user's manual
10016256_9
Page 2
PLANMECA OY Asentajankatu 6, 00880 Helsinki, Finland, tel. +358 20 7795 500
fax +358 20 7795 555, e-mail: sales@planmeca.com, www.planmeca.com
Page 3
TABLE OF CONTENTS
1
INTRODUCTION ..............................................................................................1
2
SYMBOLS ON PRODUCT LABELS ................................................................2
3
ASSOCIATED DOCUMENTATION .................................................................2
4
WARNINGS AND PRECAUTIONS ..................................................................3
5
MAIN PARTS ...................................................................................................5
5.1
5.2
5.3
5.4
5.5
General view of the system ............................................................................................ 5
General view of the X-ray unit ........................................................................................ 6
Exposure switch ............................................................................................................. 7
Patient supports ............................................................................................................. 8
Emergency stop button .................................................................................................. 8
6
SWITCHING THE X-RAY UNIT ON .................................................................9
7
PROGRAM PACKAGES ................................................................................10
8
CONTROL PANEL .........................................................................................11
8.1
8.2
8.3
8.4
8.5
8.6
8.7
8.8
8.9
8.10
8.11
8.12
8.13
8.14
8.15
8.16
9
PANORAMIC EXPOSURE .............................................................................61
9.1
9.2
9.3
9.4
10
General about control panel ......................................................................................... 11
Selecting panoramic exposure program ...................................................................... 12
Selecting temporomandibular joint (TMJ) exposure program ...................................... 17
Selecting sinus exposure program ............................................................................... 22
Selecting Cross sections exposure program (optional) ............................................... 25
Entering patient name .................................................................................................. 29
Selecting patient size ................................................................................................... 30
Selecting kilovolt and milliampere values .................................................................... 30
Selecting jaw size and shape ....................................................................................... 32
Adjusting unit height ..................................................................................................... 33
Returning C-arm to patient entry position .................................................................... 33
Opening/closing temple supports ................................................................................. 33
Adjusting layer light position ........................................................................................ 34
Switching DEC on/off (optional) ................................................................................... 34
Switching Autofocus on/off ........................................................................................... 35
Settings ........................................................................................................................ 35
Patient positioning ........................................................................................................ 62
Taking an exposure ..................................................................................................... 66
Taking an exposure with DEC ..................................................................................... 67
Taking an exposure with Autofocus ............................................................................. 67
TEMPOROMANDIBULAR JOINT EXPOSURE .............................................70
10.1 Double TMJ exposure (lateral, PA or lateral-PA) ......................................................... 70
10.2 Multi-angle TMJ exposure (3 angles lateral) ................................................................ 77
11
SINUS EXPOSURE ........................................................................................82
11.1 Patient positioning ........................................................................................................ 83
11.2 Taking an exposure ..................................................................................................... 85
12
CROSS-SECTIONAL EXPOSURE (OPTIONAL) ..........................................87
12.1 Patient positioning ........................................................................................................ 88
12.2 Taking an exposure ..................................................................................................... 92
13
CLEANING .....................................................................................................95
User’s Manual
Planmeca ProOne i
Page 4
TABLE OF CONTENTS
14
SERVICE ........................................................................................................95
15
DISPOSAL .....................................................................................................96
16
HELP MESSAGES .........................................................................................98
17
ERROR MESSAGES ...................................................................................101
18
TECHNICAL SPECIFICATIONS ..................................................................102
18.1 Dimensions ................................................................................................................ 103
The manufacturer, assembler, and importer are responsible for the safety, reliability and
performance of the unit only if:
- installation, calibration, modification and repairs are carried out by qualified
authorized personnel
- electrical installations are carried out according to the appropriate requirements such
as IEC 60364
- equipment is used according to the operating instructions.
Planmeca pursues a policy of continual product development. Although every effort is
made to produce up-to-date product documentation, this publication should not be
regarded as an infallible guide to current specifications. We reserve the right to make
changes without prior notice.
COPYRIGHT PLANMECA
PUBLICATION NUMBER 10016256 Revision 9
Released 27 June 2011
ii Planmeca ProOne
User’s Manual
Page 5
INTRODUCTION
1
INTRODUCTION
The Planmeca ProOne X-ray unit produces digital X-ray
images for the diagnosis of dentomaxillofacial anatomy.
The unit is allowed to be used only under supervision of a
dental/health care professional.
This manual describes how to operate the Planmeca
ProOne X-ray unit. Please read these instructions
thoroughly before using the unit.
