Giraffe™ and Panda™ Scale (Shekel*) Service Manual
*This manual applies only to the following scale part numbers: •
Giraffe Non-EU Scale: M1125056
•
Panda Non-EU Scale: M1125055
•
Giraffe EU Scales: M1147621, M1147623, M1147624, M1147625, M1147626, M1147628, M1147629, M1147630, M1147631, M1147632, M1147633, M1147634, M1147635, M1147636, M1147637, M1147638
•
Panda EU Scales: M1147670, M1147687, M1147688, M1147689, M1147690, M1147691, M1147692, M1147693, M1147694, M1147695, M1147696, M1147697, M1147698, M1147699, M1147700, M1147701
© 2014-2016 General Electric Company All rights reserved. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE Representative for the most current information. Giraffe are registered trademarks owned by Datex-Ohmeda, Inc. GE and GE Monogram are trademarks of General Electric Company. All other company and product names mentioned may be trademarks of the companies with which they are associated. Information contained in this document is proprietary to GE Healthcare. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form, in whole or in part, by any means electronic, mechanical or otherwise, including photocopying and recording, for any purpose without written permission of GE Healthcare. All illustrations are provided as examples only. Your product features and configuration may be different than those shown. Names of persons, institutions and places and related information are fictitious; any similarity to actual persons, entities or places is purely coincidental. Existing artwork or images that you may desire to scan may be protected by copyright law. Be sure to obtain permission for use of existing artwork. Specifications in this manual are subject to change without notice. Contact GE Healthcare for the most current information. Trademarks
This product carries the Trademark of General Electric Company. All the trademarks of component parts used by GE in the manufacture of this product are the property of their respective owners. The GE logo, Carestation, Giraffe, and Panda are registered trademarks of General Electric Company. All brand or product names are trademarks or registered trademarks of their respective companies or organizations. Warranty
This Product is sold by GE Healthcare under the warranties set forth in the following paragraphs. Such warranties are extended only with respect to the purchase of this Product directly from GE Healthcare or GE Healthcare’s Authorized Dealers as new merchandise and are extended to the Buyer thereof, other than for the purpose of resale. For a period of twelve (12) months for the product from the date of original delivery to Buyer or to Buyer’s order, but in no event for a period of more than two years from the date of original delivery by GE Healthcare to an GE Healthcare Authorized Dealer, this Product, other than its expendable parts, is warranted to be free from functional defects in materials and workmanship and to conform to the description of the Product contained in this operation manual and accompanying labels and/or inserts, provided that the same is properly operated under the conditions of normal use, that regular periodic maintenance and service is performed and that replacements and repairs are made in accordance with the instructions provided. This same warranty is made for a period of thirty (30) days with respect to expendable parts such as batteries. The foregoing warranties shall not apply if the Product has been repaired other than by GE Healthcare or in accordance with written instructions provided by GE Healthcare, or altered by anyone other than GE Healthcare, or if the Product has been subject to abuse, misuse, negligence, or accident. GE Healthcare’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the above warranties is limited to repairing or replacing, free of charge, at GE Healthcare’s option, a Product, which is telephonically reported to the nearest GE Healthcare Regional Service Office and which, if so advised by GE Healthcare, is thereafter returned with a statement of the observed deficiency, not later than seven (7) days after the expiration date of the applicable warranty, to the designated GE Healthcare Service Center during normal business hours, transportation charges prepaid, and which, upon GE Healthcare’s examination, is found not to conform with above warranties. GE Healthcare shall not be otherwise liable for any damages including but not limited to incidental damages, consequential damages, or special damages. There are no express or implied warranties that extend beyond the warranties hereinabove set forth. GE Healthcare makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof.
Table of Contents
Language Disclaimer for Service Manuals ... 5 Safety and Regulatory Information... 13 Warnings...14 Cautions...14 Symbols...14 Regulatory ...16 Standards...16 User Responsibility...16
About this Manual ... 17 Scope ...17 Intended Audience...17 Conventions...18
Chapter 1: Functional Description... 19 Chapter 2: Installation ... 21 2.1 Tool Requirements...21 2.2 Installation Procedure...21
Chapter 3: Service Maintenance and Checkout... 23 3.1 Maintenance and Checkout Procedures Schedule...23 3.2 Tool Requirements...23 3.3 Maintenance and Checkout Procedures...23 3.3.1 Visual Inspection ...23 3.3.2 Center Weight Check...24 3.3.3 Off Center Weight Check ...24 3.3.4 Load Cell Check...25 3.3.5 Scale Calibration...25
Chapter 4: Diagnostics and Troubleshooting... 29 4.1 Troubleshooting Symptoms ...29 4.2 Hardware Check...31 3
Table of Contents
4.3 Weight Check...31 4.4 Service Screen...31
Chapter 5: Repair and Replacement Procedures ...33 5.1 Control Board Replacement ...33 5.2 Load Cell Replacement ...35
Chapter 6: Service Parts ...39 6.1 Scale Part Numbers List...39 6.2 Giraffe Scale Service Parts ...40 6.3 Panda Scale Service Parts ...41
Appendix A: Specifications ...43 Index...45
4
Giraffe and Panda Scale (Shekel) Service Manual
Language Disclaimer for Service Manuals
12
Giraffe and Panda Scale (Shekel) Service Manual
Safety and Regulatory Information
The service information is important for the safety of both the patient and operator and also serves to enhance equipment reliability.
!
Warning: Before servicing the device, read through this entire manual. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. This device should be serviced only by authorized service personnel. Additional precautions specific to certain procedures are found in the text of this manual.
The information contained in this service manual pertains only to those models of products which are marketed by GE Healthcare as of the effective date of this manual or the latest revision thereof. This service manual was prepared for exclusive use by GE Healthcare service personnel in light of their training and experience as well as the availability to them of parts, proper tools, and test equipment. Consequently, GE Healthcare provides this service manual to its customers purely as a business convenience and for the customer’s general information only without warranty of the results with respect to any application of such information. Furthermore, because of the wide variety of circumstances under which maintenance and repair activities may be performed and the unique nature of each individual’s own experience, capacity, and qualifications, the fact that a customer has received such information from GE Healthcare does not imply in any way that GE Healthcare deems said individual to be qualified to perform any such maintenance or repair service. Moreover, it should not be assumed that every acceptable test and safety procedure or method, precaution, tool, equipment, or device is referred to within, or that abnormal or unusual circumstances may not warrant or suggest different or additional procedures or requirements. This manual is subject to periodic review, update, and revision. Customers are cautioned to obtain and consult the latest revision before undertaking any service of the equipment.
!
Warning: The user or service staff should dispose of all the waste properly as per federal, state, and local waste disposal regulations. Improper disposal could result in personal injury and environmental impact.
Do not use malfunctioning equipment. If the system is under warranty, contact GE technical support at the number on the back of the manual PRIOR to performing any repairs on the system.
13
Safety and Regulatory Information Warnings
Warnings
!
Warning: If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe use of equipment.
!
Warning: Complete the checkout procedure before putting the unit in operation. If the unit fails any portion of the checkout procedure it must be removed from use and repaired.
!
Warning: Do not service the device while it is in clinical use.
Cautions Caution: Always perform the checkout procedure after performing any type of repair and before placing the scale back in service. Caution: Genuine parts manufactured or sold by GE Healthcare must be used for all part replacements or repairs. Caution: U.S. Federal law restricts this device to sale by, or on the order of, a licensed medical practitioner. Caution: This device is for professional use only.
Symbols This section identifies the symbols that are displayed on the devices. Equipment configurations can differ. Labels or symbols may not appear on every device. Symbol
Description Consult manual. Serial number.
UDI Model
Unique Device Identifier. The UDI is a unique marking for identification of the medical device. Model or part number. Manufacturer. Date of manufacture.
14
Giraffe and Panda Scale (Shekel) Service Manual
Safety and Regulatory Information Symbols
Symbol
Description Do not immerse scale. European Representative. This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
III M
Weighing Precision Class III. The scale is part of a verified weighing system. Temperature range (on device and package).
Fragile, handle with care (on package).
Keep dry (on package).
%
Relative humidity range (on package).
Pressure range (on package). This end up (on package).
Beginning in September 2016, some medical devices sold in the USA must be labeled with a Unique Device Identifier (UDI). The UDI label will be located on or adjacent to the serial number label on the device. Example of UDI label format:
Giraffe and Panda Scale (Shekel) Service Manual
15
Safety and Regulatory Information Regulatory
Regulatory GE Healthcare has declared that this product conforms with the European Council Directive 93/42/EEC Medical Device Directive when it is used in accordance with the instructions provided in this Manual.
Standards This device is designed to meet requirements of: • • • • • • • • •
ANSI/AAMI ES60601-1 CSA C22.2 #60601-1 IEC 60601-1 IEC 60601-1-2 IEC 60601-2-19 IEC 60601-2-21 2002/96/EC WEEE directive 21 CFR CH-1 Section 1020.30 (n) ISO 19054 with an exception claim to clause 5.2.7.2. Slight deformation of rail occurs after application of 500N to wide surface of DIN rail.
User Responsibility This Product will perform in conformity with the description thereof contained in this manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should such repair or replacement become necessary, GE Healthcare recommends that a telephone or written request for service advice be made to the nearest GE Healthcare Regional Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by GE Healthcare and by GE Healthcare trained personnel. The Product must not be altered without GE Healthcare’s prior written approval. The user of this Product shall have the sole responsibility for any malfunction that results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than GE Healthcare.
16
Giraffe and Panda Scale (Shekel) Service Manual
About this Manual
Scope This service manual describes how to install, repair, and maintain the Shekel Scale.
!
Warning: After performing any repair or calibration, always perform the Service Checkout Procedure before putting the device back into service. Caution: Servicing of this device in accordance with this service manual should never be undertaken in the absence of proper tools, test equipment and the most recent revision to this service manual which is clearly and thoroughly understood.
Intended Audience The intended users for this service manual are authorized service personnel. The information contained in this service manual pertains only to those models of products which are marketed by GE Healthcare as of the effective date of this manual or the latest revision thereof. This service manual was prepared for exclusive use by GE Healthcare service personnel in light of their training and experience as well as the availability to them of parts, proper tools and test equipment. Consequently, GE Healthcare provides this service manual to its customers purely as a business convenience and for the customer’s general information only without warranty of the results with respect to any application of such information. Furthermore, because of the wide variety of circumstances under which maintenance and repair activities may be performed and the unique nature of each individual’s own experience, capacity, and qualifications, the fact that a customer has received such information from GE Healthcare does not imply in anyway that GE Healthcare deems said individual to be qualified to perform any such maintenance or repair service. Moreover, it should not be assumed that every acceptable test and safety procedure or method, precaution, tool, equipment or device is referred to within, or that abnormal or unusual circumstances, may not warrant or suggest different or additional procedures or requirements. This manual is subject to periodic review, update and revision. Customers are cautioned to obtain and consult the latest revision before undertaking any service of the equipment.
17
About this Manual Conventions
Conventions This section introduces the various types of warnings, cautions, information notes, and symbols used in this manual to alert you to possible safety hazards and to provide you with additional information.
!
Warning: A Warning statement is used when the possibility of injury to the patient or the operator exists. Caution: A Caution statement is used when the possibility of damage to the equipment exists. Sensitive to electrostatic discharge caution: An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that the part(s) are sensitive to electrostatic discharge and that static control procedures must be used to prevent damage to the equipment.
Note:
A note provides additional information to clarify a point in the text.
Important: An Important statement is similar to a note, but is used for greater emphasis.
18
Giraffe and Panda Scale (Shekel) Service Manual
Chapter 1: Functional Description
The Giraffe™ and Panda™ infant scale utilizes the latest developments in transducer technology and integrated circuits to provide a highly reliable and accurate weighing scale. The scale is built with two load cells and an electronic circuit, which calculate the weight value and communicate it to the bed unit (warmer or incubator) it is connected to. The scale is connected to the bed unit with a cable for data communication and getting power. GE scales that are sold in European Union (EU) member states have been designed and manufactured in accordance with the non-automatic weighing instruments (NAWI) directive, 2009/23/EC. These scales must be calibrated by an accredited local test authority. The bed service screen displays two scale parameters labeled SR and SC, which are related to the weight applied to the top of the scale. Note: •
The service screen displays SR and SC values only if the scale is connected to the bed unit prior to entering the service screen.
SR (Scale Counts Raw) The SR value represents the number of times the scale has been calibrated.
•
SC (Scale Counts Calibration) The SC parameter is the weight value after the scale’s digital processing of the applied weight. On a properly-calibrated scale, each SC count represents 1 g of applied weight. When a weight is placed on the scale, the SC value increases (1 count per 1 g of applied weight), and the increased value is added to the SC value when no weight was applied. For example, if the SC value corresponding to a no-weight condition is 10,0000, placing a 5 kg test weight on the scale results in an SC value equal to 15,000. SC values can range from 0 to 40,000, however, offset and correction factors limit this to a smaller range.
19
Chapter 1: Functional Description
20
Giraffe and Panda Scale (Shekel) Service Manual
Chapter 2: Installation
2.1 Tool Requirements One Phillips Screw Driver
2.2 Installation Procedure 1.
Use a Phillips screw driver to remove the 4 scale lock screws. Discard the screws.
FIGURE 2-1. Removing the 4 Scale Lock Screws 2.
Place the scale on the bed, and connect the scale cable to the bed. Make sure the red marker on the plug is properly aligned to the socket.
FIGURE 2-2. Connecting the Scale Cable to the Bed
21
Chapter 2: Installation Installation Procedure 3.
For NAWI scales, place the Min/Max weight labels (provided in scale label kit M1124633) on the bed as shown in Figure 2-3.
5 cm
® Giraffe OmniBed Giraffe® Incubator
Min = 300 g MAX 8.00 kg Max1 = 5 kg, e1 = 5 g MIN 0.30 kg = 8kgkg, Max2 =eMax = 0.01 e2 = 10 g
® Giraffe Warmer ® Panda Warmer
FIGURE 2-3. Where to Place the Min/Max Weight Labels 4.
22
Place the clear plate and mattress on the scale.
Giraffe and Panda Scale (Shekel) Service Manual
Chapter 3: Service Maintenance and Checkout
3.1 Maintenance and Checkout Procedures Schedule Table 3-1 includes planned maintenance procedures and checkout procedures required after product installation, repair, or maintenance. For good care and maintenance of the product, perform planned maintenance procedures at specific time intervals. These procedures must be performed by authorized service personnel. TABLE 3-1. Maintenance and Checkout Procedures Schedule Performed after First Installation
Performed as Planned Maintenance
Performed after Repair, Parts Replacement, or Calibration
Visual Inspection
Yes
Annually
Yes
Center Weight Check
Yes
No
Yes
Off Center Weight Check
Yes
No
Yes
Load Cell Check
No
No
Yes
Scale Calibration
No
Annually
Yes
Checkout Procedure Name
Caution: Table 3-1 shows the minimum frequencies required for maintenance. Always follow hospital and local regulations for required frequencies.
3.2 Tool Requirements One 5 kg Certified Test Weight (GE Part Number: 6600-0209-800)
3.3 Maintenance and Checkout Procedures 3.3.1 Visual Inspection 1. 2.
Examine the scale parts for evident signs of damage. Examine the scale connector to make sure it is tightly assembled. Check for bent pins. If any of the parts are damaged replace them. Ensure there is no mechanical interference between the scale cable and the unit.
3.
Ensure the scale cable is not crossing over a corner tray of the bed. Move the corner tray if necessary.
FIGURE 3-1. Cable Not Crossing a Tray (Correct)
FIGURE 3-2. Cable Crossing a Tray (Incorrect)
23
Chapter 3: Service Maintenance and Checkout Maintenance and Checkout Procedures
3.3.2 Center Weight Check 1.
Turn off the bed. Clear all objects from the bed, and ensure only the clear plate remains on the scale.
