Thermo Scientific
T-865 Fixed-Angle Ultracentrifuge 52266-15 Instruction Manual March 2010
Instruction Manual
35 Pages

Preview
Page 1
Thermo Fisher Scientific
T-865 Instruction Manual 52266-15
March 2010
© 2009 Thermo Fisher Scientific Inc. All rights reserved.
ULTRACRIMP< is either a registered trademark or a trademark of Thermo Fisher Scientific. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc. The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser. Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly. This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents. Release history: 52266-15 printed in March 2010.
For Research Use Only. Not for use in diagnostic procedures.
T Table of contents Thermo Scientific T-865 Fixed-Angle Ultracentrifuge Rotor ... ii Important Safety Information ... iii Chapter 1
DESCRIPTION ... 1-1 Rotor Description ... 1-2 Rotor Specifications ... 1-2 Accessories ... 1-2
Chapter 2
OPERATION ... 2-1 Prerun Safety Checks ... 2-2 Compartment Loads in Excess of Design Mass ... 2-2 Critical Speed ... 2-3 Rotor Precool ... 2-3 Relative Centrifugal Force (RCF) Determination ... 2-3 Calculation of Sedimentation Time in Aqueous (Non–gradient) Solutions ... 2-7 Chemical Compatibility ... 2-8 Rotor Balancing ... 2-8 Rotor Loading and Sealing ... 2-9 Centrifuge/Rotor Log Book ... 2-11
Chapter 3
CARE and MAINTENANCE ... 3-1 Corrosion ... 3-2 Cleaning ... 3-2 Storage ... 3-4 Overspeed Decal Replacement ... 3-4 Service Decontamination Policy ... 3-5
Chemical Compatibility Chart ...A-1 Contact Information ...B-1
Thermo Scientific
T-865
i
P
This manual is a guide for the use of
Thermo Scientific T-865 Fixed-Angle Ultracentrifuge Rotor Data herein has been verified and is believed adequate for the intended use of the rotor. Because failure to follow the recommendations set forth in this manual could produce personal injury or property damage, always follow the recommendations set forth herein. Thermo does not guarantee results and assumes no obligation for the performance of rotors or other products that are not used in accordance with the instructions provided. This publication is not a license to operate under, nor a recommendation to infringe upon, any process patents. Publications prior to the Issue Date of this manual may contain data in apparent conflict with that provided herein. Please consider all data in this manual to be the most current. NOTES, CAUTIONS, and WARNINGS within the text of this manual are used to emphasize important and critical instructions. WARNING informs the operator of a hazard or unsafe practice that could result in personal injury, affect the operator's health, or contaminate the environment. CAUTION informs the operator of an unsafe practice that could result in damage of equipment. NOTE highlights essential information.
CAUTION and WARNING are accompanied by a hazard symbol and appear near the information they correspond to.
© 1998, 1999, 2008, 2010 by Thermo Fisher Scientific
Thermo Scientific
T-865
ii
Preface
Important Safety Information Certain potentially dangerous conditions are inherent to the use of all centrifuge rotors. To ensure safe operation of this rotor, anyone using it should be aware of all safe practices and take all precautions described below and throughout this manual.
WARNING When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the materials and the hazards associated with them in the event leakage occurs during centrifugation. If leakage does occur, neither the centrifuge nor the rotor can protect you from particles dispersed in the air. To protect yourself, we recommend additional precautions be taken to prevent exposure to these materials, for example, use of controlled ventilation or isolation areas. Always be aware of the possibility of contamination when using radioactive, toxic, or pathogenic materials. Take all necessary precautions and use appropriate decontamination procedures if exposure occurs. Never use any material capable of producing flammable or explosive vapors or creating extreme exothermic reactions. Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor failure. Always reduce (derate) rotor speed as instructed in this manual whenever: • the rotor speed/temperature combination exceeds the solubility of the gradient material and causes it to precipitate. • the compartment load exceeds the maximum allowable compartment load specified (average fluid density greater than 1.2 g/ml). See page 2-2. Failure to reduce rotor speed under these conditions can cause rotor failure..
iii
T-865
Thermo Scientific
Preface
CAUTION Do not expose aluminum rotor components to: strong acids, bases, or alkaline laboratory detergents; liquid chlorine bleach; salts (chlorides) of heavy metals such as cesium, lead, silver, or mercury. Use of these material with aluminum can cause a chemical reaction that initiates corrosion. Do not operate or precool the rotor at the critical speed, as this will have a detrimental effect on centrifuge component life. See pages 2-2 and 2-3. Operate the rotor only when it is symmetrically balanced as described in this manual. Operating the rotor out of balance can cause damage to the centrifuge drive assembly. Rotors with covers must be operated with the cover in position and locked in place. Always maintain the rotor in the recommended manner. The rotor and all accessories must be clean and inspected prior to each run: do not use rotors showing signs of corrosion or cracking. See Chapter 3, Care and Maintenance. Do not autoclave the aluminum rotor cover or expose any aluminum rotor parts to temperatures in excess of 100°C.
Thermo Scientific
T-865
iv
1
DESCRIPTION This manual contains information required to operate and maintain the Thermo Fisher Scientific T-865 Fixed-Angle Ultracentrifuge Rotor.. If you require additional information regarding operation or maintenance, please contact Thermo Fisher Scientific for assistance. In the United States, call Thermo Fisher Scientific toll-free 1-866-9THERMO; outside the United States, contact the nearest Thermo Fisher Scientific office (see back cover) or your local representative for Thermo Fisher Scientific products. Thermo Fisher Scientific product information is available on our internet web site at http:// www.thermo.com.
Contents
• “Rotor Description” on page 1-2 • “Rotor Specifications” on page 1-2 • “Accessories” on page 1-2
Thermo Scientific
T-865
1-1
1 DESCRIPTION Rotor Description
Rotor Description The T-865 is an 8-place fixed-angle rotor that can be used at speeds up to 65 000 rpm.1 The rotor body is machined from a titanium forging for strength and corrosion resistance. The aluminum lid and locking nut are given a blue anodized finish for surface protection.The rotor is tightly sealed during operation by two O-rings - one that fits inside the rotor lid and another around the bottom of the locking nut. The eight tube compartments are bored at a 23.5° angle to the axis of rotation. A disc with alternative black and reflective segments on the bottom of the rotor provides overspeed protection.
Rotor Specifications Table 1-1. Rotor Specifications Rotor Type
Fixed Angle
Maximum Speed
65 000 rpm*
Relative Centrifugal Force (RCF) at Maximum Speed:
*
at rminimum (9.10 cm)
429 459
at raverage (6.47 cm)
305 340
at rmaximum (3.84 cm)
181 222
K Factor at Maximum Speed
51.7
Critical Speed
1200 rpm
Tube Size ( Maximum)
25 mm x 97 mm
Capacity per Tube Compartment (Nominal)
36 ml
Number of Tube Compartment
8
Total Rotor Capacity
288 ml
Maximum Tube Compartment Mass
66 g
Tube Angle
23.5°
Rotor Diameter
19.5 cm (7.75 inch)
Rotor Weight (empty)
8.3 kg (18.3 lb)
With tubes filled with a homogenous solution having an average density of 1.2 g/ml or less.
Accessories a. Accessories Supplied The accessories supplied with the T-865 Rotor, Catalog No. 54305 (rotor and complete accessories), are listed in Table 1-1. Those items indicated by an asterisk (*) are not supplied with Catalog No. 51411, which is a T-865 with basic accessories only. 1
Speed in revolutions per minute (rpm) is related to angular velocity, ω, according to the following:
2π ω = (rpm) ⎛⎝ ------⎞⎠ = ( rpm ) ( 0.10472 ) 60 Where ω = rad/s. All further references to speed in this manual will be designated as rpm.
