OLYMPUS
BRONCHOVIDEOSCOPE Operation Manual
102 Pages
Preview
Page 1
OPERATION MANUAL
INSTRUCTIONS
Symbols
1
Important Information - Please Read Before Use
2
BRONCHOVIDEOSCOPE
Chapter 1
Checking the Package Contents
15
OLYMPUS BF-H1200 OLYMPUS BF-1TH1200
Chapter 2
Instrument Nomenclature and Specifications
17
Chapter 3
Preparation and Inspection
25
Chapter 4
Operation
49
Chapter 5
Troubleshooting
71
Appendix
Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover, for reprocessing information.
79
Contents
Contents Symbols ... 1 Important Information - Please Read Before Use ... 2
Intended use ... 2 Contraindications ... 2 Applicability of endoscopy and endoscopic treatment ... 2 Instruction manual ... 3 User qualifications ... 3 Instrument compatibility ... 4 Reprocessing before the first use/reprocessing and storage after use ... 4 Spare equipment ... 4 Maintenance management ... 5 Prohibition of improper repair and modification ... 5 Signal words ... 5 Precautions ... 6 Precautions for disappeared or frozen endoscopic image ... 12 Examples of inappropriate handling ... 14
Chapter 1 Checking the Package Contents ... 15 1.1
Checking the package contents ... 15
Packaged items ... 16
Chapter 2 Instrument Nomenclature and Specifications ... 17 2.1
Nomenclature and functions ... 17
2.2
Specifications ... 22
Control section, insertion section ... 18 Endoscope connector ... 20 Environment ... 22 Specifications ... 23
Chapter 3 Preparation and Inspection ... 25 3.1
The workflow of preparation and inspection ... 25
3.2
Preparation of the equipment ... 27
3.3
Inspection of the endoscope ... 28
3.4
Inspection of accessories ... 35
3.5
Attaching accessories to the endoscope ... 38
Inspection of the endoscope ... 29 Inspection of the bending mechanism ... 32 Inspection of the insertion tube rotation mechanism ... 34 Inspection of the single use suction valve (MAJ-209) ... 35 Inspection of the single use biopsy valve (MAJ-210) ... 36 Inspection of the mouthpiece (MA-651) ... 37 Attaching the single use suction valve (MAJ-209) ... 38 BF-H1200, BF-1TH1200 OPERATION MANUAL
i
Contents
Attaching the single use biopsy valve (MAJ-210) ... 39
3.6
Inspection of ancillary equipment ... 40
3.7
Connection of the endoscope and ancillary equipment ... 40
3.8
Inspection of the endoscopic system ... 43
Connection to the video system center (CV-1500) (when it is used) ... 40 Connection to the light source (when NOT using the video system center CV-1500) ... 42 Connection of the suction tube ... 43
Inspection summary ... 43 Inspection of the ancillary equipment ... 44 Inspection of the endoscopic image ... 44 Inspection of the remote switches ... 45 Inspection of the water feeding function ... 46 Inspection of the suction function ... 47 Inspection of the instrument channel ... 48
Chapter 4 Operation ... 49 4.1
Precautions ... 49
4.2
Insertion ... 52
4.3
Using EndoTherapy accessories ... 58
4.4
Withdrawal of the endoscope ... 69
4.5
Transportation of the endoscope ... 70
Holding and manipulating the endoscope ... 52 Insertion of the endoscope ... 53 Observation of the endoscopic image ... 55 Angulation of the distal end ... 55 Operation of the insertion tube rotation ... 56 Feeding fluids ... 56 Suction ... 57 Insertion of EndoTherapy accessories into the endoscope ... 60 Operation of EndoTherapy accessories ... 62 Withdrawal of EndoTherapy accessories ... 62 High-frequency cauterization treatment ... 62 Argon plasma coagulation (APC) ... 64 Laser cauterization ... 66 Ultrasonic observation ... 67 Bronchoalveolar lavage ... 68
Transporting within the healthcare facility ... 70 Transporting outside the healthcare facility ... 70
Chapter 5 Troubleshooting ... 71
ii
5.1
Troubleshooting ... 71
5.2
Troubleshooting guide ... 72
Image quality or brightness ... 72 Water feeding ... 73 Suction ... 74 BF-H1200, BF-1TH1200 OPERATION MANUAL
Contents
EndoTherapy accessories ... 74 Others ... 75
5.3
Withdrawal of the endoscope with an irregularity ... 75
5.4
Returning the endoscope for repair ... 77
Withdrawal when the WLI, NBI, and RDI endoscopic images appear on the monitor ... 76 Withdrawal when any endoscopic image of the WLI, NBI, or RDI does not appear on the monitor ... 76 Withdrawal when no endoscopic image appears on the monitor or a frozen image cannot be restored ... 77
