CARESCAPE ONE Installation and maintenance checkforms Software version 3
CARESCAPE ONE English 3rd edition 2102480-003 © 2017-2018 General Electric Company. All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice. For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies, Inc. and GE Healthcare Finland Oy.
2
CARESCAPE ONE
2102480-003 2018-09-04
Contents About the check forms ... 5 Revision history ... 5 Planned maintenance schedules... 5 Service requirements... 6 Manufacturer responsibility ... 7 CARESCAPE ONE system installation check ... 9 CARESCAPE ONE system maintenance check... 13
2102480-003
CARESCAPE ONE
3
4
CARESCAPE ONE
2102480-003
About the check forms This document contains installation and maintenance check forms for CARESCAPE ONE. Use the check forms during installation and maintenance check procedures. Record the results and activities on the check forms. Perform the installation check after the hardware installation and the platform configuration are completed, before taking the monitor into clinical use. Perform the maintenance check after any corrective maintenance and during the periodic maintenance. After a completed checkout archive the signed check form and provide a signed copy to the customer. WARNING
SAFETY HAZARD. To avoid risks to personnel and patient, or damage to the equipment, only perform maintenance procedures described in this manual. Unauthorized modifications can lead to safety hazards.
Refer to the related service manual for detailed information regarding the checkout procedures. Refer to the user’s manual for important safety information regarding the whole system. The list below indicates the products (brands, models and descriptions as applicable) with which these check forms are to be used: ●
CARESCAPE ONE
●
CARESCAPE Dock F0
●
CARESCAPE ECG
●
CARESCAPE Pressure
●
CARESCAPE Temperature
●
CARESCAPE SpO2
●
CARESCAPE SpO2 - Masimo
●
CARESCAPE SpO2 - Nellcor
●
CARESCAPE CO2
Revision history Revision
Description
1st Edition
Initial release.
2nd Edition
Updated title page software version, ECG functional check acceptance criteria, and EC Rep contact information on the back cover.
3rd Edition
Updated CARESCAPE Planned maintenance schedules and the installation Electrical safety tests sections. Updated the Electrical safety tests - Corrective maintenance check and the Electrical safety tests - Planned maintenance check sections. Deleted the installation Parameter functional check section and the EC REP and CE Mark symbols from the last page.
Planned maintenance schedules See the user’s manual for the instructions about the cleaning and care schedules and about the regular calibration check intervals.
2102480-003
CARESCAPE ONE
5
Service personnel must perform the planned maintenance check as follows: Device
Planned maintenance schedule
● CARESCAPE ONE
Every two years after installation.
● CARESCAPE Dock F0 ● CARESCAPE CO2 ● CARESCAPE SpO2 - Masimo ● CARESCAPE SpO2 - Nellcor ● CARESCAPE ECG
Every seven years after installation.
● CARESCAPE Pressure ● CARESCAPE SpO2 ● CARESCAPE Temperature
NOTE
The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an Equipment Maintenance Agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
NOTE
The planned maintenance check must be performed to the whole patient monitoring system, including all the connected devices. See the related service manuals for information about the planned maintenance checks for the connected devices
Service requirements Follow the service requirements listed below. ●
Refer servicing of the equipment to qualified service personnel only. Service personnel servicing this product must have an appropriate technical qualification, or equivalent work experience, and be familiar with the service requirements described in this manual and in any related service documentation. Service training for the product is recommended.
●
Any unauthorized attempt to repair equipment under warranty voids that warranty.
●
It is the user's responsibility to report the need for service to GE or to one of their authorized agents.
●
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
●
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.
WARNING
6
Do not perform any service activities on the monitor in the patient vicinity while a patient is being connected to the monitor.
CARESCAPE ONE
2102480-003
CAUTION
DISPOSAL. At the end of its service life, the product described in this manual, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of each product. If you have any questions concerning disposal of a product, please contact GE or its representatives.
Manufacturer responsibility GE is responsible for the effects on safety, reliability, and performance of the equipment only if:
2102480-003
●
Assembly operations, extensions, readjustments, modifications, servicing, or repairs are carried out by authorized service personnel.
●
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
●
The equipment is used in accordance with the instructions for use.
●
The equipment is installed, maintained and serviced in accordance with the instructions provided in the related technical manuals.
