Giraffe Display Repair Addendum For future reference, add to your Giraffe OmniBed or Incubator Service Manual. The existing electro-luminescent display used in the Giraffe Incubator and Giraffe OmniBed accounts for a significant portion of the material cost for the product. This old technology display is also becoming increasingly hard to source (lead times are 15+ weeks) and further cost increases are expected. The Giraffe Incubator and OmniBed have been redesigned to include a new display that has a lower cost and is widely available. This new design has replaced the previous design on all production units going forward. The following repair procedures and illustrated parts are to be used in place of the existing service manual content for the new display components. The contents of this addendum will be incorporated into the service manuals in the next revision.
Conventions in this Addendum A Warning statement is used when the possibility of injury exists.
SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that the part(s) are sensitive to electrostatic discharge and that static control procedures must be used to prevent damage to the equipment.
© 2009 by General Electric Company. All rights reserved.
M1181029 003
i
Giraffe Display Repair Addendum
Repairing the Display Note: Reassembly for all procedures in this section is the reverse of those provided, unless otherwise noted. 1. If you are planning to remove or replace the bezel, touch panel, or digital encoder, then raise the canopy to its highest position, power off the unit, and unplug the unit. Note: All other replaceable display assembly components are accessible without moving the canopy. 2. Using a 2 mm hex key, remove the four button-head screws and lock washers that secure the back panel to the display module. Remove the back panel.
3. To remove the display driver board, use a 5.5 mm nut driver to remove the four KEPS nuts that secure the board to the standoffs.
© 2009 by General Electric Company. All rights reserved.
M1181029 003
1
Giraffe Display Repair Addendum 4. Pull the board away from the mounting standoffs enough to access and remove all harness/cable connectors as shown.
5. To remove the LED display board, use a 2 mm hex key to remove the two button-head screws and washers that hold the board in place. Then pull the right side of the board until it is approximately 60 degrees from the mounted position. It may be necessary to rock the board back and forth several times to release it from the self locking standoffs on the left side of the board.
2
M1181029 003
© 2009 by General Electric Company. All rights reserved.
Giraffe Display Repair Addendum 6. To remove the display interface board, disconnect the harness/cable connectors as follows. Unlock the display data cable connector on the display interface board by flipping the locking mechanism (brown door) toward you and pull the cable out and away from the connector. Disconnect the display backlight connector. The backlight connection may be different depending on which of the three display types is installed. Working several times around the board, gently pull or pry the board off of each self locking standoff, until the board is free.
Note: Metal contacts go toward the board. 7. To remove the EL display, remove the three KEPS Nuts that hold the display mounting bracket in place. Lift the display mounting bracket and EL display as a unit and place on an ESD work surface. To disconnect the display data cable from the EL display, slide the locking mechanism (black) away from the connector and remove the cable.
Note: Metal contacts go toward the board. Note: During replacement, align the metal display bracket tab with the slot in the display cage. Also, the yellow switch on the interface board should be set based on the manufacturer of the display. For details, refer to “EL Display Service Kit” on page 9. 8. Consider removing the four standoffs to make the next step easier.
© 2009 by General Electric Company. All rights reserved.
M1181029 003
3
Giraffe Display Repair Addendum 9. Lay the display face up on the work surface. Gently lift or pry the EL assembly from the display mounting bracket. Avoid getting fingerprints on the new display and be sure it is clean before installation.
Warning: Metal edges on the display bracket may be sharp. Important: During replacement, take note of the manufacturer of the display in the service kit. The yellow switch on the interface board should be set based on the manufacturer of the display. For details, refer to “EL Display Service Kit” on page 9. 10. To remove the bezel, first remove the encoder knob by pulling firmly as shown. The standoffs that hold the back cover also secure the bezel to the front of the module housing. Use a 6 mm wrench to remove the standoffs. Remove the bezel.
11. To replace the touch panel, it may be advisable to remove the cage. If you choose to not remove the cage, skip this step and proceed to step 12. For each side, do the following: Slide the decorative strips up about ½" in the groove on the outside of the upright to access the screw that secures the lower upright end cap. Use a 2.5 mm hex key to remove the screw, then remove the cap. Slide the decorative strip down to access the screw that secures the upper upright cap and remove the screw and decorative strip. Lift the top inner and outer cap up and away from the display cage just
4
M1181029 003
© 2009 by General Electric Company. All rights reserved.
Giraffe Display Repair Addendum enough to allow clearance to remove it. Using a 3 mm hex key, remove the four socket head mounting screw on either side of the Display Cage. Remove the Display Cage and place on a flat work surface.
12. Peel the old touch panel off the display cage, be sure to remove all adhesive residue from the cage before installing a new Touch panel. Caution: Once the membrane panel has been applied, it can not be repositioned without damage. 13. Remove the protective backing from the new touch panel. Do not allow the adhesive on the back of the new touch panel to come in contact with any surface. While keeping the adhesive backing from contacting the Display Cage, carefully align the two holes in the touch panel with the alignment pins on the Display Cage as shown. Firmly press the adhesive backing against the Display Cage, working out any bubbles that may have formed.
© 2009 by General Electric Company. All rights reserved.
M1181029 003
5
Giraffe Display Repair Addendum 14. To remove the encoder, while holding the encoder from the back, use an 11 mm socket remove the hex nut and washer as shown. Note: It may be necessary to fold the touch panel overlay back to access the hex nut. Be careful not to disturb the adhesive backing.
Note: Reassembly for all procedures in this section is the reverse of those provided, unless otherwise noted.
6
M1181029 003
© 2009 by General Electric Company. All rights reserved.
