INSTRUCTION MANUAL
ENGLISH
ESPAÑOL
Automatic Blood Pressure Monitor with Printer Model HEM-705CP
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 2
TITLE TABLE OF CONTENTS TITLE
Before Using the Monitor Introduction ... 3 Important Safety Notes ... 4 Before Taking a Measurement ... 5 Operating Instructions Know Your Unit ... 6 Battery Installation ... 7 Connecting the Printer ... 8 Loading the Paper ... 9 Setting the Date and Time ... 10 Applying the Arm Cuff ... 12 Taking a Measurement ... 14 Using the Memory Function ... 16 Using the Printer ... 17 Care and Maintenance Care and Maintenance ... 20 Error Indicators and Troubleshooting Tips ... 21 Quick Reference Guide ... 23 FCC Statement ... 25 Warranty Information ... 26 Specifications ... 27
2
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 3
INTRODUCTION Thank you for purchasing the Omron® HEM-705CP Intellisense® Automatic Blood Pressure Monitor with Printer
Fill in for future reference DATE PURCHASED: ______ SERIAL NUMBER: ________
Staple your purchase receipt here.
Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use. The HEM-705CP comes with the following components: • Monitor • Standard Cuff (9''-13'' arm circumference) • Printer • Printer Tape • Instruction Manual printed in English and Spanish
The HEM-705CP Intellisense® Automatic Blood Pressure Monitor with Printer is intended for home use.
SAVE THESE INSTRUCTIONS 3
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 4
IMPORTANT SAFETY NOTES
To assure the correct use of the product basic safety measures should always be followed including the precautions listed below:
þ þ þ þ þ þ þ
Read all information in the instruction book and any other literature included in the box before using the unit.
Contact your physician for specific information about your blood pressure. Follow the instructions of your healthcare provider.
Operate the unit only as intended. Do not use for any other purpose. The unit is intended for use in measuring blood pressure and pulse rate in adult patient population. Do not use on infants. Do not use a cellular phone near the unit. It may result in operational failure.
Do not plug or unplug the adapter power cord with wet hands.
Changes or modifications not approved by Omron Healthcare will void the user warranty. Do not disassemble or attempt to repair the unit or components.
4
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 5
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, smoking, and exercising for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement. 2. Stress raises blood pressure. Avoid taking measurements during stressful times. 3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
5. Sit in a chair with your feet flat on the floor. Rest your left arm on a table so that the cuff is at the same level as your heart. 6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.
8. Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement. You may need to increase the wait time depending on your individual physiological characteristics.
5
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 6
KNOWTITLE YOUR UNIT TITLE Main Unit:
Heart Symbol
AC Adapter Jack
Adjust Date/Time
Set Date/Time
Systolic Blood Pressure
Printer Jack Cuff Jack
ON/OFF Button
Diastolic Blood Pressure
START Button
Pulse Rate Date/Time Display
Memory Button
Components: Printer
Printer Cover Printer Cable Plug
Power/Operation Light
Print Graph
Paper Feed
Print All Data
Print Current Data/Stop Printing
Arm Cuff
Air Plug (Connects to Air Jack)
Air Tube
6
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 7
BATTERY INSTALLATION/REPLACEMENT TITLE TITLE Main Unit:
1. Press the t OPEN indicator on the battery cover and slide the cover off in the direction of the arrow.
Printer Unit:
2. Install 4 “AA” size batteries so the + (positive) and (negative) polarities match the polarities of the battery compartment as indicated. 3. Replace the battery cover.
BATTERY REPLACEMENT Low Battery Indicator
When the Low Battery Indicator appears on the display screen remove all the batteries. Replace with four new batteries at the same time. Long-life alkaline batteries are recommended. NOTE: Measurement values stored in the memory will not be deleted during battery replacement. After replacing the batteries, reset the date and time. Refer to page 9, Setting the Date and Time.
7
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 8
TITLE CONNECTING THE PRINTER TITLE 1. Remove the printer cord from around the printer unit.
2. Connect the printer to the blood pressure monitor. Insert the cable plug with the indicator facing upward.
CAUTION: Inserting the cable plug in any other position will damage the monitor.
8
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 9
LOADING THE PRINTER PAPER This printer uses OMRON® thermal paper, part number 90TRP. Use only this thermal paper or thermal paper of similar size, width 38mm, roll diameter not to exceed 26mm. NOTE: For information on where to purchase printer paper visit our website at www.omronhealthcare.com or call customer service at 1-800-634-4350.
