CARESCAPE B850, B650, B450, module frames, and acquisition modules Installation and maintenance checkforms Software version 3
CARESCAPE B850, B650, B450, module frames, and acquisition modules English 3rd edition 2109358-006 © 2018 General Electric Company. All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice. For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies, Inc. and GE Healthcare Finland Oy.
2
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006 2018-11-28
Contents About the check forms ... 5 Planned maintenance schedules... 6 Service requirements... 6 Manufacturer responsibility ... 7 CARESCAPE B850 system installation check ... 9 CARESCAPE B850 maintenance check ... 15 CARESCAPE B650 installation check ... 19 CARESCAPE B650 maintenance check ... 25 CARESCAPE B450 installation check ... 29 CARESCAPE B450 maintenance check ... 35 Patient Data Module maintenance check ... 39 E-PiCCO module maintenance check ... 45 E-COPSv, E-COP modules maintenance check... 49 E-PP, E-PT modules maintenance check ... 53 E-MASIMO module maintenance check ... 57 E-NSATX module maintenance check ... 59 E-ENTROPY module maintenance check... 61 E-EEGX module and N-EEGX headbox maintenance check... 63 E-BIS module maintenance check ... 65 E-NMT module maintenance check... 67 CARESCAPE Respiratory Modules maintenance check ... 71 E-miniC module maintenance check... 75 F5, F7 Module frames maintenance check ... 77
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
3
4
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
About the check forms This document contains installation and maintenance check forms for CARESCAPE B850, B650, B450, module frames, and acquisition modules. Use the check forms during installation and maintenance check procedures. Record the results and activities on the check forms. Perform the installation check after the hardware installation and the platform configuration are completed, before taking the monitor into clinical use. Perform the maintenance check after any corrective maintenance and during the periodic maintenance. After a completed checkout archive the signed check form and give it to the customer. WARNING
SAFETY HAZARD. To avoid risks to personnel and patient, or damage to the equipment, only perform maintenance procedures described in this manual. Unauthorized modifications can lead to safety hazards.
Refer to the related service manual for detailed information regarding the checkout procedures. Refer to the user’s manual for important safety information regarding the whole system. The list below indicates the products (brands, models and descriptions as applicable) with which these check forms are to be used:
2109358-006
●
CARESCAPE B850 MBC313
●
CARESCAPE B650 MBB313
●
CARESCAPE B450 MBA313
●
CARESCAPE D19KT VER01
●
Frame F5–01
●
Frame F7–01
●
E-COP-01
●
E-COPSv-01
●
E-PiCCO-00
●
E-sCAiO-00
●
E-sCAiOV-00
●
E-sCAiOVX-00
●
E-sCO-00
●
E-sCOV-00
●
E-sCOVX-00
●
E-miniC-00
●
E-MASIMO-00
●
E-NSATX-00
●
E-PP-00
●
E-PT-00
●
E-ENTROPY-01
●
E-NMT-01
●
E-EEGX MJD101
CARESCAPE B850, B650, B450, module frames, and acquisition modules
5
●
N-EEGX MJS101
●
E-BIS-01
●
Patient Data Module
Planned maintenance schedules See the user manual for the instructions about the cleaning and care schedules and about the regular calibration check intervals. Service personnel must perform the planned maintenance check as follows: ●
For each monitor and module frame every two years after installation.
●
For each PDM, E-miniC and Respiratory module every 12 months after installation.
●
For all other modules every two years after installation.
The planned maintenance parts must be replaced according to the replacement interval specified in each maintenance check form. WARNING
Planned maintenance should be carried out at recommended interval. Failure to implement the recommended maintenance schedule may cause equipment failure and possible health hazards.
NOTE
The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an Equipment Maintenance Agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
NOTE
The planned maintenance check must be performed to the whole patient monitoring system, including all the connected devices. See the related service manuals for information about the planned maintenance checks for the connected devices
Service requirements Follow the service requirements listed below.
