SCAN OPTICS
SO-5800 MICROSCOPE
ASSEMBLY MANUAL
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 2 of 25
CONTENTS INTRODUCTION ...
3
PACKING LIST ...
5
MECHANICAL ASSEMBLY
6 6 9 11 13 15 17 19
... Floor stand base and pillar ... Pantograph arm ... X-Y unit ... Microscope head and suspension bracket ... Eyepieces, camera, handles and spare lamp module ... Monitor arm and bracket ... Binocular assistant microscope ...
CABLE AND POWER ASSEMBLY
...
22
MECHANICAL SETTINGS ...
24
FIGURES Figure A1
SO-5800 Microscope
...
4
Figure A2
Assembling the floor stand base ...
7
Figure A3
Assembled floor stand base
...
7
Figure A4
Inserting the pillar in the floor stand base ...
8
Figure A5
Fixing the pillar to the floor stand base
8
Figure A6
Placing the friction washer and arm bearing mounts
Figure A7
Attaching the pantograph arm
... ...
10
...
10
Figure A8
Attaching the X-Y unit ...
12
Figure A9
Attaching the microscope head and suspension bracket ...
14
Figure A10
Attaching eyepieces, handles and camera
...
16
Figure A11
Attaching the monitor arm and bracket ...
18
Figure A12
Binocular assistant microscope in RHS orientation
...
21
Figure A13
Binocular assistant microscope in LHS orientation
...
21
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 3 of 25
INTRODUCTION Please read the following information carefully before installing and using the Scan Optics SO-5800 Ophthalmic microscope. Scan Optics is responsible for the safety, reliability and performance of the equipment only if it is used in accordance with these instructions. This microscope is designed for use by a certified practitioner, for magnified observation of patients, and for use in an operating theatre as an observation aid during surgery. Environmental storage and packing conditions of 60-95% relative humidity and 10-40 °C, are recommended for this product. No parts or accessories supplied with this microscope are supplied in a sterile condition. Apart from those instructions within this manual, there are no user-serviceable parts in this microscope. Scan Optics will retain the discretion to advise whether any repairs may be taken out by external qualified technical personnel, or whether part(s) of the microscope must be returned to the manufacturer’s premises for service or repairs to be carried out under warranty or otherwise. Where appropriately qualified technical personnel are identified by a user, and ratified by Scan Optics, then Scan Optics will make available on request any information which will assist in repairing the equipment.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 4 of 25
PANTOGRAPH ARM X-Y POSITIONING UNIT
HORIZONTAL ARM
MICROSCOPE HEAD ASSEMBLY
POWER SUPPLY
MONITOR
PILLAR
LOWER PANEL FLOOR STAND
FOOT CONTROL
Figure A-1: SO-5800 Microscope
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 5 of 25
PACKING LIST Assembly Manual (this manual) User Manual Pillar and horizontal arm assembly, with mounting screws (M10x25) x 4 (includes power supply) Pantograph arm assembly (includes main power cable and video signal cable) Suspension bracket (with mounting screws) with attached spare lamp module Monitor arm assembly Floor stand legs (2 interlocking) Floor stand joiners (top and bottom), with 4 mounting screws Box 1: Microscope head assembly Box 2: X-Y unit with top bearing and knob Box 3: Monitor with mounting bracket in original packaging including: Box Video Accessories including: Monitor power cable Monitor remote control Video camera assembly Video camera power cable Box 4: Binocular assistant microscope assembly Box 5: Accessories: Mains cable Set sterilisable knob covers (6 x ‘zoom’, 2 x ‘focus’, 6 x ‘handle’) Microscope cover Tool box including: eyepieces x 2, lens cloth Set socket keys: 9 total; 1.5 - 10 mm Arm bearing mounts x 2 with 8 mounting screws and friction washer Elbow lock knob Microscope side handles with locking pins x2 Microscope front handle with mounting screw Box 6: Foot control
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 6 of 25
MECHANICAL ASSEMBLY NOTE:
MINIMUM 2 PERSONS REQUIRED
Floor stand base and pillar: refer figures A-2 to A-5 1. Remove the floor stand base from the crate 2. Place the top joiner upside-down on the floor and insert the upper and lower
interlocking legs as shown 3. Place the bottom joiner over the crossed legs and fix the assembly together
using the four M8x70 socket head cap screws and 6 mm socket key provided. Tighten the screws alternately to ensure even pressure is applied and that the assembly is seated square and true. 4. The floor stand base is now assembled but upside-down. Place the floor
stand base the right way up on the floor and lock the two wheel stops by applying foot pressure. 5. Unpack the pillar/horizontal arm assembly from the crate and remove the four
M10 x 25 socket head cap screws and washers from the pillar legs. 6. Insert the pillar pins into the floor stand base holes so that the front panel
faces out in between the legs with wheel locks. Unlock the horizontal arm and rotate it until it is in line with the front panel. Then lock it in to place again. 7. Take care with the next step:
Use the side handles to carefully lean the pillar and base back, then carefully chock up the elevated end. While one person holds the pillar flush against the base, the other can insert the four M10 x 25 socket head cap screws and washers from underneath the floor stand base assembly into the threaded holes in the end of the pillar pins. Tighten alternately with the 8 mm socket key until secure. The entire assembly may now be placed back upright on the floor.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 7 of 25
MOUNTING SCREWS BOTTOM JOINER LOWER LEG
UPPER LEG
TOP JOINER
Figure A-2: Assembling the floor stand base
Figure A-3: Assembled floor stand base
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 8 of 25
Figure A-4: Inserting the pillar in the floor stand base
Figure A-5: Fixing the pillar to the floor stand base
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 9 of 25
Pantograph arm: refer figures A-6 and A-7 1. Identify and remove the arm bearing mounts from the box.
Remove the pantograph arm assembly from the crate. Place the friction washer over the top exposed shaft at one end of the pantograph arm. The top side may be identified as the plain side; the bottom side features the sliding knob safety stop assembly. Place the arm bearing mounts in their correct orientation over the ends of the shaft on the pantograph arm. Note that the lower arm bearing mount must be supported by hand from underneath to prevent it dropping down. With the arm bearing mounts in place insert the pantograph arm in to the end of the horizontal arm assembly. Note that some shuffling movement will help to move the arm in to place. When the arm bearing mounts are flush against the horizontal arm, insert the 8 M6 x 16 countersunk screws and tighten them alternately using the 4 mm socket key. 2. Identify and remove the elbow lock knob from the box. Screw the knob in
underneath the end of the horizontal arm.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 10 of 25
Figure A-6: Placing the friction washer and arm bearing mounts
Figure A-7: Attaching the pantograph arm
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 11 of 25
X-Y Unit: refer figure A-8 1. Identify and remove the X-Y unit from the box. Remove the top bearing and
knob assembly from the X-Y unit by unscrewing the knob. Make sure that the set of washers remains inside the counter bore in the top bearing. 2. This important safety step is crucial:
Lower the pantograph arm so that the shaft of the X-Y unit may be inserted from underneath the conical bearing surface at the free end of the pantograph arm. Support the X-Y unit from underneath and hold it firm against the bearing surface. Tighten the 3 M6 x 12 safety screws (which are already in place) located at the end of the arm using the 5 mm socket key. While holding the X-Y unit, check that the safety screws are functioning by demonstrating that the X-Y unit will not drop to the floor when support from underneath is removed. 3. Place the top bearing over the shaft of the X-Y unit from above the
pantograph arm, ensuring that the slot locates on the transverse pin in the shaft, and that the friction washers are in place. Then screw the knob up on the shaft to set the friction for rotational movement of the X-Y unit. Rotate the X-Y unit until the front panel (identified by the Scan Optics label) faces out in line with the pantograph arm.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Figure A-8: Attaching the X-Y unit
Issue Number 2.0
Reference SO-5800-AM-2
Page 12 of 25
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 13 of 25
Microscope head and suspension bracket: refer figure A-9 1. Identify and remove the microscope head assembly from the box.