1. Open the printer cover in the direction as indicated by the arrow symbols in the illustration.
2. Peel back the glued edge of the printer paper. Cut off the first 4 inches of the paper with a pair of scissors. 3. Insert the edge of the printer paper in paper feeding slot. Press the FEED button until the paper extends approximately 3 inches. 4. Insert the paper through the printer cover. Make sure the paper extends out from the top. 5. Close the printer cover.
9
4 in
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 10
SETTING THE DATE AND TIME
The Blood Pressure Monitor automatically stores up to 28 measurement values with the date and time. Set the monitor to the current date and time before taking a measurement for the first time. If the current date and time is not set, a default date and time will display.
1. SETTING THE YEAR When the batteries are installed, the 12:00 AM will display.
Press and hold the SET button for 2 seconds to adjust the year. The year flashes on the display. 2. SETTING THE YEAR The year can be set between 2001 and 2030. When the display reaches 2030, it will return to 2001. Press the ADJUST button to advance by increments of one year. Press the SET button to set the current year. The month flashes on the display.
NOTE: Press and hold the ADJUST button to advance the value quicker. 10
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 11
SETTING THE DATE AND TIME TITLE
3. SETTING THE MONTH Press the ADJUST button to advance by increments of one month. Press the SET button to set the current month. The day flashes on the display.
4. SETTING THE DAY Press the ADJUST button to advance by increments of one day. Press the SET button to set the current day. The hour flashes on the display.
5. SETTING THE HOUR Press the ADJUST button to advance by increments of one hour. Press the SET button to set the current hour. The minutes flash on the display.
6. SETTING THE MINUTE Press the ADJUST button to advance by increments of one minute. Press the SET button to set the current minute. The unit will be ready to take a reading.
7. Press the ON/OFF button to turn the display off.
11
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 12
TITLE TITLE APPLYINGTITLE THE ARM CUFF
1. Make sure the air plug is securely inserted in the main unit.
P O O D CB L M ATI M - 7 11 AU TOO D EL H E M
E SU R RES
I TO R O Y TE AD U R REEAS M
E L AC I ES REPATTER B S G RES P RO
IC TO L SY S
L I STO D IA
M O N
C
E/ m i n P U LS
2. Remove tight-fitting clothing from your upper arm. 3. Sit in a chair with your feet flat on the floor. Place your arm on a table so the cuff is level with your heart. APPLYING THE CUFF ON THE LEFT UPPER ARM 4. Put your arm through the cuff loop. The Green Marker on Bottom Edge 1/2 Inch the cuff should lie over the brachial artery on the inside of the arm. The air tube runs down the inside of your arm. The bottom of the cuff should Green Marker be approximately 1/2'' above your elbow.
12
Cuff Tube
O N /O
FF
RT STA
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 13
APPLYINGTITLE THE ARM CUFF TITLE TITLE 5. Pull the cuff so the top and bottom edges are tightened evenly around your arm. 6. When the cuff is positioned correctly, press the sewn hook material FIRMLY against the pile side of the cuff.
7. Make certain the cuff fits snugly around your arm. The cuff should make good contact with your skin. NOTE: You should be able to fit your index finger between the cuff and your arm easily, so you can pull the cuff off and on.
8. Relax your arm and turn your palm upward. 9. Be sure there are no kinks in the air tubing.
13
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 14
TITLE TITLE TAKING A TITLE MEASUREMENT
The monitor automatically determines your ideal inflation level. In rare circumstances when a higher inflation may be necessary, the monitor automatically re-inflates the cuff up to 30mmHg higher than the initial inflation and restarts the measurement.
The monitor detects the pulse during inflation. Do not move your arm and remain still during the measurement. 1. Press the ON/OFF button
All display symbols appear on the screen. The cuff starts to inflate automatically. NOTE: To stop the inflation or measurement, push the
START/STOP button. The monitor will stop inflating, start deflating, and will turn off.
2. Press the START button
As the cuff begins to inflate, the monitor automatically determines your ideal inflation level. Because this monitor detects the pulse even during inflation, do not move your arm and remain still until the entire measurement completes.
3. Inflation stops automatically and the measurement is started.
Deflation
Inflation
As the cuff deflates, decreasing numbers appear on the display. The Heart Symbol flashes at every heartbeat. 14
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 15
TITLE TAKING A TITLE MEASUREMENT TITLE
4. When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed.
5. Press the ON/OFF button to turn the monitor off.
NOTE: The monitor will automatically turn off after five minutes.