Appendix ... 79 Combination equipment ... 79
System chart ... 79 Reprocessing equipment ... 81 Compatible EndoTherapy accessories ... 82
EMC information ... 88
BF-H1200, BF-1TH1200 OPERATION MANUAL
iii
Contents
iv
BF-H1200, BF-1TH1200 OPERATION MANUAL
Symbols
Symbols The meaning(s) of the symbol(s) shown on the component packaging, the back cover of the instruction manual, and/or the instrument are as follows: Symbol Blue
Description
Symbol
Description
Follow instructions for use.
Consult instructions for use.
Caution
Endoscope
TYPE BF applied part
Single use only
Lot number
Use by (expiration date)
Not Made with Natural Rubber Latex
keep away from sunlight
keep dry
Sterilization lot number
Sterilized using irradiation
Manufacturer
Date of Manufacture
Authorized representative in the European Community
Serial number
Translation
Importer (into European Union)
Ingress protection rating is 7.
Medical device
Non-sterile
Do not resterilize
Do not use if package is damaged
BF-H1200, BF-1TH1200 OPERATION MANUAL
1
Important Information - Please Read Before Use
Important Information - Please Read Before Use
Intended use This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the airways and tracheobronchial tree.
Contraindications None known.
Applicability of endoscopy and endoscopic treatment If there are official standards on the applicability of endoscopy and endoscopic treatment that are defined by the healthcare facility’s administrations or other official institutions, such as academic societies on endoscopy, follow those standards. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their nature, extent, and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
2
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
Instruction manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals for all equipment that will be used during the procedure and use the equipment as instructed. Note that the complete instruction manual set for this endoscope consists of this manual and the “REPROCESSING MANUAL” with your endoscope model listed on the cover. It also accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, contact Olympus.
Terms used in this manual NBI (Narrow Band Imaging) observation: This is optical-digital observation using narrowband light.
RDI (Red Dichromatic Imaging) observation: This is optical-digital observation using red illumination light.
WLI (White Light Imaging) observation: This is observation using RGB illumination.
Image sensor: The image sensor is a device that converts light into electrical signals.
User qualifications If there are official standards for user qualifications to perform endoscopy and endoscopic treatment that are defined by the healthcare facility’s medical administrators or other official institutions, such as academic societies on endoscopy, follow those standards. If there are no official qualification standards, the operator of this instrument must be a physician approved by the medical safety manager of the healthcare facility or person in charge of the department (department of internal medicine, etc.). The physician should be capable of safely performing the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
BF-H1200, BF-1TH1200 OPERATION MANUAL
3
Important Information - Please Read Before Use
Instrument compatibility Refer to “Combination equipment” on page 79 to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient or operator injury and/or equipment damage. This instrument complies with the EMC standard for medical electrical equipment, edition 4 (IEC 60601-1-2: 2014). When connecting to an instrument that complies with a previous edition of the EMC standard for medical electrical equipment edition, the EMC characteristics could be vulnerable.
Reprocessing before the first use/reprocessing and storage after use
This instrument was not reprocessed before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion reprocessing manual. Improper and/or incomplete reprocessing or storage can pose an infection control risk, cause equipment damage, or reduce performance.
