Solar
Giraffe, Lullaby and Panda Neonatal Systems
Giraffe OmniBed and Giraffe Incubator Servo O2 Sensor Kit Instructions Rev A Dec 2013
Instructions
11 Pages
Preview
Page 1
Instructions for Field Use For future reference add this instruction to your Giraffe Omnibed (6600-0343-000) or Incubator (6600-0356-000) Service Manual
Giraffe® OmniBed® and Giraffe Incubator
Servo O2 Sensor Kit Instructions Kit (6600-0120-850) Includes:
2 replacement Servo Oxygen Sensors (6600-1278-600) 1 Kit instructions (this document)
Instructions Check your Servo O2 board revision (see picture for revision level location on the board). To access the Servo O2 board follow the instructions in the Service Manual under the Repair Procedures section “Servo Oxygen Board Repairs”.
Servo O2 Software Revision can be found here
Servo O2 Board Revision can be found here
Page 1
M1236601
A
Released
Mitchell, Barbara J
10-Dec-2013
Approval details stored in the electronic master in the Global Agile PDM System
Page 1
of 11
Instructions for Field Use Check your Servo O2 software revision by powering up the unit. After the Giraffe logo appears, the screen indicates the current Servo software revision level installed. Alternatively, you may find the revision level on the U6 EPROM on the Servo Oxygen board, refer to picture above. Follow actions described in table below: Servo O2 Board p/n 6600-1043-700 (in kit 6600-0233-850) Revision 8 or lower 9 or higher
GSOCS Software EPROM p/n 6600-0841-600 Revision 1.40 & lower 1.60 & higher
9 or higher
1.50 or 1.51
Servo O2 Board and Software Revision Combination Required Action(s) Replace only the O2 sensors. Replace only the O2 sensors. Order Software Chip : 6600-0255-850 Replace O2 sensors AND Software Chip (6600-0255-850)
IMPORTANT: Failure to follow the required actions in the table above may cause the unit to operate improperly. If your configuration is not listed in the table above, call Technical Support. IMPORTANT: Only install the oxygen sensors supplied with this kit or oxygen sensors that have a discard date of 2012-12 or later. Install sensors per the product Service Manual. RECOMMENDATION: To avoid frequent calibration prompts, it is recommended to match the Servo O2 software revision 1.60 or higher with display board software revision 2.03 or higher. (The display software revision is displayed when the unit is turned on. Display software revision 2.03 or higher may be obtained by ordering EPROM Kit part number 6600-0234-850 and installed using the instruction provided in the product Service Manual.)
Page 2
M1236601
A
Released
Mitchell, Barbara J
10-Dec-2013
Approval details stored in the electronic master in the Global Agile PDM System
Page 2
of 11
World Headquarters GE Healthcare 9900 West Innovation Drive Wauwatosa, WI 53226-4856 USA Tel 1 800 345 2700
EC Representative GE Medical Systems SCS 283 Rue de la miniere 78530 BUC - FRANCE
Europe, Middle East, Africa
Germany
GE Healthcare P.O. Box 900 FIN-00031 GE Finland Tel +358 10 39411 Fax +358 9 146 3310
GE Medical Systems Information Technologies GmbH Munzinger Str. 3-5 79111 Freiburg Tel. 49 761 4543 570 Fax 49 761 4543 571 Service 0800 4343258
Latin America Representatives
Asia Representative
GE Healthcare 3350 SW 148 Avenue Suite 301 Miramar, Florida, 33027 USA Tel + 1 954 744 5600
GE Healthcare Shanghai GE (China) Hi-tech Park No1 Huatuo Road, Zhangjiang Hi-tech Park Pudong, Shanghai, P.R.China 201203 上海GE中国科技园 地址:中国上海市浦东张江高科技园华佗路1号, 201203 Tel + (8621) 38777888 Fax + (8621) 38777402
Brazil Only GE Healthcare Clinical Systems Equipamentos Médicos Ltda Av. Paulista, 37 - 13º andar CEP: 01311-902 - Cerqueira César São Paulo, SP - Brasil Tel +55 11 3053 2500 Fax +55 11 3053 2573
Ohmeda Medical, a Division of Datex-Ohmeda, Inc., a General Electric Company 8880 Gorman Road Laurel MD 20723 USA
Australia
1300 722 229
China
800 810 8188
India
1 800 425 7255
Korea
(02) 1544 4564
South Eastern Asia
(65) 6277 3444
Giraffe® OmniBed® and Giraffe Incubator Operation and Maintenance Manual Printed in USA Copyright 2012 by General Electric Company All rights reserved.
M1236601 Revision A Language: English
Printing instructions: Black on white 8.5” x 11” paper size Two (2) sides print Three (3) hole punch Stapled
Page 11 M1236601
A
Released
Mitchell, Barbara J
10-Dec-2013
Approval details stored in the electronic master in the Global Agile PDM System
Page 11