6
●
Refer servicing of the equipment to qualified service personnel only. Service personnel servicing this product must have an appropriate technical qualification, or equivalent work experience, and be familiar with the service requirements described in this manual and in any related service documentation. Service training for the product is recommended.
●
Any unauthorized attempt to repair equipment under warranty voids that warranty.
●
It is the user's responsibility to report the need for service to GE or to one of their authorized agents.
●
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
●
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
WARNING
Do not perform any service activities on the monitor in the patient vicinity while a patient is being connected to the monitor.
CAUTION
DISPOSAL. At the end of its service life, the product described in this manual, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of each product. If you have any questions concerning disposal of a product, please contact GE or its representatives.
Manufacturer responsibility GE is responsible for the effects on safety, reliability, and performance of the equipment only if:
2109358-006
●
Assembly operations, extensions, readjustments, modifications, servicing, or repairs are carried out by authorized service personnel.
●
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
●
The equipment is used in accordance with the instructions for use.
●
The equipment is installed, maintained and serviced in accordance with the instructions provided in the related technical manuals.
CARESCAPE B850, B650, B450, module frames, and acquisition modules
7
8
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
CARESCAPE B850 system installation check Customer:
Monitor type: B850 MBC313
Service record #:
Software version:
Service engineer
Service Manual: 2109358–001
Monitor S/N:
Other devices connected to the monitor system Device
Serial number
Device
Serial number
Prior to testing, verify that all equipment is calibrated via “Cal” labeling and record Cal Due Dates. Measuring equipment used Equipment/ tool
Manufacturer
PASS = Test passed
Model/ Type/ Part No
Calibration due date
Serial number/ ID
N.A.= Test is not applicant, a component or a feature is not part of tested system. Pass
Visual inspection
FAIL = Test failed Fail
Comments
All product labeling, markings and symbols are intact and readable. The monitor and the connected devices are undamaged. The monitor and the connected peripheral devices are properly mounted with specified mounting solutions. The cables between the CPU and the connected devices are intact, properly connected and secured to the right connectors. The acquisition modules are properly connected and locked. Electrical safety tests Date of manufacture of the device: Is it less than 12 months since the device was manufactured? Yes:
Yes: you do not have to perform the electrical safety tests. To continue the installation check, mark the electrical safety test completion N.A. and proceed to Functional check.
No:
No: perform the electrical safety tests.
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
9
Pass
Electrical safety tests
Fail
Monitor Power outlet is correctly wired. Power cord and plug are undamaged and all conductors are properly connected. Ground integrity test Performed ground integrity test:
a.) Ground continuity test b.) Impedance of protective ground connection
Ground integrity task
Observed result ohms
Acceptance criteria
Without power cord
≤ 0.1 ohms
With power cord
≤ 0.2 ohms
Earth leakage current test
Observed result mA
Acceptance criteria
Normal polarity, Neutral closed (NC)
≤ 5 mA
Normal polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral closed (NC)
≤ 5 mA
Touch current test
Observed result μA
Acceptance criteria
Normal polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Normal polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Normal polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Electrical safety tests Multiparameter module type:
Observed result μA (d.c.)
S/N:
ECG
Acceptance criteria
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
SpO2/ IP
Patient (source) leakage current tests, using a test body. Normal polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 10 μA (d.c.)
Normal polarity, Neutral open, GND closed, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
10
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
Electrical safety tests Multiparameter module type:
Observed result μA (d.c.)
S/N:
ECG
Acceptance criteria
Pass
Fail
Pass
Fail
SpO2/ IP
Normal polarity, Neutral closed, GND open, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND open, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral open, GND closed, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 10 μA (d.c.)
Patient (sink) leakage current test. Normal polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 50 μA (d.c.)
Electrical safety tests Module type:
Observed result μA (d.c.)
Acceptance criteria
S/N:
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
11
Patient (source) leakage current tests, using a test body. Normal polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 10 μA (d.c.)
Normal polarity, Neutral open, GND closed, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Normal polarity, Neutral closed, GND open, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND open, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral open, GND closed, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 10 μA (d.c.)
