Service Manual
44 Pages
Preview
Page 1
Service Manual
RapidVac
TM
Smoke Evacuator
Part Number: 1075303
Preface This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed. It is intended as a guide for servicing the RapidVac Smoke Evacuator only. Additional users information is available in the RapidVac Smoke Evacuator User’s Guide. Equipment covered in this manual: RapidVac Smoke Evacuator-110 V and 220 V Device is compliant with the European Communities Council Directive 93/42/EEC, Medical Device Directive.
Conventions Used in this Guide Warning Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Caution Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury. Notice Indicates a hazard which may result in product damage. Important Indicates an operating tip or maintenance suggestion.
ii
RapidVac Smoke Evacuator Service Manual
Limited Warranty Covidien warrants each covered product listed below to be free from defects in material and workmanship for normal use and service for the period(s) set forth below. Covidien’s obligation under this warranty is limited to the repair or replacement, at its sole option, of any product, or part thereof, which has been returned to it (or its authorized distributor) within the applicable time period shown below after delivery of the product to the original purchaser, and which examination discloses, to Covidien’s satisfaction, that the product is defective. This limited warranty does not apply to any product, or part thereof, which has been repaired or altered in a way so as, in Covidien’s judgment, to affect its stability or reliability, or which has been subjected to misuse, neglect, or accident. The warranty periods for Covidien products are as follows: RapidVacTM Smoke Evacuator
One year from date of shipment
Notwithstanding any other provision herein or in any other document or communication, Covidien’s liability with respect to this limited warranty and the products sold hereunder shall be limited to the aggregate purchase price for the products sold to the customer. This limited warranty is non-transferable and runs only to the original purchaser of the covered product(s). There are no warranties which extend beyond the terms hereof. Covidien disclaims any liability hereunder or elsewhere in connection with the sale of products and for any form of indirect, tort, or consequential damages. This limited warranty and the rights and obligations hereunder shall be construed under and governed by the laws of the State of Colorado, USA. The sole forum for resolving disputes arising under or relating in any way to this limited warranty is the District Court of the County of Boulder, State of Colorado, USA. Covidien reserves the right to make changes in covered products built or sold by it at any time without incurring any obligation to make the same or similar changes to equipment previously built or sold by it. THE OBLIGATION TO REPAIR OR REPLACE A DEFECTIVE OR NONPERFORMING PRODUCT IS THE SOLE REMEDY OF THE CUSTOMER UNDER THIS LIMITED WARRANTY. EXCEPT AS EXPRESSLY PROVIDED HEREIN, COVIDIEN DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES, WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
RapidVac Smoke Evacuator Service Manual
iii
Table of Contents Preface... ii Conventions Used in this Guide... ii Limited Warranty... iii Table of Contents... v
Chapter 1. Introduction Parts Shipped with the Smoke Evacuator... 1-2 About the RapidVac Smoke Evacuator... 1-2 Features... 1-3 RapidVac+ System... 1-3 Operating Modes... 1-3 Service Personnel Safety... 1-4 General... 1-4 Maintenance... 1-6
Chapter 2. Controls, Indicators, and Receptacles Front Panel... 2-2 Power On/Off Switch... 2-3 Airflow Controls... 2-3 Turbo Button... 2-3 Filter-Life Indicator... 2-4 RapidVac+ Button... 2-4 Footswitch Jack... 2-4 Service-Required Indicator... 2-4 Rear Panel... 2-5 Power-Cord Receptacle... 2-5 Remote Activator Jack... 2-5 Equipotential Grounding Lug... 2-6
Chapter 3. Technical Specifications Performance Characteristics... 3-2 Dimensions and Weight... 3-2 Filter... 3-2 Airflow... 3-2 Safety... 3-2 Power... 3-3 Operating Parameters... 3-3 Transport and Storage... 3-3 EMC for the RapidVac Smoke Evacuator... 3-4 Electromagnetic Compatibility (IEC 60601-1-2)... 3-4
RapidVac Smoke Evacuator Service Manual
v
Summary of EMC Immunity Test Results... 3-4 FCC Compatibility... 3-6 Description of Symbols... 3-7 Front Panel... 3-7 Rear Panel... 3-8 Standards and IEC Classifications... 3-9
Chapter 4. Setup and Tests Cleaning the Smoke Evacuator... 4-2 Testing the Smoke Evacuator... 4-3 Checking the Remaining Filter Life... 4-3 Changing the Filter... 4-4 Testing the Airflow Controls... 4-4 Installing and Testing the Footswitch (optional)... 4-5 Installing and Testing the Remote Switch Activator (optional)... 4-5 Installing and Testing the RapidVac Interlink Cable (optional)... 4-6
Chapter 5. Troubleshooting Inspecting the RapidVac Smoke Evacuator... 5-2 Correcting Specific Malfunctions... 5-2
Chapter 6. Return Policy and Procedure Responsibility of the Manufacturer... 6-2 Obtaining a Return Authorization Number... 6-2 Returning the Smoke Evacuator for Service... 6-3 Technical Service... 6-3
Chapter 7. RapidVac Schematic
vi
RapidVac Smoke Evacuator Service Manual
C h a p te r 1 Introduction This chapter provides an overview of the features and functions of the RapidVac Smoke Evacuator. Caution Read all warnings, cautions, and instructions provided with this system before use. Read the instructions, warnings, and cautions provided with smoke evacuator accessories before use. Specific instructions are not included in this manual.
