Smiths Medical

CADD Pump Systems

CADD-Solis Quick Reference Guide July 2010

Quick Reference Guide

2 Pages

Changing the battery  Alarms and troubleshooting continued  Alarms and troubleshooting continued  Stop the pump  Troubleshooting  Downstream occlusion. Clear occlusion between pump and patient.  1. 2. 3. 4. 5. 6.  Press Stop/Start. “Stop Pump?” displays. Press Yes. Remove the used batteries. Insert the new batteries. Press the power switch to turn the pump on. The screen displays “Do you want to start a new patient?” Press No. 7. Press Stop/Start to start the pump. 8. “Start Pump?” displays. Press Yes.  Screensaver The screensaver allows the pump to conserve battery power when not in an edit mode and if no keypad buttons have been pressed for 30 seconds. The pump displays a blank screen. Press any button on the keypad to turn the display on.  Alarms and troubleshooting Alarm Conditions High Priority Alarm If the pump is running, it always stops when a high priority alarm is activated. Accompanied by a red screen, it continues until acknowledged or until the condition that triggered the alarm goes away.  Medium Priority Alarm This alarm does not stop the pump. Accompanied by an amber screen, it continues until acknowledged or until the condition that triggered the alarm goes away.  Low Priority Alarm A low priority alarm does not stop the pump. Accompanied by a blue screen, the alarm automatically clears after 5 seconds or until the condition that triggered the alarm goes away.  Informational Message This alarm does not stop the pump. This message appears in the status bar. It is displayed for 5 seconds and is generally silent, requiring no acknowledgement.  Screen is blank and alarm is sounding Alarm Priority High. The pump has lost power and is no longer delivering. The pump was delivering and the batteries were removed or the battery door was opened. Clear this alarm by replacing the batteries or closing the battery door. Then turn the pump back on or the alarm stops after the power has been off for a minimum of 2 minutes.  Air-in-line detected. Press “acknowledge” then prime tubing. Alarm Priority High. The pump is stopped and can not run. The air detector has detected air in the fluid path; the fluid path may contain air bubbles. Acknowledge the alarm. Then, if the fluid path contains air bubbles, close the clamps, disconnect the fluid path from the patient, and follow the instructions for priming to remove the air.  Battery depleted. Pump stopped. Alarm Priority High. Install 4 new AA batteries or a fully charged rechargeable battery pack. In order to start delivery, good batteries must always be installed, even when an external source of power is connected. If appropriate, restart the pump.  Battery low, replace battery. Alarm Priority Low. Change the rechargeable battery pack or the 4 AA batteries soon.  Delivery limit reached. Or, delivery limit reached and partial dose delivered. Pump’s status bar reads “KVO = 0” Alarm Priority Low. The programmed delivery limit has been reached, and the pump is not delivering fluid. This alarm occurs when the continuous rate or a PCA dose has caused the delivery limit to be exceeded. Acknowledge the alarm (the alarm automatically clears after 5 seconds).  Pump’s status bar reads “Del Limit” Alarm Priority Low. The programmed delivery limit has been reached, and the pump is delivering fluid at the KVO rate of 0.1mL/hr. This alarm occurs when the continuous rate or a PCA dose has caused the delivery limit to be exceeded. Acknowledge the alarm (the alarm automatically clears after 5 seconds).  Alarm Priority High. The pump has detected high pressure, which may be resulting from a downstream blockage, kink in the fluid path, or a closed tubing clamp. Delivery pauses and resumes if the occlusion is removed. Remove the obstruction or stop the pump to silence the alarm for 2 minutes, then remove the obstruction and restart the pump.  CADD -Solis ®  Ambulatory Infusion System Protocol Programming using Code or Key  Reservoir volume low.  Alarm Priority Medium or Low (depending on how the alarm is programmed in Admin Settings). Level of fluid in the reservoir is low. Prepare to install a new reservoir, if appropriate.  Reservoir volume is zero. Pump stopped.  B  A  C  Alarm Priority High. The reservoir volume has reached 0.0 ml. The pump stops and can not run. Acknowledge the alarm. Install a new fluid container. Reset or edit the value of the reservoir volume.  Upstream occlusion. Clear occlusion between pump and reservoir.  Alarm Priority High. Fluid is not flowing from the fluid container to the pump, which may be resulting from a kink, a closed clamp, or air bubble in the tubing between the fluid container and pump. Delivery is paused and will resume if the occlusion is removed. Remove the obstruction to resume operation. The alarm clears when the occlusion is removed. You will be required to acknowledge this alarm after it clears if it has occurred and cleared more than 3 times within 15 minutes.  K D I  E F  J  G H  Customer and Clinical Services 1-800-258-5361 www.smiths-medical.com  Smiths Medical ASD, Inc. St. Paul, MN 55112, USA Phone: 1-214-618-0218 Toll-Free USA: 1-800-258-5361  EC Authorized Representative Smiths Medical International Ltd. WD24 4LG, UK Phone: +44 (0) 1923 246434  Smiths Medical Canada Ltd. Markham, Ontario, Canada, L3R 4Y8 Phone: 905-477-2000 Toll-Free: 1-800-387-4346  www.smiths-medical.com Smiths Medical is part of the global technology business Smiths Group plc. Product(s) described may not be licensed or available for sale in Canada or other countries outside of the United States. CADD, CADD-Solis and the Smiths Medical design mark are trademarks of the Smiths Medical family of companies. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. ©2010 Smiths Medical family of companies. All rights reserved. 7/10 IN193506  A B C D E F G H I J K  Battery Compartment Display Indicator Lights USB Port Blue AC Power Light AC Power Jack Remote Dose Cord Jack Cassette Cassette Latch Keypad Power Button
File Type: PDF
File Size: 1.12 MB
File Name: cadd-solis_quick_reference_guide_july_2010.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions