OPHTHALMIC ECHOGRAPH
BIOMETRY AND PACHYMETRY
2003
SERVICE MANUAL
XE AX2 MT AN
Directive 93/42/EEC
OCTOBER 2009
QUANTEL MEDICAL Head Office: 21, rue Newton, Z.I. du Brezet, 63039 Clermont-Ferrand cedex 2 - FRANCE Tel : +33 (0) 473 745 745 - Fax : +33 (0) 473 745 700 E-mail: [email protected] Web site: www.quantel-medical.com
SERVICE MANUAL
WARNINGS
WARRANTY : * Any adjusment and calibration procedure described in this manual shall be performed by QUANTEL MEDICAL authorized technicians only or by personnel who has been previously trained by QUANTEL MEDICAL. * Any faulty part will be replaced by a strictly identical one provided by QUANTEL MEDICAL. * For safety reasons, QUANTEL MEDICAL considers any electronic board as a whole partial assembly : * Defective boards must be returned to QUANTEL MEDICAL for replacement. * Warranty will be void if a board has been damaged during maintenance or attempt to repair. Warranty will be void unless these recommendations are strictly respected. WARNING: Special training delivered by QUANTEL MEDICAL is required for technicians prior to performing maintenance on the equipment in order to ensure optimum adjustment and maximum reliability of the whole instrument.
IMPORTANT: Before performing any maintenance or repair on the equipment, ensure that all the system/user configuration and parameters are saved in order to be restored after if needed.
Rev 10-2009
P-3
SERVICE MANUAL
TABLE OF CONTENTS Page
Rev 10-2009
WARNINGS ... 3 TABLE OF CONTENTS ... 4
X
1- TECHNICAL SPECIFICATIONS ... 5 1-1 Electrical requirements ... 5 1-2 Environmental conditions ... 5 1-3 Housing material ... 5 1-4 Spare parts ... 6
X
2- INSIDE ACCESS ... 7 2-1 External view ... 7 2-2 How to open the case ... 8 2-3 Inside view ... 9 2-4 How to close the case ... 9
3- BOARD DESCRIPTION ... 10 3-1 Block diagram ... 10 3-2 CPU board ... 11 3-3 Replacing a CPU board ... 12 3-4 Replacing a video interface board ... 13 3-5 Replacing an LCD screen ... 14
4- SERVICE MODE ... 15 4-1 Hidden page access ... 15 4-2 Field description ... 15 4-3 Activating the Post-refractive feature ... 19 4-4 General reset ... 19
5- PROBE CALIBRATION ... 20 5-1 Biometry probe ... 20 5-2 Pachymetry probe ... 20
6- SOFTWARE UPGRADE ... 21 6-1 Software version compatibility ... 21 6-2 Upgrading to V5.0 ... 21 6-3 Upgrading to V4.04 ... 21 6-4 Upgrading to V3.xx, 2.xx or 1.xx ... 21 6-5 How to install a new software version ... 22
7- TROUBLESHOOTING ... 28 APPENDIX A- EVOLUTION OF THE CONFIGURATION ... 29 VALIDATION ... 37
P-4
X
Rev 10-2009
SERVICE MANUAL
1 - TECHNICAL SPECIFICATIONS 1-1 ELECTRICAL REQUIREMENTS - Power supply:
External module with automatic voltage adaptation, no selection is needed. Input Voltage:
- References: or or - Compliance:
Frequency: Mains consumption: AXIS-II consumption:
110 to 230 Vac ± 10% single phase without earth 50 or 60 Hz 20 VA 5.4 W (450mA on 12Vdc)
Class II Power Supply:
Reinforced Isolation.