You need a PC with the Planmeca Romexis/Dimaxis
imaging software to save, view and modify the
radiographs. The Romexis/Dimaxis software has a
separate manual which should be used in conjunction
with this manual.
0537
Planmeca ProOne X-ray unit fulfils the requirements of
Directive 93/42/EEC (Class IIb).
CAUTION FOR US USERS:
Federal law restricts this device to sale by or
on the order of a dentist.
NOTE The X-ray unit’s software version is shown briefly on
the control panel when the X-ray unit is switched on.
This manual is valid for software version 1.7.0.0.r or
later. This software version is compatible with
Romexis software version 2.4.0.R or later and Dimaxis
software version 4.5.0.0.r or later.
The display values shown in this manual are only
examples and should not be interpreted as recommended
values unless otherwise stated.
The Planmeca ProOne X-ray unit features preset
exposure values for all imaging programs. The exposure
values have been preset at the factory to automatically
correspond with the selected patient size. However, the
exposure values required to produce good X-ray images
vary considerably according to the build and age of the
patient. The preset exposure values are average values
and they are only meant to guide the user. Users are
encouraged to develop their own radiographic techniques
based on these values.
Make sure that you are fully acquainted with the
appropriate radiation protection measures and these
instructions before using the X-ray unit.
User’s Manual
Planmeca ProOne 1
Page 6
SYMBOLS ON PRODUCT LABELS
2
SYMBOLS ON PRODUCT LABELS
Type B applied part (Standard IEC 60601-1)
Attention, consult accompanying documents
(Standard IEC 60601-1)
Separate collection for electrical and electronic equipment
according to Directive 2002/96/EC (WEEE).
Alternating current (Standard IEC 60417)
Warning: dangerous voltage
(Standard IEC 60417)
Electrostatic sensitive device (Standard IEC 60417)
3
ASSOCIATED DOCUMENTATION
The Planmeca ProOne X-ray unit is supplied with the
following manuals:
- User’s Manual
- Installation Manual
- Technical Manual
These manuals are intended to be used in conjunction
with the documentation for the Planmeca Romexis/
Dimaxis imaging software. The imaging software package
contains the following manuals:
- User’s Manual
- Installation Manual
2
Planmeca ProOne
User’s Manual
Page 7
WARNINGS AND PRECAUTIONS
4
WARNINGS AND PRECAUTIONS
CAUTION This X-ray unit may be dangerous to both
patient and operator unless safe exposure
values are used and correct operating
procedures are observed.
CAUTION Modification of the X-ray unit is strictly
prohibited.
CAUTION Do not connect items which are not
specified as part of the system.
CAUTION Do not connect a multiple portable socket
outlet (MPSO) or extension cord to the
system.
CAUTION If the X-ray unit shows any signs of oil
leakage, disconnect the unit from mains and
contact your service technician for help.
CAUTION The X-ray unit is not suitable for use in the
presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
CAUTION Never use a defective or damaged X-ray unit.
Contact your service technician for help.
NOTE It is very important that the place where the X-ray unit
is to be used and the position from which the user is
to operate the unit are correctly shielded. Since
radiation safety requirements vary from country to
country and state to state it is the responsibility of the
user to ensure that all local safety requirements are
met.
NOTE If the X-ray unit has been stored at temperatures below
+10°C for more than a few hours, time must be allowed
for the unit to reach room temperature before turning
it on.
NOTE Ensure efficient air conditioning in the X-ray room. It is
recommended to keep the room temperature between
+20°C and +25°C at all times.
CAUTION
LBL-X-099
LASER RADIATION DO NOT STARE
INTO BEAM
1mW
CLASS II
635nm LASER
PRODUCT
AVOID
EXPOSURELaser radiation
is emitted from
this aperture.
LBL-X-100
CLASS 1 LASER PRODUCT
APPAREIL À LASER DE CLASSE 1
EN 60825-1/A2-2001
NOTE If the X-ray unit is not connected to an Uninterruptible
Power Supply (UPS), disconnect the unit from mains
during lightning storms.
NOTE FOR US USERS:
The patient positioning lights are class II laser
products (21 CFR § 1040.10).
NOTE FOR EUROPEAN USERS:
Class 1 laser product (Standard EN 60825-1: 1994).
The patient positioning lights are class 1, inherently
safe laser lights.
NOTE EMC requirements have to be considered, and the
equipment must be installed and put into service
according to the specific EMC information provided in
the accompanying documents.
NOTE Portable and mobile RF communications equipment
can affect the Planmeca ProOne X-ray unit.
