Result Sheet Test instructions / Service Card IPM-L Workstation Critical Care Part number: Location:
Serial number:
Department:
Cust. invent. no.:
Maintenance interval:
Other / Delivery date: Key
Applies to Test instructions / Service Card IPM-L Revision 21.0
Test
= OK = Spare part used
!
= Error / Report
/
= Accessory not available
-
= Not applicable
Test
Result
Result
1 Device configuration
3.1.5
1.1
3.1.5.1
Reference value
3.1.5.2
Recurrent test
3.1.6
Device configuration
1.1.1
/ OK +
Serial numbers (if not otherwise recorded)
Device leakage current
Application components for measurement of leakage current from application component
µA µA OK
1.1.1.1
C500
1.1.1.2
V-unit
txt
1.1.1.3
Inspiratory safety valve (inspiration unit)
txt
1.1.1.4
Disposable expiratory valve (YES/NO)
txt
3.1.7.1
Reference value
µA
1.1.1.5
Adult expiratory valve (reusable)
txt
3.1.7.2
Recurrent test
µA
1.1.1.6
Additional reusable adult expiratory valve
txt
3.2
Electrical safety according to IEC 60601-1
txt
3.1.7
Applied part leakage current, mains on applied part, type BF
1.1.1.7
MCable-Mainstream CO2 (CO2 sensor) option
txt
3.2.1
Visual inspection
OK
1.1.1.8
Neonatal expiratory valve option
txt
3.2.2
Protective earth resistance
Ω
1.1.1.9
Additional neonatal expiratory valve option
txt
3.2.3
1.1.1.10
GS500 option
txt
3.2.3.1
Normal condition (N.C.) IEC 60-601-1
µA
1.1.1.11
PS500 option
txt
3.2.3.2
Single fault condition (S.F.C.) IEC 60-601-1
µA
1.1.1.12
Trolley option
txt
3.2.3.3
Normal condition (N.C.) UL 60-601-1
µA
1.1.1.13
Nebulizer ejector additional hardware option (without serial number) (Yes or No)
txt
3.2.3.4
Single fault condition (S.F.C.) UL 60-601-1
µA
1.1.1.14
Neonatal flowmeter additional hardware option (without serial number) (Yes or No)
txt
3.2.4
Earth leakage current medical device
Patient leakage current
3.2.4.1
Normal condition (N.C.) DC IEC 60-601-1
µA
3.2.4.2
Normal condition (N.C.) AC IEC 60601-1
µA
2 Maintenance parts
3.2.4.3
Single fault condition (S.F.C.) DC IEC 60601-1
µA
2.1
3.2.4.4
Single fault condition (S.F.C.) AC IEC 60601-1
µA
3.2.4.5
Normal condition (N.C.) DC UL 60-601-1
µA
3.2.4.6
Normal condition (N.C.) AC UL 60-601-1
µA
1.1.2
Software versions Maintenance intervals, overview
2.1.1
Maintenance intervals and required sets and parts
2.2
Maintenance parts by specified interval
2.2.1
V-unit, battery M7 NM2,5
dat
4 Function and condition test
2.2.2
V-unit, particle filter (O2 filter)
dat
4.1
General condition check 4.1.1
2.2.3
V-unit, air filter
dat
Medical device
OK
2.2.4
V-unit, Spirolog sensor cable harness
dat
4.1.2
Accompanying documents
OK
2.2.5
(Option) GS500, room air filter (filter mat)
dat
4.1.3
OK
2.2.6
(Option) GS500, air filter blower unit (microfilter)
dat
Rating plates, labels and notices on the medical device
2.2.7
(Option) PS500, battery set PS500 (Service) 15Ah
dat
4.1.4
Rating plates, labels and notices on the installed options
OK
2.2.8
(Option) PS500, battery set PS500 (Service) 24Ah
dat
2.2.9
(Option) Medication nebulizer, pressure regulator (O2-side)
dat
2.3
Maintenance parts as required
4.2
C500 and V-unit condition test
4.2.1
C500
OK
4.2.2
V-unit
OK
4.3
Condition test, options
4.3.1
(Option) MCable Mainstream CO2 module M11.1 (CO2 sensor)
OK
4.3.2
(Option) Neonatal flow measurement
OK
4.3.3
(Option) Drug nebulizer
OK
4.3.4
(Option) Trolley
OK
4.3.5
(Option) Gas cylinder holder
OK
3 Electrical safety 3.1
Electrical safety of V-unit or ME system according to DIN EN 62353 (IEC 62353) 3.1.1
3.1.2
Visual inspection
OK
Protective earth resistance
3.1.2.1
Maximum measured value.
Ω
4.3.6
(Option) Hose holder
OK
3.1.2.2
Maximum measured value of ME system.
Ω
4.3.7
(Option) GS500
OK
3.1.3
Protective earth resistance measuring points
OK
4.3.8
(Option) PS500
OK
3.1.4
Applied parts for device leakage current
OK
4.3.9
(Option) Cable duct
OK
Revision 21.0
Workstation Critical Care
1/4
Test
Result
4.4
Test
Fuse rating and readout of software version and operating data
Result
4.12.1 Fan
OK
4.12.2
Supply voltage
OK
4.12.3
Internal temperature
OK
4.4.1
Fuse value
OK
4.4.2
Software version of medical device
txt
4.4.3
Battery operating data of the medical device with M7.3
Ah
4.13 V-unit, function test of internal power failure backup battery, part 1
4.4.4
Battery operating data of the medical device with M7.3 Plus
4.4.4.1
Battery operating data of the medical device with M7.3 Plus
txt
4.4.4.2
Battery operating data of the medical device with M7.3 Plus
txt
4.5
Preparations for the function tests
4.5.1
4.6
C500, function tests
4.6.1
4.6.2
Preparations
4.13.1
Powerfail mode function test
OK
4.14 V-unit, PI self-tests function tests
4.14.1 V-unit self-tests
OK
4.14.2 Flush flow M12.4
OK
4.14.3 IST Airway Pressure Measurement
OK
4.14.4 Pneumatic Release Logic Test
OK
4.14.5 Pressure sensors S2 and S8 OK
4.14.5.1
S2 deviation from the determined ambient pressure
mbar
Pixel test
OK
4.14.5.2
OK
S8 - variation from the measured ambient pressure
mbar
Acoustic alarm
4.6.3
Red and yellow visual alarms
OK
4.14.6 Expiratory flow measurement
4.6.4
Rotary encoder lighting
OK
4.7
V-unit, function test of module M1.3 (gas dosage module)
4.7.1
V-unit self-tests
OK
4.7.2
AIR proportional valve step response time
OK
4.7.3
O2 proportional valve step response time
OK
4.7.4
Zeroing all flow sensors
OK
4.7.5
Relative flow difference S2/S3
L/min
4.7.6
Relative flow difference S1/S3
L/min
4.7.7
Leakage LPSV V3
L/min
4.7.8
Valve calibration for workstations with gas supply unit GS500
4.7.9
Comparator shut-off pressure
4.7.12 Proportional valve AIR 5 L/min
A
4.7.13 Proportional valve O2 120 L/min
A
4.7.14 Proportional valve O2 5 L/min
A
4.7.15 System Leakage Test (M1V3)
OK
4.7.16 Leakage, AIR proportional valve
mbar
OK
OK
4.15 V-unit, function test of internal power failure backup battery, part 2 4.15.1 Battery parameters
OK
4.16 V-unit, test of device internal temperature
4.16.1 Testing
OK
4.17 V-unit, function test of piezo
4.17.1 Function check
OK
4.18 V-unit, function test of nurse call
4.18.1 Nurse call with nurse call box (8417370)
OK
4.18.2 Nurse call with test adapter (7911671)
OK
4.19 V-unit, function test of expiratory valve heater
4.19.1 Function check
mbar
4.20 Option, V-unit, Function test of nebulizer
4.7.10 Comparator switch-on pressure
mbar
4.7.11 Proportional valve AIR 120 L/min
A
4.20.1 Function check
OK OK
4.21 Option, V-unit, function test of drug nebulizer (white housing)
4.21.1 Function check
OK
4.22 Option, V-unit, function test of Mcable-Mainstream CO2 module M11.1 (CO2 sensor)
4.22.1 Function check
4.7.17 Leakage, O2 pipeline supply non-return valve
L/min
4.23 Option, V-unit, Function test of neonatal expiratory valve
4.7.18 Leakage, O2 proportional valve
mbar
4.7.19 Leakage, AIR pipeline supply non-return valve
L/min
4.24 Option, V-unit, Function test of neonatal flow sensor
4.7.20 Comparative pressure measurement, Ptank and inspiratory pressure sensor
OK
4.7.21 Comparative flow measurement M1.3
OK
4.8
V-unit, function test of module M12.4 (Pato standalone)
4.8.1
Calibration
OK
4.8.2
Measurement accuracy
OK
4.9
V-unit, function test of module M4.1 (pressure/flow temperature monitoring)
4.23.1 Function check 4.24.1 Function check
OK
OK OK
4.25 Option, V-unit, Function test of GS500
4.25.1 Function check
OK
4.26 Medical device, function tests in operation mode
4.26.1 Self-test
OK
4.26.2 Device checks
OK
4.26.3 Expiratory valves leak test
OK
4.26.4 Function tests in ventilation mode
OK
4.26.5 V-unit display function test
OK
4.9.1
Calibration of pressure sensors
OK
4.9.2
Offset of the inspiratory pressure sensor
mV
OK
4.9.3
Offset of the expiratory pressure sensor
mV
4.26.6 Function test of acoustic alarm, only for SW2.40 and SW2.41
4.9.4
Temperature measurement of module M4.1
OK
4.26.7 Function test of Silence key
OK
4.9.5
Calibration of expiratory flow sensor
OK
4.9.6
Reference voltage of module M4.1
OK
4.10 V-unit, function test of module M21 East
4.10.1 Function check
OK
4.11 V-unit, function test of safety devices
4.27 Option, PS500 function test 4.27.1 Function check
4.27.1.1
Function test after disconnecting power plug
OK
4.27.1.2
Function test after removing fuse F1
OK
4.27.1.3
Fuse rating of fuse F1
OK
4.27.1.4
Function test after inserting fuse F1 and removing fuse F2
OK
4.11.1 Emergency expiratory valve SV2
OK
4.11.2 Emergency respiratory valve SV3
OK
4.27.1.5
Fuse rating of fuse F2
OK
4.11.3 Non-return valve in expiratory valve
OK
4.27.1.6
Function test of mains power supply
OK
4.11.4 Expiratory resistance of adult expiratory valve
OK
4.28 Medical device, final action
4.12 V-unit, function test of the power supply unit module M7.3 / M7.3 Plus
Revision 21.0
4.28.1 Device status report
OK
4.28.2 Error messages and remedies
OK
Workstation Critical Care
2/4
Test
Result 4.28.3 Device handover with test label
OK
5 Test equipment
5.1 Test equipment subject to mandatory calibration
5.2
Test equipment not subject to mandatory calibration
Revision 21.0
OK
Workstation Critical Care
3/4
Report:
Test has been performed according to the test instructions. Name: ... :__________________________________________________________ Date/Signature:... :__________________________________________________________
Revision 21.0