SSL20-7612
Trade Mark:
ESS36US12-1E TR30M120-01E03
Trade Mark: Trade Mark:
ARTESYN Technologies XP POWER CINCON
Power supply and enclosure designed per IEC 60601-1 European directive: 93/42/CEE Electromagnetic compatibility: IEC 60601-1-2
1-2 ENVIRONMENTAL CONDITIONS The temperature of the room where the device is operated must be within the following range: 50°F < T° < 95°F (10 °C < T° < 35 °C) The relative humidity must not exceed 95% without condensation. The Device storage and transport temperature must be within the range: - 4°F < T° < 158°F
(-20 °C < T° < 70 °C)
1-3 HOUSING MATERIAL Upper frame: Lower frame: LCD front panel:
Rev 10-2009
Polystyrene PS CHOC complies with UL 94 VO standard Painted steel sheet Polycarbonate
P-5
SERVICE MANUAL
1 - TECHNICAL SPECIFICATIONS 1-4 SPARE PARTS Important: when ordering a spare part, the device serial number and the software version must be provided. PARTS LIST REFERENCE
DESCRIPTION
XE AX2 CABLE RS232
RS232 Cable
XE AX2 CPU
Main Board for V1.xx
XE AX2 CPU V2
Main Board for V2.xx
XE AX2 CPU V3
Main Board for V3.xx
XE AX2 CPU V5
Main Board for software V4.xx and V5.xx
XE AX2 LCD
B&W LCD screen
XE AX2 LCD COLOR
Color LCD screen
XE AX2 PCB VIDEO
Video Interface Board for V1.xx & V2.xx
XE AX2 PCB VIDEO V3
Video Interface Board for V3.xx
XE AX2 PCB VIDEO V5
Video Interface Board for V4.xx &V5.xx
XE AX2 PLEXI
Plexiglas Screen
XE AX2 FAV CLV
Plastic housing with Keyboard
XL VIT BOUTON
Knob
ACCESSORIES/OPTIONS LIST REFERENCE
DESCRIPTION
XE AX2 EXT POW
Power Supply : SSL20-7612 / Artesyn
XE AX2 PRB BIO
TP-01-b probe
XE AAA PAM
Hand-piece for biometry probe
XE PRB BIO L
Biometry probe with laser pointer
XE AAA PRB PK
Pachymetry probe
XE AX2 OPT PK
Pachymetry option
XE AAA PED XE AX2 TEST BLOC
Pedal Test block (Biometry)
XE AX2 PSD PK
Pachymetry probe holder + test block
XE CLV USB AZ
External USB Keyboard (AZERTY)
XE CLV USB QW
External USB Keyboard (QWERTY)
XE HUB USB
USB Hub
XE CLE USB
USB storage Key
XE AX2 HOUSSE
Dust Cover
XE AX2 VALISE
Axis-II Bag
P-6
Rev 10-2009
SERVICE MANUAL
2 - INSIDE ACCESS 2-1 EXTERNAL VIEW
Pachymetry probe holder and test-block Color LCD screen Biometry probe test-block
Handle
Biometry probe on its holder
Screen Function Keys Pedal key Function keys
Knob : Digital Potentiometer
Rev 10-2009
Built-in Alpha-numeric Keyboard
P-7
SERVICE MANUAL
2 - INSIDE ACCESS 2-2 HOW TO OPEN THE CASE - From the bottom panel, unscrew the 4 screws as shown. Axis-II Bottom view
4 screws to be removed
- Put the device on the normal position and lift up the upper part of the case verticaly as shown. Hood
P-8
Rev 10-2009
SERVICE MANUAL
2 - INSIDE ACCESS 2-3 INSIDE VIEW Video interface board
Keyboard ribbon cable
CPU board
Backlight cable
LCD screen
Ribbon cable between the LCD screen and the video interface board
Ribbon cable between the video interface board and the CPU board
2-4 HOW TO CLOSE THE CASE - Put the cover back from the top side (1) first then the other side (2). Put the device up-side-down and tighten the 4 screws (3). 1
2
3
4 screws to be tightened
Rev 10-2009
P-9
SERVICE MANUAL
3 - BOARD DESCRIPTION 3-1 BLOCK-DIAGRAM
Footswitch RS 232 Ports USB Ports #2 #1 #2 #1
Liquid Crystal Display
Interface Board
Keyboard
Flash memory Hard Disk Microcontroller (68328)
RAM Patient Data
Clock : Date & Time
A. D. Converter
Knob
Programmable Logic Device (PLD)
Pulser
Pulser
Receiver Receiver
Main Board
Mains : 100 to 230 V ~
External Power Supply Class II / 12 V TP-01-b Probe
P - 10
P1 Probe
Rev 10-2009
SERVICE MANUAL
3 - BOARD DESCRIPTION 3-2 CPU BOARD Probe connector : Lemo / 4 pins
BNC connector for Pachymetry probe
USB #2 RS232 ports #2 #1
USB #1 Power switch
Pedal connector
12 VDC
RAM
Buzzer
Microprocessor
ACEX (EPLD) XEAX2CPUV5 Analog part (Biometry filter) Analog part (Pachymetry filter)
Ribbon Cable
1
Link to the Keyboard Interface via the ribbon cable with 20 wires. Important : pin # 1 is red
Rev 10-2009
P - 11
SERVICE MANUAL
3 - BOARD DESCRIPTION 3-3 REPLACING A CPU BOARD 2 1
Remove the 4 screws
Unplug the power and ground connectors
XEAX2CPUV5 4
Pull the board in this direction
5
To replace the board, perform the steps above in reverse order
3
Disconnect the ribbon cable
Warning: the faulty board must be returned to QUANTEL MEDICAL as a whole without unmounting or opening any component. Failure to comply will void the equipment warranty.
