PHYSIO CONTROL
Beds, Chairs, Stretchers and Stools
Transport Stretcher Model ST104 Maintenance Manual Rev A Sept 2006
Maintenance Manual
69 Pages
Preview
Page 1
Mai nt enance Manual
Medi cal
Tr anspor tSt r et cher ModelST104 I mpor t ant I nf or mat i on Fi l ei ny our ma i nt ena nce r ecor ds
Forpar t sort echni cal assi st ancecal l 8003270770( opt i on2)
Table of Contents Introduction Specifications... 3 Warning / Caution / Note Definition... 3 Summary of Safety Precautions... 4 Preventative Maintenance Checklist... 5 Cleaning... 6 Service Information Caster Cover Installation and Removal... 7 Caster Removal... 7 Brake Rod Removal... 8 Side Control Brake Rod Removal... 8 Release Pedal Adjustment... 9 Foot End Release Pedal Replacement... 9 Foot End Release Pedal Rod Removal... 9 Brake Ring Removal... 10 Fifth Wheel Assembly Removal... 10 Litter Top Removal... 11 Jack Descent Rate Adjustment... 11 Removal of Excess Air from the Hydraulic System... 11 Head End Hydraulic Jack Removal... 12 Foot End Hydraulic Jack Removal... 12 Pneumatic Fowler Adjustment... 13 Folddown Siderail Latch Adjustment... 13 Quick Reference Replacement Parts List... 14 Assembly Drawings and Parts Lists Retractable Fifth Wheel Base Assembly... 15 Fifth Wheel Assembly... 16, 17 Standard Brake Assembly (Fifth Wheel Base)... 18, 19 Brake Rod Assembly... 20 Drive Link Assembly... 21, 22 Caster Assembly... 23 Side Control Brake Assembly... 24 − 26 Brake Rod / Side Control Assembly... 27 Side Control Brake Sub−Assembly... 28 Standard Brake Assembly, Colored Components... 29 Side Control Brake Assembly, Colored Components... 29
Table of Contents Assembly Drawings and Parts Lists (Continued) Dual Side Hydraulics Assembly (Fifth Wheel Base)... 30−32 Standard Hydraulics Assembly, Colored Components... 33 End Control Hydraulics Assembly, Colored Components... 33 Lowering Pedal Assembly... 34 Head End Pump Pedal Assembly... 35 Non−Constant Descent Jack... 36 Jack Assembly, Non−Constant... 37 Jack Base Assembly, Variable Descent Jack... 38 Base Labeling Assembly, Standard Brake Assembly... 39 Base Labeling Assembly, 4−Sided Brake Assembly... 40 Base Labeling Assembly, Colored Components... 41 Base Hood Department Labels... 41 Litter Assembly... 42−44 Siderail Assembly... 45 Pneumatic Fowler Assembly... 46−48 Foot End Push Handles... 49 Head End Push Handles... 50 Push Handle Assembly... 51 Removable I.V. Pole Assembly... 52 2−Stage I.V. Assembly... 53−54 I.V. Pole Latch... 55 3−Stage I.V. Assembly... 56−58 Defibrillator Tray Assembly... 59 Foot Extension/Defibrillator Tray Assembly... 60 Foot Board... 61 O2 Bottle Holder... 61 O2 Bottle Retainer... 62 Mattresses and Siderail Pads... 63, 64 Warranty Obtaining Parts and Service... 65 Service Contract Coverage... 65 Return Authorization... 66 Freight Damage Claims... 66
Introduction INTRODUCTION This manual is designed to assist you with the maintenance of the Model 737 Transport Stretcher. Read it thoroughly before using the equipment or beginning any maintenance on it.
SPECIFICATIONS Maximum Weight Capacity
500 pounds
Overall Stretcher Length Width
83” / 30”
Minimum Maximum Bed Height
21.5” / 36”
Fowler Angle
0_ to 90_
Trendelenburg Reverse Trendelenburg
+18_ to −18_
Minimum Under Stretcher Clearance
6” (15 cm) nominal 1.75” (4.5 cm) under the hydraulic cyclinders and 5th wheel
Stryker reserves the right to change specifications without notice.
WARNING / CAUTION / NOTE DEFINITION The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING Alerts the reader about a situation, which if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards.
CAUTION Alerts the reader of a potentially hazardous situation, which if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. This includes special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse. NOTE This provides special information to make maintenance easier or important instructions clearer.
Safe Working Load
Return to Table of Contents
3
Summary of Safety Precautions Before operating this stretcher, it is important to read and understand all information in this manual. Carefully read and strictly follow the warnings and cautions listed on this page.
WARNING Always apply the caster brakes when a patient is getting on or off the stretcher. Push on the stretcher to ensure the brakes are securely locked. Always engage the brakes unless the stretcher is being moved. Injury could result if the stretcher moves while a patient is getting on or off the stretcher. Be sure the siderail latching mechanism is working properly at all times. If it is not, refer to your stretcher maintenance manual for ”Siderail Latch Adjustment”. When lowering the siderail to the collapsed position, keep extremities of patients and staff away from the siderail spindles or injury could occur. Keep fingers/hands clear of area around the Fowler release handle and the Fowler frame when lowering. Injury could result if care is not taken when lowering the Fowler. If the stretcher is equipped with the optional foot end IV pole, the IV pole must be in the raised position when the foot extension/defibrillator tray is installed. If the IV pole is not raised, the foot extension will not function properly and injury could occur.
CAUTION To avoid damage, remove any equipment that may be in the way before raising or lowering the litter height. To avoid injury or damage to the equipment, do not allow the siderail to lower on its own. The push handles were designed for use while transporting the stretcher. Avoid using other parts of the stretcher as push/pull devices because damage could occur. If the Pneumatic Fowler is difficult to operate, refer to the stretcher maintenance manual for “Pneumatic Fowler Adjustment” To avoid damage, the weight of the I.V. bags should not exceed 40 pounds. To avoid damage while transporting the stretcher, verify the I.V. pole is at a low enough height to allow it to safely pass through door openings and under light fixtures. The unit cannot be raised (hydraulics on base) with a patient lift under the stretcher. The hood may not be used for stepping. Note Clean hood storage area regularly. The bottom of the brake rings should be cleaned regularly to prevent wax and/or floor remnant buildup.
Return to Table of Contents
4
Preventative Maintenance CHECKLIST All fasteners secure Siderails move and latch properly Engage the brake pedal and push on the stretcher to ensure all casters lock securely All casters secure and swiveling properly Body restraints working properly I.V. pole intact and operating properly Oxygen bottle holder intact and operating properly Fowler operates and latches properly Trendelenburg/Reverse Trendelenburg operating properly No rips or cracks in mattress cover Transfer boards intact and operating properly Ground chain intact No leaks at hydraulic connections Hydraulic jacks holding properly Hydraulic drop rate set properly Hydraulic oil level sufficient Lubricate where required Accessories and mounting hardware in good condition and working properly
Serial No. ______________ ______________ ______________ ______________ ______________ Completed By:_________________________________
Date:_____________
NOTE Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product.
Return to Table of Contents
5
Cleaning Hand wash all surfaces of the stretcher with warm water and mild detergent. Dry thoroughly. DO NOT STEAM CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN. Using these methods of cleaning is not recommended and may void this product’s warranty. Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate. (Appropriate disinfectant for nylon Velcro should be determined by the hospital.) In general, when used in those concentrations recommended by the manufacturer, either phenolic type or quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use because staining may result. The following products have been tested and have been found not to have a harmful effect WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME
DISINFECTANT TYPE
MANUFACTURER
*MANUFACTURER’S RECOMMENDED DILUTION
A33
Quaternary
Airwick (Professional Products Division)
2 ounces/gallon
A33 (dry)
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Beaucoup
Phenolic
Huntington Laboratories
1 ounce/gallon
Blue Chip
Quaternary
S.C. Johnson
2 ounces/gallon
Elimstaph
Quaternary
Walter G. Legge
1 ounce/gallon
Franklin Phenomysan F2500
Phenolic
Purex Corporation
1 1/4 ounce/gallon
Franklin Sentinel
Quaternary
Purex Corporation
2 ounces/gallon
Galahad
Phenolic
Puritan Churchill Chemical Company
1 ounce/gallon
Hi−Tor
Quaternary
Huntington Laboratories
1/2 ounce/gallon
LPH
Phenolic
Vestal Laboratories
1/2 ounce/gallon
Matar
Phenolic
Huntington Laboratories
1/2 ounce/gallon
Omega
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Quanto
Quaternary
Huntington Laboratories
1 ounce/gallon
Sanikleen
Quaternary
West Chemical Products
2 ounces/ gallon
Sanimaster II
Quaternary
Service Master
1 ounce/gallon
Vesphene
Phenolic
Vestal Laboratories
1 1/4 ounce/ gallon
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% Sodium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach to 100 parts water are not considered mild detergents. These products are corrosive in nature and may cause damage to your stretcher if used improperly. If these types of products are used to clean Stryker patient handling equipment, measures must be taken to insure the stretchers are rinsed with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the stretchers will leave a corrosive residue on the surface of the stretcher, possibly causing premature corrosion of critical components. NOTE Failure to follow the above directions when using these types of cleaners may void this product’s warranty.
REMOVAL OF IODINE COMPOUNDS This solution may be used to remove iodine stains from mattress cover surfaces. 1. Use a solution of 1−2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area. Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand on the surface. 2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.
Return to Table of Contents
6
Service Information CASTER COVER INSTALLATION AND REMOVAL
1. Double Prongs
Single Prong
Looking through the larger of the two side cut−outs, align the cover with the axle nut or bolt head, as shown. Push down on the opposite side of the cover until the single prong engages the caster horn.
Top View (Cut−Away)
2.
Push on the cover with your palm until the double prongs engage.
Properly Attached Cover
Top View (Cut−Away)
3.
To remove the wheel cover, insert a large screwdriver into the cut−out in the side of the wheel cover and into the space between the double prongs. Pry up the cover to disengage the double prongs and push sharply upward to disengage the single prong.
Top View (Cut−Away)
CASTER REMOVAL Required Tools: 9/16” Open End Wrench 1. Remove the caster cover. 2. While keeping pressure on the caster bolt with your index finger, use a 9/16” open end wrench to remove the nylock hex nut on top of the caster assembly. 3. Using the brake ring, lift up on the base assembly and pull the caster assembly down to remove it. 4. Reverse steps 1−3 to install the new caster.
Return to Table of Contents
7
Service Information BRAKE ROD REMOVAL Required Tools: Hammer
7/32” Punch
String or Bungee Cords
1. Pump the litter up to full height. 2. Lift the base hood and support it from the litter using string or bungee cords. 3. Remove the hex head cap screws connecting the brake rod supports to the base frame. 4. Remove the bolt connecting the drive link assembly to the fifth wheel cam. 5. Remove the rue ring cotter and clevis pin connecting the rod end link to the side control link. 6. Remove the rue ring cotter connecting the drive link assembly to the bearing pivot support on the base frame (under the brake ring weldment). 7. Remove the three hex washer head screws holding the brake rod assembly to the base frame. 8. Remove the slotted spring pins connecting the butterfly “V” pedals, drive link assemblies and side control link to the brake rod. 9. Reverse steps 1−8 to reinstall the brake rod. When reinstalling the brake rod supports, torque the hex head cap screws to 5−13 ft.−lbs. CAUTION When reattaching the brake rod assembly to the base frame, set the torque specs no higher than 13 ft.−lbs. or damage could occur to the bolts.
SIDE CONTROL BRAKE ROD REMOVAL Required Tools: Hammer
7/32” Punch
Needle Nose Pliers
String or Bungee Cords
1. Pump the litter up to full height. 2. Lift the base hood and support it from the litter using string or bungee cords. 3. Remove the rue ring cotter and clevis pin connecting the rod end link to the side control link. 4. Remove the four bolts holding the brake rod assembly to the base frame and remove the entire assembly. 5. Using a hammer and 7/32” punch, drive the slotted spring pin out of the butterfly “V” pedal on the patient’s left side and remove the pedal, 6. Using a hammer and 7/32” punch, drive the slotted spring pin out of the hard stop in the center of the support weldment. 7. Using a hammer and 7/32” punch, drive out the slotted spring pin connecting the side control link to the side control brake rod on the patient’s right side. 8. Pull on the butterfly “V” pedal on the patient’s right side to remove the side control brake rod from the base. 9. Reverse steps 1−8 to reinstall the brake rod. When reinstalling the assembly, torque the hex head cap screws to 5−13 ft.−lbs. CAUTION When reattaching the brake rod assembly to the base frame, set the torque specs no higher than 13 ft.−lbs. or damage could occur to the bolts.
Return to Table of Contents
8
Service Information RELEASE PEDAL ADJUSTMENT 1. Manually disengage the release pedal swivel (item J on page 31) from the release pedal assembly. 2. To increase the release rod engagement with the release valve, turn the release pedal swivel clockwise on the threaded release rod. 3. To decrease the release rod engagement with the release valve, turn the release pedal swivel counterclockwise on the threaded release rod. NOTE If the pedal swivel assembly is threaded too far onto the release rod, the release valve will be partially activated and the jack will drift.
FOOT END RELEASE PEDAL REPLACEMENT Required Tools:
Needle Nose Pliers
1. Apply the stretcher brakes. 2. Disconnect the release pedal return springs from the foot end release pedals. 3. Remove the rue ring cotters and the clevis pins connecting the foot end release pedals to the mounting bracket. 4. Rotate the pedals upward. 5. Remove the rue ring cotters connecting the head end and foot end release rods to the foot end release pedals and remove the pedals. 6. Reverse steps 2−5 to install the new pedals.
FOOT END RELEASE PEDAL ROD REMOVAL Required Tools:
Needle Nose Pliers
1. Remove the foot end release pedal (see procedure above). 2. Remove the snap in nyliners holding the foot end pedal release rods in the pedal mounting bracket. 3. Unsnap the foot end pedal release rods from the white plastic release rod brackets. 4. Dislodge the side control release pedal swivels from the studs on the side control release pedal weldments. 5. Remove the foot end pedal release rods. 6. Reverse steps 1−5 to reinstall the pedal rods.
Return to Table of Contents
9
Service Information BRAKE RING REMOVAL Required Tools: 9/16” Socket w/Extension
3/8” Drive Ratchet
Needle−Nose Pliers
String or Bungee Cord
1. Pump the litter up to full height. 2. Lift the base hood and support it from the litter using string or bungee cords. 3. Using needle−nose pliers, unhook the extension springs from the top of the base caster tubes. 4. Remove the plastic caster covers. 5. While putting pressure on the caster carriage bolt, use a 9/16” socket and a 3/8” drive ratchet to remove the caster nut on both sides of the stretcher. 6. Remove the casters. 7. Remove the brake rod (see procedure). 8. Remove the cotter pin from the clevis pin in the center of the brake ring weldment. 9. Remove the cotter pin from the bearing pivot support. 10. Remove the 3/4” nylock hex nut from the bearing pivot support. 11. Remove the drive link assembly. 12. Pull the brake ring down and out away from the stretcher base frame. 13. Reverse steps 1−12 to reinstall the brake ring.
FIFTH WHEEL ASSEMBLY REMOVAL Required Tools: 1/2” Socket
3/8” Drive Ratchet
1. Using a 1/2” socket and 3/8” drive ratchet, remove the 1/2” bolt holding the fifth wheel cam drive link and fifth wheel drive link to the fifth wheel cam. 2. Remove the two 1/2” bolts holding the fifth wheel mounting bracket to the base frame weldment. 3. Remove the fifth wheel assembly. 4. Reverse steps 1 and 2 to reinstall the fifth wheel. When reinstalling the assembly, torque the hex head cap screws to 5−13 ft.−lbs. CAUTION When reattaching the fifth wheel assembly to the base frame, set the torque specs no higher than 13 ft.−lbs. or damage could occur to the bolts.
Return to Table of Contents
10
Service Information LITTER TOP REMOVAL Required Tools:
1/2” Socket w/Extension
3/8” Drive Ratchet
Standard Screwdriver
1. Using the foot pedal, pump up the litter top to full height. 2. Remove the stretcher mattress 3. Remove the round, black hole plugs from the jack supports at each end of the litter to expose the jack support tube truss head screws. 4. Using a 1/2” socket, and a 3/8” drive ratchet, remove the truss head screws holding the jack support tubes to the jack shafts. 5. Thread a 7/16−20 fine thread bolt far enough into the top of the jack supports to separate the litter top from the jack shaft. 6. With the assistance of another person, lift the litter straight up to remove it from the jack shafts and set it aside.
JACK DESCENT RATE ADJUSTMENT Required Tools: Screwdriver
Bungee Cords (or equivalent)
Adjustment Procedure: 1. Pump the litter up to full height. 2. Lift the base hood, separating the hood from the base frame. Support the hood from the litter using bungee cords so it is out of the way. 3. The descent rate needle valve is located on the base of the jack. Turning the needle valve clockwise, with a screwdriver, will decrease the rate of descent. Turning it counterclockwise will increase the rate of descent. NOTE The larger percentage of a patient’s weight is located in the torso area. Adjust descent rate accordingly. 4. Remove the bungee cords supporting the base hood and secure the hood to the base frame. NOTE The jack descent rate is preset at the factory and adjustment is not recommended.
REMOVAL OF EXCESS AIR (VACUUM) FROM THE HYDRAULIC SYSTEM 1. Verify all hydraulic linkages are secure and operating properly. 2. Using the pump pedal, actuate the system several times to force the air through the system. The jack should now raise properly.
Return to Table of Contents
11
Service Information HEAD END HYDRAULIC JACK REMOVAL Required Tools:
1/2” Socket w/Extension
3/8” Drive Ratchet
1. Remove the litter top from the stretcher (see page 11). 2. Using a 1/2” socket with extension and a 3/8” drive ratchet, remove the two hex head screws holding the jack base to the stretcher base frame. 3. Remove the two hex head screws holding the jack reservoir clamp to the base frame and remove the clamps. 4. Lift straight up on the pump connecting rod and disconnect the pump piston from the connecting rod. 5. Disconnect the pump pedal swivel from the release pedal mounting plate. 6. Remove the head end release rod from the release valve assembly. 7. Using a 1/2” socket with extension and a 3/8” drive ratchet, remove the two hex head screws holding the jack base to the stretcher base frame. 8. Lift out the jack assembly. 9. To reinstall the jack, install the bolts on the jack and reservoir clamp but do not tighten them fully. 10. Reinstall the pump connecting rod and release rod. 11. Depress the pump pedal fully (to the floor). This will properly locate the jack onto the base frame. 12. Tighten the bolts on the jack and reservoir clamp. 13. Pump up the litter and apply weight to verify the jacks hold and do not drift. NOTE The jack descent rate is preset at the factory and adjustment is not recommended.
