OLYMPUS
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE Reprocessing Instructions
96 Pages
Preview
Page 1
REPROCESSING MANUAL
INSTRUCTIONS
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE
Chapter 1
General Policy
1
Chapter 2
Function and Inspection of the Accessories for Reprocessing
7
Compatible Reprocessing Methods and Chemical Agents
11
Reprocessing Workflow for the Endoscope and Accessories
23
Reprocessing the Endoscope (and related reprocessing accessories)
27
Chapter 6
Reprocessing the Accessories
71
Chapter 7
Reprocessing the Endoscope and Accessories Using an Automatic Thermal Cleaning/ Disinfector
87
Storage and Disposal
87
Chapter 3
OLYMPUS LTF-S300-10-3D Chapter 4
Chapter 5
Chapter 8
Accessories: • Sterilization cap (MAJ-1538)
• 3D adjustment cap (MAJ-2266)
• Channel-opening cleaning brush (MH-507)
• Instrument basket (WA05938A, sold separately)
MAJ-1538
MAJ-2266
MH-507
WA05938A
Refer to the endoscope’s companion manual, the “OPERATION MANUAL” with your endoscope model listed on the cover, for operation information.
Contents
Contents Chapter 1 General Policy ... 1 1.1
Instructions ... 1
1.2
Importance of reprocessing ... 1
1.3
Signal words ... 2
1.4
Precautions ... 2
1.5
Reprocessing before the first use ... 6
1.6
Reprocessing and storage after use ... 6
1.7
Reprocessing before patient procedure ... 6
Chapter 2 Function and Inspection of the Accessories for Reprocessing ... 7 2.1
Sterilization cap (MAJ-1538) ... 7
2.2
Instrument basket (WA05938A, sold separately) ... 9
2.3
Channel-opening cleaning brush (MH-507) ... 10
Chapter 3 Compatible Reprocessing Methods and Chemical Agents ... 11 3.1
Compatibility summary ... 11 Microbiological efficacy ... 12 Material durability ... 12 List of compatible methods validated in terms of microbiological efficacy and material durability ... 12 List of compatible methods validated in terms of material durability ... 14
3.2
Water (for reprocessing) ... 16
3.3
Detergent solution ... 17
3.4
Rinse water ... 17
3.5
Automatic thermal cleaning/disinfection ... 18
3.6
Ethylene oxide gas sterilization ... 20
3.7
Steam sterilization (autoclaving) ... 22
Chapter 4 Reprocessing Workflow for the Endoscope and Accessories ... 23 4.1
Summary of reprocessing workflow ... 23
4.2
Workflow for manually cleaning and sterilizing the endoscope and accessories ... 24
4.3
Workflow for reprocessing the endoscope and accessories using an automatic thermal cleaning/disinfector ... 25 LTF-S300-10-3D REPROCESSING MANUAL
i
Contents
Chapter 5 Reprocessing the Endoscope (and related reprocessing accessories) ... 27 5.1
Summary of reprocessing the endoscope ... 27
5.2
Preparing the equipment for reprocessing ... 28 Equipment needed ... 28
5.3
Precleaning the endoscope ... 30 Equipment needed ... 30 Preparation ... 30 Wipe the insertion section ... 31 Detach the video connector from the video system center ... 32 Detach the light guide connector from the video system center ... 32
5.4
Leakage testing of the endoscope ... 33 Equipment needed ... 33 Perform the leakage test ... 34
5.5
Manually cleaning the endoscope ... 46 Equipment needed ... 46 Immerse the endoscope in detergent solution ... 47 Clean the external surface ... 48 Remove detergent solution from the endoscope ... 53 Dry external surfaces ... 53
5.6
Sterilizing the endoscope and accessories ... 54 Equipment needed ... 54 Ethylene oxide gas sterilization ... 54 STERRAD® 100S/NX®/100NX® sterilization ... 62 V-PRO® maX sterilization ... 65
5.7
Presoaking the endoscope ... 68 Equipment needed ... 68 Presoaking the endoscope ... 69
Chapter 6 Reprocessing the Accessories ... 71 6.1
Summary of reprocessing the accessories ... 71 Equipment needed ... 72
6.2
Manually cleaning the 3D adjustment cap (MAJ-2266) ... 73 Remove detergent solution from the 3D adjustment cap ... 79 Dry external surfaces ... 79
6.3
