usg-400_ultrasonic_generator_instructions_jan_2016_.pdf
Page 1
INSTRUCTIONS
Ultrasonic Generator
Labels and Symbols
1
Important Information — Please Read Before
Use
3
Chapter 1
Inspecting the Package Contents
13
Chapter 2
Nomenclature and Functions
15
Chapter 3
Installation and Connections
29
Chapter 4
Inspection
55
Chapter 5
Operation
79
Chapter 6
Push Button Functions
95
Chapter 7
Care, Storage, and Disposal
119
Chapter 8
Troubleshooting
123
USG-400
Appendix
USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
147
Page 3
Contents
Contents
Labels and Symbols .......................................................................................................... 1
Important Information — Please Read Before Use ......................................................... 3
Intended use ..........................................................................................................................
Application of high-frequency treatment ................................................................................
Instruction manual .................................................................................................................
User qualifications .................................................................................................................
Ultrasonic generator compatibility ..........................................................................................
Repair and modification .........................................................................................................
Signal words ..........................................................................................................................
Dangers, warnings, and cautions ..........................................................................................
3
3
3
3
4
4
4
5
Chapter 1 Inspecting the Package Contents ..................................... 13
1.1
Inspecting the package contents .......................................................................... 13
Chapter 2 Nomenclature and Functions ............................................ 15
2.1
Symbols and descriptions ..................................................................................... 15
2.2
Front panel .............................................................................................................. 17
2.3
Rear panel ............................................................................................................... 18
2.4
Top panel ................................................................................................................. 19
2.5
All screen (touch screen) ...................................................................................... 20
2.6
THUNDERBEAT instrument set screen (touch screen) ...................................... 21
2.7
SONICBEAT instrument set screen (touch screen) ............................................ 22
2.8
Footswitch for THUNDERBEAT (MAJ-1870, optional) ........................................ 23
2.9
Footswitch for SONICBEAT (MAJ-1869, optional) ............................................... 24
2.10 Docking fixture (MAJ-1876, optional, this is required for the connection
of the ultrasonic and compatible electrosurgical generator) ............................. 25
2.11 Communication cable 0.25 m (MAJ-1871, optional, this cable is required
for the connection with the compatible electrosurgical generator) .................. 26
2.12 Communication cable 10 m (MAJ-1872, optional, this cable is required for
the connection with the compatible high flow insufflation unit) ....................... 27
2.13 Adapter for UHI-2/3 (MAJ-1873, optional, this adapter is required for the
connection with the compatible high flow insufflation unit) .............................. 28
USG-400 INSTRUCTION MANUAL
i
Page 4
Contents
Chapter 3 Installation and Connections ............................................. 29
3.1
Installation workflow .............................................................................................. 29
3.2
Installation of the ultrasonic generator ................................................................ 30
3.3
Connection of compatible electrosurgical generator
(when using the THUNDERBEAT instrument) ..................................................... 31
3.4
Connection to power supply (Ultrasonic Generator) .......................................... 40
3.5
Automatic mist & smoke evacuation system/function
(when using the compatible high flow insufflation unit) .................................... 42
High flow insufflation unit (UHI-4) ........................................................................................ 42
High flow insufflation unit (UHI-2, UHI-3) ............................................................................. 44
When adding the compatible electrosurgical generator to the compatible high flow
insufflation unit and the ultrasonic generator ....................................................................... 46
3.6
Connection of footswitch (optional) ..................................................................... 49
3.7
Connection of the THUNDERBEAT and/or SONICBEAT instrument
(various options available) .................................................................................... 52
Chapter 4 Inspection ............................................................................ 55
4.1
Verification workflow .............................................................................................. 55
4.2
Power verification ................................................................................................... 57
Checking the illumination of power switch, touch screen, and push buttons ....................... 58
Checking the start screen and start tone ............................................................................. 58
Checking the touch screen display ...................................................................................... 59
4.3
Verification of communication cable connection between compatible
electrosurgical generator and ultrasonic generator
(when using the THUNDERBEAT instrument) ..................................................... 60
When the communication indicator is illuminated ................................................................ 60
If the communication indicator is not illuminated ................................................................. 61
4.4
Verification of touch screen and push buttons ................................................... 61
Checking the touch screen operation .................................................................................. 62
Checking the push button operations .................................................................................. 66
4.5
Verification of the THUNDERBEAT and/or SONICBEAT instrument
connection(s) .......................................................................................................... 67
Verify the symbol for the connected THUNDERBEAT and/or SONICBEAT instruments .... 68
4.6
Verification of footswitch connection ................................................................... 71
Checking the illumination of footswitch indicator(s) ............................................................. 71