Attach the suspension bracket to the microscope so that the long arm of the bracket points upwards and the short arm points towards the eyepiece end of the microscope. Tighten the four M6 x 30 socket head cap screws evenly using the 5 mm socket key so that there is an equal gap on either side of the tilt adjuster. Check that the head assembly can now be safely lifted from the top end of the suspension bracket. 2. With the head assembly and bracket placed securely on a table or bench,
manoeuvre the floor stand towards the table so that when the pantograph arm is lowered, the front of the X-Y unit faces towards the location of the microscope eyepieces. Lower the pantograph arm down so that the protrusion underneath the X-Y unit aligns with the shape of the suspension bracket, and that the 4 tapped holes on the underneath of the X-Y unit align with the 4 clearance holes on the suspension bracket. Fix the suspension bracket to the X-Y unit by screwing in the four M6 x 30 socket head cap screws from underneath. Tighten the screws evenly using the 5 mm socket key so that there is an equal gap on either side of the suspension bracket. Check that the head assembly can now be safely lifted by lifting the entire pantograph arm.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 14 of 25
Figure A-9: Attaching the microscope head and suspension bracket
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Eyepieces, handles and camera
Page 15 of 25
: refer figure A-10
1. Remove the eyepiece blanks from the microscope head and replace them
with the eyepieces from the box. Keep the eyepiece blanks in a safe place for later use. 2. Identify and remove the microscope side handles and mounting bracket pins
from the box. Insert the handles in the holes located on either side of the microscope head, ensuring that the transverse holes are located vertically. Secure the handles by inserting the locating pins. 3. Remove the black cap between the eyepieces and lamp house by loosening
the side screws. Keep the cap in a safe place for later use. Fix the camera assembly in location by tightening the side screws again. 4. Note that the side handle and guide handle may be fitted with sterilisable
covers. In addition the manual knobs for zoom and focus may also be fitted with sterilisable covers.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 16 of 25
Figure A-10: Attaching eyepieces, handles and camera.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 17 of 25
Monitor arm and bracket: refer figure A-11 1. Attach the monitor arm to the back of the pillar using the four M5 x 12
countersunk screws, and tighten using the 3 mm socket key provided. 2. Slide the monitor over the arm in the desired orientation. Tighten the bottom
2 screws on the black plastic tube adapter to set the friction and adjust the monitor angle as appropriate
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 18 of 25
Figure A-11: Attaching the monitor arm and bracket
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 19 of 25
Binocular assistant microscope: refer figures A-12 and A-13 1. The binocular assistant microscope may be mounted on either side of the
main microscope, and allows for non-coaxial viewing from a perspective 90° rotated from the image in the main microscope. The binocular assistant microscope must first be adjusted to suit the left or right side before it is mounted on that side of the microscope. 2. Hold the microscope in your hand so that the black knob above the rack and
pinion system is closest, and the eyepieces are located on the other side of the rack and pinion system. A black horizontal fitting will then protrude either to the left or right side of the microscope head, with a silver shaft running transversely through the fitting.. 3. If the black fitting protrudes to the left side, the microscope is set to fit on the
left side of the main microscope. If the black fitting is on the right side the microscope is set to fit on the right side of the main microscope. 4. To insert the microscope in its fitting, first remove the appropriate side
handle from the main microscope by removing the silver pin and sliding the handle out. Then insert the shaft of the binocular assistant microscope mount in the hole and insert the pin to secure it. 5. Focus the main microscope on a flat target. 6. Adjust the viewing angle of the binocular assistant microscope by loosening
the side knob and tilting the assembly until the vertical centreline of the target is aligned horizontally in the binocular assistant microscope. Tighten the knob to secure the microscope in position. 7. If the target is already centred vertically in the binocular assistant microscope,
no further adjustment will be required. To make a small adjustment for vertical alignment, the microscope may swivel about its tilted axis. To make an adjustment, loosen the top knob, swivel the microscope until the image is centred and tighten the top knob again. 8. The binocular assistant microscope may be focussed independently from the
main microscope, but once in focus, it will remain in focus so long as the main microscope is focussed.
Issue Number 2.0
Reference SO-5800-AM-2
SO-5800 Ophthalmic Microscope
Assembly Manual
Page 20 of 25
9. To swap the binocular assistant microscope fitting from left to right hand or
vice versa: Ÿ Remove the microscope from its mounting hole on the side of the microscope Ÿ Loosen the top knob until the microscope may be rotated free of the retaining lip, then rotate it through a full 180°. Tighten the knob again. Ÿ Loosen the side knob until the horizontal fitting can be rotated free of the steel pins, then rotate it through a full 180°. Tighten the knob again. Ÿ The microscope is now ready to be mounted on the other side. 10. Note that the focussing handles on the binocular assistant microscope may
be fitted with sterilisable covers. These are the same type as used on the manual zoom control on the main optical head.
Issue Number 2.0
Reference SO-5800-AM-2