Sterilizing the 3D adjustment cap (MAJ-2266) ... 80 Ethylene oxide gas sterilization ... 80 Steam sterilization (autoclaving) ... 81 STERRAD® 100S/NX®/100NX® sterilization ... 82 V-PRO® maX sterilization ... 82
6.4
Sterilizing the channel-opening cleaning brush (MH-507) ... 83 Steam sterilization (autoclaving) ... 83
ii
LTF-S300-10-3D REPROCESSING MANUAL
Contents
Chapter 7 Reprocessing the Endoscope and Accessories Using an Automatic Thermal Cleaning/Disinfector ... 85 7.1
Reprocessing the endoscope and accessories using a thermal cleaning/disinfector ... 85
Chapter 8 Storage and Disposal ... 87 8.1
Precautions for storage and disposal ... 87
8.2
Storing the sterilized endoscope and accessories ... 88
8.3
Disposal ... 88
LTF-S300-10-3D REPROCESSING MANUAL
iii
Contents
iv
LTF-S300-10-3D REPROCESSING MANUAL
1.1 Instructions
Chapter 1 1.1
General Policy
Ch.1
Instructions • This manual contains the reprocessing methods recommended by Olympus for the endoscope and accessories listed on the front cover. • This instruction manual contains essential information on reprocessing the endoscope and accessories safely and effectively. • Before reprocessing, thoroughly review this manual and the manuals of the reprocessing equipment and chemicals that will be used for reprocessing. Reprocess all the devices as instructed. • Note that the complete instruction manual set for the endoscope and accessories consists of this manual and the “OPERATION MANUAL” with your endoscope model listed on the cover. Both manuals accompanied the endoscope at shipment. • Keep this manual and all related manuals in a safe and accessible location (e.g., in the reprocessing area). • If you have any questions or comments about any information in this manual, or if a problem that cannot be solved occurs while reprocessing, contact Olympus.
1.2
Importance of reprocessing
The medical literature reports incidents of cross-contamination resulting from improper reprocessing. It is strongly recommended that all individuals engaged in reprocessing closely observe all instructions given in this manual and the manuals of all ancillary equipment, and have a thorough understanding of the following items: • Professional health and safety policies of your hospital • Instruction manuals for the endoscope, accessories, and all the other reprocessing equipment • Structure and handling of the endoscope and accessories • Handling of pertinent chemicals When selecting appropriate methods and conditions for cleaning and sterilization, follow the policies at your institution, applicable national laws and standards, and professional society guidelines and recommended practices, in addition to the instructions given in this manual.
LTF-S300-10-3D REPROCESSING MANUAL
1
1.3 Signal words
Ch.1
1.3
Signal words
The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
1.4
Indicates additional helpful information.
Precautions WARNING • An insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who contact them. • The front panels and/or the touch panels of the video system center and other equipment may cause an infection control risk. Perform proper reprocessing as described in their respective instruction manuals. Also, a tap and/or basin that medical personnel come in contact with may cause an infection control risk as well. Perform proper replacement and/or reprocessing. • All sterilization methods require thorough cleaning prior to the instruments being reprocessed. If the instruments are not adequately cleaned prior to sterilization, these processes will be ineffective. Immediately after each patient procedure and before sterilization, thoroughly clean the endoscope and the accessories used with the endoscope. • All accessories used with the endoscope during the patient procedure must be reprocessed after each patient procedure, even if the accessories were not used during the patient procedure. Insufficient reprocessing of these components may pose an infection control risk to patients and/or operators.
