INSTRUCTIONS FOR USE GRASPING FORCEPS
O0121 O0122 O0123 O0124 O0126
O0125 O0127 O0129
Contents 1 General information... 5 1.1 Manufacturer... 6 2 Technical data... 7 3 Use... 8 4 Reprocessing... 10
General information
1 General information Intended use O0121, O01222, O0125 Grasping forceps for endoscopic diagnosis and treatment in urologic applications. O0123, O0124 Biopsy forceps for endoscopic diagnosis and treatment in urologic applications. O0126 Cannula for endoscopic diagnosis and treatment in urologic applications. O0127 Ureter stone dislodger for endoscopic diagnosis and treatment in urologic applications. O0129 Scissors for endoscopic diagnosis and treatment in urologic applications. Do not use it for any other purpose. Medical use This product is to be used only in a medical facility by trained medical personnel. NOTE Instruction manual set The complete instruction manual set for this product consists of this product-specific instruction manual (delivered together with this product) and the system-related instruction manual, entitled “System Guide Endoscopy” (supplied with Olympus telescopes). Immediately contact an Olympus representative if either manual is missing or lost. Keep the instruction manual set in a safe, accessible location.
5
General information WARNING! Read all instruction manuals thoroughly Before use, read this manual, the System Guide Endoscopy, and the manuals for all other equipment which will be used during the procedure. An insufficient understanding of the dangers, warnings, cautions, and information in these manuals can result in death, serious injury, or equipment damage. CAUTION! Wear When used as intended this product is more or less subject to wear depending on the intensity of use. In case a product has externally visible defects contact the responsible Olympus representative at once. Later guarantee claims are forfeited. Any guarantee claims are void by previous use.
1.1 Manufacturer Olympus Winter & Ibe GmbH Kuehnstr. 61 22045 Hamburg Germany
6
Technical data
2 Technical data Diameter O0121, O0123, O0126... 7 Fr. O0122, O0124, O0125, O0129... 9 Fr. O0127... 10 Fr. Working length O0121, O0122, O0123, O0124, O0125, O0126, O0127, O0129 ... 330 mm
7
Use
3 Use WARNING! Infection control risk Properly reprocess the product before use following the instructions in this manual and in the System Guide Endoscopy. Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel. CAUTION! Risk of damaging the instrument Hand instruments may be damaged if excess force is applied or if the instrument is used in a manner for which it is not intended. In order to avoid damage, make sure that the force applied to the handle is adequate and reasonable for the intended application. Inserting a surgical instrument
•• In the case of surgical instruments with jaws: close the jaws. •• Insert the distal end of the surgical instrument into the sealing cap on the working channel of the working insert or bridge. •• Open the instrument stopcock. •• Advance the surgical instrument into the working channel until the distal end appears in the endoscopic image.
8
Use CAUTION! Risk of damaging the instrument Make sure that the instrument stopcock is open and the jaws are closed when withdrawing the instrument. Never withdraw the instrument with excessive force.
9
Reprocessing
4 Reprocessing WARNING! Infection control risk Properly reprocess the product before first and each subsequent use following the instructions in this manual and in the System Guide Endoscopy. Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel. Observe the following instrument-specific instructions. Information in this section The following reprocessing information supplements that in the System Guide Endoscopy. However, if the information given in this section differs from that in the System Guide Endoscopy, use the information given here. Compatible reprocessing methods This product is compatible with the reprocessing methods listed in the table. For details of the procedures, refer to the System Guide Endoscopy. + -
compatible incompatible O0121, O0122, O0123, O0124, O0125, O0126, O0127, O0129
10
Manual or automatic cleaning
+
Ultrasonic cleaning
+
Disinfection or high-level disinfection
+
Steam sterilization (autoclaving)
+
Heat sterilization (dry)
-
Gas sterilization (ETO)
+
Gas sterilization (formaldehyde)
+
STERRAD 50/100/100S
+
STERIS
-
Reprocessing CAUTION! Risk of damaging the product Do not use incompatible reprocessing methods. Product damage may occur. Disassembling
•• Disassemble the instrument before reprocessing. Lubrication
•• Lubricate the areas indicated by arrows. •• Use only Olympus A0273. Apply lubricant sparingly.
11
W7512700_2-0 2016-03-09 en