2.
Ensure the bed is level.
3.
Turn on the bed, and go to the service screen.
4.
Record the SC value on the screen, and name it SC1. The SC1 value must be in the 9900 to 11000 range.
5.
Place a 5 kg weight on the center of the scale on the bed. (See location A in Figure 3-3.)
6.
Record the SC value on the screen, and name it SC2.
7.
Subtract SC1 from SC2. The value must be 5000 +/- 10.
3.3.3 Off Center Weight Check 1.
Turn off the bed. Clear all objects from the bed, and ensure only the clear plate remains on the scale.
2.
Ensure the bed is level.
3.
Turn on the bed, and go to the service screen.
4.
Record the SC value on the screen, and name it SC3. The SC3 value must be in the 9900 to 11000 range.
5.
Place a 5 kg weight on one corner of the scale on the bed. (See location B1 in Figure 3-3.)
6.
Record the SC value on the screen, and name it SC4.
7.
Subtract SC3 from SC4. The value must be 5000 +/- 10.
8.
Place a 5 kg weight on the second corner of the scale on the bed. (See location B2 in Figure 3-3.)
9.
Record the SC value on the screen, and name it SC5.
10. Subtract SC3 from SC5. The value must be 5000 +/- 10. 11. Place a 5 kg weight on the third corner of the scale on the bed. (See location B3 in Figure 3-3.) 12. Record the SC value on the screen, and name it SC6. 13. Subtract SC3 from SC6. The value must be 5000 +/- 10. 14. Place a 5 kg weight on the fourth corner of the scale on the bed. (See location B4 Figure 3-3.) 15. Record the SC value on the screen, and name it SC7. 16. Subtract SC3 from SC7. The value must be 5000 +/- 10.
FIGURE 3-3. Weight Placement Locations
24
Giraffe and Panda Scale (Shekel) Service Manual
Chapter 3: Service Maintenance and Checkout Maintenance and Checkout Procedures
3.3.4 Load Cell Check SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that the part(s) are sensitive to electrostatic discharge and that static control procedures must be used to prevent damage to the equipment. 1.
Perform the scale calibration procedures described in section 3.3.5, "Scale Calibration."
2.
Turn off the bed. Make sure no mechanical interference (loose screws, etc.) exists, and the scale cable is routed properly.
3.
Turn on the bed, and go to the service screen.
4.
Check the displayed SC values, and confirm that the values do not change more than 5 counts total.
5.
If the SC values change more than 5 counts, 1 or both load cells might be defective. Perform the following steps: a.
Remove the scale top tray and the control board cover to access the load cell connectors on the control board.
b.
Disconnect 1 load cell from the control board at a time, and check the SC value. If the SC values change 5 counts or fewer after disconnecting the load cell, the disconnected load cell is defective and must be replaced.
3.3.5 Scale Calibration Note:
For EU Member States, this scale has been designed and manufactured in accordance with the non-automatic weighing instruments (NAWI) directive, 2009/23/EC. To ensure complete compliance to the Directive, have the unit calibrated by an accredited local test authority. If you are unable to identify a test authority, contact your GE representative or distributor for assistance.
Periodic re-verification (calibration) of the scale must be performed in accordance with the National Regulations Governing Legal Metrology. The required re-verification frequency will vary by country. Contact your local weights and measures authority for more information. Note:
One 5 kg Certified Test Weight (GE Part Number 6600-0209-800) is required to perform the calibration procedure.
3.3.5.1 Giraffe Incubator and OmniBed™ Carestation™ 1.
Clear all objects from the bed, and ensure only the mattress and clear plate remain on the scale.
2.
Ensure the bed is level.
3.
In user mode, touch Setup on the Home screen. The Setup screen appears.
4.
Touch the data baby name field Enter here. The keyboard appears.
5.
If this is the first time accessing the application, enter Pass0000 and go to step 6. If not, enter your password (password + PIN). The password is not case sensitive and may be typed in upper or lower case.
6.
Wait for Password Verified to be displayed.
7.
Turn off the Standby switch to turn off the device.