Spare equipment Be sure to prepare another endoscope to avoid interruption of the examination due to equipment failure or malfunction.
4
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
Maintenance management The probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase. In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each healthcare facility should periodically inspect the items specified in this manual following applicable regulations, guidelines, etc. required of you. An endoscope with an observed irregularity should not be used, but should be inspected by following Section 5.2, “Troubleshooting guide”. If the irregularity is still observed after inspection, contact Olympus.
Prohibition of improper repair and modification This instrument does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury and/or equipment damage may result. Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner.
Signal words The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
BF-H1200, BF-1TH1200 OPERATION MANUAL
5
Important Information - Please Read Before Use
Precautions Follow the warnings and cautions given below when handling this endoscope. This information is to be supplemented by the warnings and cautions given in each chapter.
WARNING • Do not use this endoscope for any purpose other than its intended use. Patient or operator injury and/or equipment damage may result. • After using this endoscope, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. Using improperly or incompletely reprocessed or stored instruments may cause patient cross-contamination and/or infection. • Before endoscopy, remove any metallic objects (watch, glasses, necklace, etc.) from the patient. Performing high-frequency cauterization treatment while the patient is wearing metallic objects may cause burns on the patient in areas around the metallic objects. • Do not strike, hit, or drop the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope. Also, do not bend, pull, or twist the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force. The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations. It could also cause parts of the endoscope to fall off inside the patient. • Never perform angulation control forcibly or abruptly. Never forcefully pull, twist, or rotate the angulated bending section. Patient injury, bleeding, and/or perforation may result. It may also become impossible to straighten the bending section during an examination. In addition, the angulation of the endoscope may not return during an examination, and this product may not be pulled out. Take extra caution for the branch of the trachea. • Never insert or withdraw the endoscope’s insertion section while the bending section is locked in position. Patient injury, bleeding, and/or perforation may result. • Never operate the bending section, perform suction, insert or withdraw the endoscope’s insertion section, rotate the insertion section, or use EndoTherapy accessories while no endoscopic image is observed or the endoscopic image is frozen. Patient injury, bleeding, and/or perforation may result. • Never insert, withdraw, or rotate the endoscope’s insertion section with excessive force or while an optimum field of view cannot be obtained. Patient injury, bleeding, and/or perforation may result. If it is difficult to insert the endoscope, do not forcibly insert the endoscope; stop the endoscopy. Forcible insertion can result in patient injury, bleeding, and/or perforation.
6
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • Never insert or withdraw the insertion section abruptly or with excessive force. Patient injury, bleeding, and/or perforation may result. • For CV-1500 connected Do not touch the light guide on the endoscope connector immediately after removing it from the video system center because it is extremely hot. Operator or patient burns can result. • For CLV-290 connected Do not touch the light guide on the endoscope connector immediately after removing it from the light source because it is extremely hot. Operator or patient burns can result. • For CV-1500 connected Although the illumination light emitted from the distal end of the endoscope is required for endoscopic observation, it may also cause alteration of living tissues such as protein denaturation of living tissue and perforation of the tissue through improper usage. Observe the following warnings for illumination. Always set the minimum required brightness. The brightness of the image on a monitor may differ from the actual brightness at the distal end of the endoscope. Make sure to use the automatic brightness control function. This function can better maintain the illumination level. Refer to the instruction manual for the video system center for further details. When the endoscope will not be used for a long period, be sure to turn OFF the lamp so that the endoscope is not illuminated unnecessarily. Always maintain a suitable distance necessary for adequate viewing while using the minimum level of illumination for the minimum amount of time. Do not use close stationary viewing or leave the distal end of the endoscope close to the mucous membrane for a long time without necessity.
