Page 1
Service Bulletin No: 2016-001
Date of issue: 23 FEB 2016 Products affected: BabyPAC B100 & B10
PneuPAC transport ventilator BabyPAC B100 and BabyPAC B10 Service & Calibration Equipment Smith’s Medical has recently made changes to the instruction on test equipment (ratePAC) used during testing and calibration of pneuPAC BabyPAC B100 and B10 Models. The ratePAC (fig 1) device was required to measure both Inspiratory and Expiratory Time in seconds. Unfortunately due to recent changes in regulatory and CE marking requirements, company decided to stop manufacturing ratePAC.
Figure 1: Ratepac Smiths Medical International Ltd
Bramingham Business Part, Enterprise Way, Luton, Beds. LU3 4BU www.Smiths-Medical.com
Alternatively, TSI Certifier FA Plus 4080 (manufactured by TSI) has been tested and validated as a suitable replacement for ratePAC. This test device (TSI Certifier FA Plus 4080) contains more test applications compared to ratePAC. Advantage of the TSI Certifier FA Plus 4080 is that several service tests can be performed simultaneously. For instance Inspiratory & Expiratory tests can be performed at once, in addition the test equipment offers a reduction of service time. With the optional 4073 kit you will be able to check the oxygen concentration.
Figure 2: TSI Certifier FA Plus 4080 Details below show instructions on how to connect the TSI FA Plus to the babyPAC and the settings required to perform required tests i.e. How to access the main menu and setting the required trigger points.
TSI certifier FA plus setup
1. Connect patient hose from BabyPAC to the test lung. 2. Connect the pressure tube from the baby test lung to the + (positive) pressure port on high flow module connected to the Certifier FA plus 4080.
Smiths Medical International Ltd
Bramingham Business Part, Enterprise Way, Luton, Beds. LU3 4BU www.Smiths-Medical.com
Test Lung
Patient Hose
TSI Certifier FA Plus 4080
Pressure Tube TSI High Flow module
Figure 3: TSI Certifier FA Plus connected to the BabyPAC
3. Turn the Certifier FA Plus Test System on. 4. A screen will appear with the TSI logo for few seconds. 5. One of either of screens shown below will appear on the screen (see figure 4 and 5)
Figure 4: TSI Certifier FA Plus graph view
Figure 5: TSI Certifier FA Plus parameter view
6. Allow the pressure transducer and flow sensor to warm up for 1 minute. 7. Perform the pre-test calibration as per TSI manual. To access the manual refer to http://service.tsi.com 8. Several options can be found on both views. Select the middle one amongst the three options at the bottom of the screen. Smiths Medical International Ltd
Bramingham Business Part, Enterprise Way, Luton, Beds. LU3 4BU www.Smiths-Medical.com
9. Another window will appear (see fig 6) with the Trigger option. 10. Select “Pressure (Advanced users)” option in Trigger type. 11. Set the START and END trigger point to 5 cmH2O.
Figure 6: Interface Module 12. Press OK to validate. 13. Click on any parameter by Tapping on the screen. This will open a new window. 14. Choose the desired parameters on the left box. 15. Highlight by selecting and click on the right arrow key to add it to the right box. The parameters on the right box will be the one displayed on the actual view. 16. Follow the test routine in babyPAC service manual to complete the testing using TSI Certifier FA Plus.
Please note, Customers can continue to use ratePAC provided it is in the working condition and Smiths Medical will continue to provide service subject to parts availability. For any further question or queries, contact our Technical Support Team on +44 (0) 845 850 0445 or email [email protected] . For other features related to the TSI Certifier FA Plus please visit http://service.tsi.com