User’s Manual
Planmeca ProOne 3
Page 8
WARNINGS AND PRECAUTIONS
NOTE External equipment intended for connection to signal
input, signal output or other connectors, shall comply
with relevant IEC standard (e.g. IEC 60950 for IT
equipment and the IEC 60601 series for medical
electrical equipment). In addition, all such
combinations - systems - shall comply with the
standard IEC 60601-1-1, Safety requirements for
medical electrical systems. Equipment not complying
to IEC 60601 shall be kept outside the patient area
(more than 2m (79 in.) from the X-ray unit).
Any person who connects external equipment to
signal input, signal output or other connectors has
formed a system and is therefore responsible for the
system to comply with the requirements of IEC 606011-1. If in doubt, contact your service technician or local
representative for help.
NOTE Contact your service technician if you notice a
decrease in image quality.
NOTE Never place or hang any objects on any part of the Xray unit.
NOTE Make sure that neither you nor your patient can get
caught or hooked up on any part of the X-ray unit.
Keep loose items of clothing, hair and jewellery tucked
away safely.
NOTE Do not attempt to remove any covers from the X-ray
unit. X-ray unit covers may be removed by service
personnel only.
NOTE Do not touch the arm structures (C-arm, upper arm)
when the X-ray unit is moving.
4
Planmeca ProOne
User’s Manual
Page 9
MAIN PARTS
5
MAIN PARTS
5.1
General view of the system
Exposure switch
ProOne_013
X-ray unit
NOTE:
PC MUST BE PLACED
OUTSIDE PATIENT AREA
Patient area
2
79 m
in.
Romexis/Dimaxis
imaging software
Ethernet
User’s Manual
Planmeca ProOne 5
Page 10
MAIN PARTS
5.2
General view of the X-ray unit
Upper arm
C-arm
Temple supports
Moving column
Patient support
Patient support table
On/off switch
(on the back of the column)
Emergency stop button
Patient handles
Control panel
ProOne_015
Stationary column
6
Planmeca ProOne
User’s Manual
Page 11
MAIN PARTS
5.3
Exposure switch
The exposure switch can be mounted on the wall, or it can
be hung from the hook located on the back of the moving
column if a protected area is within reach.
When the X-ray unit is ready to take an exposure a green
indicator light will come on on the exposure button.
Additionally, the status bar at the bottom of the control
panel display will turn green.
Exposure
indicator light
AMBER = EXPOSURE
Exp_sw_wall6.eps
Exposure button with
indicator light
GREEN = READY
When you take an exposure an amber indicator light will
come on on the exposure switch. It indicates that the Xray unit is generating radiation. Additionally, a radiation
warning symbol will be shown on the control panel display
and you will hear a radiation warning signal.
When you take an exposure you must press and hold
down the exposure button for the whole duration of the
exposure. If you remove your finger from the exposure
button before the exposure cycle is completed radiation is
interrupted, the C-arm will stop moving and a help
message will appear on the control panel. The help
message must be cleared before the X-ray unit can be
used again.
User’s Manual
Planmeca ProOne 7
Page 12
MAIN PARTS
5.4
Patient supports
TEMPLE?SUPPORTAI
Lip1.eps
Chin1.eps
Bite2.eps
Bite piece
Chin cup
Chin2.eps
Chin support
Temple supports
Chin rest
5.5
Emergency stop button
The emergency stop button is located on the underside of
the patient support table arm. Press the button to stop the
X-ray unit operating in an emergency. When the
emergency stop button is pressed down, all movements
of the X-ray unit are blocked and the unit will not generate
radiation.
ProOne_016
A help message will appear on the control panel display.
Release the stop button to resume normal operation.
8
Emergency stop button
Planmeca ProOne
User’s Manual
Page 13
SWITCHING THE X-RAY UNIT ON
6
SWITCHING THE X-RAY UNIT ON
The on/off switch is located on the back of the moving
column. When you switch the X-ray unit on the main
display will be shown on the control panel and a blue light
will come on on the left side of the display. Additionally,
the X-ray unit will carry out a self-test which will last a few
seconds.
ProOne_014
The X-ray unit is then ready for use.
NOTE To prolong the lifetime of the Planmeca ProOne X-ray
unit, always switch the X-ray unit off when it is not in
active use.
On/off switch
User’s Manual
Planmeca ProOne 9
Page 14
PROGRAM PACKAGES
7
PROGRAM PACKAGES
The Planmeca ProOne X-ray unit offers a wide selection
of exposure programs for various diagnostic purposes.
The basic program package is included as standard and
the other program packages are available as options.
The basic program package includes basic panoramic,
double TMJ and sinus programs for general maxillofacial
diagnostics. The advanced program package consists of
special panoramic, TMJ, sinus and cross sections
programs for specific diagnostic needs. The panoramic
bitewing program can be purchased as part of the
advanced program package or separately.
Each exposure program has a child setting which reduces
the image size and patient dose.
In the table the programs are divided according to the
sales packages. In the following chapters the programs
will be presented according to the program type.
Program package
Contents
Basic programs (standard)
•
Standard panoramic program
•
Double lateral TMJ program
•
Double PA TMJ program
•
PA rotational sinus program
•
Interproximal panoramic program
Advanced programs (optional)
10
•
Orthogonal panoramic program
•
Panoramic bitewing program
•
Double lateral-PA TMJ program
•
3 angles lateral TMJ program (left or right)
•
Lateral non-rotational sinus program (left or right)
•
Midsagittal non-rotational sinus program (left or
right)
•
Cross sections manual
•
Cross sections automatic
Panoramic bitewing (optional)
•
Panoramic bitewing program
Segmenting (optional)
•
Horizontal and vertical segmenting for panoramic
programs
DEC (optional)
•
Dynamic Exposure Control for panoramic
programs
Planmeca ProOne
User’s Manual
Page 15
CONTROL PANEL
8
CONTROL PANEL
8.1
General about control panel
NOTE Never allow patients to touch the display while
positioning them in the X-ray unit or carrying out the
exposure. Touching the display during exposure will
interrupt radiation, stop the C-arm movement, and
cause a help message to appear on the control panel.
The help message must be cleared before the X-ray
unit can be used again.
NOTE The contents of the displays depend on the
configuration of the X-ray unit. The displays shown
here are from an X-ray unit that features all currently
available imaging programs and functions.
To make a selection on the control panel, simply touch the
display with your finger or a soft stylus on a text field or on
an icon. For example, to select a program touch the
program field. You will hear a signal tone when a field is
activated.
NOTE Do not use sharp objects to operate the control panel.
Only the text fields and icons on the display are touchsensitive, and touching areas outside them will not
activate an action.
To accept a selection, touch the green check mark button
in the bottom right corner of the display. To cancel a
selection, touch the red cross button in the bottom left
corner of the display.
Cross = Cancel
Check mark = Accept
The display will automatically switch to stand-by mode if
you do not touch the display or the exposure button for
more than 30 minutes. A blue power light on the left side
of the display indicates that the X-ray unit is switched on
even though the display is dark. The display will switch on
as soon as you touch it again.
DAP
value
You can set the X-ray unit so that the current time and/or
date will be shown at the bottom of the display. You can
choose in which format (e.g. 22.10.2008 for date and
13:49 for time) they will be shown. Refer to section
“Setting time and date (U1100)” on page 37 for more
information.
Time & date
User’s Manual
You can set the X-ray unit so that the estimated DAP
(Dose Area Product) value will be shown on the main
display for the exposure you are going to take. The DAP
value indicates the highest radiation dose the patient will
be exposed to during the exposure. Refer to section
“Program presets (P2200)” on page 46 for more
information.
Planmeca ProOne 11
Page 16
CONTROL PANEL
You can set the X-ray unit so that a preview of the image
will be shown on the control panel after exposure. This
option allows you to evaluate the image immediately after
you have taken it. Refer to section “Functions after
exposure (P2300)” on page 48 for more information.
NOTE The control panel can be adjusted to respond to your
individual finger touch. Refer to section “Operational
settings (U1300)” on page 39 for more information.
NOTE You can switch practice mode on, if you wish to
practice or demonstrate the functions of the X-ray unit
without radiation. Refer to section “Operational
settings (U1300)” on page 39 for more information.
8.2
Selecting panoramic exposure program
Program
field
To select a panoramic exposure program touch the
program field on the main display. The main display is the
display that is shown when the X-ray unit is switched on.
The Select program display appears. Select the required
panoramic exposure program by touching the appropriate
field on the display. The selected program will be
highlighted.
NOTE The patient positioning lights will be automatically
switched on when you select the exposure program.
There are four panoramic exposure programs: Standard,
Interproximal, Orthogonal and Bitewing.
NOTE The panoramic program “Standard” is included in the
basic program package. Other panoramic programs
are optional.
NOTE When the child setting (smallest patient size) is
selected, the image size and patient dose are
automatically reduced.
12
Planmeca ProOne
User’s Manual
Page 17
CONTROL PANEL
Standard panoramic program
Direction of beam
The Standard panoramic program has a traditional path
and angles of the beam.
Image size for adult setting
1 cm
layer laser beam
Interproximal panoramic program
Direction of beam
Image size for child setting
In the Interproximal program the basic imaging geometry
is the same as in the standard panoramic program but the
angle of the X-ray beam is more parallel to the
interproximal contacts of the teeth.
This program is used for caries detection.
Image size for adult setting
1 cm
layer laser beam
Image size for child setting
Orthogonal panoramic program
NOTE This program is optimized for interproximal imaging
Direction of beam
and a shadow of the opposite side teeth may therefore
be visible in the radiograph.
In the Orthogonal program the basic imaging geometry is
the same as in the standard panoramic program but the
angle of the X-ray beam is more perpendicular to the jaw.
1 cm
User’s Manual
layer laser beam
This program is useful for periodontal studies.
Planmeca ProOne 13
Page 18
CONTROL PANEL
Image size for adult setting
Image size for child setting
NOTE This program is optimized for orthogonal imaging and
a shadow of the opposite side teeth may therefore be
visible in the radiograph.
The Bitewing program produces bitewing-like images
from premolar and molar areas including parts of maxilla,
mandible and rami. The bottom of the maxillary sinus as
well as the mandibular canal and the mental foramen are
also visible.
Bitewing
Direction of beam
The bitewing program uses improved interproximal
angulation projection geometry.
Image size for adult setting
layer laser beam
Image size for child setting
NOTE This program is optimized for interproximal imaging
and a shadow of the opposite side teeth may therefore
be visible in the radiograph.
14
Planmeca ProOne
User’s Manual
Page 19
CONTROL PANEL
Accept the selection and return to the main display by
touching the green check mark button in the bottom right
corner of the display.
Cancel
Accept
Image sizes in panoramic programs
Program
Patient size
Real image size
(WxH) mm
Image size
on receptor
(WxH) mm
Magnification
Standard panoramic
Interproximal
panoramic
Orthogonal panoramic
Panoramic bitewing
User’s Manual
Adult
226 x 103
288 x 134
1.27
Child
176 x 88
223 x 112
1.27
Adult
115 x 48
146 x 60
1.27
Child
74 x 40
93 x 50
1.27
Planmeca ProOne 15
Page 20
CONTROL PANEL
8.2.1
Selecting exposure area for panoramic exposures (optional)
NOTE Segmenting is an optional feature. Refer to section
“Program licences (P2100)” on page 44 for more
information.
This optional segmenting function allows you to take
panoramic exposures of different jaw segments. It will
reduce the radiation dose as only diagnostically
interesting areas need to be X-rayed.
Touch the exposure area field on the main display.
Exposure
area
field
The Select exposure area display appears. The display
shows an image which is divided into two horizontal and
five vertical segments.
Deselected segments no exposure
Touch the corresponding arrow button(s) to deselect the
segment(s) that should not be exposed. The segments
which will not be exposed are shown faded. The
segments which will be exposed are shown in full colour.
Touching the arrow of a deselected segment again will
change the colour of the segment back to normal.
Segments to be exposed
Cancel
16
Planmeca ProOne
NOTE The image shown on the display is only an example.
The exposed image will not equal the example image.
Accept
Accept the selections you have made, and return to the
main display by touching the green check mark button in
the bottom right corner of the display.
User’s Manual
Page 21
CONTROL PANEL
With the optional Panoramic bitewing program you can
deselect right or left side by touching the corresponding
arrow button on the control panel.
Patient’s right
Cancel
8.3
Patient’s left
Accept
Accept your selection and return to the main display by
touching the green check mark button in the bottom right
corner of the display.
Selecting temporomandibular joint (TMJ) exposure program
Program
field
To select a temporomandibular joint (TMJ) exposure
program touch the program field on the main display. The
main display is the display that is shown when the X-ray
unit is switched on.
The Select program display appears. Select the required
TMJ exposure program by touching the appropriate field
on the display. The selected program will be highlighted.
NOTE The patient positioning lights will be automatically
switched on when you select the exposure program.
User’s Manual
Planmeca ProOne 17
Page 22
CONTROL PANEL
There are four TMJ exposure programs: Double lateral,
Double PA, Double lateral-PA and 3 angles lateral.