FOOT END HYDRAULIC JACK REMOVAL (BASE WITH DUAL CONTROLS) Required Tools: 1/2” Socket
3/8” Drive Ratchet
Pliers
1. Remove the litter top from the stretcher (see page 11). 2. Lift the base hood off the base frame. 3. Remove the two hex washer head screws and washers connecting the pump pedal link to the foot end pump pedal assembly and pump connecting rod. 4. Remove the foot end release rod from the release valve on the jack assembly by dislodging the release pedal swivel from the pins on the release pedal weldment. 5. Dislodge the jack pump piston from the pump connecting rod. 6. Remove the two hex washer head screws holding the reservoir clamp. 7. Remove the jack assembly. 8. Reverse steps 1−7 to install the new jack. NOTE The jack descent rate is preset at the factory and adjustment is not recommended.
Return to Table of Contents
12
Service Information PNEUMATIC FOWLER ADJUSTMENT Required Tools: 5/32” Hex Allen Wrench
1/2” Open End Wrench
1. Refer to the pneumatic Fowler assembly drawing on page 47 for parts reference. 2. For easier access, raise the Fowler to 75_ or higher. 3. Using a 1/2” open end wrench, loosen the hex nuts (item E) in the actuator arms on the end of the trip bar (item M). 4. To adjust the Fowler, use a 5/32” hex Allen wrench to turn the Allen screws (item J) 1 to 2 turns counter−clockwise if the Fowler will not move or 1 to 2 turns clockwise if the Fowler will not hold its position. 5. Retighten the hex nuts. Be sure the Fowler travels from flat up to 90_ and down again and holds its position when weight is applied before returning the stretcher to service.
FOLDDOWN SIDERAIL LATCH ADJUSTMENT Required Tools: 1/8” Hex Allen Wrench
WARNING The siderail latches are preset at the factory, and do not normally need adjustment. If adjustment must be done, it is important to follow the procedure below. If adjustment is not done properly, injury to the patient or user could result.
1. Using a 1/8” hex Allen wrench, adjust the hex Allen screw located on the latch assembly opposite the latch. Turning the screw clockwise will DECREASE the amount of ”play” in the latching mechanism. Turning counterclockwise will INCREASE the amount. NOTE The amount of ”play” in the siderail, when in full up engaged position, should be approximately 1/8 to 3/16 inches.
CAUTION Too much ”play” when the siderail is in the full up engaged position will give the siderail the appearance of being unstable and could also cause premature wearing of the latch system. Too little ”play” will obstruct the latch and keep it from engaging completely in the full up position, which may result in damage to the latch and/or injury to the patient or user.
Return to Table of Contents
13
Quick Reference Replacement Parts List
PART NAME
PART NUMBER
Caster Assembly (1 complete)... 0753−010−020 Defibrillator Tray/Foot Extension... 0785−045−400 Foot Board/Chartholder Combination... 0785−045−500 Foot Board (1 per package)... 0946−029−001 Foot Board (2 per package)... 0946−029−100 Hydraulic Jack Assembly, Variable Descent... 0753−002−170 I.V. Pole, Standard Removable... 0390−025−000 I.V. Pole Kit, Permanent 2−Stage, Head and Foot End... 0721−400−001 I.V. Pole, Permanent 3−Stage, Head End... 0721−061−000 I.V. Pole, Permanent 3−Stage, Foot End... 0721−062−000 Mattress, 3” x 26”, Standard... 1059−326−140 Mattress, 4” x 26”, Standard... 1059−426−140 Mattress, 3” x 26”, Enhanced Comfort... 1001−048−106 Mattress, 4” x 26”, Enhanced Comfort... 1001−047−106 Mattress, 4” x 26”, Ultra Comfort... 1210−040−004 Mattress, 5” x 26”, Ultra Comfort... 1210−040−005 Oxygen Bottle Holder, Upright... 1020−030−000 Paint, Touch−Up, Gloss Black, Bottle w/Brush... 7000−001−322 Paint, Touch−Up, Gloss Black, Spray Can... 7000−001−319 Paint, Touch−Up, Gray, Bottle w/Brush... 7000−001−320 Paint, Touch−Up, Gray, Spray Can... 7000−001−317 Pneumatic Cylinder... 1010−031−078 Restraint Strap, Ankle... 0946−043−100 Restraint Straps, Body... 0390−019−000 Restraint Strap, Wrist... 0946−044−100 Siderail Pad Set... 1010−052−000 NOTE The parts and accessories listed on this page are all currently available for purchase. Some of the parts identified on the assembly drawings pages in this manual may not be individually available for purchase. Please call Stryker Customer Service at 1−800−327−0770 for availability and pricing.
