Instruction Manual
31 Pages
Preview
Page 1
®
Oxford Elevate
English
Manufacturer’s Contact Details
Oxford
®
EUROPE Joerns Healthcare Limited Drakes Broughton Business Park, Worcester Road Drakes Broughton, Pershore, Worcestershire WR10 2AG United Kingdom Tel: 0844 811 1156 • Fax: 0844 811 1157 [email protected] • www.joerns.co.uk
Contents 1. Oxford Elevate Patient Lift ...
3
2. Introduction: About Your Lift ...
4
3. Assembly and Commissioning Instructions ...
5
4. Sling Guide ...
9
5. Lifting with the Oxford Elevate... 11 6. Safety Precautions... 12 7. Operating Instructions... 13 8. Charging Instructions ... 16 9. Digital Weigh Scale ... 18 10. Maintenance schedule & Daily Check List ... 22 11. Technical Specifications ... 23 12. Warranty ... 26
2 294000.10065 Rev. B
®
Oxford Elevate
English
1. The Oxford Elevate Patient Lift
Push Handle
Digital Weigh Scale (optional)
Spreader Bar (Cow-horns) Handgrips (US only)
Battery/ Control Pack Electric Actuator
Sling Attachment Points
Mast Extrusion Adjustable Knee Pad
Rear Castor (braked)
Removable Foot Plate
Front Castor (non-braked)
Note: For lifts sold outside of the US, dedicated hand grips are not incorporated. Patients are therefore encouraged to use the cow-horns if additional support is required.
3 294000.10065 Rev. B
®
Oxford Elevate
English
2. Introduction: About Your Lift The Oxford Elevate is an electrically operated patient lift. Each Elevate lift is fully assembled, load tested and certified before being packed/shipped. The packing consists of a strong, purpose built carton that is used for both export and domestic markets to ensure the safe arrival of the lift. A number of documents are supplied in a wallet, and packed with each lift and should be kept safely for future reference. • USER MANUAL • DEALER GUARANTEE CARD (Europe only) • CUSTOMER SATISFACTION CARD (Europe only) • PACKING CHECK LIST The TEST CERTIFICATE is an important document and will be required for your insurance records. It is valid for six months and after it has expired the lift should be inspected and serviced per the maintenance schedule. Servicing and periodic testing can be carried out by your authorised supplier. Please ensure your lift is included in their maintenance schedule. If you are at all unsure what your local market servicing requirements are, please check with your dealer, Joerns Healthcare and/or a local government agency (Europe only). The Oxford Elevate is suitable for the following CATEGORIES of lift within the working parameters of the lifts specified in the TECHNICAL SPECIFICATIONS. • Category A - Wheelchair • Category D - Toilet/Shower Chair
The CE mark:
4 294000.10065 Rev. B
The Oxford Elevate carries the CE mark and complies with the following EC directives: • Medical Device Directive (93/42/EEC) • EMC Directive (89/336/EEC) (Electrics only) • Low Voltage Directive (73/23/EEC) (Electrics only)
®
Oxford Elevate
English
3. Assembly and Commissioning Instructions CARTON CONTENTS Place the carton in a clear working area and open carefully. The carton contains: • • • • • • •
OXFORD ELEVATE LIFT WALLET CONTAINING DOCUMENTS HAND CONTROL BATTERY PACK CHARGING LEAD DESK TOP CHARGER/STAND SPIRIT LEVELLING DEVICE
WARNING The Oxford Elevate is heavy and will need to be lifted with care. You may need assistance to lift the Oxford Elevate from the carton.
ASSEMBLY Remove the Elevate from the carton and place on the floor.
5 294000.10065 Rev. B
®
Oxford Elevate
English
1. Tighten the mast assembly with the mast-locking device, which is located at the front of the chassis.
2. Turn the locking knob until hand tight. 3. Line the handle assembly up to the mast, and attach using the fasteners provided (these can be found in the mast). The wrench needed to attach the handle is kept in the user instruction wallet for safe keeping.