4.7
Verification of the high-frequency (RF bipolar) output
(when using the THUNDERBEAT instrument) ..................................................... 72
Verifying the output level ..................................................................................................... 73
Verifying the output .............................................................................................................. 74
ii
4.8
Verification of the alarm system ........................................................................... 76
4.9
Procedure after verification ................................................................................... 78
USG-400 INSTRUCTION MANUAL
Page 5
Contents
Chapter 5 Operation ............................................................................. 79
5.1
Operating steps ...................................................................................................... 79
5.2
Turning the ultrasonic generator ON .................................................................... 80
5.3
Automatic mist & smoke evacuation system/function
(when using the compatible high flow insufflation unit) .................................... 81
5.4
Output setting ......................................................................................................... 83
Output mode list .................................................................................................................. 86
Correlation between cutting speed and output level in SEAL & CUT mode ........................ 88
5.5
Energy output ......................................................................................................... 90
Energy output ...................................................................................................................... 92
5.6
Steps after use ........................................................................................................ 94
Chapter 6 Push Button Functions ...................................................... 95
6.1
Function list ............................................................................................................ 95
6.2
Push button hierarchy ........................................................................................... 96
6.3
Probe check ............................................................................................................ 97
6.4
Select procedure .................................................................................................. 100
6.5
Menu ...................................................................................................................... 102
Save Procedure: Saving/overwriting of output level .......................................................... 103
Delete Procedure: Deleting a procedure ............................................................................ 110
Select Language ................................................................................................................. 112
Touch Tone ......................................................................................................................... 114
Software Version ................................................................................................................ 115
Service ................................................................................................................................ 116
Output Volume Control ....................................................................................................... 116
Brightness Control .............................................................................................................. 117
Chapter 7 Care, Storage, and Disposal ............................................ 119
7.1
Care ....................................................................................................................... 119
7.2
Storage .................................................................................................................. 121
7.3
Disposal ................................................................................................................ 121
USG-400 INSTRUCTION MANUAL
iii
Page 6
Contents
Chapter 8 Troubleshooting ................................................................ 123
8.1
Troubleshooting ................................................................................................... 123
8.2
Error screen .......................................................................................................... 124
8.3
Troubleshooting guide ......................................................................................... 124
Types of error codes ..........................................................................................................
Pattern 1 Target error code: U001-U014 ..........................................................................
Pattern 2 Target error code: U501-U502, U507 ................................................................
Pattern 3 Target error code: U504 ....................................................................................
Pattern 4 Target error code: U508, U513 .........................................................................
Pattern 5 Target error code: U509, U510, U511, U512, U514, and U601 ........................
Pattern 6 Target error code: U503, U505, U506 ...............................................................
What to do when no error code is displayed .....................................................................
What to do when an error code is displayed .....................................................................
8.4
124
125
126
127
131
132
133
134
140
Returning the ultrasonic generator for repair .................................................... 146
Appendix ............................................................................................... 147
System chart .................................................................................................................. 147
Specifications ................................................................................................................. 149
Environment ...................................................................................................................... 149
Specifications ..................................................................................................................... 149
EMC information ............................................................................................................ 153
Alarm information .......................................................................................................... 158
Tone information ............................................................................................................ 158
License information of Open Source Software ........................................................... 159
GNU GPL ..........................................................................................................................
End User License Agreement of Open Source Software modules ....................................
Others ................................................................................................................................
newlib ................................................................................................................................
iv
USG-400 INSTRUCTION MANUAL
159
159
160
160
Page 7
Labels and Symbols
Labels and Symbols
Safety related labels and symbols are shown below. If labels or symbols are missing or illegible,
contact Olympus.
View of docking plug
with cover.
View of docking plug
without cover.
Consult instructions
for use
Caution
Crossed-out wheeled
bin symbol
(Only EU model)
Blue
Refer to instruction manual/booklet
Intelligent Tissue
Monitoring ID
sticker
Potential equalization
terminal
Serial number
symbol
Electrical rating label
Serial number plate
Fuse rating
Manufacturing name label
CSA/UL marking
(Only USA model)
CE marking
(Only EU model)
Electrical rating label
POWER INPUT : 120 V
50/60 Hz 360 VA
: 1 min/1 min
INTERVAL
120 V (USA model)
USG-400 INSTRUCTION MANUAL
POWER INPUT : 220 - 240 V
50/60 Hz 360 VA
INTERVAL
: 1 min/1 min
220 - 240 V (EU model)
1
Page 8
Labels and Symbols
Back cover of this instruction manual
Symbol
Description
Manufacturer
Authorized representative in the European Community
Fonts displayed on the touch screen
Ricoh bitmap fonts designed by RICOH Company, Ltd. are used.
Package
Symbol
Description
Keep away from sunlight
Keep dry
+70C
-25C
2
Temperature limitation
USG-400 INSTRUCTION MANUAL
Page 9
Important Information — Please Read Before Use
Important Information — Please Read Before Use
Intended use
The Ultrasonic Generator (USG-400) is intended to be used with the Electrosurgical Generator
(ESG-400), the THUNDERBEAT Transducer (TD-TB400), the SONICBEAT Transducer (TD-SB400),
the THUNDERBEAT, and / or the SONICBEAT for open, laparoscopic (including single-site surgery),
and endoscopic surgery to cut (dissect) or coagulate soft tissue or to ligate (seal and cut) vessels.
Application of high-frequency treatment
Before proceeding to high-frequency treatment, study well the diagnosis given and the expected
prognosis, the properties and purpose of the treatment, its risks and effects, and the possible
alternative treatments.
Particularly, balance the risks of the treatment with the potential benefits of a given procedure.
Instruction manual
This instruction manual contains essential information on using the ultrasonic generator safely and
effectively. Before use, thoroughly review this manual and the manuals for all instruments that will be
used during the procedure. Use the instrument as instructed.
Keep all instruction manuals in a safe, accessible location.
If you have any questions or comments about any information in this manual, contact Olympus.
User qualifications
This manual does not explain or discuss clinical surgical procedures. Therefore, the health care
professional using the ultrasonic generator must be a licensed physician or medical personnel under
the supervision of a licensed physician and must have received sufficient training in clinical procedures
and the use of ultrasonic and electrosurgical energy.
USG-400 INSTRUCTION MANUAL
3
Page 10
Important Information — Please Read Before Use
Ultrasonic generator compatibility
The ultrasonic generator should be used in combination with the ancillary instrument listed in “System
chart” on page 147.
Using incompatible instruments can result in injury, equipment damage, and/or decreased product
performance.
This equipment complies with EMC standard for medical electrical equipment, edition 2
(IEC 60601-1-2: 2001) and edition 3 (IEC 60601-1-2: 2007). However, when connecting to an
instrument that complies with EMC standard for medical electrical equipment, edition 1
(IEC 60601-1-2: 1993), the whole system complies with edition 1.
When the ultrasonic generator is combined with the compatible electrosurgical generator, the EMC
information shall comply with the compatible electrosurgical generator.
Repair and modification
The ultrasonic generator does not contain any user-serviceable parts. Do not disassemble, modify, or
attempt to repair it. Patient or user injury and/or equipment damage can result.
If an irregularity appears to be minor, refer to Chapter 8, “Troubleshooting”. If the irregularity cannot be
resolved, do not use the ultrasonic generator and contact Olympus.