CARESCAPE ONE
7
8
CARESCAPE ONE
2102480-003
CARESCAPE ONE system installation check Customer:
Monitor type: CARESCAPE ONE
Service record #:
Software version:
Service engineer
Service Manual: 2098697–001
Monitor S/N:
Other devices connected to the monitor system Device
Serial number
Device
Serial number
Prior to testing, verify that all equipment is calibrated via “Cal” labeling and record Cal Due Dates. Measuring equipment used Equipment/ tool
PASS = Test passed
2102480-003
Manufacturer
Model/ Type/ Part No
Serial number/ ID
N.A.= Test is not applicable, a component or a feature is not part of tested system.
CARESCAPE ONE
Calibration due date
FAIL = Test failed
9
Visual inspection Pass
Visual inspection
Fail
Comments
Check that all product labeling, markings and symbols are intact and readable. Check that the monitor and the connected devices do not have any visible damage. Check that the monitor and the connected peripheral devices are properly mounted with specified mounting solutions. Check that all mounting solution fasteners are intact, properly tightened, and correctly secured. Check that the power cord and power supply are properly secured. Check that the CARESCAPE Parameters are properly connected.
Electrical safety tests Electrical safety tests - Installation check The manufacturer has performed the electrical safety test for the monitor and measurement devices during final inspection. You do not have to perform the electrical safety tests during the installation checkout.
System functional check System functional check
Acceptance criteria
Alarm light
Power status indicator is lit. The blue, yellow and red alarm light blinks. The blue pause audio alarm indicator blinks.
Start-up
Green on/standby key is lit. Start-up screen briefly displays. Software package and version displays on welcoming screen are correct. Speaker gives an audible beep. Normal monitoring screen displays and there are no error messages displayed. Battery is fully charged.
Display: Picture quality
Text is readable, images are clear and brightness is adequate.
Display: Touchscreen control
Selected active parameter window opens the related menu.
CARESCAPE Parameter identification
Connected CARESCAPE Parameters display.
Status of connected devices
Host serial number and active software version are correct. Ordered licenses are enabled. CARESCAPE Parameters connected to the CARESCAPE ONE are correctly identified. Power line frequency is correctly configured.
10
CARESCAPE ONE
Pass
Fail
2102480-003
System functional check
Acceptance criteria
CARESCAPE Dock F0 is charging a battery
CARESCAPE Dock F0 is charging a battery.
Pass
Fail
Test completion
2102480-003
CARESCAPE ONE
11
12
CARESCAPE ONE
2102480-003
CARESCAPE ONE system maintenance check Customer:
Monitor type: CARESCAPE ONE
Service record #:
Software version:
Service engineer
Service Manual: 2098697–001
Monitor S/N:
Other devices connected to the monitor system Device
Serial number
Device
Serial number
Prior to testing, verify that all equipment is calibrated via “Cal” labeling and record Cal Due Dates. Measuring equipment used Equipment/ tool
PASS = Test passed
2102480-003
Manufacturer
Model/ Type/ Part No
Serial number/ ID
N.A.= Test is not applicable, a component or a feature is not part of tested system.
CARESCAPE ONE
Calibration due date
FAIL = Test failed
13
Visual inspection Pass
Visual inspection
Fail
Comments
Check that all product labeling, markings and symbols are intact and readable. Check that the monitor and the connected devices do not have any visible damage. Check that the monitor and the connected peripheral devices are properly mounted with specified mounting solutions. Check that all mounting solution fasteners are intact, properly tightened, and correctly secured. Check that the power cord and power supply are properly secured. Check that the CARESCAPE Parameters are properly connected.
Electrical safety tests Electrical safety tests - Corrective maintenance check Is the front bezel and/or top/bottom housings opened? Yes
Perform the electrical safety tests.
No
If the front bezel and/or the top/bottom housings were not opened, you do not have to perform the electrical safety tests. To continue the maintenance check, mark the Electrical safety test - Completion as N. A. (last section of the electrical safety tests), and proceed to the functional checks.
Electrical safety tests - Planned maintenance check Device
Test frequency
● CARESCAPE ONE
Perform the electrical safety tests if it has been more than two years since the electrical safety tests were last completed.
● CARESCAPE Dock F0 ● CARESCAPE CO2 ● CARESCAPE SpO2 - Masimo ● CARESCAPE SpO2 - Nellcor ● CARESCAPE ECG ● CARESCAPE Pressure
Perform the electrical safety tests if it has been more than seven years since the electrical safety tests were last completed.
● CARESCAPE SpO2 ● CARESCAPE Temperature
Pass
Electrical safety tests - CARESCAPE Dock F0
Fail
Power outlet is correctly wired. Power cord and plug are undamaged.
14
CARESCAPE ONE
2102480-003
Ground leakage current test CARESCAPE ONE and CARESCAPE Dock F0
Observed result µ
Acceptance criteria
GROUND, Neutral closed, Normal polarity
< 1000µ
GROUND, Neutral closed, Reverse polarity
< 1000µ
GROUND, Neutral open, Normal polarity
< 1000µ
Chassis leakage current to the ePort connection Observed result test µ CARESCAPE ONE and CARESCAPE Dock F0
Acceptance criteria
CHASSIS, Neutral closed, Normal polarity
< 100µ
CHASSIS, Neutral closed, Reverse polarity
< 100µ
CHASSIS, Neutral closed, Reverse polarity, GND open
< 100µ
CHASSIS, Neutral open, Normal polarity
< 500µ
Chassis leakage current to the Ethernet connection test CARESCAPE ONE and CARESCAPE Dock F0
Observed result µ
Acceptance criteria
CHASSIS, Neutral closed, Normal polarity
< 100µ
CHASSIS, Neutral closed, Reverse polarity
< 100µ
CHASSIS, Neutral closed, Reverse polarity, GND open
< 100µ
CHASSIS, Neutral open, Normal polarity
< 500µ Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Patient (source) leakage current tests, using a test body. LEAD-GND, RL lead, Neutral closed, Normal polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral closed, Reversed polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral open, Normal polarity
≤ 50 μ
Patient (sink) leakage current test. LEAD-ISO, RL lead, Neutral closed, Normal polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral closed, Reversed polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral open, Normal polarity
≤ 50 μ Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (source) leakage current tests, using a test body. LEAD-GND, RL lead, Neutral closed, Normal polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral closed, Reversed polarity
≤ 10 μ
2102480-003
CARESCAPE ONE
15
Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
≤ 50 μ
LEAD-GND, RL lead, Neutral open, Normal polarity Patient (sink) leakage current test. LEAD-ISO, RL lead, Neutral closed, Normal polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral closed, Reversed polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral open, Normal polarity
≤ 50 μ Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (source) leakage current tests, using a test body. LEAD-GND, RL lead, Neutral closed, Normal polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral closed, Reversed polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral open, Normal polarity
≤ 50 μ
Patient (sink) leakage current test. LEAD-ISO, RL lead, Neutral closed, Normal polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral closed, Reversed polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral open, Normal polarity
≤ 50 μ Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (source) leakage current tests, using a test body. LEAD-GND, RL lead, Neutral closed, Normal polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral closed, Reversed polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral open, Normal polarity
≤ 50 μ
Patient (sink) leakage current test. LEAD-ISO, RL lead, Neutral closed, Normal polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral closed, Reversed polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral open, Normal polarity
≤ 50 μ Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (source) leakage current tests, using a test body. LEAD-GND, RL lead, Neutral closed, Normal polarity
≤ 10 μ
LEAD-GND, RL lead, Neutral closed, Reversed polarity
≤ 10 μ
16
CARESCAPE ONE
2102480-003
Observed result μ
Electrical safety tests - system CARESCAPE Parameter: S/N:
Acceptance criteria
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
≤ 50 μ
LEAD-GND, RL lead, Neutral open, Normal polarity Patient (sink) leakage current test. LEAD-ISO, RL lead, Neutral closed, Normal polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral closed, Reversed polarity
≤ 50 μ
LEAD-ISO, RL lead, Neutral open, Normal polarity
≤ 50 μ Observed result μ
Electrical safety tests - single CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (sink) leakage current test. DUAL, CHASSIS DUAL, Neutral closed, OUTLET OFF
≤ 50 μ
DUAL, CHASSIS DUAL, Neutral closed, Reversed polarity
≤ 50 μ Observed result μ
Electrical safety tests - single CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (sink) leakage current test. DUAL, CHASSIS DUAL, Neutral closed, OUTLET OFF
≤ 50 μ
DUAL, CHASSIS DUAL, Neutral closed, Reversed polarity
≤ 50 μ Observed result μ
Electrical safety tests - single CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (sink) leakage current test. DUAL, CHASSIS DUAL, Neutral closed, OUTLET OFF
≤ 50 μ
DUAL, CHASSIS DUAL, Neutral closed, Reversed polarity
≤ 50 μ Observed result μ
Electrical safety tests - single CARESCAPE Parameter: S/N:
Acceptance criteria
Patient (sink) leakage current test. DUAL, CHASSIS DUAL, Neutral closed, OUTLET OFF
≤ 50 μ
DUAL, CHASSIS DUAL, Neutral closed, Reversed polarity
≤ 50 μ
2102480-003
CARESCAPE ONE
17
Observed result μ
Electrical safety tests - single CARESCAPE Parameter: S/N:
Acceptance criteria
Pass
Fail
N.A.