CAUTION: Ensure that the handle is secure and in place prior to using lift.
4. Fit power pack to the lift and make sure the latch holding the pack in place is fully engaged. “Click” in place. 6 294000.10065 Rev. B
®
English
Oxford Elevate
5. Fit the hand control to the lift as shown. Once complete, check the hand control is functioning correctly by referring to page 8.
Spirit levelling device
6. A self adhesive spirit levelling device is supplied as a reference tool for use when maximum weighing accuracy is required. For maximum accuracy when weighing, centre the spirit level prior to weighing the patient.
DISASSEMBLY: The lift should not be disassembled unless for service, repair or transport if necessary. Therefore follow the assembly instructions in reverse sequence.
7 294000.10065 Rev. B
®
Oxford Elevate
English
ALWAYS CHECK THE FOLLOWING BEFORE OPERATION • The mast is fully locked into position.
• The red emergency stop button, located on the rear of the control box, is in the OFF (out) position.
• The legs of the lift open and close satisfactorily (This is done via the hand control).
• Push the up and down buttons on the hand control and confirm the boom rises and lowers.
8 294000.10065 Rev. B
®
Oxford Elevate
English
4. Sling Guide The Oxford Elevate is suitable for patients in the SITTING position. The slings suitable for this device are listed as follows: * Oxford Transport sling * Oxford Standing sling
USER GUIDE FOR STANDAID SLINGS The Transport and Standing sling is an easy fit sling, suitable for standing and supporting. It is not a general purpose sling and patients must be assessed for suitability by a qualified person. Transport Sling Attachment Loop Straps: Shoulder
Chest Supports and Buckle Leg Supports Attachment Loop Straps: Legs Feed the sling down the back of the patient leaving the top of the commode aperture at the base of the spine. Check the roll part of the sling is square across the patient’s back. Attach the support strap around the chest of the patient. This strap need not be tight. It is there to hold the sling in position whilst attaching the loops to the Standaid. It may be left off altogether if preferred. Ensure patient’s arms are on the outside of the sling. Raise the patient’s leg and feed the leg strap under and up between the legs. Ensure the sling is not twisted or creased under the thigh. Repeat this procedure for the other leg. Attach the loops on the roll parts of the sling to the white hooks at the end of the hand grips. Without crossing the loops on the leg straps, attach to left and right black knobs on the lifting fork. You would normally use the middle of the three loops. The patient is now ready to be raised from the seat. Once clear of the seat, the patient will be more comfortable if lowered back in to the sitting position.
9 294000.10065 Rev. B
®
Oxford Elevate
English
Standing Sling Attachment Loop Straps
Chest Support
Place the sling down behind patient’s back between base of shoulder blades and bottom of rib cage. Attach the support straps around the front of the patient. This strap need not be tight. It is there to hold the sling in position whilst attaching the sling to the Standaid. It may be left off altogether if preferred. Ensure that the patient’s arms are outside the sling.
NOTE: For detailed fitting instructions, please refer to the user guide supplied with each sling.
WARNING Joerns Healthcare recommends that slings be checked regularly and particularly before use for signs of fraying or damage. DO NOT use slings that are worn or damaged.
WARNING OXFORD RECOMMENDS THE USE OF GENUINE OXFORD PARTS. Oxford sling and lift products are designed to be compatible with one another. For country specific guidance on sling use and compatibility, please refer to the sling label or contact your local market distributor or Joerns Healthcare.
WARNING Refer to maximum weight capacity of lift. Sling capacity is limited by the maximum capacity of the lift.
10 294000.10065 Rev. B
®
Oxford Elevate
English
5. Lifting with the Oxford Elevate Position the Elevate so that the patient’s feet fit in the foot tray and knees rest against the knee pad. Attach the loops of the sling to the hooks at the end of the Elevate hand grips. The patient’s hands should be placed on the handgrips (see Fig. A or C). Then push the UP button on the hand control to raise the patient to the standing position (see Fig. B or D). Note: For lifts sold outside of the US, dedicated hand grips are not incorporated. Patients are therefore encouraged to use the cow-horn if additional support is required.