NOTE The temporomandibular programs “Double lateral”
and “Double PA” are included in the basic program
package and enabled by default. However, the “Double
PA” program could have been disabled by the service
personnel. The “Double lateral-PA” and “3 angles
lateral” programs are optional.
NOTE When the child setting (smallest patient size) is
selected, the image size and patient dose are
automatically reduced.
Double lateral
Double lateral:
Lateral TMJ exposures of open and closed
temporomandibular joints. The imaging angle is
adjustable (factory default angle: 17°).
1 cm
2.
Double PA
1 cm
18
Planmeca ProOne
4.
3.
1.
Double PA:
Posteroanterior TMJ exposures of open and closed
temporomandibular joints. The imaging angle is 60° and it
is not adjustable.
2.
4.
3.
1.
User’s Manual
Page 23
CONTROL PANEL
Double lateral-PA
Double lateral-PA:
Lateral and posteroanterior TMJ exposures of open or
closed temporomandibular joints.
The imaging angle for the lateral exposure is adjustable
(factory default angle: 17°). The imaging angle for the
postero-anterior exposure is 60° and it is not adjustable.
1 cm
2.
4.
3.
1.
3 angles lateral:
Three lateral multi-angle left-hand or right-hand side TMJ
exposures.
3 angles lateral, left
123
The imaging angle for image no. 2 is adjustable (three
imaging angles: factory default angle 17° ±7°). The
selected imaging angle is in image no. 2
1 cm
First exposure = Selected angle - 7°
3.
2.
1.
Second exposure = Angle selected by user
(factory default angle: 17°)
Third exposure = Selected angle + 7°
User’s Manual
Accept the selection and return to the main display by
touching the green check mark button in the bottom right
corner of the display.
Planmeca ProOne 19
Page 24
CONTROL PANEL
8.3.1
Selecting imaging position for temporomandibular joint (TMJ) exposures
Touch the image position field in the middle of the display
to select the imaging position.
The X-ray image in the top right corner of the field shows
in which order the images will appear on the radiograph.
The active exposure is shown in full colour. The number in
the top left corner indicates the number of the active
exposure.
Image
position
field
Target
position
Imaging
angle
arrows
(lateral
exposures)
Symmetric/
asymmetric
setting
On the display that appears you can adjust the imaging
position by touching the target position arrows in the
bottom right corner. The arrow pointing to the left will
move the target position forwards and the arrow pointing
to the right will move the target position backwards. The
target position on the other jaw side will move accordingly
if the icon for symmetric/asymmetric setting shows an
undivided jaw (one line in the middle). The number that is
shown on the display demonstrates the position of the
layer light and serves as a reference for later retakes.
NOTE The imaging position can also be adjusted by touching
the target position arrows on the main TMJ display.
Target position arrows
Symmetric/
asymmetric
setting
20
Planmeca ProOne
The imaging angle can be adjusted for lateral TMJ
exposures. The default imaging angle for lateral
exposures can be defined by user in the Program presets
menu, option Lateral TMJ default angle, see Section
“Program presets (P2200)” starting from page 46. To
adjust the imaging angle for the exposure you are going to
take touch the imaging angle arrows next to the degree
marking. Touching the upwards pointing arrow will
decrease the angle and touching the downwards pointing
arrow will increase the angle. The imaging angle on the
other jaw side will change accordingly if the icon for
symmetric/asymmetric setting shows an undivided jaw
(one line in the middle).
If the patient’s right and left side are asymmetric you will
need to set the target position and imaging angle (lateral
views) separately on the other side. To do this, touch the
icon for symmetric/asymmetric setting to select the
asymmetric setting and adjust the other side as required.
In the asymmetric setting the icon shows a divided jaw
(two lines in the middle).
User’s Manual
Page 25
CONTROL PANEL
NOTE In the asymmetric setting, touching the target position
arrows on the main TMJ display will adjust the imaging
position only on the patient’s left side.
The right/left buttons at the top of the display allow you to
select on which side the exposures will be taken. You can
set the unit to take TMJ exposures on one side or on both
sides. The button(s) of the selected side(s) will be
highlighted. Note that the buttons refer to the patient’s
(not your) right and left side.
Patient’s
right/left
side
NOTE In multi-angle exposures the selected imaging angle is
in image no. 2.
3 angles lateral, left
123
3.
2.
1.
Imaging angle selected by the user
1 cm
First exposure = Selected angle - 7°
Second exposure = Angle selected by the user
(factory default angle: 17°)
Third exposure = Selected angle + 7°
Accept the selection and return to the main display by
touching the green check mark button in the bottom right
corner of the display.