Signal words
The following signal words are used throughout this manual:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor
or moderate injury. It may also be used to alert against unsafe practices or potential
equipment damage.
NOTE
4
Indicates additional helpful information.
USG-400 INSTRUCTION MANUAL
Page 11
Important Information — Please Read Before Use
Dangers, warnings, and cautions
Follow the dangers, warnings, and cautions described below when handling this ultrasonic generator.
This information is supplemented by the dangers, warnings, and cautions described in each chapter.
Interference between devices
WARNING
• Use care when installing the ultrasonic generator near, on, or under another device
(a device other than the ultrasonic generator or a series of components of the
system including this ultrasonic generator). Otherwise, electromagnetic
interference may cause a malfunction or failure of the ultrasonic generator and/or
other devices.
• For safety, do not use the ultrasonic generator in combination with the following
medical electrical instrument:
Medical electrical instrument that has not been confirmed to be safe when it is
used in combination with the ultrasonic generator.
Medical electrical instrument that has not been confirmed to be safe with regard
to leakage current, etc.
• Always use the supplied power cord. Use of any other power cord may cause a
malfunction of the instrument or burns on the power cord. Additionally, the supplied
power cord is for exclusive use with the ultrasonic generator and must not be used
with other devices.
• Strictly observe the following cautions when handling the ultrasonic generator and
the transducer cord(s).
Do not loop cords, but separate them so that they will not contact the main
bodies or cords of other medical electrical instrument (electrocardiograph,
endoscopic video camera, etc.). Otherwise, the high-frequency (RF bipolar)
signal generated from the ultrasonic generator and the noise due to spark
discharge during coagulation may be enhanced, causing malfunction including
output shutdown of the ultrasonic generator as well as various adverse effects
to other medical electrical instrument.
Separate cords so they will not contact the patient or the metal parts of the
operating table. Also, make sure that the patient does not contact the metal part
of the operating table or other devices. Otherwise, the patient may suffer burns.
It is also recommended to use an electrically non-chargeable bed sheet.
Separate the other devices that may cause electromagnetic interference.
• To prevent malfunction, do not use the ultrasonic generator in a location exposed to
a strong electromagnetic field (near a microwave therapeutic device, short-wave
therapeutic device, MRI, wireless set or cellular phone, etc.).
USG-400 INSTRUCTION MANUAL
5
Page 12
Important Information — Please Read Before Use
CAUTION
Electromagnetic interference may occur on this instrument near equipment marked
with the following symbol or other portable and mobile RF (Radio Frequency)
communications equipment such as cellular phones. If electromagnetic
interference occurs, mitigation measures may be necessary, such as reorienting or
relocating this equipment, or shielding the location.
Temperature and humidity during operation and storage
CAUTION
Use the ultrasonic generator in compliance with the “ Environment” on page 149.
Otherwise, the ultrasonic generator may fail.
Mechanical stress on accessories
CAUTION
Do not apply excessive force to the cords/cables by bending, straining, or twisting.
Doing so may damage the cords/cables and cause the instrument to malfunction.
Damage of accessories
CAUTION
Always inspect the accessories as instructed in the instruction manuals before use.
Confirm that the insulation on the transducer cord is not damaged. Do not use an
accessory if any irregularity is observed. Otherwise, malfunction, human injury,
and/or equipment damage may result.
6
USG-400 INSTRUCTION MANUAL
Page 13
Important Information — Please Read Before Use
Grounding
WARNING
• The exterior of the ultrasonic generator should be grounded to ensure safety. To
prevent an electric shock hazard, be sure to connect the power cord plug to a
properly grounded hospital-grade 3-conductor power outlet. Do not use an adapter
for converting a 3-conductor power plug into a 2-conductor plug as this reduces the
safety.
• Use a hospital-grade 3-conductor power outlet with a sufficient capacity for all
devices connected to the same power supply. If the power capacity is insufficient, a
fire hazard may result and tripping of the circuit breaker of the hospital may shut
down the power supply to the ultrasonic generator as well as to other devices
connected to the same power supply.
