PACHYMETER ECHOGRAPH
ON AC MEAS.
OK
TEST
SERVICE MANUAL XE PCK2 MT AN
NOVEMBER 2003
QUANTEL MEDICAL SA 21, rue Newton, Z.I. du BREZET 63039 Clermont-Ferrand Cedex 2 - FRANCE Tel : +33 (0) 473 745 745 - Fax : +33 (0) 473 745 700 e-mail : [email protected] web site : www.quantel-medical.fr
SERVICE MANUAL
WARNINGS
WARRANTY : - For safety reasons, QUANTEL MEDICAL considers the instrument CPU Board as whole part. Defective CPU Board will be returned to QUANTEL MEDICAL as a whole for replacement. Any faulty component will be replaced by a strictly identical component. - QUANTEL MEDICAL shall not be liable for any intervention or attempt to repair the "CPU Board" should any incident or breakdown occur. The guarantee of the equipment will be void if the equipment is opened (even partially), modified or repaired in any way by persons who are not authorised by QUANTEL MEDICAL. In case of user or equipment problems, call QUANTEL MEDICAL Service department : Quantel Medical SA Head Office : 21, rue Newton, Z.I. du BREZET 63039 Clermont-Ferrand Cedex 2 - FRANCE Tel : +33 (0) 473 745 745 Fax : +33 (0) 473 745 700 E-mail : [email protected] Quantel Medical Inc. 601, Haggerty Lane BOZEMAN MT 59715 USA Tel : +1 888 660 6726 Fax : +1 406 586 2924 E-mail : [email protected]
Rev 11-2003
P -2
SERVICE MANUAL
T ABLE OF CONTENTS Page Rev 11-2003 WARNINGS
P-2
TABLE OF CONTENTS
P-3
1- TECHNICAL SPECIFICATIONS 1-1 Frequency 1-2 ANTI Interference 1-3 Power Supply 1-4 Stickers
P-4 P-4 P-4 P-4 P-5
2- OVERALL BLOCK DIAGRAM
P-8
3- DIAGRAM OF THE CPU BOARD 3-1 Schematic 3-2 Layout
P-9 P-9 P - 10
4- INSTRUMENT BOX 4-1 Material 4-2 Dimensions and Weight
P - 11 P - 11 P - 11
5- ACCESSORIES AND SPARE PARTS
P - 12
X
6- MAINTENANCE P - 13 6-1 How to open the POCKET-II box P - 13 6-2 Pachymetry probe calibration P - 14 6-3 Checking procedure P - 16
X X
VALIDATION SHEET
X
Rev 11-2003
X
X
X
P - 18
P -3
SERVICE MANUAL
1- TECHNIC AL SPECIFIC ATIONS TECHNICAL 1-1 FREQUENCY - Probe transducteur : 20 MHz - Programmable Logic device (PLD)Clock:32MHz - Microprocessor Clock : 8 MHz 1-2 ANTI INTERFERENCE The POCKET-II equipment consumes very low current and cannot generate any significant EMI effect. The POCKET-II equipment is powered by direct current (DC) voltage. 1-3 POWER SUPPLY The POCKET-II equipment requires four 1,5V AA batteries in order to operate. An optional external power supply may be used to replace the batteries. This external power supply can be found in any country with the appropriate regulation. The voltage and current required are 9V DC and 500 mA. Do not remove the batteries before connecting to the external power supply (in order to safeguard the configuration data) . external power supply placement
9V
The external power supply connector is positionned on the upper left side of the POCKET-II instrument. Rev 11-2003
P -4
SERVICE MANUAL
1- TECHNIC AL SPECIFIC ATIONS TECHNICAL 1-4 STICKERS Upper panel 9V
1 Rear panel
2
