Respiratory Diagnostic Tips #1 Invacare Concentrator Alarm Modes For more technical information, see “The TECHNICAL Zone” @www.invacare.com CAUTION: Please read and understand all instructions or details carefully.
ALARM MODES 3LX, 5LX, & 5LXO2 (BEFORE S/N 97J0000)
3LX, 5LX, 5LXO2/O2T, & 6LXO2/O2T/H (AFTER S/N 97J0000)
INTERNAL LED ON P.C. BOARD
INTERNAL LED FLASHES ON P.C. BOARD
ALARM High Pressure
RED On
GREEN Off
Low Pressure
On
On
Short Beep, Long Pause
0
0
Main Power Loss
Time-Out (30 Seconds)
On
On
Continuous
1
1
Low Pressure Major Leak
Low O2
On
Off
Continuous
1
2
High Pressure No Switching
Continuous
1
3
Time-Out Failure Compressor
Continuous
1
4
Pilot Valve 1 Coil
Continuous
1
5
Pilot Valve 2 Coil
Continuous
2
1
73% Shutdown Low O2
Continuous
2
3
P.E. Valve Coil
Continuous
2
4
EPROM Failure on P.C. Board
Continuous
2
5
RAM Failure on P.C. Board
Continuous
3
1
Oxygen Sensor Failure on P.C. Board
Continuous
3
2
Mode “F” Auto Tune Failure
Continuous
3
3
Mode “O” Auto Tune Failure
All alarms have compressor shutdown, horn ON and IRC5LXO2 ONLY - RED Panel LED ON.
Sens O2 Alarm Thresholds O2 Purity Over 85% 73% to 85% Below 73%
LED Indicators
ALARM 3 Beeps at Start-up
RED GREEN 0 0 System OK
Green Yellow Red - Indicator light Continuous Alarm Sieve-GARD TM Compressor Shutdown
All alarms have compressor shutdown, horn ON and 5LXO2/5LXO2T, 6LXO2/6LXO2T/H ONLY - RED Panel LED ON.