Electric shock hazard
WARNING
• Do not drop or spill liquids on the ultrasonic generator. If liquid gets on or into the
ultrasonic generator, immediately stop using it and contact Olympus. Continued
use may deteriorate the insulation and could cause an electric shock incident.
• Never moisten the power plug as this may cause an electric shock hazard. If it is
moistened, dry it completely before use.
• Do not set up or control the ultrasonic generator with wet hands. Otherwise, an
electric shock or burns may result.
Burns of the patient and/or surgeon
WARNING
• During surgery, the THUNDERBEAT and/or SONICBEAT instrument should be
placed in an electrically insulated location when not in use. The THUNDERBEAT
and/or SONICBEAT instrument should never be placed on the patient when not in
use. Doing so may cause patient and/or user burns.
• When the compatible electrosurgical generator is used in conjunction with the
ultrasonic generator, keep the THUNDERBEAT and/or SONICBEAT instrument
connected to the generator not in use away from the surgical site while the other
generator is in operation. Otherwise, patient or user injury may result due to the
concentration of electric current.
• When using the compatible electrosurgical generator together with the ultrasonic
generator, connect the two generators to separate power outlets. Otherwise, burns
of the patient may result.
USG-400 INSTRUCTION MANUAL
7
Page 14
Important Information — Please Read Before Use
WARNING
• Avoid contact between two skin parts (between an arm and flank, for example) by
inserting a non-conductive material such as a blanket or piece of dry gauze
between them or positioning the patient to avoid contact. Be sure to withdraw
metallic accessories (including rings, wristwatch, necklace, and hairpins) from the
patient and ensure the patient is not in contact with any metallic object such as a
surgical bed.
• When using an electrocardiograph or other physiological monitoring instrument
simultaneously with the ultrasonic generator on a patient, any monitoring
electrodes should be placed as far away as possible from surgical sites. If placed
too close, high-frequency (RF bipolar) signals or spark discharge noise from the
ultrasonic generator may interfere with the operation of an electrocardiograph or
other physiological monitoring instrument. Needle monitoring electrodes should not
be used, as they may cause patient burns. Physiological monitoring instrument
incorporating high-frequency (RF bipolar) current limiting devices is recommended.
Impacts on the heart
DANGER
• Do not use the ultrasonic generator in combination with a product other than the
compatible instrument shown in the “System chart” on page 147. Otherwise, an
electrical shock may result.
• When treating the heart or working in close proximity to the heart, set the output
power to the minimum effective level. Otherwise, the spark discharge during
coagulation may stimulate the heart.
Impacts on cardiac pacemaker or defibrillator
DANGER
Before using the ultrasonic generator with a patient wearing a cardiac pacemaker,
cardioverter defibrillator, or other active implanted device, consult a cardiovascular
specialist and the manufacturer of the device, and make sufficient preparation to
ensure safety. The use of high-frequency energy may have a serious effect on the
patient due to malfunction or failure of the cardiac pacemaker, etc.
8
USG-400 INSTRUCTION MANUAL
Page 15
Important Information — Please Read Before Use
Use of cardioverter defibrillator
WARNING
Always have a defibrillator ready in case of cardiac emergency. Be sure to withdraw
all endoscopic instruments from the body cavity before using the defibrillator.
Flammable anesthetic and gases
DANGER
• Do not install the ultrasonic generator in the following locations:
In a place with high oxygen concentration.
In an atmosphere of oxidized substance such as nitrous oxide (N2O).
In a place where flammable anesthetic gas is used.
• The ultrasonic generator is not designed to be explosion-proof and cannot be used
in a flammable atmosphere. Otherwise, an explosion or fire hazard may result.
• Do not activate the ultrasonic generator while directing oxygen to the cauterized
region. Otherwise, the cauterized region may combust.
Flammable disinfectant
WARNING
After wiping the ultrasonic generator with gauze moistened with disinfectant
alcohol, be sure to dry it completely before use. Otherwise, a fire hazard may
result.