PACHYMETER Echograph Model : POCKET-II
Class II 0459
Type BF
QUANTEL MEDICAL 21, rue Newton, Z.I. du BREZET 63039 CLERMONT-FERRAND cedex 2 - FRANCE Tel: +33 (0) 473 745 745 Fax: +33 (0) 473 745 700
3
4
- Power supply : 6 Vdc = 4 Alkaline LR6 batteries - Option : External power supply Complying with IEC 065 standard Maximum : 9 Vdc ; 500 mA (Does not recharge batteries)
SN 002
Lower panel
5 Rev 11-2003
P -5
SERVICE MANUAL
1- TECHNIC AL SPECIFIC ATIONS TECHNICAL
1 Plate with serigraphy : 9V
Size : 23 x 93 mm - external power supply placement
9V
- serial port to PC placement - parallel port for printer placement
2 Identification sticker : PACHYMETER Echograph Model : POCKET-II
Class II 0459
Type BF
QUANTEL MEDICAL 21, rue Newton, Z.I. du BREZET 63039 CLERMONT-FERRAND cedex 2 - FRANCE Tel: +33 (0) 473 745 745 Fax: +33 (0) 473 745 700 - Power supply : 6 Vdc = 4 Alkaline LR6 batteries - Option : External power supply Complying with IEC 065 standard Maximum : 9 Vdc ; 500 mA (Does not recharge batteries)
Size 60 x 50 mm
Rev 11-2003
P -6
SERVICE MANUAL
1- TECHNIC AL SPECIFIC ATIONS TECHNICAL
3 Special sticker for USA : CAUTION : FEDERAL USA LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN OR AN OPTOMETRIST LICENCED BY THE LAW OF THE STATE IN WHICH HE PRACTISES TO USE OR ORDER THE USE OF THIS DEVICE
EXPLOSION HAZARD. DO NOT USE IN PRESENCE OF DANGER: FLAMMABLE ANESTETICS OR OXIGEN. RICH ATMOSPHERE ELECTRIC SHOCK HAZARD. DO NOT OPEN. REFER CAUTION: SERVICING TO QUALIFIED SERVICE PERSONNEL.
Size 59 x 15 mm
4 Serial number sticker : Size :33 x 12 mm
SN 002
5 Type BF symbol : Size : 13 x 13 mm
Rev 11-2003
P -7
SERVICE MANUAL
2- OVERALL BLOCK DIAGRAM
Sampler Level Control Counter
Receiver
Probe
32Mhz clock
Control of 8 Time delays
Measurment Control Counter
Display Control
Numerical display
Parallel Port Printer Control
Printer
Serial Port
PC
Pulser
Programmable Logic Device (PLD) : This hardware circuit realizes fast sub-functions on request of the micro-processor.
8 Mhz clock
Repplaceable batteries (4x1.5V)
Power Supply
5 Volts
Microprocessor
User Buttons
Footswitch
Plug for 9 Vdc External power suppply
Rev 11-2003
P -8
SERVICE MANUAL
3- DIAGRAM 3-1 SCHEMATIC
REFER TO THE TREATMENT CARD REF : 3 101 0109 Rev B
Rev 11-2003
P -9
SERVICE MANUAL
3- DIAGRAM 3-2 LAYOUT
REFER TO THE TREATMENT CARD REF : 4 152 0109 Rev A.
Rev 11-2003
P -10
SERVICE MANUAL
4- INSTRUMENT BOX 4-1 MATERIAL The box is in ABS plastic in accordance with the UL94V0 Standard.
4-2 DIMENSIONS AND WEIGHT Width : 100 mm Depth : 200 mm Front panel
Height : 45 mm Weight : 460 g Rear Panel PACHYMETER Echograph Model : POCKET-II
Class II 0459
Type BF
QUANTEL MEDICAL 21,rueNewton,Z.I.duBREZET 63039CLERMONT-FERRANDcedex2 - FRANCE Tel: +33 (0) 473 745 745 Fax: +33 (0) 473 745 700 - Power supply : 6 Vdc = 4 Alkaline LR6 batteries - Option : External power supply Complying with IEC 065 standard Maximum : 9 Vdc ; 500 mA (Does not recharge batteries)
ON AC MEAS.