Giraffe Display Repair Addendum
Illustrated Parts
2
1
3 4
6
5
7
30
4 8 9 10
29
11
28 27 3 26
10 12 13 14
25
15 24
22
23
21
16 20
17 19
18
The following table describes each part in the above diagram:
Table 1: Parts Item #
Part #
Description
1
6600-1134-500
Rear Cover
2
6600-1070-400
Stand Off, M3 X 2.25
3
6600-0713-431
Lock Washer, M3 X 3.2 ID 0.4 Internal Tooth
4
M1074118
Nut, M3 Hex KEPS Stainless Steel
5
6600-0213-850
PCA Display Driver
6
6600-0725-700
EL Display Harness
7
M1182174
Service Kit Interface Board
8
M1176979
Display Mounting Bracket, Sheet Metal
© 2009 by General Electric Company. All rights reserved.
M1181029 003
7
Giraffe Display Repair Addendum
Table 1: Parts
8
Item #
Part #
Description
9
M1182173
Service Kit EL Display (For details, refer to “EL Display Service Kit” on page 9.)
10
M1179275
Rotary Encoder Switch (includes hex nut and washer)
11
6600-1276-500
Nut Plate
12
M1176658
Touch Panel Assembly (Giraffe OmniBed)
12
M1177948
Touch Panel Assembly (Giraffe Incubator)
13
6600-0715-403
Screw, M2.5 x 8 Socket Head
14
6600-1132-500
Alarm Lens
15
6600-1736-501
East Seal
16
6600-1480-500
Control Knob
17
6600-1065-400
Power Fail Lens
18
6600-1131-500
Display Bezel
19
6600-1736-502
West Seal
20
6600-1066-400
Light Pipe
21
6600-0707-409
Screw, M4 x 10 Socket Head
21
6600-0712-403
Flat Washer, M4
21
6600-0713-403
Lock Washer, M4 Split Ring
22
M1172128
Display Cage, GP, Sheet Metal
23
6600-0215-850
PCA LED
24
6600-0739-700
LED Display Harness
25
6600-0706-405
Screw, M3 x 8 Button Head
25
6600-0712-402
Flat Washer, M3
25
6600-0713-402
Lock Washer, M3 Split Ring
26
M1179278
Harness FFC
27
M1177289
Stand Off, M3 X 1.73
28
6600-0703-700
Wire Harness Display Data
29
6600-0706-700
Wire Harness Display Power
30
6600-0706-401
Screw, M3 x 10 Button Head
M1181029 003
© 2009 by General Electric Company. All rights reserved.
Giraffe Display Repair Addendum
EL Display Service Kit The EL Display Service Kit (M1182173) may include one of three possible displays. To distinguish between the displays, refer to the following photo and descriptions:
Table 2: Three Possible Displays Photo
Description The AUO displays have a sticker with the AUO logo visible inside the hole in the upper left corner on the back of the display. The AUO displays have a cable from the top left of the display, which will be connected to the lower connector on the interface board.
The Sharp displays say SHARP in lower left corner on the back of the display. The Sharp displays have a harness from the top left of the screen connected to the hinged connector in the upper left corner of the interface.
The Optrex displays are recognizable by the harness exiting from the bottom left corner of the display. This harness will be fed under the interface board to the lower connector on the interface board.
© 2009 by General Electric Company. All rights reserved.
M1181029 003
9
Giraffe Display Repair Addendum The yellow switch on the interface board should be set based on the manufacturer as shown below.
Table 3: Switch Settings
10
Manufacturer
Switch Setting
Optrex
1
AUO
2
Sharp
3
M1181029 003
Photo
© 2009 by General Electric Company. All rights reserved.
Giraffe Display Repair Addendum
© 2009 by General Electric Company. All rights reserved.
M1181029 003
11
World Headquarters
EC Representative
Datex-Ohmeda Ltd. Ohmeda House 71 Great North Road Hatfield Hertfordshire AL9 5EN Tel +44 1707 263570 Fax +44 1707 260065
GE Healthcare 9900 West Innovation Drive Wauwatosa, WI 53226-4856 USA Tel 1 800 345 2700
Europe, Middle East, Africa
Germany
GE Healthcare P.O. Box 900 FIN-00031 GE Finland Tel +358 10 39411 Fax +358 9 146 3310
GE Medical Systems Information Technologies GmbH Munzinger Str. 3-5 79111 Freiburg Tel. 49 761 4543 570 Fax 49 761 4543 571 Service 0800 4343258
Latin America Representatives
Asia Representative
GE Healthcare 3350 SW 148 Avenue Suite 301 Miramar, Florida, 33027 USA Tel + 1 954 744 5600
GE Healthcare Shanghai GE (China) Hi-tech Park No1 Huatuo Road, Zhangjiang Hi-tech Park Pudong, Shanghai, P.R.China 201203 上海 GE 中国科技园 地址:中国上海市浦东张江高科技园华佗路 1 号 , 201203 Tel + (8621) 38777888 Fax + (8621) 38777402
Brazil Only
GE Healthcare Clinical Systems Equipamentos Médicos Ltda Australia: Av. Paulista, 37 - 13º andar China: CEP: 01311-902 - Cerqueira César India: São Paulo, SP - Brasil Korea: Tel +55 11 3053 2500 Fax +55 11 3053 2573
Ohmeda Medical, a Division of Datex-Ohmeda, Inc., a General Electric Company 8880 Gorman Road Laurel MD 20723 USA
South Eastern Asia:
1300 722 229 800 810 8188 1 800 425 7255 (02) 1544 4564 (65) 6277 3444
Giraffe Display Repair Addendum Printed in USA Copyright 2009 by General Electric Company All rights reserved.
M1181029 003 Printing instructions: Black on white 8.5" x 11" paper size Two (2) sides print Three (3) hole punch Stapled