STANDING SLING:
Fig. A
Fig. B
TRANSPORT SLING:
Fig. C
Fig. D
WARNING! Always utilise the foot tray when standing or seating a patient. Once the patient is in a standing position, the foot tray may be removed for gait training. 11 294000.10065 Rev. B
®
Oxford Elevate
English
6. Safety Precautions Please read and follow the safety precautions listed below. The operation and use of Oxford patient lifts is simple and straightforward. Following these few basic safety precautions will make lifting operations easy and trouble free.
READ AND UNDERSTAND THE USER INSTRUCTION MANUAL BEFORE USING YOUR “ELEVATE”
WARNING! • ALWAYS plan your lifting operations before commencing. • ALWAYS carry out the DAILY CHECK LIST before using the lift. • ALWAYS familiarise yourself with the operating control and safety features of a lift before lifting a patient. • DO NOT use a sling unless it is recommended for use with the lift. • ALWAYS check the sling is suitable for the particular patient and is of the correct size and capacity. • NEVER use a sling which is frayed or damaged. • ALWAYS fit the sling according to the instructions provided (user instructions). • DO NOT exceed the safe working load. • ALWAYS check the safe working load of the lift is suitable for the weight of the patient. • ALWAYS carry out lifting operations according to the instructions in the user manual. • ALWAYS utilise the foot tray when standing or sitting a patient. • NEVER disconnect or bypass a control or safety feature because it seems easier to operate the lift. • DO NOT lift a patient with the castor brakes on. Always let the lift find the correct centre of gravity. • DO NOT attempt to manoeuvre the lift by pushing on the mast, boom or patient. • ALWAYS manoeuvre the lift with the handle / foot push pad provided. • ALWAYS lower the patient to the lowest comfortable position before transfers. • DO NOT push a loaded lift at speeds which exceed a slow walking pace (3 Km/hour 0.8 metres/second). • DO NOT push the lift over uneven or rough ground. Particularly if loaded. 12 294000.10065 Rev. B
®
Oxford Elevate
English
• DO NOT attempt to push/pull a loaded lift over a floor obstruction. • NEVER force an operating/safety control. All controls are easy to use and do not require excessive force. • The MINIMUM load required to manually lower the hoist using the mechanical emergency down function is 30 kg. • DO NOT park a loaded lift on ANY sloping surface. • DO NOT use electric lifts in a shower. • DO NOT charge an electric lift in a bathroom or shower room. • DO NOT lift a patient unless you are trained and competent to do so. • YOUR lift is for patient lifting. DO NOT use it, or allow it to be used, for any other purpose. • DO NOT bump the lift down steps, loaded or unloaded. • DO NOT attempt to negotiate a loaded lift on a slope which exceeds 1:12 (approximately 5 degrees). • DO NOT attempt to negotiate a slope without a second helper being present. • DO NOT use in a wet or corrosive environment such as poolside locations.