Patient (sink) leakage current test. Normal polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 50 μA (d.c.)
Reversed polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 50 μA (d.c.)
Electrical safety test
Pass
N.A.
Fail
Pass
N.A.
Fail
Test completion Functional check
Acceptance criteria
Alarm light
1. Primary display: ● Power indicator is lit. ● Blue, yellow and red alarm lights
illuminate. ● Pause audio alarm indicator illuminates.
2. Secondary display, if connected: ● Power indicator is lit. ● Blue, yellow and red alarm lights
illuminate. ● Pause audio alarm indicator illuminates.
Start-up
Monitor starts up normally without error messages.
-
Displays: test both primary display and secondary display, if connected.
1. Picture quality: Text is readable, images are clear and brightness is adequate.
-
2. Touchscreen control: Touchscreen is correctly calibrated and operates correctly. 3. Trim Knob: Selected menu opens and selected activity starts.
12
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
Functional check
Acceptance criteria
Status of connected devices
Host device and all connected devices are identified and configured correctly.
IX printers
Test page is printed to the selected printer.
Insite RSvP connectivity
The patient monitor is connected to the remote service server.
USB remote control: test both a hard key and the Trim Knob.
Selected menu opens and selected activity starts.
Mouse
Selected menu opens and selected activity starts.
Keyboard
Keyboard language configuration is correct.
Barcode reader
Data is correctly populated.
MC Network communication
Network symbol and the other monitor’s parameter data is correctly shown on the screen.
Citrix connection
The program is launched correctly.
Pass
N.A.
Fail
-
-
Test completion Notes
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
13
14
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
CARESCAPE B850 maintenance check Customer:
Monitor type: B850 MBC313
Service record #:
Software version:
Service engineer:
Module type:
S/N:
Service Manual: 2109358-001
Start date:
Corrective maintenance
Planned maintenance
S/N:
Prior to testing, verify that all equipment is calibrated via “Cal” labeling and record Cal Due Dates. Measuring equipment used Equipment/ tool
Manufacturer
PASS = Test passed
Model/ Type/ Part No
Calibration due date
Serial number/ ID
N.A.= Test is not applicant, a component or a feature is not part of tested system.
FAIL = Test failed
Replacement of planned maintenance parts
Replaced
N.A.
CPU battery (every 6th year) Pass
Visual inspection
Fail
Comments
All product labeling, markings and symbols are intact and readable. The monitor and the connected devices are undamaged. The monitor and the connected peripheral devices are properly mounted with specified mounting solutions. The cables between the CPU and the connected devices are intact, properly connected and secured to the right connectors. The acquisition modules are properly connected and locked. Pass
Electrical safety tests
Fail
Monitor Power outlet is correctly wired. Power cord and plug are undamaged and all conductors are properly connected. Ground integrity test Performed ground integrity test:
a.) Ground continuity test b.) Impedance of protective ground connection
Ground integrity task
Observed result ohms
Acceptance criteria
Without power cord
≤ 0.1 ohms
With power cord
≤ 0.2 ohms
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
Pass
Fail
15
Earth leakage current test
Observed result mA
Pass
Fail
Pass
Fail
Pass
N.A.
Fail
Pass
N.A.
Fail
Acceptance criteria
Normal polarity, Neutral closed (NC)
≤ 5 mA
Normal polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral closed (NC)
≤ 5 mA
Touch current test
Observed result μA
Acceptance criteria
Normal polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Normal polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Normal polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Electrical safety test Test completion Functional check
Acceptance criteria
Alarm light
1. Primary display: ● Power indicator is lit. ● Blue, yellow and red alarm lights
illuminate. ● Pause audio alarm indicator illuminates.
2. Secondary display, if connected: ● Power indicator is lit. ● Blue, yellow and red alarm lights
illuminate. ● Pause audio alarm indicator illuminates.
Start-up
16
Monitor starts up normally without error messages.