2
LTF-S300-10-3D REPROCESSING MANUAL
1.4 Precautions
Ch.1
WARNING • The results of sterilization depend on various factors. These factors include how the equipment was packaged and the placing and loading of the package in the sterilization device. Verify the sterilization process using biological and/or chemical indicators. Follow the guidelines for sterilization issued by national authorities, professional organizations, and infection control professionals, including the frequency of the above verification, as well as the instruction manual for the sterilization device. • Establish an internal system of identifying contaminated versus reprocessed endoscopes and accessories to prevent both mix-ups and cross-contamination. Some national or professional guidelines recommend separating dirty (contaminated) area, clean area, and storage area. Touching a reprocessed endoscope and/or accessories with contaminated gloves or placing them on a contaminated hanger or surface, including letting them touch the floor, will recontaminate them. • Prior to each patient procedure, confirm that the endoscope and accessories have been properly reprocessed and stored. If there are any doubts or questions, reprocess them again before the patient procedure, following the instructions given in this manual. • Perform a leakage test on the endoscope after each precleaning procedure. Do not use the endoscope if a leak is detected. Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions. Use of a leaking endoscope may also pose an infection control risk. Leakage tester
Figure 1.1 • Store alcohol in an airtight container. Alcohol stored in an open container may cause a fire hazard and may result in a loss of efficacy due to evaporation.
LTF-S300-10-3D REPROCESSING MANUAL
3
1.4 Precautions
Ch.1
WARNING • The accessories listed on the front cover of this manual cannot be refurbished or repaired and are intended to be replaced once they show any signs of wear and tear. Should any irregularity be observed, use a replacement accessory instead. Using defective accessories may cause equipment malfunction, reduce the efficacy of reprocessing, present a risk to patients and/or operators, or damage the endoscope and/or accessories. • Patient debris and used reprocessing chemicals pose infection control risks. To guard against contact with dangerous chemicals and potentially infectious material, wear appropriate personal protective equipment during reprocessing. Such protective equipment should include appropriate eyewear, face mask, cap, moisture-resistant clothing, shoe covers, and chemical-resistant gloves that fit properly and are long enough to prevent skin exposure. • The reprocessing room must be adequately ventilated to minimize the risks from chemical vapors. • Always remove contaminated personal protective equipment before leaving the reprocessing area to prevent contamination from spreading. • The instructions provided in this manual are not valid for Olympus devices repaired by a non-Olympus facility. The Olympus recommended reprocessing procedures have not been validated for reprocessing devices repaired by a non-Olympus facility. In the event that your device has been repaired by a non-Olympus facility, contact that repair facility for instructions regarding reprocessing. • Prions, which are the pathogenic agents of the Creutzfeldt-Jakob disease (CJD) cannot be destroyed or inactivated by the reprocessing methods stated in this instruction manual. When using the endoscope and accessories on patients with CJD or variant Creutzfeldt-Jakob disease (vCJD), be sure to use them for such patients only, or immediately dispose of them after use in an appropriate manner to prevent the usage of exposed devices on other patients. For methods to handle CJD, follow the respective guidelines in your country.
4
LTF-S300-10-3D REPROCESSING MANUAL
1.4 Precautions
Ch.1
WARNING • The endoscope and accessories may be damaged by published methods for destroying or inactivating prions. For information on the durability of Olympus equipment against a particular reprocessing method, contact Olympus. In general, Olympus cannot guarantee the effectiveness, safety, and durability of reprocessing methods not described in this reprocessing manual. If you choose to use a reprocessing method not recommended in this manual, the local institution and/or physicians must assume responsibility for its safety and efficacy. Make sure to carefully inspect each piece of endoscopic equipment for irregularities (damage) prior to each patient procedure. Do not use the equipment if any irregularity is found. • Good quality control practices typically require appropriate documentation. Items such as local SOPs (Standard Operating Procedures), confirmation of operator training, etc. should be documented as performed. • Do not perform the endoscope and accessories using reprocess method that does not meet the specified condition. It can be damage the equipment and hindering effective removal organic debris. It may result failure to achieve sterilization of the endoscope.
CAUTION • Before immersing the endoscope in reprocessing fluids, confirm that the sterilization cap (MAJ-1538) is not attached to the endoscope. If the sterilization cap is attached, the reprocessing fluids will be able to penetrate into the endoscope, and it can be damaged. • Store spare accessories in their original packaging to prevent damage. • To prevent damage, do not apply excessive force to the endoscope and accessories during reprocessing. • Vapors from alcohol may damage electronic devices such as computers. Properly manage the quality and durability of the devices used in reprocessing rooms and the ventilation performance of the rooms.
LTF-S300-10-3D REPROCESSING MANUAL
5
1.5 Reprocessing before the first use
Ch.1
1.5
Reprocessing before the first use
New endoscopes, repaired endoscopes, accessories, and the carrying case for endoscopes are not reprocessed prior to shipping from Olympus, regardless of whether those instruments are for new purchase, demo, or loaner purposes. Reprocess all such endoscopes and accessories received from Olympus according to the instructions given in this manual before storage and before using them in a patient procedure.