8.
Turn on the Standby switch within one minute to turn on the device. The device restarts in service mode and is open to the Setup screen.
Giraffe and Panda Scale (Shekel) Service Manual
25
Chapter 3: Service Maintenance and Checkout Maintenance and Checkout Procedures 9.
In service mode, touch Calibration in the left-navigation area. The display area is populated.
10. Touch Cal Scale. 11. When the screen prompt REMOVE THE WEIGHT appears, verify the weight is removed and touch Next. A series of instructions appears on-screen. The screen prompts INITIALIZING for a few seconds. 12. When the screen prompts PLACE TEST WEIGHT AND PRESS NEXT, place the weight and press Next. The screen prompts MEASURING for a few seconds. 13. When the ENTER TEST WEIGHT prompt appears, type the test weight to the nearest gram and touch Enter. The screen prompts MEASURING for a few seconds. 14. When the ENTER TEST WEIGHT prompt appears, type the test weight to the nearest gram and touch Enter. For a successful calibration, the screen prompts CALCULATING for a few seconds. 15. When the screen prompts EXIT ONLY, RESTORE DEFAULT, SAVE AND EXIT, touch SAVE AND EXIT. Note: If SAVE AND EXIT is grayed out, the calibration was not successful. Repeat the calibration. 16. Turn off the power to exit the service mode. 3.3.5.2 Giraffe Incubator and OmniBed 1.
Clear all objects from the bed, and ensure only the mattress and clear plate remain on the scale.
2.
Turn off the bed. Ensure the bed is level.
3.
Place the test weight on the center of the scale.
4.
Enter the service screen on a Giraffe by turning on the bed while holding the override key (>37).
5.
On the second service screen, select Cal Scale.
6.
Remove the weight and push the knob at the screen prompt REMOVE THE WEIGHT AND PUSH KNOB. The screen will prompt INITIALIZING for a few seconds.
7.
Replace the weight and push the knob at the screen prompt PLACE TEST WEIGHT AND PUSH KNOB. The screen will prompt MEASURING for a few seconds.
8.
When the screen prompts ENTER TEST WEIGHT, dial in the test weight to the nearest gram. Press the knob to enter. The screen prompts CALCULATING for a few seconds.
9.
When the screen prompts SAVE AND EXIT EXIT ONLY RESTORE DEFAULT, select and enter SAVE AND EXIT.
10. Power cycle the unit to exit service mode and enter operation mode. 11. Verify the accuracy of the scale by performing a weighing cycle using the test weight. The measured value should be the test weight +/- 10 g. 12. Perform the checkout procedures described in Chapter 3, “Service Maintenance and Checkout.”
26
Giraffe and Panda Scale (Shekel) Service Manual
Chapter 3: Service Maintenance and Checkout Maintenance and Checkout Procedures 3.3.5.3 Giraffe and Panda Warmer 1.
Clear all objects from the bed and ensure only the mattress and clear plate remain on the scale.
2.
Turn off the bed. Ensure the bed is level.
3.
Enter service screen by turning on the bed while holding the Help and Alarm Silence keys until you hear 2 tones (approximately 5 seconds).
4.
On the Main Menu, select Scale Calibration and then Manual Calibration.
5.
Use the up/down arrow keys to set the gram value of the known test weight, and press the Enter key.
6.
When prompted, clear all objects from the bed.
7.
Press the Start key. Scale Obtaining Tare is displayed.
8.
When Apply Test Weight is displayed, place the weight on the center of the bed and press the OK key.
9.
If calibration is successful, Calibration Successful is displayed. If not, press the Recal key and repeat the procedure.
10. Do one of the following: – Press the Save key to save the new calibration constants. – Press the Abort key to retain the original setting. 11. Power cycle the unit to exit service mode and enter operation mode. 12. Verify the accuracy of the scale by performing a weighing cycle using the test weight. The measured value should be the test weight +/- 10 g. 13. Perform the checkout procedures described in Chapter 3, “Service Maintenance and Checkout.”
Giraffe and Panda Scale (Shekel) Service Manual
27