1-2
T-865
Thermo Scientific
1 DESCRIPTION Accessories
Note In order to seal the ULTRACRIMP< tubes (supplied with Catalog No. 54305) you must have an ULTRACRIMP<Sealing Tool, Catalog No. 03920. To order replacement accessories, telephone 1-800-522-7746 in the United States. Outside the United States, contact the local representative, distributor or agent for Thermo Fisher Scientific products. Be sure to provide the catalog number and description of the part included in Table 1-1, plus the rotor model and serial number. Table 1-2.Accessories Supplied Quantity
*
Catalog Number *
Description
ULTRACRIMP< tubes, Polyallomer, 35 ml (Pkg/25)**
8
03989
8
03996*
1
*
03935
Tube Racks, 4-place (1 pair)
1
*
12826
ULTRACRIMP< Tube Removal Tool
1
51942
Rotor Stand
1
65937
Vacuum Grease
1
61556
Lubricant
1
51345
Overspeed Decal, 65 000 rpm, (extra)
2
64666
Rotor Cover O-Ring, Viton® (extra)
2
64667
Locking Nut O-Ring, Viton® (extra)
1
52392
Rotor Cover Tool
1
52384
Centrifuge/Rotor Log Book
1
52266
Instruction Manual
Rotor Cap
Not supplied with Catalog No. 51411 (basic accessories only).
**
Each package of tubes is supplied with an equal number of tube caps and tube plugs.
b. Optional Multipiece Sealing Assembly If you prefer, you can use the T-865 Rotor with thinwall polyallomer tubes and aluminum sealing cap assemblies. These sealing assemblies consist of a tube plug, tube plug cover, sealing nut, and postfill screw. When the assembly is tightened, the cover crimps the top of the tube to form the seal. To use these assemblies, you must purchase all items separately: the T-865 Rotor with basic accessories only, Catalog No. 51411; a Tool Kit, Catalog No. 52610; the tubes; and the sealing assembly components. Table 1-2 lists the catalog number and description of the tubes and sealing assembly components required, and Table 1-3 lists the parts supplied in the Tool Kit. If you have the tools listed, you need not purchase the kit; or, if necessary, any of the tools listed can be ordered separately by the catalog number given in the table.
Thermo Scientific
T-865
1-3
1 DESCRIPTION Accessories
Table 1-3.Optional Multipiece Sealing Assembly Components Catalog Number
Description
03141
Tubes, 36.0 ml, Polyallomer (Pkg/25)
52572
Sealing Cap Assembly (each); 8 required
Table 1-4.Tool Kit (Catalog No. 52610) for use with Multipiece Sealing Assembly Catalog Number
Description
51834
Tube Cap Vise
52013
Tube Removal Tool
66298
Tube Cap Wrench, 3/4 inch
Note A Delrin< adapter (Catalog No. 00399) is available that fits the tube compartments of the T-865 Rotor. With this adapter, the rotor can also be used with 11.5 ml tubes and 10 ml tubes or bottles to a maximum speed of 30 000 rpm. For ordering information, please see Thermo Fisher Scientific Rotor and Tube Guide on our internet web site at http:// www.thermo.com
c. Rotor Cover Removal Tool The rotor cover of the T-865 Rotor is designed to fit the rotor body tightly to ensure a positive seal; however, because of this tight fit, it is sometimes difficult to remove. If you have difficulty removing the rotor cover, a removal tool is available (Catalog No. 52652; complete instructions for use of the tool are supplied with it).
