Smiths Medical

CADD Pump Systems

CADD-Legacy PCA Model 6300 Quick Reference Card Lock Level 1

Quick Reference Card

2 Pages

AlArmS & TrouBlESHooTING  AlArmS & TrouBlESHooTING Continued  The CADD-legacy PCA pump signals with audible alarms and messages that appear on the display when a condition requires your attention.  Programming Incomplete Two-tone alarm when starting pump  ®  res Vol low Three beeps  Level of fluid in the reservoir is low. Prepare to install new reservoir.  reservoir Volume Empty Two-tone alarm  Reservoir volume has reached 0.0 ml. Press NEXT or STOP/START to silence alarm. Install a new reservoir, if appropriate.  low Bat 3 two-tone beeps every five minutes  Batteries are low, but pump is operable. Change batteries soon.  Battery Depleted Two-tone alarm  Batteries are depleted and cannot operate pump; install new batteries immediately.  Battery removed Pump Won’t run Two-tone alarm  With AC adapter attached, batteries have been removed while pump was running, or you have tried to start pump with depleted batteries. Install new batteries.  High Pressure Two-tone alarm  High pressure is being caused by an obstruction in the fluid path between the pump and the patient. Remove the obstruction to resume operation. Or press STOP/START or NEXT to stop the pump and silence the alarm for two minutes, then remove obstruction and restart pump.  upstream occlusion Two-tone alarm  CADD-legacy PCA ®  infusion Pump  Value Not Saved A value was not saved by pressing ENTER/CLEAR. Press NEXT to resume programming.  Model 6300  Error An error has occurred. Remove pump Two-tone alarm from service.  Quick Reference Card for Clinicians  lock level 1 (ll1)  No message The pump was running when batteries were Displayed removed or the batteries were removed Two-tone alarm within 15 seconds after stopping the pump. Batteries must be reinstalled.  A  ImPorTANT: Always stop pump before removing batteries. No Disposable, Clamp Tubing Two beep (long-short) alarm  B  The disposable was removed, or the cassette is not aligned with the pump or is damaged, or a malfunction of the pump sensor(s) is occurring. Clamp the tubing immediately. Press STOP/START or NEXT to silence the alarm.  C G D  Service Due The pump is due for service. Two-tone alarm  H  Fluid is not flowing from the reservoir to the pump. Check for a kink or air bubble in the tubing between the reservoir and pump. Remove the occlusion to resume operation. Press STOP/START or NEXT to stop the pump and silence alarm for two minutes, then remove the occlusion and restart the pump.  Screen Displays Current Pump Status Two-beep (long-short) alarm  The cassette is not aligned with the pump, or is damaged, or a malfunction of the pump sensor(s) is occurring. Reposition the pump to silence the alarm. If repositioning the pump does not silence the alarm within two minutes, the pump will display “No Disposable, Clamp Tubing.”  Air In line Detected Two-tone alarm  Air in fluid path. Press NEXT or STOP/ START to silence alarm. Make sure tubing is threaded in air detector properly. If the fluid path contains air, close clamps, disconnect from patient and prime fluid path.  Stopped Three beeps every five minutes  Pump is stopped. Start pump, if appropriate.  remote Dose Cord removed Two beeps when pump stopped, two-tone alarm when pump is running  Verify all programming screens by pressing ENTER/CLEAR before moving to next screen or starting pump.  Remote dose cord removed while pump was running.  E  I  ®  ®  F  WArNING: This Quick reference Card should be used by clinicians only. Do not permit patients to have access to this card, as the information would allow access to all programming and operating functions. Assistance with the CADD-legacy PCA pump is available to clinicians 24-hours-a-day by calling 1-800-258-5361. ®  Smiths medical ASD, Inc.  EC Authorized representative  A Display shows programmed values and messages B AC Adapter Jack used to plug in AC Adapter C Accessory Jack used to plug in Remote Dose Cord D Indicator Light indicates AC power in use E Air Detector F Cassette (part of reservoir or administration set that attaches to pump) G Threaded Mounting Hole for use with Polemount Bracket Adapter (back of pump) H Battery Compartment (back of pump) I Cassette Latch attaches cassette to pump (side of pump) KEYPAD STOP/START  Stops and starts the infusion  ENTER/CLEAR  Enters or clears displayed value  Smiths medical Canada ltd.  PRIME  Fills tubing with fluid  Markham, Ontario, Canada, L3R 4Y8 Phone: 905-477-2000 Toll-Free: 1-800-387-4346  LOCK  Displays or changes Lock Level (security level)  NEXT  Advances to next programming screen  ´ Î  Increases or decreases displayed values or scrolls through menu items  ON/OFF  Turns the pump on or off (low power)  DOSE  Delivers demand dose  St. Paul, MN 55112, USA Phone: 1-214-618-0218 Toll-Free USA: 1-800-258-5361  Smiths Medical international Ltd. TN25 4BF, UK Phone: +44 (0) 1233 722100  www.smiths-medical.com Smiths medical is part of the global technology business Smiths Group plc. Product(s) described may not be licensed or available for sale in Canada or other countries outside of the United States. CADD, CADD-Legacy, the Medication Cassette Reservoir design and the Smiths Medical and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. ©2010 Smiths Medical. All rights reserved. 10/10 IN19883  Read the entire Operator’s Manual before operating the CADD-Legacy ambulatory infusion pump. Failure to properly follow warnings, cautions and instructions could result in death or serious injury to the patient. ®
File Type: PDF
File Size: 881 KB
File Name: cadd-legacy_pca_model_6300_quick_reference_card_loc.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions