchial tree, female reproductive and urinary organs.
FG-21L-1 This instrument has been designed to be used with Olympus endoscopes in retrieving foreign bodies from within the digestive tract. 2
GRASPING FORCEPS
Important Information - Please Read Before Use
Instruction Manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment which will be used during the procedure and use the instruments as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus.
User Qualifications The operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique. This manual, therefore, does not explain or discuss clinical endoscopic procedures.
Instrument Compatibility Refer to the tables in Section 2.2, “Specifications” on page 7 to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient injury or equipment damage.
GRASPING FORCEPS
3
Important Information - Please Read Before Use
Reprocessing and Storage This instrument was not sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions in Chapter 4, “Reprocessing”. After using this instrument, reprocess and store it according to the instructions in Chapter 4, “Reprocessing” and Chapter 5, “Storage”. Improper and/or incomplete reprocessing can present an infection control risk, cause equipment damage or reduce performance.
Repair and Modification This instrument contains no user-serviceable parts. Do not disassemble, modify or attempt to repair; patient or user injury and equipment damage can result.
Signal Words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage. Indicates additional helpful information. 4
GRASPING FORCEPS
Chapter 1 Checking the Package Contents
Chapter 1 1.1
Checking the Package Contents
Checking the Package Contents Match all items in the package with the components shown below. Inspect each item for damages. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions in Chapter 4, “Reprocessing”.
Grasping Forceps
Instruction Manual
GRASPING FORCEPS
5
Chapter 2 Instrument Nomenclature and Specifications
Chapter 2
2.1
Instrument Nomenclature and Specifications
Nomenclature and Functions
Lot Number (five digits) – Identification Number (four digits) The forceps can be individually recognized by an Identification Number. Ring The color of the Ring indicates the minimum instrument channel diameter required for the endoscope to be compatible.
Slider Push the Slider to open the Grasping Jaws. Pull to close. A green reference label indicates that the instrument is autoclavable. Handle Model Reference Label Indicate the product number.
Insertion Portion /Working Length Made from a metallic coil.
6
GRASPING FORCEPS
Forceps Cap Prevents damage to the distal end of during transportation. (Exclude FG-4L-1)
Grasping Jaws Used to grasp foreign body or tissue.
Chapter 2 Instrument Nomenclature and Specifications
Distal End
FG-4L-1, FG-25C-1
Rubber (Non latex) Prevents a grasped object from slipping.
FG-14P-1, FG-26C-1
Sheet Prevents the rubber chips from fastening when the instrument is transported.
FG-20P-1, FG-21L-1
2.2
FG-32C/L-1, FG-38SX-1 (Have a sharper tooth than FG-14P-1, FG-26C-1)
Specifications The compatible Olympus endoscopes are listed in the following table. New endoscopes released after the introduction of this instrument may also be compatible for use in combination with this instrument. For further details, contact Olympus. Use this instrument only in combination with products recommended by Olympus. If combined with products not recommended by Olympus, injury, malfunction or equipment damage may result. GRASPING FORCEPS
7
Chapter 2 Instrument Nomenclature and Specifications
Operating Environment Ambient Temperature
10 – 40C (50 – 104F)
Relative Humidity
30 – 85%
Air Pressure
700 – 1060 hPa (0.71 – 1.08 kgf/cm2) (10.1 – 15.4 psia)
Specifications Model
FG-14P-1
Shape of the
Rat Tooth
Grasping Jaws
Maximum Insertion Portion
ø 1.85
Diameter (mm) Working Length (mm) Opening Width (mm)
8
GRASPING FORCEPS
1900 3.0
Chapter 2 Instrument Nomenclature and Specifications
Compatible
Model and
Working length
Olympus
Length
less than 1450 mm;
Endoscopes
EF, GIF, OGF, GF, GTF, JF, TJF,
(All of these
CHF, CF (Exclude L-length),
parameters
PCF (Exclude L-length), OSF, BF,
should be met.)
ENF, CYF, HYF Channel
ø 2, ø 2.2 (Blue);
Inner
ø 2.6 (Green);
Diameter
ø 2.8 (Green or Yellow);
(mm)
ø 3.2 (Yellow);
(Color Code)
ø 3.7, ø 4.2, ø 6 (Orange); ø 5.5 (Pink)
GRASPING FORCEPS
9
Chapter 2 Instrument Nomenclature and Specifications
Model
FG-26C-1
Shape of the
Rat Tooth
Grasping Jaws
Maximum Insertion
ø 2.2
Portion Diameter (mm) Working Length
1050
(mm) Opening Width
3.5
(mm) Compatible
Model and
Working length
Olympus
Length
less than 600 mm;
Endoscopes
CHF, OSF, BF
(All of these
Channel
ø 2.6 (Green);
parameters
Inner
ø 2.8 (Green or Yellow)
should be met.)
Diameter (mm) (Color Code) Other
10
GRASPING FORCEPS
Exclude BF-XT20/30
Chapter 2 Instrument Nomenclature and Specifications
Model
FG-20P-1
FG-21L-1
Shape of the
Rubber Tips (Non latex)
Grasping Jaws
Maximum Insertion Portion
ø 1.85
ø 2.4
1900
1650
4.8
7.3
Diameter (mm) Working Length (mm) Opening Width (mm) Compatible
Model and
Working length
Working length
Olympus
Length
less than
less than
Endoscopes
1450 mm;
1200 mm;
(All of these
EF, GIF, OGF,
EF, GIF, OGF,
parameters
GF, GTF, JF,
GF, GTF, CF
should be met.)
TJF, CHF, CF
(Exclude I-,
(Exclude
L-length), OSF
L-length), PCF (Exclude L-length), OSF, BF, ENF, CYF, HYF Channel
ø 2,ø 2.2
ø 2.8, ø 3.2
Inner
(Blue);
(Yellow);
Diameter
ø 2.6 (Green);
ø 3.7, ø 4.2,
(mm)
ø 2.8 (Green or
ø 6 (Orange)
(Color Code)
Yellow); ø 3.2 (Yellow); ø 3.7, ø 4.2, ø 6 (Orange); ø 5.5 (Pink)
GRASPING FORCEPS
11
Chapter 2 Instrument Nomenclature and Specifications
Model
FG-32C-1
Shape of the
FG-32L-1
Sharp Tooth
Grasping Jaws
Maximum Insertion
ø 2.6
Portion Diameter (mm) Working Length
1050
(mm) Opening Width
1650 4.7
(mm) Compatible
Model and
Working length
Working length
Olympus
Length
less than
less than
Endoscopes
600 mm;
1200 mm;
(All of these
OSF, BF
EF, GIF, OGF,
parameters
GF, GTF, CF
should be met.)
(Exclude I-, L-length), OSF, BF Channel
ø 2.8,ø 3.2
ø 2.8, ø 3.2
Inner
(Yellow)
(Yellow);
Diameter
ø 3.7, ø 4.2,
(mm)
ø 6 (Orange)
(Color Code)
12
GRASPING FORCEPS
Chapter 2 Instrument Nomenclature and Specifications
Model
FG-4L-1
FG-38SX-1
Shape of the
W Shape
Sharp Tooth
Grasping Jaws
Maximum Insertion
ø 1.85
Portion Diameter (mm) Working Length (mm) Opening Width (mm)
1650
700
13.0
5.0
Compatible
Model and
Working length
Working length
Olympus
Length
less than
less than
Endoscopes
1200 mm;
400 mm;
(All of these
GIF, OGF,
CHF, ENF,
parameters
CHF, CF
CYF, HYF
should be met.)
(Exclude I-, L-length), OSF, ENF, CYF, HYF Channel
ø 2, ø 2.2
ø 2, ø 2.2
Inner
(Blue);
(Blue);
Diameter
ø 2.6 (Green);
ø 2.6 (Green)
(mm)
ø 2.8, ø 3.2
(Color Code)
(Yellow); ø 3.7, ø 6 (Orange)
Other
Direction of View: 0 (Forward
–
viewing) only.
GRASPING FORCEPS
13
Chapter 2 Instrument Nomenclature and Specifications
Model
FG-25C-1
Shape of the
W Shape
Grasping Jaws
Maximum Insertion
ø 2.2
Portion Diameter (mm) Working Length
1050
(mm) Opening Width
13.0
(mm) Compatible
Model and
Working length less than
Olympus
Length
600 mm;
Endoscopes
CHF, OSF, BF
(All of these
Channel
ø 2.6 (Green); ø 2.8 (Green or
parameters
Inner
Yellow); ø 3.2 (Yellow);
should be met.)
Diameter (mm) (Color Code) Other
Direction of View: 0 (Forward viewing) only.
Medical Devices
This device complies with the
Directive
requirements of Directive 93/42/EEC concerning medical devices. Classification: Class I
14
GRASPING FORCEPS
Chapter 3 Preparation, Inspection and Operation
Chapter 3
Preparation, Inspection and Operation •
Before each case, prepare and inspect this instrument as instructed below. Inspect other equipment used with this instrument as instructed in their respective instruction manuals. Should the slightest irregularity be suspected, do not use the instrument; contact Olympus. Damage or irregularity may compromise patient or user safety as well as punctures, hemorrhages or mucous membrane damage and may result in more severe equipment damage.
•
This instrument was not sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions in Chapter 4, “Reprocessing”. Do not use an instrument that has not been cleaned and sterilized. This poses an infection control risk and can cause tissue irritation.
GRASPING FORCEPS
15
Chapter 3 Preparation, Inspection and Operation
16
•
Before use, confirm that the instrument is not corroded, dented or discolored ; do not use it if any of these conditions are observed. If the instrument is damaged, part of it could fall off inside the patient, or it could become impossible to open or close.
•
If the instrument does not operate properly during the procedure, or if it becomes difficult to open or close the grasping jaws, stop the procedure immediately and withdraw the instrument into the endoscope’s channel. If this is not possible, carefully remove it together with the endoscope to avoid causing patient injury.
•
If a part of the instrument falls off inside the patient, use a spare instrument to retrieve it.
•
Do not coil the Insertion Portion in a diameter of less than 15 cm. This could damage the Insertion Portion.
•
Never use excessive force to open or close the Grasping Jaws. This could damage the instrument.
GRASPING FORCEPS
Chapter 3 Preparation, Inspection and Operation
3.1
Preparation Equipment and Personal Protective Equipment Prepare all equipment and personal protective equipment which will be used with the instrument in accordance with their respective instruction manuals. Appropriate personal protective equipment may include: Eye wear, a face mask, moisture-resistant clothing and chemical-resistant gloves, etc.
Spare Instrument Always have a spare instrument available.
Reprocessing Equipment Prepare reprocessing equipment as described in Section 4.2, “Required Reprocessing Equipment” for immediate reprocessing after use.
GRASPING FORCEPS
17