Fail
Pass
Fail
Patient (sink) leakage current test. DUAL, CHASSIS DUAL, Neutral closed, OUTLET OFF
≤ 50 μ
DUAL, CHASSIS DUAL, Neutral closed, Reversed polarity
≤ 50 μ Pass
Electrical safety test - Completion Test completion
System functional check System functional check
Acceptance criteria
Alarm light
Power status indicator is lit. The blue, yellow and red alarm light blinks. The blue pause audio alarm indicator blinks.
Start-up
Green on/standby key is lit. Start-up screen briefly displays. Software package and version displays on welcoming screen are correct. Speaker gives an audible beep. Normal monitoring screen displays and there are no error messages displayed. Battery is fully charged.
Display: Picture quality
Text is readable, images are clear and brightness is adequate.
Display: Touchscreen control
Selected active parameter window opens the related menu.
CARESCAPE Parameter identification
Connected CARESCAPE Parameters display.
Status of connected devices
Host serial number and active software version are correct. Ordered licenses are enabled. CARESCAPE Parameters connected to the CARESCAPE ONE are correctly identified. Power line frequency is correctly configured.
CARESCAPE Dock F0 is charging a battery
CARESCAPE Dock F0 is charging a battery.
Test completion
Parameter functional check
18
CARESCAPE ONE
2102480-003
ECG functional check Tasks
Acceptance criteria
Normal sinus rhythm
ECG II, VI, and aVL waveforms are shown and are noise-free.
Observed result
Pass
N.A.
Fail
Observed result
Pass
N.A.
Fail
Heart rate is 80 ± 5 BPM Audible QRS tone is heard. All 12 ECG leads are available and are noise-free. Pacemaker detection
Pacemaker spikes are shown on the ECG waveform. Heart rate is 80 ± 1 BPM
Asystole detection
Asystole alarm is shown on the screen. Heart rate is 80 ± 1 BPM and is a normal sinus rhythm.
Leads off detection without the RA/R leadwire
Lead II disappears from the screen and RA/R lead off message appears. Lead II is replaced by Lead III on the screen after the RA/R leadwire is disconnected. Lead III is replaced by Lead II on the screen after the RA/R leadwire is reconnected.
Test completion Impedance respiration functional check Tasks
Acceptance criteria
Respiration rate
RESP waveform is shown. RR value is 20 ± 5 breaths per minute.
Apnea detection
RESP waveform is shown as a flat line. An Apnea alarm is activated. RR value is 20 ± 5 breaths per minute.
Test completion
2102480-003
CARESCAPE ONE
19
Invasive pressure functional check Tasks
Acceptance criteria
Zeroing
Zeroing followed by a Zeroed message is shown.
Static pressure
Pressure line appears as a flat line.
Observed result
Pass
N.A.
Fail
Observed result
Pass
N.A.
Fail
Observed result
Pass
N.A.
Fail
Pressure value is 200 ± 4 mmHg/26.7 ± 0.5 kPa. Pressure waveform
Pressure waveform and Sys/Dia (Mean) values 120/80 mmHg / 16/10.7 kPa are shown in the related parameter window.
Test completion Temperature functional check Tasks
Acceptance criteria
T1 temperature channel
The temperature value matches the configured simulator value ± 0.1°C/ ± 0.2°F and that there are no error messages shown on the monitor screen.
T2 temperature channel
The temperature value matches the configured simulator value ± 0.1°C/ ± 0.2°F and that there a no error messages shown on the monitor screen.
Test completion SpO2 functional check Tasks
Acceptance criteria
SpO2 value
The SpO2 value is shown in the parameter window.
Plethysmographic waveform
A plethysmographic waveform is shown on the monitor screen.
Test completion
20
CARESCAPE ONE
2102480-003