RapidVac Smoke Evacuator Service Manual
1-1
Parts Shipped with the Smoke Evacuator
Parts Shipped with the Smoke Evacuator When unpacking the smoke evacuator, verify that the following parts are included: • Power cord • User’s guide and service manual Upon initial receipt, inspect the smoke evacuator for dents, cracks, or damage that may have occurred during shipment. If damage is present, notify the freight carrier for assessment. Additional items available from Covidien: • Replacement RapidVac Filters • RapidVac Footswitch • RapidVac Remote Switch Activator • RapidVac Interlink Cable • Laparoscopic Smoke Evacuator Tubing Kit • Hoses, tubing, adapters, wands and other accessories • Smoke pencil
About the RapidVac Smoke Evacuator Indications for use for the RapidVac Smoke Evacuator include: To remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures. The RapidVac Smoke Evacuator is designed with a vacuum motor. The motor is used to draw surgical smoke from the surgical site through the vacuum tubing and into the filter where the surgical smoke is processed by a series of filters. A single disposable filter is used to simplify the installation and removal during filter changes. The filter is completely enclosed to protect the healthcare personnel from potential contamination during filter changes. One RapidVac SEA3700 filter contains four different stages within to capture the smoke plume: • The first stage filtration is a prefilter whose function is to trap and remove gross particulate. • The second stage filtration is ULPA (Ultra Low Penetration Air) grade filter whose high-tech design captures particulates and micro-organisms from .1 to .2 microns at an efficiency of 99.999%. • The third stage filtration is comprise of virgin activated carbon. • The fourth stage filtration is a woven fiberglass filtration media used to reduce the amount of activated carbon fines from migrating out of the filter.
1-2
RapidVac Smoke Evacuator Service Manual
About the RapidVac Smoke Evacuator
RapidVac Smoke Evacuator features include: • Filter-life monitoring for its four-stage, 25-hour filter
Introduction
Features
• Standby, RapidVac+, footswitch, or continuous operation modes • Adjustable flow • Compatibility with all Covidien electrosurgery pencils and generators
RapidVac+ System The RapidVac Smoke Evacuator incorporates the RapidVac+ mode, which controls the flow of the smoke evacuator when using a Covidien electrosurgical pencil. When the electrosurgical pencil is activated, the smoke evacuator operates in the preselected flow setting. When the pencil is deactivated, the airflow reduces to a low-flow purge setting. Two Covidien accessories link the electrosurgical generator and smoke evacuator to enable the RapidVac+ mode: • RapidVac Remote Switch Activator • RapidVac Interlink Cable
Operating Modes Four operating modes are available: standby, footswitch, RapidVac+, and continuous. • Standby mode powers up the unit without activating the motor or vacuum flow. • Footswitch mode is used to turn the smoke evacuator on or off with the optional footswitch pedal. • RapidVac+ mode is used to control the vacuum flow simultaneously with activation of an electrosurgical pencil. This mode is controlled with the optional interlink cable or electrosurgery sensor. • Continuous mode is used to operate the smoke evacuator using continuous airflow.