OK
TEST
SN 002
Rev 11-2003
P -11
SERVICE MANUAL
5- ACCESSORIES & SPARE PARTS
IDENTIFICATIO N
CODE
ACCESSORIES Pachym etry probe reference P1
XE AAA PRB PK
User M anual
XE PCK2 M E
Technical Manual
XE PCK2 M T AN OPTION
Footswitch
XE AAA PED SPARE PAR TS
Pocket-II plastic box
XE POC BOIT2
Pocket-II front panel + test block
XE POC LEXA N2
Pocket-II CPU board
XE POC CPU2
Pocket-II LCD screen and Board
XE POC LCD2
Pocket-II transpor t case
XE POC VALISE2
Rev 11-2003
P -12
SERVICE MANUAL
6- MAINTENANCE 6-1 HOW TO OPEN THE POCKET BOX The POCKET-II plastic box is composed of 2 parts. 1) Unscrew the 2 large headed metal screws on the underside of the POCKET-II and remove the back panel of the battery compartment 2 screws
2) Unscrew the 4 screws and remove the inferior part of the POCKET-II plastic box. 4 screws
Rev 11-2003
P -13
SERVICE MANUAL
6- MAINTENANCE 6-2 PACHYMETRY PROBE CALIBRATION
ON AC MEAS.
OK
NOTE : The parameters in the screens below are given for examples. You can find parameters slightly differents. Firstly, ensure that the probe is connected and the tip is clean and dry. - From this screen: Pres. MEAS M1 OK
- Press
3 seconds.
- The configuration screen appears: Freeze MEDIUM
- Press simultaneously these two keys 3s:
ON AC
- Speed selection appears on the screen: OK
Speed 1620 m/s
- Press - This message appears on the screen: CalUnp 1878 2554
- Press simultaneously these two keys : - This message appears on the screen : Unplug the probe
- Unplug the probe. Rev 11-2003
P -14
SERVICE MANUAL
6- MAINTENANCE OK
- Press - CalUnp with the new parameters is displayed : CalUnp 1911 2557 OK
- Press - This message appears on the screen: CalWat 2021 1876
- Press simultaneously these keys : -This message appears on the screen: Probe in water
- Plug the probe on the pocket-II and put the probe in a glass of water like indicate below
BE CARREFULL that the end of the probe is not in contact with the bottom of the container OK
- Press - CalWat with the new parameters is displayed: CalWat
- The calibration is done.
2052 1842
OK
- In order to save this calibration press
3 s.
Rev 11-2003
P -15
SERVICE MANUAL
6- MAINTENANCE 6-3 CHECKING PROCEDURE The measurement acquisition may be checked with the test-block : TEST Test-block 1- Dampen the probe extremity with water and place it perpendicular on the Pocket-II test block like indicated below.
Do not put water on the test block 2- Press “MEAS”. The acquisition should be fast : if necessary, change slightly the position of the probe to obtain an acquisition. You ear a repetitive sound while the position of the probe is incorrect. You must be perpendicular like the opposite diagram.
90°
Test-block on the Pocket-II
Note : Make a series of measurments to have a representative Standard Deviation (SD). Rev 11-2003
P -16
SERVICE MANUAL
3- Results of the test : if the procedure has been succesfull and the position of the probe is correct, the readings of the series of measurements must be repetitive to ± 20 µm. As some microns represent a very thin thickness, some parameters influencing the test-block itself : - the variation of the thickness with the ambiant temperature. - possible dust. - Incorrect positioning. If you suspect any problem, please call for Service.
Rev 11-2003
P -17
SERVICE MANUAL
VALIDATION SHEET
Software Version
1.04
Publishing Date
November 2003
Revision Date
November 2003
Written by F. LOYAL Checked by C. CHABRIER Validated by P. QUERO