Return to Table of Contents
14
0853−006−110 Retractable 5th Wheel Base Assembly
Item A B C D
Return to Table of Contents
Part No.
Part Name
0023−288−000 0023−305−000 0753−006−148 (page 16)
Hex Washer Hd. Screw Hex Hd. Cap Screw Cam Bearing Fifth Wheel Assembly
15
Qty. 2 1 1 1
0853−006−130 Fifth Wheel Assembly
Return to Table of Contents
16
0853−006−130 Fifth Wheel Assembly Item A B C D E F G H J K L M N P R T U V W Y AA AB AC AD AE AF AG AH
Part No.
Part Name
0003−083−000 0011−360−000 0016−035−000 0023−288−000 0025−050−000 0029−007−000 0029−009−000 0753−006−074 0753−006−075 0753−006−097 0753−006−106 0753−006−108 0753−006−115 0753−006−120 0753−006−133 0753−006−142 0753−006−143 0753−006−149 0753−006−152 0753−006−153 0753−006−198 0753−006−223 0753−006−277 0853−006−126 0853−006−131 0853−006−227 0853−010−045 1210−001−147
Hex Hd. Cap Screw Washer Nylock Hex Nut Hex Washer Hd. Screw Rivet Dual Lock Dual Lock Torsion Spring Torsion Spring Drive Shaft Bearing Dampener Thrust Washer Bearing Bumper Mounting Pin Cam Pivot Block Fifth Wheel Cam Cam Bearing 5th Wheel Ramp Spring Spacer Roller Stem Drive Pin Drive Shaft Roller 5th Wheel Bracket 5th Wheel Weldment Return Spring Hook Hood Standoff Wheel
17
Qty. 1 2 1 5 2 1 1 1 1 2 1 1 2 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
0853−006−130 Fifth Wheel Assembly Item A B C D E F G H J K L M N P R T U V W Y AA AB AC AD AE AF AG AH
Return to Table of Contents
Part No.
Part Name
0003−083−000 0011−360−000 0016−035−000 0023−288−000 0025−050−000 0029−007−000 0029−009−000 0753−006−074 0753−006−075 0753−006−097 0753−006−106 0753−006−108 0753−006−115 0753−006−120 0753−006−133 0753−006−142 0753−006−143 0753−006−149 0753−006−152 0753−006−153 0753−006−198 0753−006−223 0753−006−277 0853−006−126 0853−006−131 0853−006−227 0853−010−045 1210−001−147
Hex Hd. Cap Screw Washer Nylock Hex Nut Hex Washer Hd. Screw Rivet Dual Lock Dual Lock Torsion Spring Torsion Spring Drive Shaft Bearing Dampener Thrust Washer Bearing Bumper Mounting Pin Cam Pivot Block Fifth Wheel Cam Cam Bearing 5th Wheel Ramp Spring Spacer Roller Stem Drive Pin Drive Shaft Roller 5th Wheel Bracket 5th Wheel Weldment Return Spring Hook Hood Standoff Wheel
17
Qty. 1 2 1 5 2 1 1 1 1 2 1 1 2 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Standard Brake Assembly (Fifth Wheel Base) Assembly part number 0853−003−205 (reference only)
Return to Table of Contents
18