RapidVac Smoke Evacuator Service Manual
1-3
Service Personnel Safety
Service Personnel Safety General Warning During initial set up, inspect the smoke evacuator for any damage that may have been caused during shipping. If damaged, do not use or attempt to repair. Call Covidien for service assistance. The RapidVac Smoke Evacuator filter is a disposable component that captures potentially hazardous particles. Handle used filters as you would any biohazardous material. Dispose of used filters according to your local codes and regulations, and follow your institution’s procedures for disposal. The smoke evacuation filter has a life of approximately 25 hours of use and should not be used beyond the specified time. If the replace-filter indicator illuminates during operation, the smoke evacuator will continue to run until it is turned off. Turning power off to the smoke evacuator will require replacement of the filter before subsequent use. Failure to change the filter accordingly may result in decreased efficiency in smoke evacuation and contamination of the motor, and sound absorbing media within the smoke evacuator. The RapidVac Smoke Evacuator filter is disposable. Dispose of used filters according to your local codes and regulations, and follow your institution’s procedures for disposal. Do not operate this device in potentially explosive environments, such as in the presence of flammable anesthetics. Refer routine servicing to qualified biomedical technical personnel. Use only with the power cord provided and always plug into a grounded outlet. Do not use two- or three-prong adapters with the smoke evacuator power cords. The power-cord assembly should be checked periodically for damaged insulation or connectors. Do not use damaged cords. Use of extension cords may result in fire hazards. This device is not intended for the evacuation of fluid. If fluid is expected to be aspirated to the smoke evacuation filter, a fluid-collection device must be installed with the vacuum hose assembly. Failure to install a fluid-collection device could cause filter blockage and electrical damage. Contact Covidien Service for additional information. Do not connect a wet power cord to the wall receptacle. Connect the smoke evacuator power cord to a properly grounded receptacle. Plug the power cord directly into the power receptacle without any adapter plugs. Use of power plug adapters may result in electric shock. If the smoke evacuator becomes wet, either from a leaking tube or from being sprayed, unplug the smoke evacuator from the Mains outlet. Wipe dry or allow to air dry before proceeding.
1-4
RapidVac Smoke Evacuator Service Manual
Service Personnel Safety
Do not remove any covers or panels exposing the internal components of the smoke evacuator. Refer to a Covidien representative for service. The smoke evacuator produces a strong vacuum. Adjust the airflow and the position of the inlet end of the wand or tubing to prevent patient injury and to prevent suction of surgical materials and surgical specimens.
Introduction
Warning
If the smoke evacuator is activated while the airflow is set to a high speed, it may produce a sudden, strong suction action. Check the airflow setting before activating the smoke evacuator to prevent patient injury and to prevent suction of surgical materials and surgical specimens.
Caution Do not stack equipment on top of the smoke evacuator or place the smoke evacuator on top of electrical equipment. These configurations are unstable and/or do not allow for adequate cooling. Use only RapidVac SEA3700 ULPA Filters in the RapidVac Smoke Evacuator. Filters from other manufacturers may cause damage to the system, thereby voiding the warranty. Provide as much distance as possible between the smoke evacuator and other electronic equipment (such as monitors). An activated smoke evacuator may cause interference with them. Read all warnings, cautions, and instructions provided with the RapidVac Smoke Evacuator before using. To maximize patient safety, the tubing or wand should not come in direct contact with tissue. Otherwise, patient injury may result. Do not block either the tubing or the filter. If either becomes occluded or significantly restricted, the motor/blower may overheat and cause the unit to shut down. Care must be exercised in the installation of hoses, adapters and suction canisters. Failure to follow the procedures outlined in this manual may result in overheating of the motor and may void the unit warranty.
Notice Using any unapproved filter or accessory with the smoke evacuator may cause damage and will void the warranty. The RapidVac Smoke Evacuator functions properly only when the ULPA filter is installed properly. If the filter is installed incorrectly, the amber replace-filter indicator flashes and the Service indicator is illuminated. Fluids may damage the filters. When evacuating smoke and incidental fluids, always use the appropriate tubing, Connect the power cord to a wall receptacle having the correct voltage. Otherwise, product damage will result.
RapidVac Smoke Evacuator Service Manual
1-5
Service Personnel Safety
Notice Power line voltage below 100 VAC (110 V SEA3690) and 200 VAC (220 V SEA3695) will significantly reduce airflow. Connecting multiple lengths of tubing together may cause the smoke evacuator to overheat. The RapidVac Smoke Evacuator has been specially designed to fit on Covidien mounting carts only. Do not install the smoke evacuator on a cart other than a Covidien cart. Installing the smoke evacuator inside a cart with improper ventilation may result in overheating or may adversely affect the stability of the cart. Periodically check the cart brackets that hold the smoke evacuator in place to ensure that the screws are securely fastened. Stacking the smoke evacuator on top of a generator may cause an unstable condition and/or cause the smoke evacuator to overheat. Do not use sterile accessory products if sterile packaging has been compromised.