1-4
T-865
Thermo Scientific
2
OPERATION Contents
• “Prerun Safety Checks” on page 2-2 • “Compartment Loads in Excess of Design Mass” on page 2-2 • “Critical Speed” on page 2-3 • “Rotor Precool” on page 2-3 • “Relative Centrifugal Force (RCF) Determination” on page 2-3 • “Calculation of Sedimentation Time in Aqueous (Non–gradient) Solutions” on page 2-7 • “Chemical Compatibility” on page 2-8 • “Rotor Balancing” on page 2-8 • “Rotor Loading and Sealing” on page 2-9 • “Centrifuge/Rotor Log Book” on page 2-11
Thermo Scientific
T-865
2-1
2 OPERATION Prerun Safety Checks
Prerun Safety Checks To ensure safe performance of the rotor, before every run you should: a. read the Safety Information Page in the front of this manual. b. make sure each tube compartment is clean and that there is no sign of corrosion. c. be sure the rotor itself is clean and shows no sign of corrosion or cracking. Also, make sure there are no scratches or burrs around the rim of the rotor. d. check the centrifuge chamber and drive spindle to be sure that they are clean and free of scratches and burrs. e. verify that the proper overspeed decal is firmly attached to the bottom of the rotor; the decal should have 14 black segments (see page 3-4 for Overspeed Decal Replacement procedure). f.
check the chemical compatibility of all materials used (see Appendix A).
g. inspect the rotor cover O-rings for cracks, tears, or abrasions; replace if necessary. Be sure the O-rings are properly lubricated. h. make sure the rotor cover is on and properly tightened. i.
be sure that the proper environment has been selected for operation; for example, controlled ventilation or isolation, if required.
j.
check the top speed capability of the tube (or bottle) being used. CAUTION When using a tube or bottle assembly other than those supplied by Thermo Fisher Scientific, be sure to check the top speed capability; when in doubt, do a test run for the desired application. If using a Thermo Fisher Scientific tube (or bottle) assembly other than those supplied with the rotor, refer to the Product Guide for the maximum speed. Exceeding the top speed capability of the tube (or bottle) can result in its breakage.
Compartment Loads in Excess of Design Mass The maximum operating speed (65 000 rpm) is based on the recommended design mass that has been established for this rotor, representing the maximum mass that each tube compartment can contain at top speed. The total contents of each compartment (including specimen, tube, rotor cap, tube cap, and tube plug) should not exceed the recommended figure unless the rotor speed is reduced proportionately. (If using optional multipiece sealing assembly, include the weight of the complete assembly.) Strict adherence to the maximum allowable compartment mass or reduced speed is required to prevent rotor failure. Observe the WARNING on the Safety Information Page in the front of this manual. The design mass for each compartment of the T-865 Rotor is 66 grams at 65 000 rpm. This figure is based on the use of a thinwall polyallomer tube filled with a liquid at 1.2 specific gravity, plus all sealing assembly components. If the total compartment mass is greater than 66 grams, use the following formula to determine the reduced speed:
66 g Reduced Speed = 65000 -------------------------------Actual Weight
2-2
T-865
Thermo Scientific
2 OPERATION Critical Speed
Critical Speed The critical speed is that speed at which any rotor imbalance will produce a driving frequency equal to the resonant frequency of the rotating system (that is, the rotor and the centrifuge drive). At this speed, the rotor may produce large amplitude vibrations which can be felt in the centrifuge frame. Mass imbalance will contribute to increased vibration intensity at the critical speed. Avoid operating the rotor at the critical speed, which is approximately 1200 rpm for the T-865 Rotor. Operation at the critical speed will have a detrimental effect on centrifuge component life. Observe the CAUTION on the Safety Information Page in the front of this manual.
Rotor Precool If samples are routinely processed around 4°C or below, the rotor should be stored in a refrigerator or a cold room. If this is not possible, the rotor can be precooled in a Thermo Fisher Scientific Ultracentrifuge. Refer to the ultracentrifuge instruction manual for precooling directions. Be careful not to precool the rotor at its critical speed (read paragraph on page 2-2, Critical Speed).