BF-H1200, BF-1TH1200 OPERATION MANUAL
7
Important Information - Please Read Before Use
WARNING • For CLV-290 connected Although the illumination light emitted from the distal end of the endoscope is required for endoscopic observation, it may also cause alteration of living tissues such as protein denaturation of living tissue and perforation of the tissue through improper usage. Observe the following warnings for illumination. Always set the minimum required brightness. The brightness of the image on a monitor may differ from the actual brightness at the distal end of the endoscope. When using a light source and video system center that are compatible with the light source’s automatic brightness control function, make sure to use the automatic brightness control function. This function can better maintain the illumination level. Refer to the instruction manual for the light source and the video system center for further details. When the endoscope will not be used for a long period, be sure to turn OFF the light source or activate the light shield function (standby mode, etc.) so that the endoscope is not illuminated unnecessarily. Always maintain a suitable distance necessary for adequate viewing while using the minimum level of illumination for the minimum amount of time. Do not use close stationary viewing or leave the distal end of the endoscope close to the mucous membrane for a long time without necessity. • Do not connect the endoscope connector while the electrical contacts are wet and/or dirty, which may result in an electric shock, causing severe damage to the endoscope and compromising patient and/or operator safety. • If the endoscopic image becomes dimmer during the procedure, it may indicate that blood or mucus is adhering to the light guide lens on the distal end of the endoscope. Immediately withdraw the endoscope from the patient, remove blood or mucus, and confirm that the light guide lens has no irregularities before using again. If you continue to use the endoscope with its obstructed light guide lens, the temperature at the distal end of the endoscope may rise, which may cause patient injury or operator and/or patient burns. • When the endoscopic image does not appear on the monitor, the image sensor may have been damaged. Turn the video system center OFF immediately. Continued power supply in such a case will cause the distal end of the endoscope to become hot and could cause operator and/or patient burns. • Do not rely on the NBI or RDI observation mode alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention.
8
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • When performing transnasal insertion with the endoscope, follow the warnings below. The shape and size of the nasal cavity and its suitability for transnasal insertion may vary from patient to patient. No endoscope, including this one, can always be inserted transnasally into all patients. Before proceeding, always be sure to confirm that transnasal insertion is possible with the patient by considering both the size of the patient’s nasal cavity as well as the size of the endoscope’s insertion section. Patient injury can result or the endoscope could become lodged and difficult to withdraw. Transnasal insertion is accompanied by the risk of inflammation of the nasal cavity. If this happens, the nasal passage will be constricted, making it more difficult to withdraw the endoscope. In this case, do not use force to withdraw the endoscope because patient injury, bleeding, and/or perforation may result. Transnasal insertion is accompanied by the risk of bleeding in the nasal cavity. Be sure to be prepared to deal with any bleeding. When withdrawing the endoscope, observe the inside of the nasal cavity to ensure that there is no bleeding. Even when the endoscope has been withdrawn without bleeding, do not allow the patient to blow his or her nose strongly because this could cause it to start bleeding. Before transnasal insertion, apply the appropriate pretreatment and lubrication to the patient to enlarge the nasal cavity. Otherwise, patient injury can result or the endoscope could become lodged and difficult to withdraw. When applying a pretreatment agent through a tube, insert the tube into the same path as the path planned for the endoscope’s insertion. Otherwise, the treatment will have no effect. The effects of the pretreatment agent and lubricant will decrease the longer the procedure lasts. Apply the pretreatment agent or lubricant as required during the procedure - for example, when withdrawal seems to be difficult. Transnasal insertion of the endoscope should be performed carefully. If resistance to insertion is felt, or the patient reports pain, stop the insertion immediately. Patient injury can result or the endoscope could become lodged and difficult to withdraw. If it becomes impossible to withdraw the transnasally inserted endoscope, pull the distal end of the endoscope out of the mouth, cut the flexible tube using wire cutters, and after ensuring that the cut section will not injure the body cavity or nasal cavity of the patient, withdraw the endoscope carefully. Therefore, always prepare wire cutters in advance. • When using the electronic zoom function of the video system center, never insert or withdraw the endoscope’s insertion section or use EndoTherapy accessories while the image is electronically zoomed. Patient injury, bleeding, and/or perforation can result. BF-H1200, BF-1TH1200 OPERATION MANUAL
9
Important Information - Please Read Before Use
WARNING • The bending section will only bend to the UP or DOWN direction. To insert or withdraw, operate the endoscope by considering the direction in which the bending section is angulated. Patient injury, bleeding, and/or perforation can result.