CARESCAPE B850, B650, B450, module frames, and acquisition modules
-
2109358-006
Functional check
Acceptance criteria
Displays: test both primary display and secondary display, if connected.
1. Picture quality: Text is readable, images are clear and brightness is adequate.
Pass
N.A.
Fail
-
2. Touchscreen control: Touchscreen is correctly calibrated and operates correctly. 3. Trim Knob: Selected menu opens and selected activity starts. Status of connected devices
Host device and all connected devices are identified and configured correctly.
IX printers
Test page is printed to the selected printer.
Insite RSvP connectivity
The patient monitor is connected to the remote service server.
USB remote control: test both a hard key and the Trim Knob.
Selected menu opens and selected activity starts.
Mouse
Selected menu opens and selected activity starts.
Keyboard
Keyboard language configuration is correct.
Barcode reader
Data is correctly populated.
MC Network communication
Network symbol and the other monitor’s parameter data is correctly shown on the screen.
Citrix connection
The program is launched correctly.
-
-
Test completion Notes
Used service parts
Signature
2109358-006
Date
CARESCAPE B850, B650, B450, module frames, and acquisition modules
17
18
CARESCAPE B850, B650, B450, module frames, and acquisition modules
2109358-006
CARESCAPE B650 installation check Customer:
Monitor type: B650 MBB313
Service record #:
Software version:
Service engineer
Service Manual: 2109358-002
Monitor S/N:
Other devices connected to the monitor system Device
Serial number
Device
Serial number
Prior to testing, verify that all equipment is calibrated via “Cal” labeling and record Cal Due Dates. Measuring equipment used Equipment/ tool
Manufacturer
PASS = Test passed
Model/ Type/ Part No
Serial number/ ID
N.A.= Test is not applicant, a component or a feature is not part of tested system. Pass
Visual inspection
Calibration due date
FAIL = Test failed Fail
Comments
All product labeling, markings and symbols are intact and readable. The monitor and the connected devices are undamaged. The monitor and the connected peripheral devices are properly mounted with specified mounting solutions. The cables between the patient monitor and the connected devices are intact, properly connected and secured to the right connectors. The acquisition modules are properly connected and locked. The pivoting module frame and battery door are properly locked. Electrical safety tests Date of manufacture of the device: Is it less than 12 months since the device was manufactured? Yes:
Yes: you do not have to perform the electrical safety tests. To continue the installation check, mark the electrical safety test completion N.A. and proceed to Functional check.
No:
No: perform the electrical safety tests.
2109358-006
CARESCAPE B850, B650, B450, module frames, and acquisition modules
19
Pass
Electrical safety tests
Fail
Monitor Power outlet is correctly wired. Power cord and plug are undamaged and all conductors are properly connected. Ground integrity test Performed ground integrity test:
a.) Ground continuity test b.) Impedance of protective ground connection
Ground integrity task
Observed result ohms
Acceptance criteria
Without power cord
≤ 0.1 ohms
With power cord
≤ 0.2 ohms
Earth leakage current test
Observed result mA
Acceptance criteria
Normal polarity, Neutral closed (NC)
≤ 5 mA
Normal polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral open (SFC)
≤ 10 mA
Reversed polarity, Neutral closed (NC)
≤ 5 mA
Touch current test
Observed result μA
Acceptance criteria
Normal polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Normal polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Normal polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND open Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral open, GND closed Single Fault Condition (SFC)
≤ 500 μA EN /IEC ≤ 300 μA UL
Reversed polarity, Neutral closed, GND closed Normal condition (NC)
≤ 100 μA
Electrical safety tests Multiparameter module type:
Observed result μA (d.c.)
S/N:
ECG
Acceptance criteria
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
SpO2/ IP
Patient (source) leakage current tests, using a test body. Normal polarity, Neutral closed, GND closed, Normal condition (NC)
≤ 10 μA (d.c.)
Normal polarity, Neutral open, GND closed, Single Fault Condition (SFC)
≤ 50 μA (d.c.)
20