Thermo Scientific
Heratherm Compact Microbiological Incubator Operating Instructions May 2010
Operating Instructions
39 Pages

Preview
Page 1
© 2010 Thermo Fisher Scientific Inc. All rights reserved. This Instruction Manual is copyright protected. Rights resulting thereof, particularly reprint, photomechanical or digital postprocessing or reproduction, even in part, are only allowed with the written consent of Thermo Electron LED GmbH. This regulation does not apply to reproductions for in-plant use. The contents of this document are subject to change without notice. Concerning translations into foreign languages, the German version of these operating instructions is binding. Trademarks Heratherm® is a registered trademark of Thermo Scientific. Thermo Scientific is a brand owned by Thermo Fisher Scientific Inc. All other trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers. Thermo Electron LED GmbH Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany Thermo Electron LED GmbH is an affiliate to: Thermo Fisher Scientific Inc. 81 Wyman Street Waltham, MA 02454 USA Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc. The content of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser. Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly. This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
Heratherm IMC 18
Thermo Scientific
T Table of contents Chapter 1
Safety Information ... 1-1 Basic operating precautions ... 1-1 Safe usage instructions ... 1-2 Guarantee ... 1-2 Explanation of safety information and symbols... 1-2 Intended use of the device... 1-4
Chapter 2
Delivery of the device... 2-1 Packaging ... 2-1 Acceptance inspection... 2-1 Scope of supply ... 2-1
Chapter 3
Installation ... 3-1 Environmental conditions ... 3-1 Room ventilation ... 3-2 Installation... 3-2 Transport ... 3-2
Chapter 4
Description of unit ... 4-1 Device overview... 4-1 Mains connection ... 4-2 Useable space components... 4-2 Condensation tray... 4-3
Chapter 5
Installation procedures... 5-1 Insert the perforated shelves ... 5-1 Preparing the useable space... 5-1 Mains connection ... 5-1
Chapter 6
Operation ... 6-1 Preparing the device ... 6-1 Starting Operation ... 6-1
Chapter 7
Handling and control... 7-1 Powering up and down... 7-2 Setting the temperature... 7-2 Interior light... 7-2
Chapter 8
Shut-down... 8-1 Shutting the device down ... 8-1
Thermo Scientific
Heratherm IMC 18
i
Table of contents
Chapter 9
Cleaning and disinfection ... 9-1 Empty the condensation tray ... 9-1 Cleaning exterior surfaces... 9-1 Wipe / spray disinfection... 9-1
Chapter 10 Maintenance ... 10-1 Inspections and checks ... 10-1 Chapter 11 Disposal ... 11-1 Overview of materials used... 11-2 WEEE Conformity ... 11-2 Chapter 12 Technical Data... 12-1 Chapter 13 Contact Information ... 13-1
ii
Heratherm IMC 18
Thermo Scientific
1
Safety Information Basic operating precautions These operating instructions describe the Heratherm Incubator IMC 18. The Heratherm Incubator IMC 18 has been manufactured to the latest state of the art and been tested thoroughly for flawless functioning prior to shipping. However, the device may present potential hazards, particularly if it is operated by inadequately trained personnel or if it is not used in accordance with the intended purpose. Therefore, the following must be observed for the sake of accident prevention: • The Heratherm Incubator IMC 18 is only to be used by trained and authorised personnel. • The present operating instructions, applicable safety data sheets, plant hygiene guidelines and the corresponding technical rules issued by the operator shall be used to create written procedures targeted at personnel working with the subject matter device, detailing: • the disinfection measures to be employed for the device and the accessories used with it, • not to be used in a humid atmosphere, • the protective measures which are to be taken when processing certain agents, • action to be taken in case of accidents. • Maintenance work on the device is only to be carried out by trained and authorised specialists. • The contents of these operating instructions are subject to change at any time without further notice. • Concerning translations into foreign languages, the German version of these operating instructions is binding. • Keep these operating instructions close to the device so that safety instructions and important information are always accessible. • Should you encounter problems that are not detailed adequately in these operating instructions, please contact Thermo Fisher Scientific immediately for your own safety (see chapter Contact Information).
Thermo Scientific
Heratherm IMC 18
1-1
Safety Information Safe usage instructions
Safe usage instructions Observe the load of the stiffening shelves (refer to chapter technical data). Do not charge the floor of the interior as there is a risk of the inserted item overheating. Distribute the items evenly and do not place them too close to the interior walls, so as to ensure a good temperature distribution.
Guarantee Thermo Fisher Scientific warrants the operational safety and functions of the Heratherm Incubators IMC 18 only under the condition that: • the device is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions, • the device is not modified, • only original spare parts and accessories that have been approved by Thermo Fisher Scientific are used. The warranty is valid from the date of delivery of the device to the operator.