CAUTION • For CV-1500 connected Do not pull the universal cord during an examination. The endoscope connector will be pulled out from the output socket of the video system center and the endoscopic image will disappear. • For CLV-290 connected Do not pull the universal cord during an examination. The endoscope connector will be pulled out from the output socket of the light source and the endoscopic image will disappear. • Do not coil the insertion tube or universal cord with a diameter of less than 12 cm. Equipment damage may result. • Do not attempt to bend or twist the endoscope’s insertion section with excessive force. The insertion section may be damaged. • Do not apply shock to the distal end of the endoscope, in particular the objective lens surface at the distal end. An abnormal endoscopic image may result. • If the endoscope is dropped or the distal end of the endoscope receives a hard impact, the endoscope may be damaged even if no visible damage of the lens on the distal end can be found. In this case, stop using the endoscope, and contact Olympus. • Do not twist or bend the bending section with your hands. Equipment damage may result. • Do not squeeze the bending section forcefully. The covering of the bending section may stretch or break and cause water leakage. • Do not put or press the endoscope connector on the insertion section when transporting or reprocessing. The insertion section may be damaged. • For CV-1500 connected Turn the video system center ON only when the endoscope connector is connected to the video system center. In particular, confirm that the video system center is OFF before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor. • For CLV-290 connected Turn the video system center ON only when the endoscope connector is connected to the light source. In particular, confirm that the video system center is OFF before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor.
10
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
CAUTION • The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling, or twisting them with excessive force can break the switches and/or cause water leakage. • For CV-1500 connected Do not hit or bend the electrical contacts on the endoscope connector. The connection to the video system center may be impaired and a faulty contact can result. • For CLV-290 connected Do not hit or bend the electrical contacts on the endoscope connector. The connection to the light source may be impaired and a faulty contact can result. • When using an endotracheal tube with the endoscope, select the tube that gives a sufficient gap between the insertion section of the endoscope and itself. A narrow gap may make it difficult for a patient to breathe and/or damage the endoscope. • Before inserting the endoscope with an endotracheal tube into the patient, confirm that the insertion section of the endoscope can be inserted into the endotracheal tube smoothly by running it back and forth over the entire length of the insertion section and that the tube does not damage the endoscope. Any protrusions may damage the bending section cover or strip the external surface of the insertion section. When using lubrication, make above confirmation before applying lubrication. • Be sure that this endoscope is not used adjacent to or stacked with other equipment (other than the components of this endoscope or system) to avoid electromagnetic interference. • This endoscope supports radio communication function (receive frequency: 13.56 MHz) that identifies instruments. Electromagnetic interference may occur on this endoscope near equipment marked with the following symbol or other portable and mobile RF (Radio Frequency) communications equipment such as cellular phones. If electromagnetic interference occurs, mitigation measures may be necessary, such as moving the RF communications equipment away, reorienting or relocating this endoscope, or shielding the location.
NOTE • This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center. • When the endoscope gets strong static electricity, noise may be observed in the endoscopic image. This does not indicate a malfunction.