USG-400 INSTRUCTION MANUAL
9
Page 16
Important Information — Please Read Before Use
Risk of treatment and complications
WARNING
• If any irregularity is observed during the use of the ultrasonic generator,
immediately stop using it and take proper measures by referring to Chapter 8,
“Troubleshooting”. Also, refer to the instruction manual for the THUNDERBEAT
and/or SONICBEAT instrument in use. If the irregularity cannot be resolved,
contact Olympus.
• Always maintain an alternative means to complete a procedure to avoid interruption
of treatment.
• The ultrasonic generator and the THUNDERBEAT and/or SONICBEAT instrument
combined with it are designed for use in cutting and coagulation of tissue during
surgical procedures. Do not use the THUNDERBEAT and/or SONICBEAT
instrument for other purposes, as this may lead to perforation, tearing, bleeding,
burns, pain and infection, as well as equipment damage.
• Use this equipment properly. If not properly used, it may cause deformation,
breakage, patient and/or operator injury, malfunction of the cardiac pacemaker,
burns of the surgeon, surgical staff and/or patient, electric shock, abnormal output
or malfunction, perforation, bleeding, postoperative bleeding, tissue damage and
infection to the surgeon, surgical staff and/or patient.
• Should any unintended output be observed during surgery, immediately stop using
the ultrasonic generator and turn OFF the power switch.
Neuromuscular stimulation
WARNING
Patients may feel a neuromuscular stimulus when a spark discharges from the
THUNDERBEAT instrument or a metallic object such as forceps occurs during
activation. The neuromuscular stimulus is caused by low frequency components
generated during discharging. To prevent this, minimize the discharge by activating
the output keeping the grasping section in good contact with the tissue to be
cauterized.
10
USG-400 INSTRUCTION MANUAL
Page 17
Important Information — Please Read Before Use
Intelligent Tissue Monitoring*1, *2
The Intelligent Tissue Monitoring detects a small change of impedance on ultrasonic vibrations when a
part of living tissue is divided. When it detects the change, it stops output with a stop tone*3. The
Intelligent Tissue Monitoring can be set either ON or OFF on the touch screen display. When ON is
selected, the Intelligent Tissue Monitoring is enabled.
WARNING
• The Intelligent Tissue Monitoring is not guaranteed to operate optimally under all
procedures. The output stop function of the Intelligent Tissue Monitoring does not
work in all cases because it depends on several factors such as patient condition,
living tissue condition, living tissue type, how the instrument handle is gripped and
how a part of living tissue is grasped. Once a part of living tissue is divided, stop the
output immediately regardless of absence of the stop tone. Otherwise, the
THUNDERBEAT/SONICBEAT instrument may be damaged.
• Regardless of the setting of the Intelligent Tissue Monitoring, do not use the
THUNDERBEAT instrument to seal a blood vessel with a diameter over 7 mm or
the SONICBEAT instrument to seal a blood vessel with a diameter over 5 mm.
Otherwise, sufficient sealing may not be achieved.
• Regardless of the setting of the Intelligent Tissue Monitoring, squeeze the control
handle firmly until the control handle touches the grip handle to stop and grasp a
blood vessel in the middle of the grasping section when treating the blood vessel.
Otherwise, incomplete sealing may cause bleeding.
• Regardless of the setting of the Intelligent Tissue Monitoring, sufficient sealing may
not be achieved to treat thick blood vessels or certain blood characteristics.
Consider the combination use of suture, ligation or the metal clips in advance.
• Regardless of the setting of the Intelligent Tissue Monitoring, sufficient sealing may
not be achieved if the THUNDERBEAT instrument or the SONICBEAT instrument
is used to treat a patient with hypertension, coronary disease, arteriosclerosis,
diabetes, hepatocirrhosis or a patient with blood vessel irregularities such as
calcification. To ensure high sealing capability, use the instrument to seal healthy
normal vessels.
*1 The Intelligent Tissue Monitoring is available with software version 2.00 or later. To check the current
software version, refer to Section 6.5, “Menu”.