Explanation of safety information and symbols Safety information and symbols used in the operating instructions
Indicates a hazardous situation which, if not avoided, will result in death or serious injuries.
Indicates a hazardous situation which, if not avoided, could result in death or serious injuries.
Indicates a situation which, if not avoided, could result in damage to property.
1-2
Heratherm IMC 18
Thermo Scientific
Safety Information Explanation of safety information and symbols
Is used for useful hints and information regarding the application.
Additional symbols for safety information
Wear safety gloves!
Wear safety goggles!
Harmful liquids!
Electrical shock!
Fire hazard!
Contamination hazard!
Thermo Scientific
Heratherm IMC 18
1-3
Safety Information Intended use of the device
Symbols on the device / Applicable standards Observe operating instructions
VDE test mark This device has been tested by the VDE on the basis of DIN EN/ISO/IEC 61010-1 and 61010-2-010 and bears the VDE GS symbol for tested safety. It is subjected to production monitoring by the VDE. CE-conformity mark: confirms the conformity with the EU directives having validity for this product, this being confirmed in the EC Declaration of Conformity for this product. Mark of conformity USA/Canada This device has been tested by the Canadian Standards Association (CSA) on the basis of CAN/CSA-C22.2 No. 61010-1 and CAN/CSA-C22.2 No. 61010-2-010, UL 61010-1 and IEC 61010-2-010 and bears the cCSAus symbol for tested safety. It is subject to production monitoring by the CSA.
Intended use of the device Correct use The Heratherm Incubator IMC 18 is a laboratory device in the standard version and serves the preparation and cultivation of microbiological cultures and micro-organisms in safety levels L1 and L2. The device enables the special physiological ambient conditions which these cultures require to be simulated due to it exactly controlling the useable space temperature. The device is only to be used in indoor locations.
Incorrect use To avoid the risk of explosion do not load the device with tissue, material, or liquids that: • are easily flammable or explosive, • release vapor or dust that forms combustible or explosive mixtures when exposed to air, • release poisons, • create a humid atmosphere, • release dust, • have an increased biological risk potential (safety level L3, L4). The devices are not to be stacked.
1-4
Heratherm IMC 18
Thermo Scientific
Safety Information Intended use of the device
A use of the device which is not intended can be hazardous.
Thermo Scientific
Heratherm IMC 18
1-5
Safety Information Intended use of the device
1-6
Heratherm IMC 18
Thermo Scientific
2
Delivery of the device Packaging The Heratherm Incubator IMC 18 is delivered in a rugged packaging box. All packaging materials can be separated and are reusable: Packaging materials Packaging carton:
Recycled paper
Foam elements:
Styrofoam (CFC-free)
Packaging film:
Polyethylene
Acceptance inspection After the device has been delivered, check the delivery immediately for: • completeness, • possible damage.
If components are missing or damage is found on the device or the packaging, in particular damage caused by humidity and/or water, please notify the carrier as well as Thermo Scientific Technical Support immediately.
Scope of supply
Thermo Scientific
Quantity of components supplied
Pieces
Perforated shelf
2
Power cord, version US/JP
1
Power cord, version EMEA/APAC
4
Mains adapter 12 V DC
1
Instruction Manual
1
Heratherm IMC 18
2-1
Delivery of the device Scope of supply
2-2
Heratherm IMC 18
Thermo Scientific
3
Installation Environmental conditions The Heratherm Incubator IMC 18 device must only be operated in a location that meets all of the specific ambient condition requirements listed below:
Location requirements • Cited in a dry area within a closed space which is free from draughts. • Dust loading should be in keeping with the degree of contamination, which is Degree 2 according to IEC EN 61010-1. Using the incubator in an atmosphere with electrically conductive dust is prohibited. • The minimal distance to adjacent surfaces must be observed on all sides (see Section 3.3). • The operating room must be equipped with appropriate ventilation. • Solid, level, fire-proof surface. • Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of the device and its accessories. • The electrical circuitry of the device has been designed for an operating height of up to 2000 m above sea level. • The permissible ambient temperature is between +18 °C and 32 °C (+64.4 °F and + 89.6 °F) for an incubation temperature of between 17 °C and 40 °C (62.6 °F and 104 °F). • Relative humidity up to 80% (maximum; preferably 60-70%), non condensing. • Dew is to be avoided, e.g. after changes of location or transportation. Should condensation exist, wait until the moisture has evaporated completely before connecting the device to a power source and powering up. • Avoid direct exposure to sunlight. • Devices that produce excessive amounts of heat must not be placed near the device. • Storage temperature: When the device is placed in intermediate storage, make sure that the ambient temperature is between 20 °C and 60 °C (68 °F and 140 °F) and the maximum relative humidity does not exceed 90%, non-condensing.