P - 12
Rev 10-2009
SERVICE MANUAL
3 - BOARD DESCRIPTION 3-4 REPLACING A VIDEO INTERFACE BOARD
1
Remove the knob by pulling it vertically then unscrew the nut underneath
3
Unplug the connectors
2
Remove the screw
XEAX2PCBVIDEOV5 4
Remove the board
5
To replace the board, perform the steps above in reverse order
Warning: the faulty board must be returned to QUANTEL MEDICAL as a whole without unmounting or opening any component. Failure to comply will void the equipment warranty.
Rev 10-2009
P - 13
SERVICE MANUAL
3 - BOARD DESCRIPTION 3-5 REPLACING AN LCD SCREEN Board notches
3
Unplug the ribbon cable connector
4
Pull the board in this direction
Note:Ribbon blue mark is up
to remove it from the notches
1
Remove the 2 screws
XEAX2LCDCOLOR Note:Ribbon blue mark is up 2
Unplug the cable connectors 5
To replace the board, perform the steps above in reverse order
Note: Before disconnecting the ribbon cable, note the polarizing position (blue side up) for future reconnection. On the V5 board, the ribbon cable can be disconnected from both end while on earlier version boards, the cable can be removed only on one end. Warning: the faulty board must be returned to QUANTEL MEDICAL as a whole without unmounting or opening any component. Failure to comply will void the equipment warranty.
P - 14
Rev 10-2009
SERVICE MANUAL
4 - SERVICE MODE 4-1 HIDDEN PAGE ACCESS
AXIS-II PR - P Sept/03/2007
Serial Number :####-#### ID:####
GENERAL SETUP
version:5.00
Serial Output (1) : Not Used Position of marker # 1 in Contact : +0.00 mm
User #1 User1 #2 User2 #3 User3 #4 User4 #5 User5
03/09/07
Information : Quantel Medical AXIS II V5.xx
18:53:00
Zero:16 Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 0 Pachy Option : NO KeyCode: 0
Language : English Day : Nov/19/2007 Time : 09:53:20 Appearence : Model 3
PR Option : YES
Volume adjustment (-/+): 30% Contrast Adjustment(</>): 40% Gral SETUP USER Setup
1
Pachymetry
Delete
Test
Press F1
2
Move cursor on “Language” filed and press
Back Space CAPS/Small
3 The hidden screen
is then displayed
4-2 FIELD DESCRIPTION
Information field. Use F2 (delete character), F3 (backspace) and F4 (Caps/ small) to enter characters.
Serial number, required for pachymetry activation (refer to the activation guide XE AX2PK NI). Serial Number :####-#### ID:####
03/09/07
Information : Quantel Medical AXIS II V5.xx
To be adjusted according to the manufacturer of the probe and to be used to activate the PR option (see section 4-3).