Maintenance Warning Always turn off and unplug the smoke evacuator before cleaning. The ULPA filter captures potentially hazardous particles. Handle used filters as you would any biohazard material. Dispose of filters with other operative waste materials according to the procedures for your institution. Do not reuse or re-sterilize smoke evacuator accessories labeled “disposable,” “single use only,” or “sterile.”
Notice Do not rub, touch, or clean the smoke evacuator with alcohol or caustic, corrosive, or abrasive cleaning or disinfectant compounds, solvents, or other materials that could scratch the control panel or damage the smoke evacuator. Do not autoclave, pressure sterilize, or gas sterilize the smoke evacuator. Keep the smoke evacuator away from liquids. Liquids that enter the smoke evacuator will damage internal components.
1-6
RapidVac Smoke Evacuator Service Manual
C h a p te r 2 Controls, Indicators, and Receptacles This chapter describes the front and rear panels of the RapidVac Smoke Evacuator Service Manual smoke evacuator, including all controls, indicators, receptacles, and ports.
RapidVac Smoke Evacuator Service Manual
2-1
Front Panel
Front Panel
+ -
1/4" 6 mm
3/8"
10 mm 7/8"
22 mm
Power on/off switch Airflow indicators (LED) Turbo button Airflow controls Power-on indicator (LED centered in fan symbol) Filter-life indicators (LED) RapidVac+ button Service-required indicator Footswitch jack Air filter with tubing connectors
2-2
RapidVac Smoke Evacuator Service Manual
Front Panel
Caution Connect accessories to the proper receptacle type. Otherwise, the smoke evacuator may not function properly.
The electronic system controls on the RapidVac Smoke Evacuator are easy to understand and simple to use. The control panel contains the suction on/off switch, suction power adjustment, filter-life indicator, and service indicator light. Notice
Power On/Off Switch The power switch that controls power to the RapidVac Smoke Evacuator is in the lower-left corner of the front panel. To power up the smoke evacuator, place the power switch in the on position (l). The smoke evacuator starts in standby (system power is on with no suction). When the smoke evacuator is on in standby, the LED in the center of the fan symbol on the front panel will blink continually until a fan speed is selected.
Controls, Indicators, and Receptacles
Read all instructions before installing accessories or operating this equipment. Failure to do so may result in damage to the unit and/or personal injury.
Turn the system power off by placing the power switch in the off position (O).
Airflow Controls The level of suction is controlled by the airflow controls: up-arrow ( ) and down-arrow ( ) buttons. To start the fan from standby mode, press the up-arrow button. It starts in the lowest suction setting indicated by a single illuminated airflow indicator. The suction control should be set at the lowest practical setting to completely remove the surgical smoke from the operative site. Pressing either the up- or down-arrow buttons changes the current suction. The selected level of suction is displayed by the airflow-indicator LEDs.
Turbo Button When increased smoke removal is necessary, the Turbo button may be pressed at anytime to increase suction. Turbo mode effectively increases the airflow to 100% suction flow to quickly remove the smoke plume from the surgical site.
RapidVac Smoke Evacuator Service Manual
2-3
Front Panel
Filter-Life Indicator The filter-life indicator provides a visual indication of the status of the life of the filter in use. Each illuminated LED represents five hours of remaining time on the 25-hour filter. The RapidVac Smoke Evacuator tracks the time each filter has been used allowing for an accurate display of remaining use for new or previously used filters.
Reading the Filter-Life Indicator Install an unused RapidVac filter into the system. When the smoke evacuator is on, all five of the filter-life indicator LEDs illuminate indicating the new filter has 100% filter life. The indicator will regress through subsequent LEDs until the final LED is illuminated indicating less than 5 hours of filter life remaining. At one hour remaining, the LED changes to amber. When the time remaining reaches 30 minutes, the amber LED flashes slowly.
RapidVac+ Button When the smoke evacuator is connected to a generator using the RapidVac Interlink Cable, pressing the RapidVac+ button synchronizes the smoke evacuator with the use of an electrosurgical pencil. When a pencil is activated, the smoke evacuator activates to a preselected airflow level.