Thermo Scientific
Heratherm IMC 18
3-1
Installation Room ventilation
Room ventilation Heat dissipating from the device during continuous operation may cause a change in the room climate. Therefore, the device must only be installed in rooms with sufficient ventilation. Do not install the incubator in room recesses without ventilation. When several devices are to be placed in the same room, additional ventilation may have to be provided as necessary. The room ventilation is to be of a technical form which is designed in accordance with the thermal input.
Installation When installing the device, make sure that the installation and supply connections remain freely accessible. The minimum wall clearance is 5 cm / 2 in on the sides and back. No other devices are to be placed on the device.
The cooling fan on the back of the device is not to be covered. It is not to suck-in any loose items (e.g. paper).
Transport The device is not to be lifted by the door or attachments when being transported.
3-2
Heratherm IMC 18
Thermo Scientific
4
Description of unit Device overview Front view
Figure 4-1: Front view Heratherm Incubator IMC 18 [1] Temperature display [2] Control panel [3] Door/door handle [4] Perforated shelf
Thermo Scientific
Heratherm IMC 18
4-1
Description of unit Mains connection
Rear view .
Figure 4-2: Rear view Heratherm Incubator IMC 18 [1] Fan [2] Mains adapter
Mains connection The mains connection of the device is done by connecting the enclosed mains adapter to the 100 V or 240V a.c. supply; mains fluctuation +/-10%, 50/60 Hz
Useable space components The surface of the useable space of the Heratherm Incubators IMC 18 is reduced to a minimum and therefore supports the protection against contamination and its easy and efficient elimination. All of the parts of the inner container of the Heratherm Incubator IMC 18 are made of plastic (the racks are made of stainless steel) and have a smooth and easy to clean surface. A heat exchanger element [Fig. 4-3; 1] in the rear wall serves to heat or cool the useable space as required. The useable space lighting can be switched on using the button
on the control panel.
Perforated shelf The metal shelves have an anti-tipping device. The perforated shelf is described in detail in the chapter entitled Commissioning.
4-2
Heratherm IMC 18
Thermo Scientific
Description of unit Condensation tray
Condensation tray A removable tray [Fig. 4-3; 2] is provided underneath the device which collects the condensation which occurs whilst the useable space is in use. .
Figure 4-3: Cooling or heating elements and the condensation tray in the Heratherm Incubators IMC 18
Thermo Scientific
Heratherm IMC 18
4-3
Description of unit Condensation tray
4-4
Heratherm IMC 18
Thermo Scientific
5
Installation procedures Remove the packing foil.
Insert the perforated shelves Remove the transport securing devices from the perforated shelves. Insert the perforated shelves in the grooved profiles on the side walls of the useable space.
Preparing the useable space The Heratherm Incubator IMC 18 is not delivered in a sterile condition. The following useable space components should be checked for cleanliness and disinfected prior to use: • perforated shelves, • useable space surfaces, • door seal, • useable space door.
Mains connection Electrical shock Contact with live electrical components may cause a lethal electric shock. Before connecting the device to the power supply, check the power cord and the plug for damage. Do not use damaged cables for connecting the device to the power supply! Connect the device to the on site fused mains supply with the following values: • Safety fuse (time-lag) T 16 A or circuit breaker B 16. • Connection via a fault current circuit breaker (triggering current < 30 mA) is recommended.
Thermo Scientific
Heratherm IMC 18
5-1
Installation procedures Mains connection
Connection to the power supply source 1. Before connecting the device to the power source, check to see if the power supply voltage corresponds with the specifications on the nameplate on the rear of the device. If the voltage (V) and current (A) ratings given are not as required, do not connect the device to the power source! 2. Connect the IEC connector to the socket on the rear of the device. 3. Insert the earthed plug [3] of the mains cable in a correctly earthed and fused socket.
Power sockets
The socket which the device is connected to is to be fully accessible when operating the device so as to ensure that the plug can be removed should danger occur! 4. Make sure the power cord is not subjected to tensile or compressive force.
Condensation
The device should be in the room for a sufficiently long duration before it is switched on, in order to ensure that it can reach the ambient temperature. This prevents condensation forming on live components.
5-2
Heratherm IMC 18
Thermo Scientific
6
Operation Preparing the device The device must not be released for operation before all major start-up activities have been completed (see chapter Installation procedures).
Device Check Prior to starting operation, the following device components must be checked for their correct function: • The door seal in the front frame must not be damaged. • The perforated shelves have been correctly inserted.
Starting Operation 1. Switch the device on with button
.
2. The display shows the current useable space temperature. 3. Set the target values for the temperature using buttons
and
on the control panel.
4. Wait until the device has reached the set temperature. The temperature control regulates the temperature at the set target temperature value. 5. Load the useable space.
Proper loading
To ensure sufficient air circulation and uniform heating of the samples, do not use more than 70% of the maximum surface area of the useable space. Bulky objects in the useable space that dissipate heat may impair heat distribution.
Thermo Scientific
Heratherm IMC 18