INSTRUCTIONS FOR SPECIAL CONDITIONS If your systolic pressure is known to be more than 220mmHg, press and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure. NOTE:
• Do not apply more pressure than necessary. • The monitor will not inflate above 299 mmHg.
15
START
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 16
TITLE USING THE MEMORY TITLE FUNCTION TITLE
Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory. The Blood Pressure Monitor automatically stores up to 28 measurement values. When 28 sets of measurement values are stored in the memory, the oldest record is deleted to save the most recent measurement values.
TO DISPLAY THE MEASUREMENT VALUES The date and time is alternately displayed with the measurement values. 1. Press the button down while the Heart Symbol is displayed. The most recent measurement values will be displayed.
NOTES:
• Press the button repeatedly to display the next values. • Press and hold the button to display the values faster.
2. Press the START/STOP button to turn the power off.
TO DELETE ALL VALUES STORED IN THE MEMORY
You cannot partially delete records stored in the memory. All records will be deleted. 1. Press and hold the memory button and START button simultaneously for more than 2 seconds. START
16
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 17
USING TITLE THE PRINTER TITLE
The HEM-705CP printer unit has the capability to print all the of measurement values stored in the memory of the Blood Pressure Monitor. NOTES:
• The correct date and time must be set up on the monitor before taking a reading. If the date and time is not set, the numerical values will not appear on the print out.
• The printer unit must be connected to the monitor. The monitor does not have to be ON to operate the printer unit.
• The printer unit will automatically shut off when the printing is completed. • To stop printing, press the DATA/STOP button at any time.
1. TO PRINT THE MOST RECENT MEASUREMENT VALUES Press the DATA/STOP button. The numeric results of the most recent reading will print.
Month/Date
Systolic Reading Diastolic Reading Pulse Reading
7/9
6:36 AM
SYS DIA PULSE
111mmHg 67mmHg 77/min
17
Time
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 18
TITLE USING TITLE THE PRINTER
2. TO PRINT ALL THE MEASUREMENT VALUES Press the ALL DATA button. The numeric printout will list all of the readings stored in the memory starting with the oldest reading to the most recent reading followed by the average of all the stored readings.
The stored data will print as illustrated in the following example: Systolic Reading Column
Month(s) Date
Time/Hour
3-6
D. T. 20 15 23 11 1 15 3 14 5 7 15 7 20 12 3 12 10 12 12 14 14 7 21 15 22 12 4 6 *AV.
Diastolic Reading Column
mmHg
SYS 113 117 113 110 112 110 106 110 108 111 107 111 121 112 112
DIA 75 69 71 67 66 61 66 65 61 56 64 64 62 64 65
/min
PUL 67 64 63 66 64 64 65 64 67 62 64 64 66 63 65
Pulse Reading Column
Oldest reading stored
Most recent reading Average of all stored readings
Month(s) – The month or months will print. For example: If the blood pressure readings were stored during the months of March through June, then the printout will indicate 3-6. Date – The day’s date will print. For example: If the measurement was stored on June 1st, then the printout will indicate 1 under the column ''D''. Time/Hour – The hour in military time will print. For example: If the measurement was stored at 3:00pm, then the printout will indicate 15 under the column ''T''. 18
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 19
USING TITLE THE PRINTER TITLE
3. TO PRINT A GRAPH OF ALL MEASUREMENT VALUES Press the GRAPH button. The graph will list all of the readings stored in the memory.
The stored data will print as illustrated in the following example: Month(s)
3-6 50
100
150
If reading is lower than 50 mmHg or higher than 200 mmHg, (3) will appear. Diastolic Reading
mmHg 200
Oldest reading stored
Systolic Reading
Most recent reading Average systolic reading of stored measurements
Average diastolic reading of stored measurements
19
HEM-705CP-IM-US.qxd
10/29/09
11:06 AM
Page 20
TITLE CARE ANDTITLE MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage follow the directions listed below: • Keep the monitor in the storage case when not in use.
• Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
• Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile cleaners. Do not attempt to clean the cuff. Never immerse the monitor or any components in water. • Store the monitor in a safe and dry location. Do not subject the monitor to extreme hot or cold temperatures, humidity and direct sunlight.
• Avoid subjecting the monitor to strong shocks, such as dropping the unit on the floor. • Remove the batteries if the unit will not be used for three months or longer. Always replace all the batteries with new ones at the same time.
• Use the unit consistent with the instruction provided in this manual. Use only authorized parts and accessories.
20