ONE-YEAR RESERVOIR MAINTENANCE PROTOCOL The one-year maintenance solution is a 0.3% hydrogen peroxide/distilled water solution. The below protocol should be followed annually for the Level 1® Fast Flow Fluid Warmer (H-1200, H-1025, H-1000, H-1100), as well as for the initial set up in your facility.
Level 1® Fast Flow Fluid Warmer To prepare 0.3% hydrogen peroxide/distilled water solution for the reservoir, mix 140 ml of 3% hydrogen peroxide solution and 1,260 mL of distilled water. 1. Empty reservoir of current solution 2. Place a container under the drain valve of the Level 1® Fast Flow Fluid Warmer. 3. Drain the re-circulating solution by turning the drain valve clockwise 90 degrees. 4. When all of the re-circulating solution has drained from the unit, close the drain valve. 5. Refill the unit with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water solution. 6. Insert a D-series disposable set (D-50/DI-50, D-60HL/DI-60HL, D-100/DI-100, D-300/DI-300) into the Level 1® Fast Flow Fluid Warmer. If you use a Level 1® Fast Flow Fluid Warmer for urological and/or gynecological procedures, insert an IR-series disposable set (IR-500, IR-600/IRI-600/IRI-600B, IR-700, IR-40). 7. Turn the unit on and run for a 30-minute disinfection period. 8. Switch the unit off, discard the disposable according to established hospital procedures and empty the reservoir of the re-circulating solution by turning the drain valve clockwise 90 degrees. When all of the re-circulating solution has been drained from the unit, close the drain valve. 9. Refill the reservoir with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water solution. The Level 1® Fast Flow Fluid Warmer is now ready for use. During the 12-month period, any solution lost from the reservoir through the disposable must be topped off with the 0.3% hydrogen peroxide/distilled water solution.
ONE-YEAR RESERVOIR MAINTENANCE PROTOCOL The one-year maintenance solution is a 0.3% hydrogen peroxide/distilled water solution. The below protocol should be followed annually the HOTLINE® Blood and Fluid Warmer (HL-90), as well as for the initial set up in your facility.
HOTLINE® Blood and Fluid Warmer To prepare 0.3% hydrogen peroxide/distilled water solution for the reservoir, mix 140 mL of 3% hydrogen peroxide solution and 1,260 mL of distilled water. 1. Empty solution currently in HOTLINE® Warmer reservoir. 2. Fill the reservoir with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water solution. 3. Insert a disposable set (L-70 or L-70NI) into the HOTLINE® Warmer. 4. Turn the unit on and let the solution circulate for a 30-minute disinfection period. 5. Switch the unit off and empty the reservoir. Discard the disposable according to established hospital procedures. 6. Refill the reservoir with 1.4 liters of fresh 0.3% hydrogen peroxide/distilled water solution. The HOTLINE® Blood and Fluid Warmer is now ready for use. During the 12-month period, any solution lost from the reservoir through the disposable must be topped off with the 0.3% hydrogen peroxide/distilled water solution.
Smiths Medical ASD, Inc. Anesthesia and Safety Devices Division 160 Weymouth Street Rockland MA 02370 USA Tel: 781 878 8011 Fax: 781 878 8201 www.smiths-medical.com
RECOMMENDED ONE-YEAR MAINTENANCE CHECKLIST Model: Level 1® Fast Flow Fluid Warmer Ref:
H-1000, H-1100, H-1025, H-1200
One-Year Reservoir Maintenance Protocol for the Level 1 Fast Flow Fluid Warmer* A new one-year reservoir maintenance protocol was launched in 2004 for Level 1 Fast Flow Fluid Warmer. This maintenance protocol involves using a 0.3% hydrogen peroxide/distilled water solution that maintains a microbe-free environment for one year. For details on this upgraded version of our reservoir maintenance protocol, please contact your local Smiths Medical ASD, Inc. representative or contact our Technical Support Department at 1-800-553-8351. One-Year Maintenance Tasks All testing and maintenance should be performed by qualified personnel only. Date of purchase:
____________________
Serial number:
____________________
Task
Date
Date
Date
Date
Change 0.3% Hydrogen Peroxide/Distilled Water Solution Replace O-Rings Clean Fan Filter Test Over-Temp Alarm Test Float Switch Test Disposable Interlocks Verify Temperature Calibration *The one-year maintenance protocol is in addition to the maintenance protocols listed in the Fast Flow Fluid Warmer operator’s manual. P/N: 7310129, Rev. A
Smiths Medical ASD, Inc. Anesthesia and Safety Devices Division 160 Weymouth Street Rockland MA 02370 USA Tel: 781 878 8011 Fax: 781 878 8201 www.smiths-medical.com
RECOMMENDED ONE-YEAR MAINTENANCE CHECKLIST Model: HOTLINE® Blood and Fluid Warmer Ref:
HL-90
HOTLINE Blood and Fluid Warmer One-Year Reservoir Maintenance Protocol* A new one-year reservoir maintenance protocol was launched in 2004 for the HOTLINE Warmer reservoir. This maintenance protocol involves using a 0.3% hydrogen peroxide/distilled water solution that maintains a microbe-free environment for one year. For details on this upgraded version of our reservoir maintenance protocol, please contact your local Smiths Medical ASD, Inc. representative or contact our Technical Support Department at 1-800-553-8351. One-Year Maintenance Tasks All testing and maintenance should be performed by qualified personnel only. Date of purchase:
____________________
Serial number:
____________________
Task
Date
Date
Date
Date
Change 0.3% Hydrogen Peroxide/Distilled Water Solution Replace O-Ring and Seals Lubricate O-Rings Test Alarm Switch Test Float Switch Test Disposable Switch Test Over-Temp Alarm Verify Temperature Calibration *The one-year protocol is an addition to the maintenance protocols listed in our HOTLINE® Warmer operators manual. *The one-year reservoir maintenance protocol supercedes the 6-Month maintenance protocol. P/N 7310128, Rev. A