7. Operating Instructions OPERATING CONTROLS 1. Leg adjustment Detachable Battery Pack Emergency Lower/Raise
Actuator Motor
Emergency Stop Button
Hand Control
Battery Indicator
13 294000.10065 Rev. B
®
Oxford Elevate
English
The legs on the Oxford Elevate are electrically adjustable for width. The legs can be opened to enable access around armchairs or wheelchairs. For transferring and negotiating narrow doorways and passages the lift legs should be in the closed position. Electric leg adjustment - is achieved by pressing the appropriate buttons on the handcontrol. The legs will be locked whenever the handcontrol switch is released. 2. Castors and Braking The lift has two braked castors which can be applied for parking. When lifting, the castors should be left free and un-braked. The lift will then be able to move to the centre of gravity of the lift. If the brakes are applied it is the patient that will swing to the centre of gravity and this may prove disconcerting and uncomfortable. 3. Raising and lowering the boom The movement of the boom is achieved by a powerful electric actuator, which is controlled by a simple handcontrol unit. The handcontrol has two buttons with directional arrows UP and DOWN . The actuator stops automatically at the limit of travel in both directions. The handcontrol plugs into a socket at the base of the control box. 4. Emergency Stop The red Emergency Stop Button is located on the front of the control box and is activated by pressing in. This will cut all power to the lift and only be reset by twisting the button anti-clockwise and releasing. 5. Electrical Emergency Down/Up Emergency lower (down) and raise (up) buttons are provided at the front of the control box. This is operated by inserting a ballpoint pen into the button highlighted . This will bring the boom up or down should the handcontrol fail at any time. NOTE: This is an emergency device and not intended for repetitive use. 6. Mechanical Emergency Down In the case of a complete electrical failure the electrical actuator is fitted with mechanical lowering device (RED BOSS). This will only operate when the lift is under load. The device must be pulled upwards to activate, and a slow decent will commence. Note: The MINIMUM load required to manually lower the hoist using the mechanical emergency down function is 30 kg. This is an emergency device and not intended for repetitive use. 7. Batteries The batteries are protected from deep discharge by a LOW VOLTAGE ALARM. This will sound when the batteries need recharging and the handcontrol is being operated. It will not sound independently of the handcontrol being operated. DO NOT IGNORE THIS WARNING ALARM. Complete the lifting operation and place the battery on charge (see charging instructions).
WARNING! Battery or charger is not to be opened by unauthorised personnel. (Contact your distributor or Joerns Healthcare for warranty and repairs).
14 294000.10065 Rev. B
®
Oxford Elevate
English
8. Slings The Oxford Elevate is suitable for patients in the SITTING position. The slings suitable for this device are the Oxford Standing sling and Transport sling as detailed in section 4 (Sling Guide). 9. Knee pad adjustment The Elevate has an adjustable kneepad. It is recommended for safety reasons that adjustment is carried out by holding the top of the adjustment handle. The range of travel is approx 7 inches. Adjustment is achieved by following two easy steps. STEP 1 pull handle towards you, STEP 2 either push handle downwards or pull handle upwards to select the kneepad position. Once the ideal position is found release handle and the kneepad will automatically lock into position.
10. Removable footplate The Elevate has a removable footplate. To remove the footplate, move to the front of the product, and lift the footplate upwards from the middle of the assembly (see Fig. E). Additionally you can also remove the footplate cover to aid in cleaning and disinfecting of the footplate area. This is achieved by removing the footplate and unclipping the cover from the back of the footplate (see Fig. F). NOTE: Be careful when lifting the footplate from its low-seated position.
Fig. E
Fig. F
WARNING! Always utilise the foot tray when standing or seating a patient. Once the patient is in a standing position, the foot tray may be removed for gait training. 15 294000.10065 Rev. B
®
Oxford Elevate
English
8. Charging Instructions The batteries are located in the battery pack and are charged via an off board desktop charger unit.
Removable Battery Pack
Hand Control Legs Open/Close, Boom Raise/Lower Control Box Emergency Stop, Raise/Lower
When the battery pack needs charging it is removed from the lift and fitted to an off board charging unit. 1. Remove the power pack from the lift. The pack is retained by a simple latch at the top of the power pack. Lift the latch and the power pack will be released. 2. Fit the power pack to the charging unit. The location and latching of the power pack to the charger is the same system as used on the lift. 3. Plug the charger mains plug into a suitable mains outlet and switch the mains supply ON. 4. Charging is automatic and will fully charge the batteries over a period of eight to twelve hours. Note: Even if the charger is left plugged in for extended periods it will not allow the batteries to overcharge. 5. To return the lift to service, disconnect power supply remove the power pack from the charger. Fit the power pack to the lift and make sure the latch holding the pack in place is fully engaged. “Click” in place. The charging of Oxford electric lifts is simple and straightfor ward, but it is important to follow the charging instructions closely. Please pay particular attention to the following warnings, they will help you avoid problems with discharged batteries.