1.6
Reprocessing and storage after use WARNING To maintain sterility of equipment following sterilization, use sterilization pouches and wraps according to national guidelines.
1.7
Reprocessing before patient procedure WARNING • Improper storage practices, such as not thoroughly drying external surfaces prior to storage, will lead to an infection control risk. • Improper handling, such as touching a reprocessed endoscope and/or accessories with contaminated gloves, placing a reprocessed device on a contaminated hanger or surface, allowing devices to touch the floor, etc. will recontaminate the device.
NOTE Some national or professional guidelines recommend reprocessing endoscopes prior to their first use of the day. Confirm that the endoscope and accessories have undergone proper reprocessing following their last use and that they have been stored properly. Check the storage period of reprocessed endoscopes, and check for surface contamination (e.g., dust). Check the expiration date marked on all items and check the sterilization pouches for tears or breaches. If there are any doubts or questions concerning whether a device is contaminated, reprocess it again following the instructions given in this manual.
6
LTF-S300-10-3D REPROCESSING MANUAL
2.1 Sterilization cap (MAJ-1538)
Chapter 2
Function and Inspection of the Accessories for Reprocessing
Ch.2
Certain accessories are required for reprocessing the endoscope. This chapter describes the function of these accessories. It also describes how to inspect these accessories before using them to reprocess the endoscope.
2.1
Sterilization cap (MAJ-1538)
Sterilization cap
Figure 2.1
LTF-S300-10-3D REPROCESSING MANUAL
7
2.1 Sterilization cap (MAJ-1538)
Function When performing gas sterilization (e.g., ethylene oxide gas sterilization), the sterilization cap must be attached to the venting connector on the light guide connector.
Ch.2
Venting connector
Sterilization cap Light guide connector
Figure 2.2
Inspection Confirm that the sterilization cap is free from scratches, flaws, and debris.
8
LTF-S300-10-3D REPROCESSING MANUAL
2.2 Instrument basket (WA05938A, sold separately)
2.2
Instrument basket (WA05938A, sold separately) Ch.2
Figure 2.3
Function This instrument basket is used to perform an Automatic Thermal Cleaning/Disinfector and ethylene oxide gas sterilization for the endoscope and accessories. This instrument basket is not used for STERRAD® sterilization and V-PRO® maX sterilization. This instrument basket is also used to transport the sterilized endoscope and accessories.
Inspection NOTE Refer to its instruction manual.
Check every part of the instrument basket for damage.
LTF-S300-10-3D REPROCESSING MANUAL
9
2.3 Channel-opening cleaning brush (MH-507)
2.3
Channel-opening cleaning brush (MH-507)
Ch.2 Shaft Bristles
Brush handle
Brush head
Figure 2.4
Function The channel-opening cleaning brush is used to brush the endoscope and the 3D adjustment cap (MAJ-2266).
Inspection 1 Check the brush head for loose or missing bristles. 2 Check the bristles for any damage. If the bristles are crushed, gently straighten them with your gloved fingertips.
3 Check for bends, scratches, and other damage to the shaft. 4 Visually check for debris on the shaft and/or the bristles of the brush head. If there is debris on the brush, immerse the brush in the water as described in Section 3.2, “Water (for reprocessing)” and clean the brush until no debris is observed on the brush. NOTE The channel-opening cleaning brush does not need to be reprocessed prior to its first use.
10
LTF-S300-10-3D REPROCESSING MANUAL
3.1 Compatibility summary
Chapter 3
Compatible Reprocessing Methods and Chemical Agents Ch.3
3.1
Compatibility summary
The endoscope and accessories are compatible with several methods of reprocessing. However, not all reprocessing methods are compatible with all endoscopes and all accessories. Reprocessing with incompatible methods can cause equipment damage even if the number of reprocessing cycles is small. For appropriate reprocessing methods, see Table 3.1 and 3.2. Follow the policies at your local institution when choosing which methods listed in Table 3.1 and 3.2 to employ.