*2 The Intelligent Tissue Monitoring can be set independently for the THUNDERBEAT instrument and the
SONICBEAT instrument. Before using either instrument, be sure to check the current setting by referring
to Section 5.4, “Output setting”.
*3 When the SEAL & CUT mode with the THUNDERBEAT instrument is activated and the Intelligent Tissue
Monitoring is operated, both ultrasonic output and high-frequency output are stopped. When the MAX
mode with the SONICBEAT instrument is activated and the Intelligent Tissue Monitoring is operated,
ultrasonic output is stopped.
USG-400 INSTRUCTION MANUAL
11
Page 18
Important Information — Please Read Before Use
12
USG-400 INSTRUCTION MANUAL
Page 19
1.1 Inspecting the package contents
Chapter 1
1.1
Inspecting the Package
Contents
Ch.1
Inspecting the package contents
Match all items in the packaging with the components shown below. Inspect the ultrasonic generator
and each accessory for damage. If the ultrasonic generator is damaged, a component is missing or
you have any questions, do not use the ultrasonic generator, contact Olympus.
Ultrasonic generator (USG-400)
Power cord
Instruction manual
NOTE
• Keep the instruction manual in an easily accessible location.
• The ultrasonic generator is designed for use with the dedicated instruments.
Additional compatible instruments may be available in the future. Contact Olympus
periodically for updated information.
USG-400 INSTRUCTION MANUAL
13
Page 20
1.1 Inspecting the package contents
Ch.1
14
USG-400 INSTRUCTION MANUAL
Page 21
2.1 Symbols and descriptions
Chapter 2
2.1
Nomenclature and
Functions
Symbols and descriptions
Ch.2
Front panel
Symbol
Description
Symbol
Description
Power-ON/OFF
Symbol 1 for THUNDERBEAT
instrument
Symbol 2 for SONICBEAT
instrument
Select procedure
Probe check
Menu
Touch screen
Symbol
Description
Symbol
Description
Footswitch indicator
Footswitch for THUNDERBEAT
(left: purple, right: blue)
Footswitch for SONICBEAT
(left: light gray, right: dark gray)
Plus: increment of output level,
output tone, or brightness.
Minus: decrement of output level,
output tone, or brightness.
Return
OK
Cancel
Save procedure
Delete procedure
Languages
Touch tone (ON/OFF)
Software version
Service
USG-400 INSTRUCTION MANUAL
15
Page 22
2.1 Symbols and descriptions
Symbol
Ch.2
Description
Symbol
Description
Volume
Brightness
Select procedure (in title line)
Menu (in title line)
Previous
Next
Numeric: switch to numeral
Alphabetic: switch to alphabet
Uppercase/Lowercase: switches
between uppercase and lowercase
Backspace
Caution
Communication indicator
Intelligent Tissue Monitoring*1 ON
Intelligent Tissue Monitoring*1 OFF
*1 The Intelligent Tissue Monitoring is available with software version 2.00 or later.
Rear Panel
Symbol
Description
Symbol
Volume
16
USG-400 INSTRUCTION MANUAL
Description
Footswitch
Page 23
2.2 Front panel
2.2
Front panel
1
2
3
4
5
6
Ch.2
7
No.
Nomenclature
Description
1
Power switch
Depress to turn the ultrasonic generator ON and depress again to turn it OFF.
2
Transducer socket
(THUNDERBEAT)
Connect the transducer plug of the THUNDERBEAT.
3
Transducer socket
(SONICBEAT)
Connect the transducer plug of the SONICBEAT.
4
Touch screen
Displays the connection status of THUNDERBEAT and/or SONICBEAT instrument,
footswitch, communication cables, and the compatible electrosurgical generator
connected to the ultrasonic generator. Also, used to view and modify the output levels as
well as to control other functions.
5
SELECT
PROCEDURE
push button
Press to recall a saved procedure name and display the custom settings on the “All”
screen.
6
PROBE CHECK
push button
Press to check the probe operation, connection, and cancellation of output inhibition.
