LUCAS® 2 Service Checklist Work Order #:
Type of inspection:
□ Annual □ Routine □ Post-Repair
Customer: LUCAS Serial #:
Exterior Physical Inspection 1 Device exterior condition (cleaned – no exterior damage) 2 Check Bellows and Suction Cup 3 Check Fabrics (Patient straps, Stabilization strap and carrying bag)
Pass
Maintenance Procedure Perform if necessary 1 Software version (Fill in version/checksum below OR add print-out) CPU: • Control • Protective • Charger 2 3 4 5 6
SW Ver:
NA
Remarks
Print-out added Checksum:
Remarks
J J J
Clean Electric Fan and Mesh Grill Lubricate Carry Ball Screw Lubricate Claw Lock Mechanism Test run Attach Service Label (optional)
Function Check Mechanics Tests 1 Check that the Decompression Spring moves smoothly and without unusual noise. 2 Check that the Carry Ball Nut runs smoothly and without unusual noise. 3 Check for Release Ring play when Claw Mechanism is unlocked. 4 Check that Claw Mechanism remains locked when pressure is applied. Electronics Tests 1 Internal Function Test and Adjust LED is green 2 “30:2” mode • Device runs without unusual sound. • Confirm active 30:2 LED is green. • Confirm audible alert prior to ventilation pause and intermittent LED. 3 Check that PAUSE LED is green. “Continuous” mode 4 • Confirm continuous LED is green. • Confirm LED blinks every 12 strokes. 5 “Adjust” mode • Confirm that Suction Cup runs smoothly. 6 “Charge” mode • Connect a PSU and confirm that the charging sequence starts.
Complete Complete Complete Complete Complete Pass
Remarks:
Performed by:
LUCAS® 2 Chest Compression System – Service PIP Checklist 3323774 Rev B © 2017 Physio-Control, Inc.
Fail
Date:
Fail
NA