CAUTION • The methods listed as “compatible” in Table 3.1 are compatible for routine use only when used according to manufacturer’s instructions. Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear. Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration. In general, sterilization processes are harsher on equipment than disinfection processes. Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion “OPERATION MANUAL”. • The instructions provided in this manual regarding material compatibility are not valid for Olympus devices repaired by a non-Olympus facility. Olympus repairs devices to manufacturer’s specifications using original equipment manufacturer’s (OEM) materials. The use of non-OEM materials to repair an Olympus device may affect the material compatibility of the device with certain reprocessing chemicals or methods. In the event that your device has been repaired by a non-Olympus facility, contact the repair facility for instructions regarding material compatibility. Olympus provides information regarding validated methods in terms of the following two criteria. Microbiological efficacy Material durability
LTF-S300-10-3D REPROCESSING MANUAL
11
3.1 Compatibility summary
Microbiological efficacy If it is stated that a method is “validated” in terms of microbiological efficacy, it means that instruments have been reprocessed successfully using the standardized method described in this manual.
Ch.3
Material durability If it is stated that a method is “validated” in terms of material durability, it means that the method can be adopted as a repetitive reprocessing procedure. Just because the material durability is confirmed, it does not mean that a certain degree of microbiological effectiveness is guaranteed.
List of compatible methods validated in terms of microbiological efficacy and material durability For sterilization
Steam sterilization (autoclaving) Ethylene oxide gas sterilization (gas mixture: 20% ethylene oxide gas/80% CO2 for countries other than the USA) Ethylene oxide gas sterilization (100% ethylene oxide gas)
For cleaning
Detergent solution Ultrasonic cleaning
Endoscope
Sterilization cap (MAJ-1538) 3D adjustment cap (MAJ-2266) Channel-opening cleaning brush (MH-507) compatible
Table 3.1
12
LTF-S300-10-3D REPROCESSING MANUAL
not compatible
3.1 Compatibility summary
NOTE Endoscopes and accessories that are marked by the words “AUTOCLAVE” or “AUTOCLAVABLE”, or with green markings (such as a green component or label) are compatible with steam sterilization (autoclaving).
Ch.3
Green marking
Channel-opening cleaning brush (MH-507)
Figure 3.1
LTF-S300-10-3D REPROCESSING MANUAL
13
3.1 Compatibility summary
List of compatible methods validated in terms of material durability Low temperature steam and formaldehyde sterilization*1, *2 V-PRO® maX (Flexible cycle, Non lumen cycle)*3
Ch.3
STERRAD® NX® (Advanced cycle, Standard cycle)*4, *5 STERRAD® 100NX® (Duo cycle)*4, *5 STERRAD® 100S (Long cycle*2, Short cycle)*4 Automatic thermal cleaning/ disinfector Endoscope *6 Sterilization cap (MAJ-1538)
*6
3D adjustment cap (MAJ-2266)
*6
Channel-opening cleaning brush (MH-507)
compatible
not compatible
Table 3.2 *1 This method may not be available in some areas. *2 For countries other than the USA. *3 STERIS Corporation validates the microbiological efficacy of Olympus products in sterility. As for the microbiological efficacy, contact STERIS Corporation. Also, refer to the “V-PRO® maX” instruction manual. *4 ASP (Advanced Sterilization Products) validates the microbiological efficacy of Olympus products in sterility. As for the microbiological efficacy, contact ASP. Also, refer to the “STERRAD® 100S/NX®/100NX®” instruction manual. * 5 The endoscope and accessories listed as compatible with STERRAD® NX®/100NX® can be processed with or without ALLClear Technology. For further details, contact ASP. *6 For further details, contact Olympus.
14
LTF-S300-10-3D REPROCESSING MANUAL
3.1 Compatibility summary
STERRAD® 100S/NX®/100NX® CAUTION • When performing STERRAD® 100S/NX®/100NX® sterilization, the sterilization cap (MAJ-1538) must be attached to the venting connector.
Ch.3
Sterilization cap
Connecting the sterilization cap
Figure 3.2 • Sterilization by STERRAD® 100S/NX®/100NX® system may deteriorate the adhesive of the insertion section. Depending on the circumstances, replacement of the insertion section may be required. Before use, confirm that the endoscope is free from any damage or other irregularities. For further details, contact Olympus.
LTF-S300-10-3D REPROCESSING MANUAL
15