7
MENU push button
Press to access the menu, with control functions such as the output level save/delete,
language selection, touch tone control, output volume control, brightness control and
software version functions.
USG-400 INSTRUCTION MANUAL
17
Page 24
2.3 Rear panel
2.3
Rear panel
2
1
3
4
5
1
Ch.2
11
10
9
8
7
6
No.
Nomenclature
Description
1
Threaded holes for
the fixture plates
Threaded holes are for attaching the fixture plates.
2
Volume control
Rotate to adjust only the SONICBEAT instrument output volume. Adjustment is possible
during output.
3
SONICBEAT
footswitch socket
Connect the plug of the footswitch for SONICBEAT.
4
THUNDERBEAT
footswitch socket
Connect the plug of the footswitch for THUNDERBEAT.
5
Equipotential
bonding point
When equalizing the electrical potential of the ultrasonic generator, connect the
equipotential bonding wire to this terminal.
6
AC power socket
Connect the power cord.
7
Vents (cooling fan)
Openings to allow cooling. There are also vents on each side of the ultrasonic generator.
8
Threaded holes for
the docking plug
cover
Threaded holes are for attaching the docking plug cover.
9
LINK-OUT
Connect the plug (14-pin) of a cable connected to peripheral instrument to this
connector (14-pins) connector. Refer to Section 3.5, “Automatic mist & smoke evacuation system/function
(when using the compatible high flow insufflation unit)”.
10
LINK-IN connector
(26-pin)
Connect the plug (26-pin) of a cable connected to peripheral instrument to this
connector. Refer to Section 3.5, “Automatic mist & smoke evacuation system/function
(when using the compatible high flow insufflation unit)”.
11
Intelligent Tissue
Monitoring mark
This mark is shown at software version 2.00 or later.
18
USG-400 INSTRUCTION MANUAL
Page 25
2.4 Top panel
2.4
Top panel
2
1
Ch.2
No.
Nomenclature
Description
1
Docking plug cover The docking plug is hidden underneath the cover. When the docking plug cover is
withdrawn, the docking plug is spring-loaded.
2
Docking plug
Connect the docking plug to the compatible electrosurgical generator when using the
THUNDERBEAT instrument.
USG-400 INSTRUCTION MANUAL
19
Page 26
2.5 All screen (touch screen)
2.5
All screen (touch screen)
1
2
3
4
5
6
Ch.2
7
8
Screen for switching Intelligent Tissue
Monitoring OFF
No.
Nomenclature
Screen for switching Intelligent Tissue
Monitoring ON
Description
1
Symbol
The symbol is shown on the THUNDERBEAT or SONICBEAT instrument, transducer
plug, transducer socket, and touch screen.
2
Instrument names
The instrument names are shown.
3
Footswitch indicator The footswitch indicators indicate the connection status of the footswitches. They
illuminate when a footswitch is connected.
4
Procedure name
The name of the custom procedure recalled by the select procedure button is displayed
here.
5
Communication
indicator
The communication indicator indicates the status of communications between the
compatible electrosurgical generator and ultrasonic generator. It illuminates when the
communication cable is connected.
6
Intelligent Tissue
Monitoring
indicator*1
The screen displays whether the Intelligent Tissue Monitoring is ON or OFF.
7
THUNDERBEAT
button
This button shows the THUNDERBEAT instrument output level and illuminates when the
THUNDERBEAT instrument is connected. The THUNDERBEAT instrument has two
output modes: the SEAL & CUT mode and the SEAL mode. During activation, the button
is highlighted in purple for the SEAL & CUT mode and in blue for the SEAL mode. Tap
this button to adjust the output level.
8
SONICBEAT button This button shows the SONICBEAT instrument output level and illuminates when the
SONICBEAT instrument is connected. The SONICBEAT instrument has two output
modes: the VAR mode and the MAX mode. During activation, the button is highlighted in
light gray for the VAR mode and in dark gray for the MAX mode. Tap this button to adjust
the output level.
*1 The Intelligent Tissue Monitoring is available with software version 2.00 or later.
20