16 294000.10065 Rev. B
®
Oxford Elevate
English
WARNING Battery or charger is not to be opened by unauthorised personnel. (Contact your distributor or Joerns Healthcare for warranty and repairs).
WARNING DO NOT touch battery/charger terminals.
WARNING NOTE: KEEP the batteries fully charged. Place the battery on charge whenever it is not in use. If it is more convenient to do so, place on charge every night. The charger will not allow the batteries to overcharge. NOTE: NEVER run the batteries completely flat. As soon as the audible warning sounds, complete the lifting operation in hand and place on charge. NOTE: NEVER store the power pack for long periods without regular charging throughout the storage period.
WARNING Do not charge an electric lift in a bathroom or shower room.
WARNING DO NOT leave charger switched on with battery disconnected.
17 294000.10065 Rev. B
®
Oxford Elevate
English
9. Digital Weigh Scale (Optional) The Oxford Elevate has the option of an integrated digital weigh scale.
ABOUT YOUR SCALE: The Oxford Digital Scale is a technologically advanced, microprocessor controlled, digital weighing device. It features a large, easy to read, Liquid Crystal Display (LCD) that shows the patient’s weight. The controls are simple and straightforward to use. The Oxford Digital Scale is powered by a 9-volt disposable battery, which will provide approx. 3000 weighing operations before replacement will be necessary. The Oxford Digital Scale is suitable for the following CATEGORIES of lift: Category A - Wheelchair Category D - Toilet/Shower Chair The Oxford Digital Scale is suitable for patients in the SITTING position. The CE mark (Europe Only): The Oxford Digital Scale carries the CE mark and complies with the following EC Directives: - Medical Device Directive (93/42/EEC) - EMC Directive (89/336/EEC) - Low Voltage Directive (73/23/EEC) - NAWI Directive (90/384/EEC) OPERATING INSTRUCTIONS: The Oxford scale has been designed and certified to weigh up to 200kg or 440lbs in 0.2 increments. NOTE: For customers or users in North America, your scale will measure in either Lbs or Kgs. For all other countries outside NA, your scale will weigh in Kgs.
18 294000.10065 Rev. B
®
Oxford Elevate
English
To weigh a patient follow the instructions below: 1. Assemble the sling to the cow horn and press the ON/ZERO pad. The display will show 0.0 and the word HOLD will appear in the lower left hand corner. This step prepares the scale for weighing. The sling can now be removed from the cow horn or cradle and placed around the patient. 2. Place the sling around the patient ensuring the patient’s feet are in the foot tray. Attach the sling to cow horns and raise the patient so that they are in the standing position. They must not be touching anything such as a chair or bed since this will cause an error in weighing. 3. The scale has been designed to turn itself off after 60 seconds to conserve battery life. To restart the scale, push the ON/ZERO pad (once only). The display will show the start up screen and then show the patient’s weight. When the HOLD sign appears, the weight can be recorded. NOTE: To restart the scale, be sure to push the ON/ZERO. If the scale starts and the pad is pushed again, the scale may go to zero (0.0) and the patient’s weight may be lost. This may require the reweighing of the patient. 4. The RECALL feature is designed to retain in the memory the last stable weight over 10 kg/22lbs. To access this weight, the scale must be on. Press the RECALL pad. The display will flash between “RECL” and the retained weight. MAINTENANCE: It is recommended that twice yearly inspections of the connections on the scale be scheduled and performed to assure trouble free operation. Look for wear or any irregularities that may contribute to scale failure. If any wear or irregularity is seen DO NOT USE THE LIFT and contact Joerns Healthcare to access the assembly. TECHNICAL SPECIFICATIONS: • Electronics. Microprocessor controlled. • Power. 9-volt alkaline battery PP3. • Battery life. Exceeds 70 hours continuous operation. • Display. LCD (liquid crystal display) 63 X 18 mm Characters 15 mm high. • Weighing units. Kilograms (kgs) or (Lbs and kgs North America only) • Load cell. Rated at 300 kgs/660 lbs.