Relative Centrifugal Force (RCF) Determination Relative centrifugal force (RCF) refers to the force during centrifugation that moves the particulate outward from the center of rotation. This force is proportional to the radial distance and the square of the rotor speed. The RCF value, also known as g force, is determined by the following formula:
rpm RCF = 11.17(r) ⎛⎝ ------------⎞⎠ 1000 when r
2
= the radius in centimeters from the centerline of the rotor to the point in the tube where RCF value is required
and rpm = the rotor speed in revolutions per minute Figure 2-1 shows the minimum, average, and maximum radii of the T-865. Table 2-1 gives the RCF value at each radius at speeds from 20 000 rpm to 65 000 rpm (in increments of 500 rpm). The RCF value at any other speed can be calculated by using the above formula. Note The radii values given are the actual rotor specifications. These values do not take the thickness of the tube into consideration.
Thermo Scientific
T-865
2-3
2 OPERATION Relative Centrifugal Force (RCF) Determination
Table 2-5. T-865 Rotor: RCF Values and K Factors* rmax. 9.10 cm
RCF ravg. 6.47cm
rmin. 3.84 cm
20 000
40 659
28 908
17 157
546
20 500
42 717
30 371
18 026
519
21 000
44 826
31 871
18 916
495
21 500
46 986
33 407
19 827
472
22 000
49 197
34 979
20 760
451
22 500
51 459
36 587
21 714
431
23 000
53 771
38 231
22 690
413
23 500
56 135
39 911
23 688
395
24 000
58 549
41 627
24 706
379
24 500
61 014
43 380
25 746
364
25 000
63 529
45 169
26 808
349
25 500
66 096
46 994
27 891
336
26 000
68 713
48 854
28 996
323
26 500
71 382
50 752
30 121
311
27 000
74 101
52 685
31 269
299
27 500
76 871
54 654
32 438
289
28 000
79 691
56 660
33 628
278
28 500
82 563
58 701
34 840
269
29 000
85 485
60 779
36 073
260
29 500
88 458
62 893
37 327
251
30 000
91 482
65 043
38 604
243
30 500
94 557
67 229
39 901
235
31 000
97 683
69 451
41 220
227
31 500
100 859
71 710
42 560
220
32 000
104 087
74 004
43 922
213
32 500
107 365
76 335
45 306
207
33 000
110 694
78 702
46 710
200
33 500
114 073
81 105
48 136
195
34 000
117 504
83 544
49 584
189
34 500
120 985
86 019
51 053
183
35 000
124 518
88 531
52 544
178
35 500
128 101
91 078
54 056
173
36 000
131 735
93 662
55 589
168
36 500
135 419
96 282
57 144
164
Speed (rpm)
2-4
T-865
K Factors
Thermo Scientific
2 OPERATION Relative Centrifugal Force (RCF) Determination Table 2-5. T-865 Rotor: RCF Values and K Factors* rmax. 9.10 cm
RCF ravg. 6.47cm
rmin. 3.84 cm
37 000
139 155
98 937
58 720
159
37 500
142 941
101 630
60 318
155
38 000
146 778
104 358
61 937
151
38 500
150 666
107 122
63 578
147
39 000
154 605
109 923
65 240
144
39 500
158 595
112 759
66 923
140
40 000
162 635
115 632
68 628
136
40 500
166 726
118 541
70 355
133
41 000
170 869
121 486
72 103
130
41 500
175 062
124 467
73 872
127
42 000
179 305
127 484
75 663
124
42 500
183 600
130 538
77 475
121
43 000
187 945
133 627
79 309
118
43 500
192 342
136 753
81 164
115
44 000
196 788
139 914
83 040
113
44 500
201 286
143 112
84 938
110
45 000
205 835
146 347
86 858
108
45 500
210 435
149 617
88 799
105
46 000
215 085
152 923
90 761
103
46 500
219 786
156 266
92 745
101
47 000
224 538
159 644
94 750
98.8
47 500
229 341
163 059
96 777
96.7
48 000
234 195
166 510
98 825
94.7
48 500
239 099
169 997
100 895
92.8
49 000
244 054
173 520
102 986
90.9
49 500
249 061
177 079
105 098
89.1
50 000
254 118
180 675
107 232
87.3
50 500
259 225
184 306
109 387
85.6
51 000
264 384
187 974
111 564
83.9
51 500
269 593
191 678
113 762
82.3
52 000
274 853
195 418
115 982
80.7
52 500
280 165
199 194
118 223
79.2
53 000
285 526
203 006
120 486
77.7
53 500