Zero:16 Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 0 Pachy Option : NO KeyCode: 0
Delete
PR Option : YES
Back Space CAPS/Small
To be adjusted for a right noise level in biometry mode (see section 4-2-4 “Noise level setting”) and to be used for general reset (see section 4-4 “General reset”).
Keycode to be entered to activate the pachymetry option (refer to the activation guide).
Rev 10-2009
P - 15
SERVICE MANUAL
4 - SERVICE MODE 4-2-1 Field “ZERO” and Position of Marker #1 The default value of this field is “16” and is given by the manufacturer of the biometry probe. This value can be different with another manufacturer and it depends on the size of the probe tip. Adjusting “Zero” will impact the mesured values during acquisition as follows: - decreasing ZERO by 1 will increase the mesured value by 0.04mm (with the “Position of marker #1 in contact” of the general setup page as 0.00mm). The test block has been developed as a standard in contact technique to calibrate the measurement of the eye and adjust the corresponding parameters. Follow the steps below to enter the correct value in “Zero”. 1. Make a biometry acquisition on the test-block: - from the Main page, press F1 to go to the General Setup page, - move the cursor to “Position of marker #1 in contact” - press F1 and follow the instructions on the screen as below: Put the probe on the Test-Block With a drop of water. Press "F1" to Start and then to Stop the measurements. Note : the result is corrected by the position of the first marker : +0.00 mm
Best = 10.04 mm L = 10.08 mm
2. The mesured value X (displayed as “Best”) must be 10.00 < X < 10.11mm 3. If X is not within this interval, modify ZERO so that X will fit into these limits (knowing that 1 unit of ZERO corresponds to 0.04mm of X). Ex: X=9.96mm, ZERO can be adjusted to: 12 < ZERO =< 15 Note: Instead of modifying ZERO, this correction can also be done by entering the value in the “Position of marker #1 in contact” field of the general setup page.
P - 16
Rev 10-2009
SERVICE MANUAL
4 - SERVICE MODE 4-2-2 Field “RETINA SLOPE TEST”
Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00 Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : NO KeyCode: 000000000000
Delete
Back Space
The default value of this field is “6”. It defines the sharpness of the retinal peaks in order to freeze the A-scan when the probe direction is in the visual axis, reaching the macula area. This value means that the slope is raised up by 6 digital points. It appears that the retina slope test of “4” is perfect for physicians using the Prager shell in immersion technique. This shell holds the probe in its axis and helps to keep it perpendicular to the macula.
If the physician is not using the immpersion technique, but the contact one, and has difficulties to find the visual axis, the Retina Slope Test should be set to “6”. Increasing this value will make freezing easier.
4-2-3 Field “D2/LOW DYN” The recommended value for this field is “20”. The entry range is 15 to 25dB. It corresponds to the dynamic scale to avoid distorsion during acquisition. The dynamic value is inversely proportional to the Gain value (as specified in the user setup). The display is the same as above except that the highlighted zone is on “D2/Low Dyn”.
Rev 10-2009
P - 17
SERVICE MANUAL
4 - SERVICE MODE 4-2-4 Field “NOISE”: Setting procedure Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00 Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : NO KeyCode: 000000000000
Gral RESET
Delete
Back Space
In USER Setup, select Display mode D1. Noise level: In the Acquisition page, set: G=90db TGC = -10dB Technique: Contact Unfreeze the image. The noise level must be < 1mm.
If the noise level is not correct, set the gain to obtain the right noise level and write down this value: New Gain = _________ dB Calculate the new noise value: New Noise Value = 90dB - New Gain = _________ Set this value in the Noise field (of the hidden page). Noise value = _______
4-2-5 Field “KEYCODE” This field is used to activate the pachymetry option by entering the right keycode (refer to the activation guide). Enter “0” in the field to suppress the pachymetry option. The below screens show the option that is not activated (Pachy Option: NO) and activated (Pachy Option: Yes). Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00
P - 18
Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00
Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : NO KeyCode: 0
Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : YES KeyCode: 1444606271
Gral RESET
Gral RESET
Delete
Back Space
Delete
Back Space
Rev 10-2009
SERVICE MANUAL
4 - SERVICE MODE 4-3 ACTIVATING THE POST-REFRACTIVE FEATURE
Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00 Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : NO KeyCode: 000000000000
The post-refractive feature can be activated (when “PR Option: Yes” is present) or desactivated (no “PR option” is displayed). To activate it: from the hidden screen, position the cursor on “ZERO”, press “F1”, “7”, “8” and “F4” successively. “PR Option: Yes” will then be displayed. Pressing this key sequence again will desactivate the PR feature.