WARNING Do not exceed the safe working load of the lift.
19 294000.10065 Rev. B
®
Oxford Elevate
English
A 9-volt battery powers the Oxford Electronic Digital Scale. A new battery should provide approximately 3000 readings before needing replacement. When it needs to be replaced, a “B” will appear on the display. To Replace Battery: 1. 2. 3. 4.
Push black pad on bottom of unit to left. Battery holder will slide out. Remove old battery and replace with a new one. Push holder back into scale.
OPERATION FEATURES: • All functions are accessible to the user without opening the case. • Starting or zeroing the scale is done through a single touchpad located on the front of the case. • Stable readings are achieved through digital averaging. The HOLD message is activated when weight fluctuations decrease to a predetermined range. The scale becomes active again when that range is exceeded. The held weight is stored in a non-volatile memory for later recall if required. • Display turns off after 60 seconds of inactivity to conserve battery life. • Recall of the last held weight is available via a touchpad on the front of the case. This allows a reading of the last patient’s weight, after the patient has been removed from the scale. • Battery change is made easy with external access to the battery holder. MAINTENANCE, INSPECTION AND TEST: Joerns Healthcare recommends that a thorough inspection and test of all Oxford lifts and lifting accessories, slings etc. is carried out every six months. Joerns Healthcare recommends maintenance, inspection and certified testing, is only carried out by competent and authorised service dealers. Note: These recommendations are in compliance with the requirements of 1998 No2307 Health and Safety: The Lifting Operations and Lifting Equipment Regulations 1998. This is a UK regulation. Outside the UK please check your local requirements. Important Note (Europe Requirement Only) The Oxford Weigh scale is a class 3 compliant weighing device and conforms to the NAWI Directive 90/384/EEC and has been certified by the National Weights and Measures Laboratory. NOTE: Scale is calibrated prior to shipment. This certifies the scale for twelve months. Joerns Healthcare’s advice is that you contact your local trading standards / government authority or Joerns Healthcare for your nearest authorised body who can advise upon recalibration.
20 294000.10065 Rev. B
®
English
Oxford Elevate
Initially
Before Use
Service Intervals
*THESE CHECKS SHOULD INCLUDE:
1. BOOM: Check the attachment of the boom to the mast. Make sure there is only minimal side movement of the boom and the boom is free to rotate on the boom bearing.
2. MAST: Check the operation of the mast-locking device. Make sure the mast fully engages into the socket. Check the bottom actuator.
3. ACTUATOR: The actuator should require no maintenance other than checking for correct operation and listening for unusual noise.
4. POWER PACK: Check the function of the emergency stop button and emergency lower. Note: The MINIMUM load required to manually lower the hoist using the emergency lower function is 30 kg.
5. BATTERIES: The batteries are housed in the power pack and should not require maintenance other than the regular charging as detailed in the charging instructions. Check that the connections remain clean.
6. LEG ADJUSTMENT: Check the legs operate in both full extensions (inward/outward).
7. CASTORS: Check all castors for firm attachment to the legs. Check for free rotation of the castor and the wheels.
8. CLEANING: Clean with ordinary soap and water and/or any hard surface disinfectant. Harsh chemical cleaners or abrasives should be avoided as these may damage the surface finish of the lift. Avoid wetting any of the electrical parts. 9. LOAD TEST: The load test should be carried out in accordance with the manufacturer’s test procedures. It is strongly recommended the testing is carried out by an authorised service dealer. 10. BASE AND WHEELS: Ensure base is even and level (all four wheels are on the floor). 11. SLINGS: Check for wear and fraying. 12. LUBRICATION: Lubricate pivot joints, including mast and boom connections, pedal assembly, spreader bar joint (only if required). (you can use silicone grease/spray). 13. HAND SET: Ensure plugged fully into controller.
21 294000.10065 Rev. B