290 939
206 855
122 770
76.3
54 000
296 403
210 739
125 075
74.9
Speed (rpm)
Thermo Scientific
K Factors
T-865
2-5
2 OPERATION Relative Centrifugal Force (RCF) Determination
Table 2-5. T-865 Rotor: RCF Values and K Factors* rmax. 9.10 cm
RCF ravg. 6.47cm
rmin. 3.84 cm
54 500
301 917
214 660
127 402
73.5
55 000
307 482
218 616
129 751
72.2
55 500
313 098
222 609
132 121
70.9
56 000
318 765
226 638
134 512
69.6
56 500
324 483
230 704
136 925
68.4
57 000
330 251
234 805
139 359
67.2
57 500
336 070
238 942
141 814
66.0
58 000
341 941
243 116
144 291
64.9
58 500
347 861
247 326
146 790
63.8
59 000
353 833
251 572
149 310
62.7
59 500
359 856
255 854
151 851
61.7
60 000
365 929
260 172
154 414
60.6
60 500
372 053
264 526
156 998
59.6
61 000
378 228
268 916
159 604
58.7
61 500
384 454
273 343
162 231
57.7
62 000
390 731
277 805
164 880
56.8
62 500
397 059
282 304
167 550
55.9
63 000
403 437
286 839
170 242
55.0
63 500
409 866
291 410
172 954
54.1
64 000
416 346
296 018
175 689
53.3
64 500
422 877
300 661
178 445
52.5
65 000
429 459
305 340
181 222
51.7
Speed (rpm)
*
2-6
T-865
K Factors
These values do not take the thickness of the tube into consideration.
Thermo Scientific
2 OPERATION Calculation of Sedimentation Time in Aqueous (Non–gradient) Solutions
Figure 2-1. T-865 Rotor Radii
Calculation of Sedimentation Time in Aqueous (Non–gradient) Solutions The time required to sediment a particle in water at 20°C through the maximum rotor path length (that is, the distance between rminimum and rmaximum) can be calculated using the equation:
K t = ------------S 20 ,w where: t = sedimentation time in hours K = the clearing factor for the rotor (defined on the next page) the sedimentation coefficient for the particle of interest in water at 20°C as expressed in S20, w = Svedbergs1 1
The sedimentation coefficient (S) in seconds, for a particle in a centrifugal field is defined by the equation S = (dx/dt) [1/(ω2x)]; where dx/dt = sedimentation velocity of the particle in cm/s; ω = rotor speed in rad/s; and x = the distance of the particle from the axis of rotation in centimeters. Conventionally, experimentally determined values of sedimentation coefficients are multiplied by 1013 to convert them to Svedberg units (S), so a particle with an experimentally determined sedimentation coefficient of 10-11 seconds is usually referred to in the literature as a "100 S particle." Since the value determined for the sedimentation coefficient is dependent on the density and viscosity of the solution in which centrifugation is performed, values are usually reported for the standard conditions of infinite dilution in water at 20°C, and designated S20, w.
Thermo Scientific
T-865
2-7
2 OPERATION Chemical Compatibility
The clearing (or K) factor is defined by the equation:
r maximum rotor speed 2 K = ( 253000 ) In ⎛ --------------------⎞ ÷ ⎛ -------------------------⎞ ⎝ r minimum⎠ ⎝ 1000 ⎠ Where rmaximum and rminimum are the maximum and minimum rotor radii, respectively, and rotor speed is expressed in rpm. K factors for the T-865 Rotor, at speeds from 20 000 rpm to 65 000 rpm (in increments of 500 rpm), have been listed in Table 2-1. Example: The T-865 Rotor has a K factor of 51.7 at the maximum permitted speed (65 000 rpm). If the particles to be sedimented have a sedimentation coefficient of 10S, the estimated run time required at maximum speed will be:
51.7 t = ---------- 5.17 hours = 5 hours, 10 minutes 10S Note that the calculation assumes particles in water at 20°C. If the suspending medium is denser or more viscous than water, the sedimentation time will be greater.
