CLEANING AND STERILIZATION GUIDE†
iDrive™ Ultra Powered Stapling System
MANUAL 1.
Wipe down the components with a lint-free cloth soaked in a pH-neutral detergent solution.
The recommended automatic cleaning cycle is as follows1,‡: Treatment
Time Temperature (MM:SS)
Chemical
00:45
Cold Tap Water
N/A
2.
Scrub all exposed surfaces with a soft brush to remove surgical debris, as required.
Pre-Wash
3.
Immerse the components in a warm enzymatic bath diluted per the manufacturer’s specifications and soak for ten minutes.
Enzyme Wash
04:00
Hot Tap Water
Enzymatic Cleaner (Prolystica™* 2X concentrate enzymatic cleaner or equivalent)
Rinse
00:15
Hot Tap Water
N/A
4.
Inspect the components for cleanliness.
5.
Manually agitate the devices in the enzymatic bath for a minimum of one minute.
6.
Rinse under warm running tap water for a minimum of one minute.
7.
Perform a final rinse under purified water for a minimum of one minute.
8.
Dry with a clean, soft, lint-free cloth.
AUTOMATIC 1.
Wipe down the components with a lint-free cloth soaked in a pH-neutral detergent solution.
2.
Scrub all exposed surfaces with a soft brush to remove surgical debris, as required.
3.
Rinse under warm running tap water for a minimum of one minute.
Wash
03:00
Hot Tap Water
pH Neutral Detergent (Prolystica™* 2X concentrate neutral detergent or equivalent pH neutral detergent)§
Rinse
00:15
Hot Tap Water
N/A
Thermal Rinse 01:00
Hot Purified Water heated to 180˚ F N/A (82˚ C)
Dry
High Setting 203˚ F N/A (95˚ C)
05:00
†Refer to user’s guide for complete instructions. ‡Tested to allow exposure to 203° F (95° C) for up to 5 minutes during the automated washer’s thermal rinse cycle to allow for variations in local requirements and regulations. (See note, Automatic Cleaning section, iDrive™ Ultra IFU). §Tested for material compatibility with cleaners that have pH ranges from neutral to 10.8 (such as Dr. Weigert neodisher™* Mediclean forte). Please refer to the cleaner’s manufacturer for information on the microbiological effectiveness of the cleaner. (See note, Automatic Cleaning section, iDrive™ Ultra IFU).
STERILIZATION For complete instructions, always refer to the User’s Guide.
Prevacuum (Hi Vac) Steam Sterilization Use autoclave sterilization only. Do NOT flash-sterilize. Do not use hydrogen peroxide gas plasma (Sterrad™*).
Exposure Temperature
Exposure Time
Vacuum Dry Time
132–137° C (270–279° F)
4–10 minutes
20–40 minutes
134–137° C (272–279° F)
3–10 minutes
20–40 minutes
CHARGING AND CLEANING THE BATTERY For complete instructions, always refer to the User’s Guide.
Charging 1. 2. 3.
Insert the battery pack into the battery charger, aligning the connectors. The charging cycle will start automatically when the battery pack is inserted correctly into the battery charger. The charge status indicator light panel will indicate the following: ▪ Green: battery pack is ready for use. ▪ Flashing Amber: battery pack is charging, and is not ready for use. ▪ Flashing Red: remove and reinsert battery; if flashing red persists replace battery pack.
Cleaning Clean the battery pack after each use and prior to charging. 1. Do NOT sterilize the battery pack, immerse it in liquid, or expose it to excessive moisture. 2. Wipe down all exposed surfaces of the battery pack with a lint-free cloth moistened with a detergent solution, or a pre-moistened alcohol wipe. 3. Avoid exposing the battery pack connectors to excessive moisture, conductive materials, or contaminates.
Storage Store the components at room temperature, at a humidity between 0 and 90%. For battery pack and charger store at a humidity between 0 and 85%.
1. iDrive Ultra Powered Stapling System [instructions for use]. New Haven, CT: Medtronic; 2012. © 2018 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 03/2018 – US180083 – [WF# 2281033]