Footswitch Jack A RapidVac Footswitch (SEA3745) is available for use with the RapidVac Smoke Evacuator II. A footswitch may be added to any system by simply plugging it into the footswitch jack on the front panel. The unit may be turned on or off by depressing the footswitch pedal once for each operation. For footswitching operation, the smoke evacuator should be on and the suction level to be activated from the footswitch should be selected on the control panel.
Service-Required Indicator The motor powering the RapidVac Smoke Evacuator requires regular maintenance every 500 hours. When the smoke evacuator has been in use for 500 hours, the service-required Indicator illuminates. Contact the Covidien Customer Service Department and return the smoke evacuator for service.
2-4
RapidVac Smoke Evacuator Service Manual
Rear Panel
Rear Panel 1
1
3
5
Controls, Indicators, and Receptacles
4 2
Cord brackets Carrying handle Power-cord receptacle Remote Activator jack Equipotential grounding lug
Power-Cord Receptacle The cord receptacle provides power to the smoke evacuator using a removable cord appropriate for regional power supplies. Cord can be wound on the cord brackets to store with the smoke evacuator.
Remote Activator Jack A RapidVac Interlink Cable (SEA3730) and Remote Switch Activator (SEA3741) are available for the automatic remote activation of the RapidVac Smoke Evacuator. The RapidVac Interlink Cable is compatible with the Covidien Force FXTM, Force EZTM and ForceTriadTM generators. The cable attaches to a generator’s RF activation port and to the smoke evacuator’s Remote Activator jack. When the generator is activated, the smoke evacuator automatically initiates the pre-set suction. The Remote Switch Activator (RSA) is programmed to recognize cut and coag activation currents emitted from the generator. The recognized current initiates activation of the smoke evacuator.
RapidVac Smoke Evacuator Service Manual
2-5
Rear Panel
Equipotential Grounding Lug The equipotential grounding lug provides an alternate route for electrical energy back to ground when attached to an earth-ground cable.
2-6
RapidVac Smoke Evacuator Service Manual
C h a p te r 3 Technical Specifications All specifications are nominal and subject to change without notice.
RapidVac Smoke Evacuator Service Manual
3-1
Performance Characteristics
Performance Characteristics Dimensions and Weight Width
14” (36 cm)
Depth
18.75” (48 cm) with cord wrap and handle
Height
8” (20.5 cm)
Weight
20.5 lb (9.9 kg) without filter
Filter ULPA filter
0.1-0.2 micron particulate size at 99.999% efficiency; activated charcoal
Airflow 0 - 1.25 m3/min. (0 - 44 cfm) with 7/8” x 10’ (2.2 cm x 3 m) tubing with sponge guard. Airflow measured at sea level.
Safety
3-2
110 Volt
220 Volt
Fuse: 10 AMPS, 250 V (Slo-Blo™*)
Fuse: 8 AMP, 250 V (Slo-Blo)
Power cord: 3-prong hospital grade plug
Power cord: 3-prong locally approved connector
Leakage current: (50/60 Hz)
Leakage current: (50/60 Hz)
Normal polarity, smoke evacuator in standby mode < 300 µA
Normal polarity, smoke evacuator in standby mode < 300 µA
Reverse polarity, smoke evacuator in standby mode < 300 µA
Reverse polarity, smoke evacuator in standby mode < 300 µA
RapidVac Smoke Evacuator Service Manual
Performance Characteristics
Power This unit is equipped from the factory with either a 110 VAC hospital grade NEMA 5-15 power cord or a 220 VAC CEE7/7 power cord. Should the AC power cord need replacing to match another plug configuration, the replacement plug/cable/receptacle configuration must meet or exceed the following specifications.
100 - 120 VAC Power Cord - SJT, No. 16AWG, 3 Conductor, IEC Female, 125VAC-13A, UL/IEC Approvals
220 - 240 VAC Power Cord - HOVV-F3G1.00, IEC Female, 250VAC-10A, EU Approvals
Operating Parameters Ambient temperature range
50 °F to 104 °F (10 °C to 40 °C)
Relative humidity
30% to 75%, noncondensing
Ambient temperature range
-29 °F to 149 °F (-34 °C to 65 °C)
Relative humidity
25% to 85%, noncondensing
RapidVac Smoke Evacuator Service Manual
Technical Specifications
Transport and Storage
3-3