Chemical Compatibility The critical components of the T-865 Rotor apt to come in contact with solution are: rotor body (titanium), rotor cover assembly (aluminum), rotor cap and tube cap (aluminum), tube plug (Buna N), O-rings (Viton<), plus the material of the tubes (or bottles) being used. The chemical compatibility of rotor elements and accessory materials is given in the Appendix. Because no organized chemical resistance data exists for materials under the stress of centrifugation, this data is intended to be used only as a guide. When in doubt, we recommend pretesting of sample lots.
Rotor Balancing Always balance the rotor according to the following criteria: a. observe the CAUTION on the Safety Information Page in the front of this manual; b. balance pairs of tubes containing fluid of identical specific gravity to within 0.5 gram and place them in opposite tube compartments; c. when using less than a full complement of eight tubes, the rotor can be operated at its maximum allowable speed with two, four, or six samples, provided opposing pairs are positioned as shown in figure 2-2; and d. if one, three, five, or seven samples are to be run, balance the load as above with a dummy tube that contains a solution of the same specific gravity as the sample in the opposing tube compartment.
2-8
T-865
Thermo Scientific
2 OPERATION Rotor Loading and Sealing
Figure 2-2. Compartment Loading
Rotor Loading and Sealing Prepare the ULTRACRIMP® tubes for use by following the tube filling and tube sealing procedures given in the ULTRACRIMP® Tube Sealing System Instruction Manual. Note If using the optional tubes and multipiece sealing assemblies (Catalog No. 03141 and 52572 respectively), seal the tubes and load them in the rotor according to the instructions supplied with the sealing cap assemblies. Load and seal the rotor as follows: 1. Gently place the filled and sealed ULTRACRIMP< tubes in the tube compartments, balancing the rotor as described in previous paragraph. 2. Place a rotor cap (Catalog No. 03996) in each tube compartment that contains a tube. Push the rotor cap into the tube compartment until it is properly seated; when properly seated, the edge of the cap will be approximately flush with the top of the outer edge of the tube compartment. See figure 2-3. CAUTION Do not put a rotor cap in an empty tube compartment. Rotor caps should only be installed in compartments that contain a filled tube.
Thermo Scientific
T-865
2-9
2 OPERATION Rotor Loading and Sealing
Figure 2-3. Properly Seated Rotor Cap 3. Check the locking nut O-ring and rotor cover O-ring for scuffs, cracks or breaks. Replace, if necessary (see Table 1-1 for replacement part catalog numbers). 4. Coat O-rings lightly with vacuum grease, Catalog No. 65937. 5. Check that the locking nut thread is lightly lubricated with grease, Catalog No. 61556. 6. Place the rotor cover (with O-ring) on the rotor body. Finger tighten the locking nut by turning it clockwise. Insert the rotor cover tool through locking nut and continue turning the locking nut clockwise until rotor cover is tightened securely. Remove the rotor cover tool from the locking nut. CAUTION The rotor cover tool must be removed prior to centrifuge run.
7. Carry the rotor carefully, and lower it into the centrifuge chamber slowly and vertically. Be sure it snaps in place on the drive adapter. Check that the rotor is properly seated by pulling it gently in an upward direction and noting a small amount of resistance. Perform the run as explained in the centrifuge instruction manual. The rotor cover is designed to fit the rotor body tightly to ensure a positive seal; however, because of this tight fit, it is sometimes difficult to remove. If you have difficulty removing the cover, a tool is available that is used to easily remove the cover. The catalog number of the rotor cover removal tool is 52652; complete instructions for use of the tool are supplied with it.
2-10
T-865
Thermo Scientific