4-4 GENERAL RESET
Serial Number : xxxx-xxxx ID:xxxx
03/09/07
Information : Quantel Medical AXIS II Vx.00 Zero:16 PR Option : YES Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 2 Pachy Option : NO KeyCode: 000000000000
Gral RESET
Delete
Back Space
To perform a general reset, position the cursor on “Noise” in the hidden page then press “F1: General Reset” and confirm by “Yes”. Select “F1: Setup1” (if D1 display mode is used by default for all users) or “F2: Setup2” (if D2 is used). Wait for a few seconds then the main page will be displayed. The D1 or D2 mode can be set manually in the “User setup” page.
Note: a general reset will erase all the setup and system parameters including PR and pachymetry options. Make sure you have written down all parameters of the User and General setup, and Hidden page before performing a general reset.
Rev 10-2009
P - 19
SERVICE MANUAL
5 - PROBE CALIBRATION 5-1 BIOMETRY PROBE Calibrating the biometry probe consists in performing an acquisition on the biometry test block (on the front panel) then enter the correct value of “ZERO” and “Position of Marker #1 in contact” in the hidden page. Follow carefully. the procedure as detailed in chapter 4-2-1 Note: It is recommended to check the position of Marker #1 in contact when changing probe. If the measurement is not correct, you have to perform a calibration of the probe (see chapter 4-2-1). If the probe is not correctly calibrated, the measurements of the patient eye will not be accurate and the values are incorrect.
5-2 PACHYMETRY PROBE Serial Number :####-#### ID:####
03/09/07
Information : Quantel Medical AXIS II V5.xx Zero :16 Retina Slope Test : 6 D2 / Low Dyn (dB) : 20 Noise : 0 Pachy Option : YES KeyCode:1234567890
PR Option : YES
Clean and dip the probe tip into a glass of water (30 º) (Check there are no micro bubbles on the tip). Before starting the calibration, leave the probe in the water for at least one minute, to make sure its temperature stabilizes, then press F1 to start the calibration. Calib Value: xxx y When the Calib Value is displayed press the <ESC> key to exit the calibration screen.
Calibration
Start
The pachymetry probe must be calibrated before use: 1. in the hidden page, position the cursor on “Pachy Option” field, 2. the screen will display “calibration” in front of “F4”, 3. press “F4” and follow the instructions carefully, 4. repeat the process 4 or 5 times and check that the displayed values xxx are stable (±10). Note: It is recommended to perform a new calibration when changing probe, in order to improve the accuracy of acquisition.
P - 20
Rev 10-2009
SERVICE MANUAL
6 - SOFTWARE UPGRADE 6-1 S/W VERSION COMPATIBLITY (to) UPGRADE (from)
V1.11 V2.04 V3.05 V4.04
V5.0
V1.xx
Y
-
-
-
-
V2.xx
-
Y
-
-
-
V3.xx
-
-
Y
-
-
V4.xx
-
-
-
Y
-
V5.0
-
-
-
-
Y
Note: the software version number is displayed on the main page at the top right corner. The next paragraphs detail the recommendations when upgrading to a specific version.
6-2 UPGRADING TO V5.0 Version 5.0 can be installed only on specific device hardware that has been designed for, i.e. with a pachymetry connector on the rear panel and a handle. Refer to section 6-5 for further details.
6-3 UPGRADING TO V4.04 Version 4.04 supports the external USB keyboard or a storage USB key or a USB hub to be connected to the USB port. Ensure the device has USB ports on the rear panel and no pachymetry connector. Refer to the installation guide XE AX2 NI AN-FR for further details.
6-4 UPGRADING TO V3.XX, 2.XX OR 1.XX Refer to the above table for version compatibility before upgrading to a specific version. Then carefully follow section 6-5 that details the upgrading procedure.