BF-H1200, BF-1TH1200 OPERATION MANUAL
11
Important Information - Please Read Before Use
Precautions for disappeared or frozen endoscopic image WARNING • If the endoscopic image disappears unexpectedly or the frozen image cannot be restored during an examination, immediately stop using the endoscope and withdraw it from the patient as described in Section 5.3, “Withdrawal of the endoscope with an irregularity”. Continued use of the endoscope under this condition could result in patient injury, bleeding, and/or perforation. • For CV-1500 connected Follow the precautions given below. Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination. Connect the endoscope connector to the video system center completely by pushing the endoscope connector until it clicks. Otherwise, a faulty contact can result. Do not bend, hit, pull, or twist the insertion section, bending section, control section, universal cord, and endoscope connector. The endoscope may be damaged, and water leakage and/or breakage of internal parts like the image sensor cable can result. Before connecting the endoscope connector to the video system center, confirm that the endoscope connector, including the electrical contacts, is completely dry and clean. If the endoscope is used with the electrical contacts wet and/or dirty, the endoscope and video system center may malfunction. If air bubbles emerge from the endoscope continuously during the leakage test, do not use the endoscope. Water may enter the endoscope and cause a short circuit. This may result in image sensor damage. When inserting the endoscope through the mouth, place the mouthpiece (MA-651) in the patient’s mouth as necessary before inserting the endoscope to prevent the patient from accidentally biting the insertion section. Biting the insertion section may result in a break in the cable or malfunction of the light guide.
12
BF-H1200, BF-1TH1200 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • For CLV-290 connected Follow the precautions given below. Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination. Connect the endoscope connector to the light source completely by pushing the endoscope connector until it clicks. Otherwise, a faulty contact can result. Do not bend, hit, pull, or twist the insertion section, bending section, control section, universal cord, and endoscope connector. The endoscope may be damaged, and water leakage and/or breakage of internal parts like the image sensor cable can result. Before connecting the endoscope connector to the light source, confirm that the endoscope connector, including the electrical contacts, is completely dry and clean. If the endoscope is used with the electrical contacts wet and/or dirty, the endoscope and light source may malfunction. If air bubbles emerge from the endoscope continuously during the leakage test, do not use the endoscope. Water may enter the endoscope and cause a short circuit. This may result in image sensor damage. When inserting the endoscope through the mouth, place the mouthpiece (MA-651) in the patient’s mouth as necessary before inserting the endoscope to prevent the patient from accidentally biting the insertion section. Biting the insertion section may result in a break in the cable or malfunction of the light guide.
BF-H1200, BF-1TH1200 OPERATION MANUAL
13
Important Information - Please Read Before Use
Examples of inappropriate handling Details on clinical endoscopic technique are the responsibility of trained specialists. Patient safety in endoscopic examinations and endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility. Examples of inappropriate handling are described below. • Applying suction with the distal end of the endoscope in contact with the mucosal surface, with higher suction pressure than required or with prolonged suction time may cause bleeding and/or lesions. • The endoscope has not been designed for use in retroflexed observation. Performing retroflexed observation in a narrow lumen may make it impossible to straighten the angle of the bending section and/or withdraw the endoscope from the patient. In case the patient moves due to coughing and other reasons while the endoscope is angulated in the narrow lumen, the bending section of the endoscope may be pushed into the lumen and be retroflexed. Pretreatment to control patient’s coughing reflex and other possible unexpected moves is the responsibility of trained specialists. Retroflexed observation should be performed only when the usefulness of doing so is determined to be greater than the risk that is posed to the patient. • Inserting, withdrawing, and using EndoTherapy accessories without a clear endoscopic image may cause patient injury, burns, bleeding, and/or perforation. • Inserting or withdrawing the endoscope, rotating the insertion section, applying suction, or operating the bending section without a clear endoscopic image may cause patient injury, bleeding, and/or perforation.
14
BF-H1200, BF-1TH1200 OPERATION MANUAL
1.1 Checking the package contents
Chapter 1
1.1
Checking the Package Contents
Checking the package contents
Ch.1
Match all items in the package with the components shown below. Inspect each item for damage. If the endoscope is damaged, a component is missing, or you have any questions, do not use the items; immediately contact Olympus.
Endoscope
BF-H1200, BF-1TH1200
BF-H1200, BF-1TH1200 OPERATION MANUAL
15